{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-10-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "219693",
      "product_res_number": "Z-2201-2026",
      "event_date_initiated": "2026-04-08",
      "event_date_posted": "2026-05-14",
      "recall_status": "Open, Classified",
      "res_event_number": "98750",
      "product_code": "LKK",
      "pma_numbers": [
        "P860004"
      ],
      "product_description": "Medtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application, loaded on the following devices:   \n1. PROG CT900A CLINICIAN TABLET US, Model Number CT900A; \n2. PROG CT900B CLINICIAN TABLET EU -UK, Model Number CT900B; \n3. PROG CT900C CLINICIAN TABLET UK, Model Number CT900C; \n4. PROG CT900D CLINICIAN TABLET GLOBAL, Model Number CT900D; \n5. PROG CT900E CLINICIAN TABLET GLBL SMSNG, Model Number CT900E; \n6. PROG CT900F CLINICIAN TABLET GLBL SMSNG, Model Number CT900F; implanted programmable infusion pump",
      "code_info": "Software application (app) version 1.x and 2.x, loaded on the following units:  \n1. Model Number CT900A, GTIN 00643169890923, Serial Numbers:  NPL0004445; \n\n2. Model Number CT900B, GTIN 00643169890947, Serial Numbers:  NPL1000042, NPL1000045, NPL1000046, NPL1000048, NPL1000050, NPL1000051, NPL1000053, NPL1000054, NPL1000059, NPL1000060, NPL1000062, NPL1000065, NPL1000069, NPL1000070, NPL1000071, NPL1000072, NPL1000073, NPL1000074, NPL1000075, NPL1000077, NPL1000081, NPL1000082, NPL1000083, NPL1000085, NPL1000086, NPL1000087, NPL1000090, NPL1000091, NPL1000093, NPL1000094, NPL1000098, NPL1000117, NPL1000133, NPL1000135, NPL1000137, NPL1000138, NPL1000151, NPL1000159, NPL1000160, NPL1000164, NPL1000166, NPL1000167, NPL1000169, NPL1000173, NPL1000174, NPL1000179, NPL1000180, NPL1000183, NPL1000185, NPL1000186, NPL1000187, NPL1000188, NPL1000191, NPL1000192, NPL1000193, NPL1000194, NPL1000195, NPL1000196, NPL1000230, NPL1000235, NPL1000237, NPL1000242, NPL1000244, NPL1000245, NPL1000246, NPL1000248, NPL1000249, NPL1000252, NPL1000254, NPL1000257, NPL1000258, NPL1000281, NPL1000292, NPL1000297, NPL1000301, NPL1000303, NPL1000307, NPL1000321, NPL1000322, NPL1000323, NPL1000325, NPL1000326, NPL1000332, NPL1000334, NPL1000343, NPL1000361, NPL1000362, NPL1000363, NPL1000364, NPL1000365, NPL1000366, NPL1000367, NPL1000369, NPL1000370, NPL1000381, NPL1000483, NPL1000486, NPL1000487, NPL1000501, NPL1000502, NPL1000505, NPL1000506, NPL1000507, NPL1000509, NPL1000510, NPL1000512, NPL1000515, NPL1000516, NPL1000518, NPL1000519, NPL1000528, NPL1000529, NPL1000539, NPL1000540, NPL1000555, NPL1000556, NPL1000558, NPL1000560, NPL1000561, NPL1000562, NPL1000563, NPL1000564, NPL1000571, NPL1000573, NPL1000574, NPL1000575, NPL1000576, NPL1000577, NPL1000578, NPL1000580, NPL1000585, NPL1000592, NPL1000603, NPL1000606, NPL1000612, NPL1000627, NPL1000630, NPL1000631, NPL1000632, NPL1000634, NPL1000637, NPL1000641, NPL1000652, NPL1000653, NPL1001194, NPL1001195, NPL1001201, NPL1001204, NPL1001205, NPL1001206, NPL1001207, NPL1001208, NPL1001210, NPL1001211, NPL1001212, NPL1001216, NPL1001223, NPL1001224, NPL1001225, NPL1001226, NPL1001228, NPL1001229, NPL1001230, NPL1001231, NPL1001235, NPL1001236, NPL1001237, NPL1001238, NPL1001240, NPL1001241, NPL1001243, NPL1001244, NPL1001250, NPL1001251, NPL1001254, NPL1001259, NPL1001264, NPL1001265, NPL1001272, NPL1001273, NPL1001274, NPL1001277, NPL1001282, NPL1001283, NPL1001290, NPL1001291, NPL1001306, NPL1001307, NPL1001313, NPL1001314, NPL1001328, NPL1001330, NPL1001332, NPL1001333, NPL1001334, NPL1001336, NPL1001337, NPL1001339, NPL1001341, NPL1001345, NPL1001349, NPL1001350, NPL1001351, NPL1001353, NPL1001354, NPL1001360, NPL1001361, NPL1001362, NPL1001366, NPL1001367, NPL1001369, NPL1001372, NPL1001375, NPL1001380, NPL1001381, NPL1001385, NPL1001386, NPL1001389, NPL1001398, NPL1001407, NPL1001408, NPL1001410, NPL1001412, NPL1001419, NPL1001421, NPL1001423, NPL1001430, NPL1001431, NPL1001434, NPL1001435, NPL1001436, NPL1001439, NPL1001445, NPL1001449, NPL1001458, NPL1001461, NPL1001464, NPL1001465, NPL1001468, NPL1001482, NPL1001483, NPL1001486, NPL1001499, NPL1001500, NPL1001505, NPL1001507, NPL1001510, NPL1001512, NPL1001513, NPL1001514, NPL1001518, NPL1001520, NPL1001524, NPL1001526, NPL1001531, NPL1001532, NPL1001537, NPL1001539, NPL1001540, NPL1001541, NPL1001542, NPL1001544, NPL1001545, NPL1001546, NPL1001547, NPL1001549, NPL1001550, NPL1001678, NPL1001689, NPL1001694, NPL1001695, NPL1001697, NPL1001699, NPL1001700, NPL1001701, NPL1001702, NPL1001705, NPL1001707, NPL1001708, NPL1001709, NPL1001710, NPL1001711, NPL1001716, NPL1001717, NPL1001723, NPL1001724, NPL1001727, NPL1001728, NPL1001739, NPL1001740, NPL1001741, NPL1001742, NPL1001743, NPL1001744, NPL1001745, NPL1001748, NPL1001749, NPL1001750, NPL1001754, NPL1001755, NPL1001756, NPL1001757, NPL1001759, NPL1001760, NPL1001761, NPL1001763, NPL1001764, NPL1001765, NPL1001766, NPL1001767, NPL1001769, NPL1001770, NPL1001773, NPL1001777, NPL1001784, NPL1001787, NPL1001789, NPL1001792, NPL1001794, NPL1001795, NPL1001797, NPL1001798, NPL1001799, NPL1001800, NPL1001801, NPL1001803, NPL1001804, NPL1001805, NPL1001806, NPL1001811, NPL1001812, NPL1001815, NPL1001816, NPL1001817, NPL1001820, NPL1001823, NPL1001824, NPL1001825, NPL1001829, NPL1001830, NPL1001833, NPL1001841, NPL1001842, NPL1001843, NPL1001844, NPL1001847, NPL1001875, NPL1001876, NPL1001877, NPL1001879, NPL1001880, NPL1001881, NPL1001882, NPL1001884, NPL1001885, NPL1001886, NPL1001887, NPL1001888, NPL1001889, NPL1001891, NPL1001897, NPL1001899, NPL1001900, NPL1001901, NPL1001903, NPL1001904, NPL1001905, NPL1001937, NPL1001940, NPL1001946, NPL1001947, NPL1001948, NPL1001950, NPL1001955, NPL1001957, NPL1001959, NPL1001963, NPL1001966, NPL1001968, NPL1001969, NPL1001970, NPL1001971, NPL1001972, NPL1001973, NPL1001974, NPL1001975, NPL1001976, NPL1001977, NPL1001995, NPL1001996, NPL1002004, NPL1002007, NPL1002010, NPL1002016, NPL1002020, NPL1002021, NPL1002025, NPL1002034, NPL1002035, NPL1002038, NPL1002048, NPL1002049, NPL1002050, NPL1002051, NPL1002053, NPL1002054, NPL1002064, NPL1002065, NPL1002066, NPL1002067, NPL1002068, NPL1002088, NPL1002089, NPL1002090, NPL1002091, NPL1002092, NPL1002093, NPL1002094, NPL1002095, NPL1002124, NPL1002126, NPL1002132, NPL1002137, NPL1002146, NPL1002152, NPL1002190, NPL1002191, NPL1002192, NPL1002194, NPL1002200, NPL1002203, NPL1002211, NPL1002214, NPL1002219, NPL1002221, NPL1002222, NPL1002223, NPL1002224, NPL1002225, NPL1002226, NPL1002227, NPL1002228, NPL1002229, NPL1002230, NPL1002231, NPL1002232, NPL1002233, NPL1002234, NPL1002236, NPL1002238, NPL1002239, NPL1002240, NPL1002243, NPL1002244, NPL1002245, NPL1002246, NPL1002247, NPL1002254, NPL1002255, NPL1002256, NPL1002311, NPL1002312, NPL1002313, NPL1002316, NPL1002324, NPL1002329, NPL1002330, NPL1002334, NPL1002335, NPL1002337, NPL1002338, NPL1002339, NPL1002340, NPL1002341, NPL1002342, NPL1002343, NPL1002344, NPL1002347, NPL1002349, NPL1002355, NPL1002356, NPL1002357, NPL1002358, NPL1002359, NPL1002360, NPL1002361, NPL1002362, NPL1002363, NPL1002364, NPL1002367, NPL1002368, NPL1002369, NPL1002370, NPL1002371, NPL1002373, NPL1002375, NPL1002380, NPL1002381, NPL1002382, NPL1002383, NPL1002386, NPL1002392, NPL1002396, NPL1002399, NPL1002401, NPL1002405, NPL1002408, NPL1002409, NPL1002410, NPL1002411, NPL1002412, NPL1002413, NPL1002414, NPL1002415, NPL1002416, NPL1002417, NPL1002418, NPL1002421, NPL1002427, NPL1002428, NPL1002431, NPL1002436, NPL1002438, NPL1002443, NPL1002445, NPL1002460, NPL1002461, NPL1002462, NPL1002464, NPL1002465, NPL1002466, NPL1002467, NPL1002470, NPL1002471, NPL1002474, NPL1002482, NPL1002484, NPL1002490, NPL1002491, NPL1002495, NPL1002496, NPL1002499, NPL1002500, NPL1002501, NPL1002502, NPL1002503, NPL1002505, NPL1002508, NPL1002521, NPL1002530, NPL1002537, NPL1002540, NPL1002544, NPL1002545, NPL1002548, NPL1002549, NPL1002555, NPL1002556, NPL1002557, NPL1002564, NPL1002567, NPL1002568, NPL1002572, NPL1002592, NPL1002593, NPL1002594, NPL1002595, NPL1002596, NPL1002613, NPL1002616, NPL1002617, NPL1002618, NPL1002620, NPL1002626, NPL1002628, NPL1002630, NPL1002631, NPL1002632, NPL1002636, NPL1002637, NPL1002638, NPL1002639, NPL1002640, NPL1002642, NPL1002646, NPL1002648, NPL1002650, NPL1002651, NPL1002652, NPL1002653, NPL1002654, NPL1002656, NPL1002659, NPL1002661, NPL1002666, NPL1002668, NPL1002669, NPL1002671, NPL1002673, NPL1002677, NPL1002679, NPL1002684, NPL1002685, NPL1002686, NPL1002687, NPL1002689, NPL1002690, NPL1002693, NPL1002695, NPL1002696, NPL1002698, NPL1002706, NPL1002709, NPL1002710, NPL1002711, NPL1002712, NPL1002714, NPL1002718, NPL1002719, NPL1002720, NPL1002721, NPL1002728, NPL1002729, NPL1002730, NPL1002731, NPL1002732, NPL1002733, NPL1002735, NPL1002736, NPL1002737, NPL1002739, NPL1002740, NPL1002741, NPL1002743, NPL1002753, NPL1002756, NPL1002757, NPL1002760, NPL1002761, NPL1002778, NPL1002779, NPL1002780, NPL1002781, NPL1002784, NPL1002786, NPL1002787, NPL1002789, NPL1002791, NPL1002792, NPL1002794, NPL1002796, NPL1002800; \n\n3. Model Number CT900C, GTIN 00643169890961, Serial Numbers:  NPL1000429, NPL1000430, NPL1000442, NPL1000443, NPL1000444, NPL1000445, NPL1000446, NPL1000447, NPL1000448, NPL1000451, NPL1000453, NPL1000771, NPL1000774, NPL1000776, NPL1000777, NPL1000780, NPL1000781, NPL1000794, NPL1000795, NPL1000797, NPL1000798, NPL1000799, NPL1000802, NPL1000804, NPL1000805, NPL1000806, NPL1000808, NPL1000809, NPL1000810, NPL1000811, NPL1000813, NPL1000814, NPL1000815, NPL1000817, NPL1000818, NPL1000819, NPL1000827, NPL1000830, NPL1000832, NPL1000833, NPL1000834, NPL1000835, NPL1000838, NPL1000845, NPL1000847, NPL1000854, NPL1000855, NPL1000858, NPL1001552, NPL1001562, NPL1001564, NPL1001565, NPL1001566, NPL1001570, NPL1001571, NPL1001574, NPL1001582, NPL1001583, NPL1001588, NPL1001590, NPL1001591, NPL1001592, NPL1001594, NPL1001595, NPL1001596, NPL1001601, NPL1001602, NPL1001605, NPL1001606, NPL1001607, NPL1001608, NPL1001609, NPL1001610, NPL1001611, NPL1001615, NPL1001617, NPL1001618, NPL1001620, NPL1001621, NPL1001627, NPL1001629, NPL1001630, NPL1001631, NPL1001632, NPL1001637, NPL1001638, NPL1001644, NPL1001645, NPL1001647, NPL1001649, NPL1001650, NPL1001651, NPL1001652, NPL1001656, NPL1001657, NPL1001670, NPL1001673, NPL1001675, NPL1002257, NPL1002259, NPL1002260, NPL1002262, NPL1002264, NPL1002265, NPL1002266, NPL1002267, NPL1002268, NPL1002275, NPL1002279, NPL1002280, NPL1002281, NPL1002282, NPL1002284, NPL1002285, NPL1002286, NPL1002288, NPL1002294, NPL1002295, NPL1002297, NPL1002298, NPL1002306, NPL1002307, NPL1002308, NPL1002309, NPL1002310; \n\n4. Model Number CT900D, (1) GTIN  00763000077730, Serial Numbers:  NPL3000277, NPL3000278, NPL3000279, NPL3000280, NPL3000281, NPL3000282, NPL3000283, NPL3000285, NPL3000290, NPL3000292, NPL3000293, NPL3000298, NPL3000299, NPL3000300, NPL3000301, NPL3000304, NPL3000305, NPL3000309, NPL3000310, NPL3000312, NPL3000460, NPL3000555, NPL3000563, NPL3000582, NPL3000590, NPL3000591, NPL3002090, NPL3002102, NPL3002311, NPL3002312, NPL3002313, NPL3002314, NPL3002315, NPL3002316, NPL3002318, NPL3002760, NPL3002761, NPL3002762, NPL3002763, NPL3002816, NPL3002995, NPL3002997, NPL3002998, NPL3002999, NPL3003000, NPL3003001, NPL3003002, NPL3003003, NPL3003004, NPL3003005, NPL3003009, NPL3003010, NPL3003011, NPL3003012, NPL3003013, NPL3003014, NPL3003015, NPL3003016, NPL3003018, NPL3003021, NPL3003022, NPL3003023, NPL3003024, NPL3003026, NPL3003028, NPL3003029, NPL3003030, NPL3003031, NPL3003032, NPL3003036, NPL3003037, NPL3003038, NPL3003042, NPL3003046, NPL3003048, NPL3003050, NPL3003051, NPL3003054, NPL3003059, NPL3003060, NPL3003062, NPL3003065, NPL3003067, NPL3003068, NPL3003070, NPL3003072, NPL3003073, NPL3003074, NPL3003076, NPL3003077, NPL3003078, NPL3003086, NPL3003260, NPL3003267, NPL3003268, NPL3003271, NPL3003275, NPL3003277, NPL3003280, NPL3003282, NPL3003283, NPL3003284, NPL3003402, NPL3003409, NPL3003412, NPL3003413, NPL3003415, NPL3003416, NPL3003417, NPL3003418, NPL3003419, NPL3003420, NPL3003422, NPL3003423, NPL3003425, NPL3003427, NPL3003513, NPL3003517, NPL3003518, NPL3003521, NPL3003524, NPL3003528, NPL3003529, NPL3003530, NPL3003532, NPL3003533, NPL3003534, NPL3003535, NPL3003536, NPL3003537, NPL3003538, NPL3003539, NPL3003540, NPL3003542, NPL3003543, NPL3003544, NPL3003545, NPL3003546, NPL3003547, NPL3003548, NPL3003549, NPL3003551, NPL3003552, NPL3003558, NPL3003560, NPL3003566, NPL3003567, NPL3003569, NPL3003570, NPL3003576, NPL3003577, NPL3003578, NPL3003579, NPL3003580, NPL3003581, NPL3003598, NPL3003599, NPL3003601, NPL3003745, NPL3003756, NPL3003766, NPL3003817, NPL3003821, NPL3003823, NPL3003824, NPL3003850, NPL3003885, NPL3003886, NPL3003887, NPL3003890, NPL3003891, NPL3003892, NPL3003893, NPL3003894, NPL3003895, NPL3003896, NPL3003897, NPL3003899, NPL3003900, NPL3003901, NPL3003904, NPL3003905, NPL3003906, NPL3003907, NPL3003921, NPL3003923, NPL3003924, NPL3003929, NPL3003943, NPL3003944, NPL3003951, NPL3003952, NPL3003954, NPL3004050, NPL3004052, NPL3004053, NPL3004057, NPL3004061, NPL3004062, NPL3004063, NPL3004064, NPL3004066, NPL3004067, NPL3004068, NPL3004069, NPL3004071, NPL3004072, NPL3004076, NPL3004078, NPL3004081, NPL3004082, NPL3004087, NPL3004095, NPL3004096, NPL3004102, NPL3004114, NPL3004117, NPL3004118, NPL3004139, NPL3004140, NPL3004141, NPL3004143, NPL3004144, NPL3004145, NPL3004146, NPL3004148, NPL3004149, NPL3004151, NPL3004152, NPL3004153, NPL3004340, NPL3004342, NPL3004344, NPL3004345, NPL3004352, NPL3004353, NPL3004417, NPL3004421, NPL3004436, NPL3004793, NPL3004841, NPL3004860, NPL3004871, NPL3004879, NPL3004895, NPL3004897, NPL3004905, NPL3004910, NPL3004913, NPL3004914, NPL3004918, NPL3004925, NPL3004942, NPL3004951, NPL3004952, NPL3004960, NPL3004969, NPL3004971, NPL3004992, NPL3005006, NPL3005009, NPL3005010, NPL3005018, NPL3005025, NPL3005033, NPL3005044, NPL3005053, NPL3005054, NPL3005059, NPL3005060, NPL3005061, NPL3005065, NPL3005075, NPL3005136, NPL3005187, NPL3005199, NPL3005750, NPL3005752, NPL3005760, NPL3005769, NPL3005773, NPL3005779, NPL3005782, NPL3005783, NPL3005785, NPL3005790, NPL3005799, NPL3005801, NPL3005811, NPL3005814, NPL3005817, NPL3005818, NPL3005866, NPL3005876, NPL3005889, NPL3005893, NPL3006002, NPL3006011, NPL3006027, NPL3006033, NPL3006069, NPL3006072, NPL3006084, NPL3006132, NPL3006142, NPL3006146, NPL3006147, NPL3006148, NPL3006151, NPL3006152, NPL3006153, NPL3006155, NPL3006160, NPL3006161, NPL3006167, NPL3006168, NPL3006176, NPL3006179, NPL3006181, NPL3006188, NPL3006189, NPL3006190, NPL3006192, NPL3006194, NPL3006198, NPL3006203, NPL3006205, NPL3006209, NPL3006213, NPL3006217, NPL3006221, NPL3006225, NPL3006230, NPL3006237, NPL3006238, NPL3006242, NPL3006245, NPL3006247, NPL3006248, NPL3006249, NPL3006250, NPL3006253, NPL3006254, NPL3006255, NPL3006267, NPL3006279, NPL3006280, NPL3006289, NPL3006305, NPL3006309, NPL3006310, NPL3006312, NPL3006359, NPL3006360, NPL3006361, NPL3006366, NPL3006372, NPL3006374, NPL3006375, NPL3006377, NPL3006379, NPL3006380, NPL3006381, NPL3006389, NPL3006391, NPL3006392, NPL3006393, NPL3006394, NPL3006399, NPL3006400, NPL3006406, NPL3006407, NPL3006408, NPL3006410, NPL3006413, NPL3006414, NPL3006416, NPL3006417, NPL3006433, NPL3006443, NPL3006445, NPL3006446, NPL3006448, NPL3006449, NPL3006450, NPL3006452, NPL3006453, NPL3006454, NPL3006457, NPL3006458, NPL3006461, NPL3006465, NPL3006468, NPL3006469, NPL3006475, NPL3006476, NPL3006485, NPL3006486, NPL3006487, NPL3006489, NPL3006490, NPL3006491, NPL3006492, NPL3006494, NPL3006495, NPL3006496, NPL3006498, NPL3006513, NPL3006516, NPL3006518, NPL3006519, NPL3006520, NPL3006522, NPL3006523, NPL3006524, NPL3006525, NPL3006526, NPL3006527, NPL3006531, NPL3006532, NPL3006533, NPL3006534, NPL3006536, NPL3006572, NPL3006581, NPL3006586, NPL3006587, NPL3006600, NPL3006601, NPL3006602, NPL3006604, NPL3006605, NPL3006608, NPL3006612, NPL3006620, NPL3006625, NPL3006626, NPL3006627, NPL3006634, NPL3006825, NPL3006827, NPL3006885, NPL3006886, NPL3006887, NPL3006888, NPL3006889, NPL3006900, NPL3006902, NPL3006904, NPL3006909, NPL3006911, NPL3006917, NPL3006918, NPL3006920, NPL3006921, NPL3006922, NPL3006923, NPL3006924, NPL3006930, NPL3006932, NPL3006933, NPL3006934, NPL3006936, NPL3006938, NPL3006944, NPL3006946, NPL3006947, NPL3006948, NPL3006949, NPL3006950, NPL3006951, NPL3006952, NPL3006953, NPL3006954, NPL3006955, NPL3006962, NPL3006963, NPL3006964, NPL3006966, NPL3006967, NPL3006968, NPL3006971, NPL3006972, NPL3006974, NPL3006977, NPL3006979, NPL3006982, NPL3006983, NPL3006985, NPL3006986, NPL3007004, NPL3007011, NPL3007012, NPL3007053, NPL3007268, NPL3007269, NPL3007272, NPL3007273, NPL3007274, NPL3007276, NPL3007277, NPL3007362, NPL3007363, NPL3007365, NPL3007366, NPL3007367, NPL3007368, NPL3007370, NPL3007371, NPL3007372, NPL3007384, NPL3007385, NPL3007393, NPL3007439, NPL3007451, NPL3007453, NPL3007455, NPL3007456, NPL3007457, NPL3007458, NPL3007464, NPL3007465, NPL3007583, NPL3007584, NPL3007585, NPL3007589, NPL3007592.NPL3007593, NPL3007594, NPL3007596, NPL3007601, NPL3007605, NPL3007608, NPL3007611, NPL3007617, NPL3007619, NPL3007621, NPL3007672, NPL3007677, NPL3007679, NPL3007685, NPL3007688, NPL3007776, NPL3007867, NPL3007869, NPL3007873, NPL3007875, NPL3007877, NPL3007878, NPL3007886, NPL3007887, NPL3007889, NPL3007890, NPL3007895, NPL3007900, NPL3007901, NPL3007904, NPL3007907, NPL3007908, NPL3007909, NPL3007910, NPL3007914, NPL3007916, NPL3007917, NPL3007918, NPL3007921, NPL3007924, NPL3007926, NPL3007927, NPL3007929, NPL3007930, NPL3007937, NPL3007947, NPL3007948, NPL3008029, NPL3008031, NPL3008033, NPL3008034, NPL3008036, NPL3008039, NPL3008046, NPL3008048, NPL3008053, NPL3008058, NPL3008068, NPL3010566, NPL3010568, NPL3010569, NPL3010570, NPL3010573, NPL3010578, NPL3010588, NPL3010593, NPL3010607, NPL3010628, NPL3010629, NPL3010631, NPL3010642, NPL3010645, NPL3010647, NPL3010652, NPL3010653, NPL3010658, NPL3010660, NPL3010665, NPL3010683, NPL3010688, NPL3010690, NPL3010691, NPL3010692, NPL3010782, NPL3010784, NPL3010786, NPL3010793, NPL3010800, NPL3010806, NPL3010817, NPL3010820, NPL3010824, NPL3010828, NPL3010829, NPL3010831, NPL3010833, NPL3010835, NPL3010866, NPL3010874, NPL3010876, NPL3010877, NPL3010879, NPL3010880, NPL3010887, NPL3010891, NPL3010892, NPL3010901, NPL3010902, NPL3010904, NPL3010907, NPL3010908, NPL3010910, NPL3010911, NPL3010914, NPL3010915, NPL3010927, NPL3010928, NPL3010934, NPL3010940, NPL3010945, NPL3010946, NPL3010953, NPL3010954, NPL3010955, NPL3010958, NPL3010968, NPL3010991, NPL3010996, NPL3010997, NPL3010998, NPL3011008, NPL3011104, NPL3011106, NPL3011107, NPL3011108, NPL3011109, NPL3011113, NPL3011169, NPL3011217, NPL3011219, NPL3011221, NPL3011224, NPL3011226, NPL3011227, NPL3011228, NPL3011234, NPL3011235, NPL3011236, NPL3011239, NPL3011240, NPL3011241, NPL3011242, NPL3011243, NPL3011244, NPL3011245, NPL3011246, NPL3011247, NPL3011249, NPL3011267, NPL3011269, NPL3011286, NPL3011287, NPL3011367, NPL3011426, NPL3011427, NPL3011428, NPL3011429, NPL3011430, NPL3011434, NPL3011435, NPL3011436, NPL3011437, NPL3011445, NPL3011446, NPL3011447, NPL3011448, NPL3011449, NPL3011452, NPL3011459, NPL3011460, NPL3011462, NPL3011463, NPL3011481, NPL3011482, NPL3011485, NPL3011487, NPL3011489, NPL3011490, NPL3011568, NPL3011569, NPL3011572, NPL3011573, NPL3011574, NPL3011575, NPL3011587, NPL3011589, NPL3011590, NPL3011600, NPL3011602, NPL3011605, NPL3011606, NPL3011607, NPL3011632, NPL3011633, NPL3011637, NPL3011639, NPL3011651, NPL3011652, NPL3011653, NPL3011656, NPL3011701, NPL3011703, NPL3011704, NPL3011705, NPL3011706, NPL3011707, NPL3011708, NPL3011709, NPL3011716, NPL3011718, NPL3011719, NPL3011720, NPL3011725, NPL3011734, NPL3011738, NPL3011739, NPL3011741, NPL3011749, NPL3011750, NPL3011753, NPL3011754, NPL3011756, NPL3011757, NPL3011762, NPL3013178, NPL3013655, NPL3013656, NPL3013665, NPL3013669, NPL3013670, NPL3013671, NPL3013674, NPL3013675, NPL3013676, NPL3013679, NPL3013680, NPL3013683, NPL3013684, NPL3013687, NPL3013699, NPL3013702, NPL3013712, NPL3013713, NPL3013714, NPL3013715, NPL3013717, NPL3013719, NPL3013721, NPL3013740, NPL3013743, NPL3013746, NPL3013748, NPL3013751, NPL3013754, NPL3013755, NPL3013758, NPL3013759, NPL3013760, NPL3013762, NPL3013763, NPL3013764, NPL3013766, NPL3013767, NPL3013768, NPL3013769, NPL3013773, NPL3013781, NPL3013783, NPL3013786, NPL3013795, NPL3013796, NPL3013797, NPL3013798, NPL3013799, NPL3013817, NPL3013818, NPL3013819, NPL3013821, NPL3013825, NPL3013826, NPL3013827, NPL3013830, NPL3013856, NPL3013857, NPL3013858, NPL3013860, NPL3013861, NPL3013862, NPL3013863, NPL3013864, NPL3013865, NPL3013871, NPL3013875, NPL3013876, NPL3013877, NPL3013878, NPL3013879, NPL3013890, NPL3013895, NPL3013896, NPL3013897, NPL3013901, NPL3013903, NPL3013904, NPL3013905, NPL3013906, NPL3013913, NPL3013916, NPL3013920, NPL3013923, NPL3013931, NPL3013932, NPL3013933, NPL3013935, NPL3013936, NPL3013937, NPL3013938, NPL3013939, NPL3013940, NPL3013941, NPL3013942, NPL3013943, NPL3013944, NPL3013949, NPL3013952, NPL3013953, NPL3013954, NPL3013960, NPL3013961, NPL3013962, NPL3013963, NPL3013964, NPL3013965, NPL3013966, NPL3013967, NPL3013969, NPL3013970, NPL3013971, NPL3013979, NPL3014238, NPL3014289, NPL3014291, NPL3014296, NPL3014297, NPL3014298, NPL3014299, NPL3014304, NPL3014307, NPL3014309, NPL3014311, NPL3014315, NPL3014317, NPL3014319, NPL3014321, NPL3014322, NPL3014323, NPL3014324, NPL3014326, NPL3014328, NPL3014330, NPL3014334, NPL3014335, NPL3014340, NPL3014343, NPL3014344, NPL3014457, NPL3014459, NPL3014462, NPL3014466, NPL3014467, NPL3014468, NPL3014521, NPL3014522, NPL3014526, NPL3014543, NPL3014544, NPL3014546, NPL3014573, NPL3014574, NPL3014575, NPL3014580, NPL3014584, NPL3014586, NPL3014588, NPL3014629, NPL3014630, NPL3014638, NPL3014770, NPL3014771, NPL3014772, NPL3014773, NPL3014774, NPL3014775, NPL3014776, NPL3014777, NPL3014778, NPL3014782, NPL3014784, NPL3014785, NPL3014788, NPL3014789, NPL3014791, NPL3014792, NPL3014793, NPL3014794, NPL3014796, NPL3014798, NPL3014799, NPL3014800, NPL3014869, NPL3014886, NPL3014887, NPL3014888, NPL3014889, NPL3014891, NPL3014892, NPL3014895, NPL3014971, NPL3014974, NPL3014975, NPL3014976, NPL3014977, NPL3014978, NPL3014980, NPL3014984, NPL3014988, NPL3014989, NPL3014990, NPL3014991, NPL3014997, NPL3015001, NPL3015005, NPL3015006, NPL3015007, NPL3015008, NPL3015010, NPL3015011, NPL3015014, NPL3015018, NPL3015071, NPL3015072, NPL3015074, NPL3015075, NPL3015076, NPL3015077, NPL3015078, NPL3015079, NPL3015081, NPL3015082, NPL3015083, NPL3015085, NPL3015086, NPL3015087, NPL3015088, NPL3015090, NPL3015091, NPL3015094, NPL3015095, NPL3015096, NPL3015097, NPL3015098, NPL3015099, NPL3015100, NPL3015101, NPL3015105, NPL3015107, NPL3015112, NPL3015113, NPL3015114, NPL3015116, NPL3015127, NPL3015128, NPL3015139, NPL3015140, NPL3015141, NPL3015142, NPL3015143, NPL3015144, NPL3015145, NPL3015146, NPL3015147, NPL3015148, NPL3015149, NPL3015150, NPL3015151, NPL3015152, NPL3015153, NPL3015154, NPL3015155, NPL3015156, NPL3015158, NPL3015160, NPL3015161, NPL3015162, NPL3015165, NPL3015166, NPL3015167, NPL3015168, NPL3015169, NPL3015170, NPL3015171, NPL3015178, NPL3015179, NPL3015184, NPL3015186, NPL3015187, NPL3015188, NPL3015189, NPL3015190, NPL3015261, NPL3015264, NPL3015266, NPL3015267, NPL3015270, NPL3015274, NPL3015275, NPL3015277, NPL3015278, NPL3015279, NPL3015310, NPL3015384, NPL3015385, NPL3015386, NPL3015387, NPL3015388, NPL3015389, NPL3015390, NPL3015391, NPL3015392, NPL3015393, NPL3015394, NPL3015396, NPL3015398, NPL3015399, NPL3015401, NPL3015402, NPL3015403, NPL3015404, NPL3015405, NPL3015406, NPL3015407, NPL3015512, NPL3015559, NPL3015561, NPL3015565; \n\n(2) GTIN 00763000094027, Serial Numbers:  NPL3000006, NPL3000007, NPL3000009, NPL3000010, NPL3000011, NPL3000012, NPL3000014, NPL3000016, NPL3000017, NPL3000018, NPL3000021, NPL3000023, NPL3000025, NPL3000026, NPL3000029, NPL3000030, NPL3000031, NPL3000033, NPL3000035, NPL3000040, NPL3000043, NPL3000046, NPL3000047, NPL3000048, NPL3000049, NPL3000050, NPL3000055, NPL3000057, NPL3000059, NPL3000062, NPL3000063, NPL3000068, NPL3000071, NPL3000074, NPL3000076, NPL3000077, NPL3000079, NPL3000083, NPL3000093, NPL3000094, NPL3000097, NPL3000103, NPL3000105, NPL3000109, NPL3000111, NPL3000125, NPL3000128, NPL3000129, NPL3000134, NPL3000135, NPL3000143, NPL3000184, NPL3000186, NPL3000188, NPL3000189, NPL3000196, NPL3000197, NPL3000198, NPL3000199, NPL3000202, NPL3000203, NPL3000205, NPL3000206, NPL3000208, NPL3000209, NPL3000210, NPL3000212, NPL3000213, NPL3000214, NPL3000215, NPL3000216, NPL3000219, NPL3000220, NPL3000221, NPL3000222, NPL3000223, NPL3000224, NPL3000226, NPL3000227, NPL3000229, NPL3000230, NPL3000232, NPL3000233, NPL3000236, NPL3000237, NPL3000240, NPL3000241, NPL3000242, NPL3000246, NPL3000248, NPL3000249, NPL3000250, NPL3000252, NPL3000253, NPL3000255, NPL3000258, NPL3000259, NPL3000260, NPL3000261, NPL3000262, NPL3000264, NPL3000265, NPL3002259, NPL3002261, NPL3002262, NPL3002263, NPL3002266, NPL3002270.NPL3002274, NPL3002282, NPL3002284, NPL3002288, NPL3002291, NPL3002293, NPL3002294, NPL3002296, NPL3002300, NPL3002304, NPL3002655, NPL3002659, NPL3002660, NPL3002661, NPL3002662, NPL3002663, NPL3002664, NPL3002665, NPL3002666, NPL3002667, NPL3002668, NPL3002669, NPL3002670, NPL3002671, NPL3002672, NPL3002673, NPL3002674, NPL3002675, NPL3002676, NPL3002677, NPL3002679, NPL3002680, NPL3002681, NPL3002682, NPL3002684, NPL3002685, NPL3002686, NPL3002687, NPL3002688, NPL3002691, NPL3002692, NPL3002693, NPL3002694, NPL3002696, NPL3002697, NPL3002699, NPL3002700, NPL3002701, NPL3002702, NPL3002703, NPL3002704, NPL3002705, NPL3002706, NPL3002707, NPL3002709, NPL3002710, NPL3002711, NPL3002712, NPL3002713, NPL3002715, NPL3002716, NPL3002717, NPL3002718, NPL3002719, NPL3002720, NPL3002721, NPL3002722, NPL3002723, NPL3002724, NPL3002725, NPL3002726, NPL3002727, NPL3002728, NPL3002729, NPL3002730, NPL3002731, NPL3002732, NPL3002733, NPL3002734, NPL3002736, NPL3002737, NPL3002738, NPL3002739, NPL3002741, NPL3002742, NPL3002743, NPL3002745, NPL3002746, NPL3002747, NPL3002749, NPL3002750, NPL3002751, NPL3002752, NPL3002753, NPL3002754, NPL3002755, NPL3002756, NPL3002757, NPL3002872, NPL3002873, NPL3002874, NPL3002875, NPL3002877, NPL3002878, NPL3002879, NPL3002880, NPL3002881, NPL3002882, NPL3002883, NPL3002884, NPL3002885, NPL3002887, NPL3002888, NPL3002891, NPL3002892, NPL3002894, NPL3002898, NPL3002899, NPL3002900, NPL3002901, NPL3002902, NPL3002904, NPL3002906, NPL3002907, NPL3002908, NPL3002909, NPL3002910, NPL3002911, NPL3002913, NPL3002914, NPL3002915, NPL3002918, NPL3002919, NPL3002922, NPL3002924, NPL3002925, NPL3002927, NPL3002928, NPL3002929, NPL3002935, NPL3002940, NPL3002941, NPL3002942, NPL3002949, NPL3002950, NPL3002954, NPL3002955, NPL3002960, NPL3002961, NPL3003298, NPL3003304, NPL3003308, NPL3003310, NPL3003311, NPL3003313, NPL3003315, NPL3003317, NPL3003319, NPL3003324, NPL3003331, NPL3003332, NPL3003334, NPL3003337, NPL3003338, NPL3003339, NPL3003342, NPL3003345, NPL3003346, NPL3003347, NPL3003353, NPL3003354, NPL3003355, NPL3003356, NPL3003357, NPL3003359, NPL3003694, NPL3003696, NPL3003704, NPL3003705, NPL3003708, NPL3003711, NPL3003712, NPL3003713, NPL3003715, NPL3003717, NPL3003718, NPL3003720, NPL3003723, NPL3003729, NPL3003730, NPL3003731, NPL3003735, NPL3003736, NPL3003737, NPL3003741, NPL3003742, NPL3003984, NPL3003987, NPL3003988, NPL3003989, NPL3003990, NPL3007387, NPL3007388, NPL3007390, NPL3007391, NPL3010714, NPL3010717, NPL3010721, NPL3010722, NPL3010724, NPL3010725, NPL3010726, NPL3010732, NPL3010735, NPL3010736, NPL3010739, NPL3010748, NPL3010750, NPL3010752, NPL3010753, NPL3010756, NPL3010758, NPL3010759, NPL3010760, NPL3010761, NPL3010764, NPL3010769, NPL3010770, NPL3010771, NPL3010772, NPL3010774, NPL3010779, NPL3011292, NPL3011296, NPL3011298, NPL3011304, NPL3011309, NPL3011310, NPL3011311, NPL3011313, NPL3011314, NPL3011316, NPL3011317, NPL3011319, NPL3011320, NPL3011323, NPL3011328, NPL3011329, NPL3011330, NPL3011331, NPL3011332, NPL3011333, NPL3011336, NPL3011338, NPL3011343, NPL3011344, NPL3011345, NPL3011346, NPL3011347, NPL3011349, NPL3011350, NPL3011352, NPL3011353, NPL3011357, NPL3011358, NPL3011362, NPL3011363, NPL3011365, NPL3014039, NPL3014040, NPL3014042, NPL3014043, NPL3014045, NPL3014052, NPL3014053, NPL3014054, NPL3014055, NPL3014056, NPL3014057, NPL3014058, NPL3014059, NPL3014061, NPL3014064, NPL3014065, NPL3014066, NPL3014070, NPL3014071, NPL3014072, NPL3014073, NPL3014074, NPL3014076, NPL3014079, NPL3014083, NPL3014085, NPL3014086, NPL3014087, NPL3014095, NPL3014096, NPL3014097, NPL3014098, NPL3014346, NPL3014347, NPL3014454, NPL3014455, NPL3014456, NPL3014501, NPL3014505, NPL3014506, NPL3014508, NPL3014529, NPL3014530, NPL3014537, NPL3014541, NPL3014542, NPL3017239, NPL3017240, NPL3017247, NPL3017252, NPL3017254, NPL3017257, NPL3017260, NPL3017265, NPL3017266, NPL3017267, NPL3017268, NPL3017269, NPL3017270, NPL3017272, NPL3017275, NPL3017284, NPL3017285, NPL3017286, NPL3017287, NPL3017289, NPL3017290, NPL3017932, NPL3017933, NPL3017991, NPL3017992, NPL3017993, NPL3018577, NPL3018578, NPL3018655, NPL3018709, NPL3018711, NPL3018715, NPL3018718, NPL3018719, NPL3018885, NPL3019477, NPL3019531, NPL3019532, NPL3019533, NPL3019534, NPL3019535, NPL3019537, NPL3019539, NPL3019541, NPL3019607, NPL3019660, NPL3019663, NPL3019664, NPL3019666, NPL3019668, NPL3019672, NPL3019677, NPL3019678, NPL3019769, NPL3020210, NPL3021261, NPL3021268, NPL3021270, NPL3021465, NPL3021466, NPL3021467, NPL3021468, NPL3021471, NPL3021479, NPL3021480, NPL3021481, NPL3021485, NPL3021487, NPL3021488, NPL3021490, NPL3021491, NPL3022395, NPL3022399; \n\n(3) GTIN 00763000421007, Serial Numbers:  NPL3016404, NPL3016409, NPL3016410, NPL3016411, NPL3016418, NPL3016420, NPL3016421, NPL3016422, NPL3016423, NPL3016424, NPL3016428, NPL3016429, NPL3016431, NPL3016435, NPL3016436, NPL3016437, NPL3016438, NPL3016439, NPL3016440, NPL3016441, NPL3016443, NPL3016444, NPL3016446, NPL3016447, NPL3016448, NPL3016449, NPL3016457, NPL3016458, NPL3016460, NPL3016461, NPL3016464, NPL3016465, NPL3016469, NPL3016470, NPL3016475, NPL3016477, NPL3016478, NPL3016479, NPL3016481, NPL3016482, NPL3016484, NPL3016488, NPL3016492, NPL3016493, NPL3016495, NPL3016498, NPL3016499, NPL3016501, NPL3016502, NPL3016504, NPL3016507, NPL3016508, NPL3016513, NPL3016515, NPL3016516, NPL3016517, NPL3016518, NPL3016519, NPL3016522, NPL3016524, NPL3016525, NPL3016526, NPL3016527, NPL3016528, NPL3016529, NPL3016530, NPL3016531, NPL3016535, NPL3016536, NPL3016537, NPL3016539, NPL3016582, NPL3016584, NPL3016592, NPL3016594, NPL3016600, NPL3016601, NPL3016602, NPL3016603, NPL3016606, NPL3016607, NPL3016608, NPL3016613, NPL3016617, NPL3016621, NPL3016699, NPL3016702, NPL3016712, NPL3016713, NPL3016714, NPL3016715, NPL3016716, NPL3016737, NPL3016738, NPL3016743, NPL3016744, NPL3016745, NPL3016746, NPL3016747, NPL3016748, NPL3016749, NPL3016750, NPL3016757, NPL3016761, NPL3016762, NPL3016763, NPL3016765, NPL3016766, NPL3016775, NPL3016776, NPL3016778, NPL3016780, NPL3016784, NPL3016785, NPL3016787, NPL3016788, NPL3016790, NPL3016792, NPL3016793, NPL3016795, NPL3016796, NPL3016797, NPL3016798, NPL3016800, NPL3016801, NPL3016802, NPL3016933, NPL3016995, NPL3017043, NPL3017053, NPL3017201, NPL3017214, NPL3017221, NPL3017224, NPL3017231, NPL3017232, NPL3017303, NPL3017307, NPL3017308, NPL3017309, NPL3017310, NPL3017312, NPL3017313, NPL3017315, NPL3017316, NPL3017320, NPL3017324, NPL3017325, NPL3017333, NPL3017335, NPL3017344, NPL3017346, NPL3017347, NPL3017348, NPL3017349, NPL3017350, NPL3017367, NPL3017384, NPL3017392, NPL3017393, NPL3017402, NPL3017412, NPL3017414, NPL3017424, NPL3017425, NPL3017426, NPL3017427, NPL3017430, NPL3017431, NPL3017444, NPL3017506, NPL3017610, NPL3017611, NPL3017613, NPL3017614, NPL3017620, NPL3017625, NPL3017628, NPL3017629, NPL3017630, NPL3017654, NPL3017660, NPL3017662, NPL3017663, NPL3017664, NPL3017685, NPL3017686, NPL3017692, NPL3017693, NPL3017694, NPL3017695, NPL3017704, NPL3017705, NPL3017706, NPL3017707, NPL3017717, NPL3017726, NPL3017727, NPL3017739, NPL3017744, NPL3017748, NPL3017761, NPL3017767, NPL3017768, NPL3017769, NPL3017770, NPL3017771, NPL3017774, NPL3017779, NPL3017890, NPL3017891, NPL3017895, NPL3017897, NPL3017898, NPL3017900, NPL3017922, NPL3017924, NPL3017926, NPL3017927, NPL3017928, NPL3017942, NPL3017943, NPL3017944, NPL3017995, NPL3017996, NPL3017998, NPL3018002, NPL3018004, NPL3018005, NPL3018006, NPL3018007, NPL3018008, NPL3018009, NPL3018010, NPL3018011, NPL3018012, NPL3018013, NPL3018015, NPL3018016, NPL3018017, NPL3018018, NPL3018019, NPL3018020, NPL3018021, NPL3018022, NPL3018023, NPL3018024, NPL3018025, NPL3018026, NPL3018028, NPL3018045, NPL3018046, NPL3018048, NPL3018049, NPL3018050, NPL3018052, NPL3018053, NPL3018054, NPL3018058, NPL3018059, NPL3018062, NPL3018527, NPL3018528, NPL3018532, NPL3018534, NPL3018537, NPL3018538, NPL3018562, NPL3018563, NPL3018564, NPL3018579, NPL3018602, NPL3018607, NPL3018608, NPL3018609, NPL3018613.NPL3018614, NPL3018617, NPL3018619, NPL3018623, NPL3018626, NPL3018627, NPL3018657, NPL3018659, NPL3018677, NPL3018720, NPL3018721, NPL3018726, NPL3018727, NPL3018728, NPL3018729, NPL3018735, NPL3018736, NPL3018737, NPL3018743, NPL3018747, NPL3018748, NPL3018750, NPL3018751, NPL3018756, NPL3018757, NPL3018759, NPL3018760, NPL3018761, NPL3018762, NPL3018763, NPL3018765, NPL3018769, NPL3018775, NPL3018776, NPL3018777, NPL3018778, NPL3018780, NPL3018784, NPL3018792, NPL3018808, NPL3018809, NPL3018829, NPL3018830, NPL3018831, NPL3018832, NPL3018833, NPL3018834, NPL3018835, NPL3018836, NPL3018839, NPL3018840, NPL3018842, NPL3018843, NPL3018844, NPL3018892, NPL3018894, NPL3019405, NPL3019406, NPL3019407, NPL3019408, NPL3019409, NPL3019410, NPL3019483, NPL3019484, NPL3019485, NPL3019486, NPL3019487, NPL3019488, NPL3019489, NPL3019490, NPL3019491, NPL3019492, NPL3019493, NPL3019494, NPL3019496, NPL3019497, NPL3019501, NPL3019502, NPL3019503, NPL3019543, NPL3019547, NPL3019548, NPL3019552, NPL3019558, NPL3019561, NPL3019580, NPL3019581, NPL3019582, NPL3019583, NPL3019585, NPL3019586, NPL3019610, NPL3019613, NPL3019614, NPL3019629, NPL3019630, NPL3019632, NPL3019698, NPL3019699, NPL3019701, NPL3019703, NPL3019704, NPL3019705, NPL3019706, NPL3019731, NPL3019732, NPL3019733, NPL3019734, NPL3019735, NPL3019736, NPL3019737, NPL3019742, NPL3019765, NPL3019895, NPL3019898, NPL3019974, NPL3020003, NPL3020074, NPL3020103, NPL3020104, NPL3020105, NPL3020106, NPL3020107, NPL3020108, NPL3020112, NPL3020113, NPL3020121, NPL3020127, NPL3020129, NPL3020134, NPL3020136, NPL3020137, NPL3020138, NPL3020139, NPL3020141, NPL3020151, NPL3020155, NPL3020157, NPL3020164, NPL3020165, NPL3020170, NPL3020177, NPL3020222, NPL3020252, NPL3020925, NPL3020931, NPL3020944, NPL3020954, NPL3020958, NPL3020959, NPL3020960, NPL3020964, NPL3020969, NPL3020970, NPL3020980, NPL3020982, NPL3020983, NPL3020986, NPL3020987, NPL3020988, NPL3020990, NPL3020991, NPL3021002, NPL3021003, NPL3021004, NPL3021015, NPL3021025, NPL3021030, NPL3021031, NPL3021044, NPL3021045, NPL3021047, NPL3021048, NPL3021050, NPL3021052, NPL3021155, NPL3021157, NPL3021160, NPL3021162, NPL3021163, NPL3021168, NPL3021169, NPL3021170, NPL3021227, NPL3021229, NPL3021334, NPL3021337, NPL3021338, NPL3021352, NPL3021354, NPL3021367, NPL3021425, NPL3021492, NPL3021493, NPL3021495, NPL3021496, NPL3021497, NPL3021499, NPL3021501, NPL3021559, NPL3021560, NPL3021564, NPL3021565, NPL3021567, NPL3021568, NPL3021569, NPL3022374, NPL3022380, NPL3022441, NPL3022461, NPL3022489, NPL3022490, NPL3022515, NPL3022520; \n\n5. Model Number CT900E, (1) GTIN  00763000447090, Serial Numbers:  NPL3020725, NPL3022005, NPL3022030, NPL3022054, NPL3022062, NPL3022070, NPL3022114, NPL3022119, NPL3022131, NPL3022145, NPL3022174, NPL3022231, NPL3022236, NPL3022238, NPL3022245, NPL3022250, NPL3022257, NPL3022263, NPL3022278, NPL3022279, NPL3022280, NPL3022281, NPL3022298, NPL3022299, NPL3022300, NPL3022301, NPL3022304, NPL3022305, NPL3022306, NPL3022307, NPL3022308, NPL3022309, NPL3022310, NPL3022311, NPL3022313, NPL3022314, NPL3022324, NPL3022325, NPL3022328, NPL3022329, NPL3022330, NPL3022331, NPL3022332, NPL3022333, NPL3022334, NPL3022335, NPL3022336, NPL3022337, NPL3022338, NPL3022339, NPL3022340, NPL3022341, NPL3022342, NPL3022343, NPL3022344, NPL3022347, NPL3022349, NPL3022350, NPL3022351, NPL3022352, NPL3022357, NPL3022364, NPL3022365, NPL3022712, NPL3022716, NPL3022717, NPL3022720, NPL3022721, NPL3022724, NPL3022725, NPL3022730, NPL3022731, NPL3022732, NPL3022733, NPL3022735, NPL3022740, NPL3022741, NPL3022742, NPL3022743, NPL3022744, NPL3022745, NPL3022746, NPL3022747, NPL3022748, NPL3022749, NPL3022750, NPL3022751, NPL3022753, NPL3022754, NPL3022755, NPL3022756, NPL3022757, NPL3022761, NPL3022762, NPL3022769, NPL3022772, NPL3022773, NPL3022774, NPL3022776, NPL3022781, NPL3022784, NPL3022785, NPL3022786, NPL3022787, NPL3022788, NPL3022791, NPL3022792, NPL3022812, NPL3022858, NPL3022874, NPL3022900, NPL3022912, NPL3022949, NPL3023030, NPL3023059, NPL3023131, NPL3023148, NPL3023150, NPL3023197, NPL3023277, NPL3023279, NPL3023292, NPL3023300, NPL3023319, NPL3023323, NPL3023367, NPL3023425, NPL3023427, NPL3023429, NPL3023431, NPL3023433, NPL3023434, NPL3023437, NPL3023438, NPL3023439, NPL3023440, NPL3023442, NPL3023443, NPL3023446, NPL3023448, NPL3023450, NPL3023494, NPL3023530, NPL3023554, NPL3023583, NPL3023605, NPL3023615, NPL3023640, NPL3023659, NPL3023692, NPL3023701, NPL3023702, NPL3023704, NPL3023708, NPL3023710, NPL3023716, NPL3023722, NPL3023729, NPL3023734, NPL3023738, NPL3023740, NPL3023759, NPL3023769, NPL3023775, NPL3023776, NPL3023780, NPL3023784, NPL3023785, NPL3023792, NPL3023798, NPL3023805, NPL3023830, NPL3023846, NPL3023854, NPL3023858, NPL3023869, NPL3023873, NPL3023874, NPL3023891, NPL3023893, NPL3023932, NPL3023936, NPL3023938, NPL3023941, NPL3023946, NPL3023948, NPL3023952, NPL3023955, NPL3023957, NPL3023959, NPL3023966, NPL3023973, NPL3023974, NPL3023975, NPL3024015, NPL3024018, NPL3024022, NPL3024023, NPL3024026, NPL3024028, NPL3024033, NPL3024080, NPL3024157, NPL3024188, NPL3024189, NPL3024201, NPL3024206, NPL3024211, NPL3024221, NPL3024243, NPL3024256, NPL3024258, NPL3024286, NPL3024299, NPL3024303, NPL3024311, NPL3024312, NPL3024313, NPL3024314, NPL3024317, NPL3024319, NPL3024320, NPL3024322, NPL3024323, NPL3024327, NPL3024330, NPL3024357, NPL3024390, NPL3024399, NPL3024405, NPL3024408, NPL3024422, NPL3024426, NPL3024427, NPL3024431, NPL3024436, NPL3024445, NPL3024458, NPL3024460, NPL3024472, NPL3024478, NPL3024480, NPL3024482, NPL3024484, NPL3024485, NPL3024486, NPL3024487, NPL3024492, NPL3024495, NPL3024496, NPL3024504, NPL3024511, NPL3024527, NPL3024528, NPL3024531, NPL3024533, NPL3024575, NPL3024587, NPL3024602, NPL3024633, NPL3024637, NPL3024646, NPL3024699, NPL3024709, NPL3024712, NPL3024713, NPL3024719, NPL3024720, NPL3024722, NPL3024731, NPL3024735, NPL3024747, NPL3024748, NPL3024750, NPL3024751, NPL3024754, NPL3024755, NPL3024759, NPL3024760, NPL3024764, NPL3024765, NPL3024766, NPL3024771, NPL3024773, NPL3024774, NPL3024775, NPL3024777, NPL3024778, NPL3024780, NPL3024784, NPL3024786, NPL3024792, NPL3024794, NPL3024796, NPL3024797, NPL3024808, NPL3024832, NPL3024836, NPL3024837, NPL3024841, NPL3024843, NPL3024858, NPL3024861, NPL3024871, NPL3024872, NPL3024884, NPL3024888, NPL3024889, NPL3024890, NPL3024897, NPL3024900, NPL3024902, NPL3024903, NPL3024909, NPL3024912, NPL3024917, NPL3024919, NPL3024924, NPL3024929, NPL3024933, NPL3024934, NPL3024935, NPL3024936, NPL3024948, NPL3024965, NPL3024973, NPL3024978, NPL3024989, NPL3024998, NPL3025007, NPL3025012, NPL3025013, NPL3025017, NPL3025021, NPL3025023, NPL3025025, NPL3025027, NPL3025031, NPL3025033, NPL3025042, NPL3025044, NPL3025047, NPL3025057, NPL3025058, NPL3025067, NPL3025076, NPL3025082, NPL3025097, NPL3025155, NPL3025179, NPL3025184, NPL3025197, NPL3025205, NPL3025222, NPL3025224, NPL3025226, NPL3025241, NPL3025242, NPL3025253, NPL3025257, NPL3025265, NPL3025268, NPL3025270, NPL3025271, NPL3025275, NPL3025279, NPL3025281, NPL3025284, NPL3025285, NPL3025288, NPL3025292, NPL3025293, NPL3025301, NPL3025304, NPL3025313, NPL3025316, NPL3025326, NPL3025327, NPL3025329, NPL3025331, NPL3025335, NPL3025339, NPL3025349, NPL3025372, NPL3025384, NPL3025396, NPL3025482, NPL3025509, NPL3025533, NPL3025640, NPL3025641, NPL3025649, NPL3025650, NPL3025715, NPL3025743, NPL3025760, NPL3025890, NPL3025902, NPL3025908, NPL3025917, NPL3025920, NPL3025945, NPL3025947, NPL3025949, NPL3025958, NPL3025965, NPL3026034, NPL3026053, NPL3026077, NPL3026097, NPL3026159, NPL3026177, NPL3026188, NPL3026190, NPL3026191, NPL3026196, NPL3026201, NPL3026202, NPL3026203, NPL3026204, NPL3026205, NPL3026207, NPL3026208, NPL3026210, NPL3026212, NPL3026215, NPL3026218, NPL3026220, NPL3026223, NPL3026224, NPL3026233, NPL3026245, NPL3026250, NPL3026258, NPL3026260, NPL3026262, NPL3026294, NPL3026324, NPL3026341, NPL3026351, NPL3026356, NPL3026358, NPL3026360, NPL3026362, NPL3026367, NPL3026382, NPL3026383, NPL3026390, NPL3026391, NPL3026393, NPL3026396, NPL3026397, NPL3026402, NPL3026407, NPL3026408, NPL3026414, NPL3026428, NPL3026429, NPL3026435, NPL3026438, NPL3026441, NPL3026442, NPL3026443, NPL3026447, NPL3026449, NPL3026650, NPL3026936, NPL3027060, NPL3027340, NPL3027473, NPL3027482, NPL3027519, NPL3027595, NPL3027607, NPL3027623, NPL3027627, NPL3027697, NPL3027720, NPL3027747, NPL3027748, NPL3027803, NPL3027825, NPL3027838, NPL3027845, NPL3027858, NPL3027859, NPL3030796, NPL3030799, NPL3030820, NPL3030824, NPL3030828, NPL3030836, NPL3030848, NPL3030853, NPL3030861, NPL3030865, NPL3030869, NPL3030872, NPL3030913, NPL3030916, NPL3030919, NPL3030930, NPL3030932, NPL3030933, NPL3030934, NPL3030935, NPL3030936, NPL3030941, NPL3030942, NPL3031031, NPL3031073, NPL3031075, NPL3031079, NPL3031080, NPL3031083, NPL3031084, NPL3031085, NPL3031110, NPL3031128, NPL3031144, NPL3031167, NPL3031175, NPL3031197, NPL3031205, NPL3031221, NPL3031239, NPL3031245, NPL3031247, NPL3031248, NPL3031258, NPL3031261, NPL3031266, NPL3031269, NPL3031274, NPL3031280, NPL3031284, NPL3031289, NPL3031290, NPL3031310, NPL3031324, NPL3031326, NPL3031337, NPL3031379, NPL3031404, NPL3031435, NPL3031471, NPL3031475, NPL3031481, NPL3031492, NPL3031493, NPL3031494, NPL3031495, NPL3031499, NPL3031501, NPL3031518, NPL3031525, NPL3031529, NPL3031540, NPL3031546, NPL3031548, NPL3031552, NPL3031563, NPL3031569, NPL3031570, NPL3031591, NPL3031593, NPL3031594, NPL3031595, NPL3031596, NPL3031597, NPL3031600, NPL3031603, NPL3031604, NPL3031606, NPL3031622, NPL3031627, NPL3031628, NPL3031650, NPL3031655, NPL3031660, NPL3031666, NPL3031674, NPL3031680, NPL3031690, NPL3031709, NPL3031719, NPL3031727, NPL3031734, NPL3031738, NPL3031761, NPL3031763, NPL3031793, NPL3031796, NPL3031798, NPL3031840, NPL3031848, NPL3031874, NPL3031883, NPL3031884, NPL3031892, NPL3031924, NPL3031951, NPL3031966, NPL3031968, NPL3031973, NPL3031975, NPL3031986, NPL3031990, NPL3031991, NPL3031995, NPL3031999, NPL3032036, NPL3032043, NPL3032049, NPL3032066, NPL3032072, NPL3032074, NPL3032086, NPL3032088, NPL3032089, NPL3032091, NPL3032095, NPL3032098, NPL3032108, NPL3032110, NPL3032120, NPL3032145, NPL3032151, NPL3032163, NPL3032193, NPL3032195, NPL3032200, NPL3032202, NPL3032207, NPL3032211, NPL3032214, NPL3032219, NPL3032225, NPL3032235, NPL3032236, NPL3032237, NPL3032238, NPL3032239, NPL3032240, NPL3032241, NPL3032242, NPL3032243, NPL3032246, NPL3032250, NPL3032251, NPL3032252, NPL3032253, NPL3032261, NPL3032262, NPL3032265, NPL3032281, NPL3032299, NPL3032310, NPL3032311, NPL3032320, NPL3032329, NPL3032330, NPL3032331, NPL3032332, NPL3032334, NPL3032342, NPL3032344, NPL3032355, NPL3032360, NPL3032378, NPL3032450, NPL3032457, NPL3032483, NPL3032488, NPL3032491, NPL3032497, NPL3032502, NPL3032513, NPL3032516, NPL3032568, NPL3032587, NPL3032600, NPL3032610, NPL3032611, NPL3032613, NPL3032614, NPL3032616, NPL3032622, NPL3032628, NPL3032642, NPL3032648, NPL3032650, NPL3032665, NPL3032678, NPL3032683, NPL3032687, NPL3032696, NPL3032707, NPL3032729, NPL3032747, NPL3032763, NPL3032771, NPL3032776, NPL3032781, NPL3032790, NPL3032807, NPL3032816, NPL3032834, NPL3032839, NPL3032883, NPL3032891, NPL3032894, NPL3032907, NPL3032917, NPL3032928, NPL3032939, NPL3032945, NPL3032954, NPL3032978, NPL3032980, NPL3032983, NPL3032991, NPL3032997, NPL3032998, NPL3033001, NPL3033004, NPL3033013, NPL3033017, NPL3033020, NPL3033036, NPL3033038, NPL3033039, NPL3033040, NPL3033042, NPL3033047, NPL3033049, NPL3033057, NPL3033062, NPL3033072, NPL3033077, NPL3033090, NPL3033091, NPL3033094, NPL3033107, NPL3033113, NPL3033134, NPL3033139, NPL3033143, NPL3033187, NPL3033196, NPL3033216, NPL3033218, NPL3033221, NPL3033229, NPL3033248, NPL3033249, NPL3033268, NPL3033274, NPL3033276, NPL3033277, NPL3033282, NPL3033285, NPL3033291, NPL3033339, NPL3033342, NPL3033365, NPL3033371, NPL3033379, NPL3033380, NPL3033400, NPL3033401, NPL3033402, NPL3033403, NPL3033499, NPL3033521, NPL3033560, NPL3033606, NPL3033620, NPL3033623, NPL3033624, NPL3033650, NPL3033680, NPL3033681, NPL3033687, NPL3033729, NPL3033742, NPL3033751, NPL3033753, NPL3033755, NPL3033765, NPL3033773, NPL3033791, NPL3033840, NPL3033846, NPL3033848, NPL3033862, NPL3033869, NPL3033871, NPL3033880, NPL3033883, NPL3033895, NPL3033900, NPL3033901, NPL3033902, NPL3033905, NPL3033910, NPL3033911, NPL3033914, NPL3033921, NPL3033925, NPL3033949, NPL3033952, NPL3033955, NPL3033958, NPL3033960, NPL3033962, NPL3033968, NPL3033975, NPL3033977, NPL3033979, NPL3033980, NPL3033989, NPL3033994, NPL3033995, NPL3033996, NPL3034000, NPL3034004, NPL3034007, NPL3034008, NPL3034009, NPL3034017, NPL3034018, NPL3034020, NPL3034021, NPL3034022, NPL3034046, NPL3034047, NPL3034048, NPL3034050, NPL3034058, NPL3034066, NPL3034068, NPL3034074, NPL3034075, NPL3034081, NPL3034084, NPL3034090, NPL3034093, NPL3034095, NPL3034108, NPL3034109, NPL3034112, NPL3034126, NPL3034134, NPL3034135, NPL3034142, NPL3034152, NPL3034165, NPL3034167, NPL3034168, NPL3034171, NPL3034174, NPL3034178, NPL3034179, NPL3034183, NPL3034186, NPL3034188, NPL3034191, NPL3034200, NPL3034202, NPL3034205, NPL3034206, NPL3034208, NPL3034210, NPL3034216, NPL3034226, NPL3034227, NPL3034229, NPL3034230, NPL3034231, NPL3034232, NPL3034239, NPL3034243, NPL3034244, NPL3034247, NPL3034248, NPL3034250, NPL3034252, NPL3034257, NPL3034260, NPL3034264, NPL3034269, NPL3034270, NPL3034271, NPL3034272, NPL3034273, NPL3034275, NPL3034277, NPL3034280, NPL3034281, NPL3034283, NPL3034284, NPL3034285, NPL3034287, NPL3034288, NPL3034290, NPL3034294, NPL3034301, NPL3034303, NPL3034306, NPL3034308, NPL3034310, NPL3034323, NPL3034324, NPL3034325, NPL3034326, NPL3034331, NPL3034332, NPL3034334, NPL3034338, NPL3034342, NPL3034349, NPL3034351, NPL3034360, NPL3034365, NPL3034368, NPL3034372, NPL3034374, NPL3034376, NPL3034383, NPL3034396, NPL3034397, NPL3034404, NPL3034407, NPL3034427, NPL3034431, NPL3034448, NPL3034452, NPL3034469, NPL3034477, NPL3034484, NPL3034512, NPL3034514, NPL3034515, NPL3034516, NPL3034517, NPL3034520, NPL3034521, NPL3034523, NPL3034526, NPL3034528, NPL3034535, NPL3034539, NPL3034540, NPL3034549, NPL3034550, NPL3034556, NPL3034557, NPL3034560, NPL3034566, NPL3034571, NPL3034582, NPL3034597, NPL3034605, NPL3034613, NPL3034617, NPL3034621, NPL3034648, NPL3034656, NPL3034673, NPL3034717, NPL3034740, NPL3034744, NPL3034752, NPL3034761, NPL3034774, NPL3034777, NPL3036810, NPL3036833, NPL3036840, NPL3036854, NPL3036887, NPL3036905, NPL3036926, NPL3036942, NPL3036944, NPL3036945, NPL3036946, NPL3036947, NPL3036948, NPL3036954, NPL3036957, NPL3036959, NPL3036960, NPL3036975, NPL3036985, NPL3036992, NPL3037094, NPL3037107, NPL3037112, NPL3037117, NPL3037131, NPL3037196, NPL3037216, NPL3037218, NPL3037223, NPL3037225, NPL3037240, NPL3037241, NPL3037245, NPL3037246, NPL3037251, NPL3037252, NPL3037254, NPL3037255, NPL3037262, NPL3037278, NPL3037307, NPL3037310, NPL3037311, NPL3037323, NPL3037325, NPL3037326, NPL3037327, NPL3037335, NPL3037379, NPL3037381, NPL3037382, NPL3037384, NPL3037391, NPL3037397, NPL3037398, NPL3037399, NPL3037400, NPL3037403, NPL3037405, NPL3037406, NPL3037412, NPL3037419, NPL3037420, NPL3037429, NPL3037430, NPL3037431, NPL3037440, NPL3037444, NPL3037461, NPL3037492, NPL3037494, NPL3037499, NPL3037509, NPL3037516, NPL3037526, NPL3037528, NPL3037529, NPL3037549, NPL3037559, NPL3037563, NPL3037567, NPL3037568, NPL3037569, NPL3037571, NPL3037604, NPL3037615, NPL3037621, NPL3037623, NPL3037624, NPL3037626, NPL3037662, NPL3037668, NPL3037697, NPL3037700, NPL3037707, NPL3037714, NPL3037715, NPL3037721, NPL3037729, NPL3037730, NPL3037740, NPL3037753, NPL3037756, NPL3037764, NPL3037779, NPL3037850, NPL3037880, NPL3037902, NPL3037966, NPL3037967, NPL3037996, NPL3038006, NPL3038010, NPL3038066, NPL3038078, NPL3038092, NPL3038112, NPL3038127, NPL3038177, NPL3038188, NPL3038218, NPL3038334, NPL3038335, NPL3038336, NPL3038345, NPL3038361, NPL3038372, NPL3038374, NPL3038375, NPL3038381, NPL3038385, NPL3038386, NPL3038412, NPL3038419, NPL3038422, NPL3038437, NPL3038440, NPL3038443, NPL3038446, NPL3038449, NPL3038450, NPL3038452, NPL3038456, NPL3038457, NPL3038460, NPL3038461, NPL3038462, NPL3038465, NPL3038471, NPL3038474, NPL3038478, NPL3038479, NPL3038480; \n\n(2) GTIN 00763000729141, Serial Numbers:  NPL3022833, NPL3022835, NPL3022836, NPL3022838, NPL3022839, NPL3022840, NPL3022841, NPL3022954, NPL3022973, NPL3022974, NPL3022975, NPL3023250, NPL3023254, NPL3023256, NPL3023257, NPL3023288, NPL3023459, NPL3023460, NPL3023461, NPL3023463, NPL3023464, NPL3023465, NPL3023466, NPL3023467, NPL3023471, NPL3023473, NPL3023474, NPL3024180, NPL3024181, NPL3024187, NPL3024192, NPL3024195, NPL3024224, NPL3024227, NPL3024229, NPL3024251, NPL3024290, NPL3024302, NPL3024344, NPL3025010, NPL3025032, NPL3025045, NPL3025053, NPL3025054, NPL3038210, NPL3038230, NPL3038351, NPL3038714, NPL3038718, NPL3038719, NPL3038736, NPL3038737, NPL3038738, NPL3038741, NPL3038742, NPL3038779, NPL3038780, NPL3038781, NPL3038867, NPL3039871, NPL3040070, NPL3040071, NPL3040073, NPL3040080, NPL3040083, NPL3040084, NPL3040099, NPL3040100, NPL3040101, NPL3040103, NPL3040104, NPL3040105, NPL3040106, NPL3040107, NPL3040109, NPL3040111, NPL3040112, NPL3040175, NPL3040176, NPL3040177, NPL3040178, NPL3040179, NPL3040242, NPL3040243, NPL3040245, NPL3040246, NPL3040247, NPL3040251, NPL3040252, NPL3040256, NPL3040257, NPL3040261, NPL3040262, NPL3040263, NPL3040266, NPL3040268, NPL3040271, NPL3040307, NPL3040523, NPL3040725, NPL3040728, NPL3040729, NPL3040730, NPL3040731, NPL3040732, NPL3040733, NPL3040734, NPL3040735, NPL3040736, NPL3040737, NPL3040738, NPL3040739, NPL3040741, NPL3040743, NPL3040744, NPL3040745, NPL3040746, NPL3040747, NPL3040748, NPL3040846, NPL3040855, NPL3040928, NPL3040932, NPL3040933, NPL3040935, NPL3040938, NPL3040939, NPL3044955, NPL3044969, NPL3044970, NPL3044971, NPL3044972, NPL3044973, NPL3045026, NPL3045027, NPL3045028, NPL3045029, NPL3045030, NPL3045031, NPL3045032; \n\n(3) GTIN 00763000737290, Serial Numbers:  NPL3038512, NPL3038518, NPL3038519, NPL3038520, NPL3038522, NPL3038526, NPL3038530, NPL3038531, NPL3038533, NPL3038534, NPL3038539, NPL3038543, NPL3038545, NPL3038549, NPL3038553, NPL3038556, NPL3038559, NPL3038561, NPL3038567, NPL3038575, NPL3038576, NPL3038578, NPL3038581, NPL3038585, NPL3038604, NPL3038657, NPL3038746, NPL3038747, NPL3038748, NPL3038750, NPL3038751, NPL3038752, NPL3038753, NPL3038754, NPL3038755, NPL3038756, NPL3038757, NPL3038758, NPL3038759, NPL3038760, NPL3038761, NPL3038762, NPL3038763, NPL3038764, NPL3038765, NPL3038766, NPL3038767, NPL3038768, NPL3038769, NPL3038770, NPL3038771, NPL3038775, NPL3038777, NPL3038778, NPL3038784, NPL3038786, NPL3038802, NPL3038805, NPL3038812, NPL3038827, NPL3038828, NPL3038829, NPL3038830, NPL3038831, NPL3038832, NPL3038843, NPL3038844, NPL3038846, NPL3038847, NPL3038849, NPL3038851, NPL3038854, NPL3038856, NPL3038865, NPL3038873, NPL3038875, NPL3038876, NPL3038877, NPL3038878, NPL3038879, NPL3038887, NPL3038888, NPL3038889, NPL3038891, NPL3038892, NPL3038895, NPL3038896, NPL3038898, NPL3038899, NPL3038900, NPL3038901, NPL3038902, NPL3038903, NPL3038905, NPL3038908, NPL3038909, NPL3038911, NPL3038916, NPL3038917, NPL3038918, NPL3038919, NPL3038920, NPL3038921, NPL3038922, NPL3038924, NPL3038928, NPL3038929, NPL3038946, NPL3038947, NPL3038948, NPL3038949, NPL3038951, NPL3038953, NPL3038955, NPL3038956, NPL3038957, NPL3038958, NPL3038959, NPL3038960, NPL3038965, NPL3038966, NPL3038970, NPL3038971, NPL3038972, NPL3038973, NPL3038974, NPL3038975, NPL3038976, NPL3038982, NPL3038984, NPL3038985, NPL3038987, NPL3038989, NPL3038990, NPL3038992, NPL3038994, NPL3038997, NPL3038999, NPL3039000, NPL3039002, NPL3039003, NPL3039005, NPL3039006, NPL3039007, NPL3039008, NPL3039018, NPL3039043, NPL3039061, NPL3039065, NPL3039066, NPL3039071, NPL3039072, NPL3039073, NPL3039074, NPL3039076, NPL3039084, NPL3039085, NPL3039086, NPL3039087, NPL3039088, NPL3039094, NPL3039097, NPL3039101, NPL3039109, NPL3039119, NPL3039120, NPL3039121, NPL3039127, NPL3039132, NPL3039134, NPL3039136, NPL3039137, NPL3039138, NPL3039149, NPL3039150, NPL3039151, NPL3039152, NPL3039156, NPL3039164, NPL3039168, NPL3039169, NPL3039174, NPL3039176, NPL3039177, NPL3039178, NPL3039179, NPL3039180, NPL3039182, NPL3039186, NPL3039194, NPL3039195, NPL3039196, NPL3039197, NPL3039202, NPL3039245, NPL3039246, NPL3039255, NPL3039267, NPL3039280, NPL3039334, NPL3039345, NPL3039346, NPL3039347, NPL3039363, NPL3039364, NPL3039374, NPL3039378, NPL3039388, NPL3039393, NPL3039395, NPL3039396, NPL3039404, NPL3039413, NPL3039426, NPL3039427, NPL3039428, NPL3039444, NPL3039446, NPL3039447, NPL3039448, NPL3039449, NPL3039450, NPL3039451, NPL3039452, NPL3039453, NPL3039454, NPL3039463, NPL3039464, NPL3039465, NPL3039466, NPL3039468, NPL3039472, NPL3039476, NPL3039477, NPL3039478, NPL3039510, NPL3039511, NPL3039512, NPL3039513, NPL3039514, NPL3039515, NPL3039516, NPL3039517, NPL3039518, NPL3039519, NPL3039530, NPL3039551, NPL3039555, NPL3039563, NPL3039564, NPL3039565, NPL3039566, NPL3039567, NPL3039568, NPL3039569, NPL3039570, NPL3039571, NPL3039572, NPL3039574, NPL3039575, NPL3039576, NPL3039577, NPL3039578, NPL3039580, NPL3039581, NPL3039582, NPL3039583, NPL3039584, NPL3039585, NPL3039587, NPL3039589, NPL3039590, NPL3039591, NPL3039592, NPL3039593, NPL3039594, NPL3039596, NPL3039597, NPL3039598, NPL3039599, NPL3039600, NPL3039601, NPL3039603, NPL3039604, NPL3039605, NPL3039607, NPL3039608, NPL3039609, NPL3039610, NPL3039611, NPL3039612, NPL3039613, NPL3039614, NPL3039615, NPL3039619, NPL3039620, NPL3039621, NPL3039622, NPL3039624, NPL3039625, NPL3039626, NPL3039627, NPL3039635, NPL3039636, NPL3039637, NPL3039638, NPL3039639, NPL3039640, NPL3039648, NPL3039703, NPL3039748, NPL3039749, NPL3039750, NPL3039751, NPL3039752, NPL3039754, NPL3039756, NPL3039762, NPL3039763, NPL3039764, NPL3039776, NPL3039777, NPL3039781, NPL3039785, NPL3039787, NPL3039792, NPL3039799, NPL3039801, NPL3039802, NPL3039804, NPL3039806, NPL3039807, NPL3039808, NPL3039810, NPL3039820, NPL3039822, NPL3039823, NPL3039829, NPL3039830, NPL3039832, NPL3039835, NPL3039837, NPL3039840, NPL3039843, NPL3039844, NPL3039850, NPL3039853, NPL3039857, NPL3039859, NPL3039860, NPL3039862, NPL3039865, NPL3039872, NPL3039873, NPL3039874, NPL3039877, NPL3039880, NPL3039881, NPL3039883, NPL3039885, NPL3039886, NPL3039887, NPL3039888, NPL3039890, NPL3039891, NPL3039892, NPL3039893, NPL3039894, NPL3039895, NPL3039899, NPL3039900, NPL3039901, NPL3039902, NPL3039903, NPL3039909, NPL3039913, NPL3039914, NPL3039916, NPL3039919, NPL3039920, NPL3039921, NPL3039923, NPL3039924, NPL3039925, NPL3039926, NPL3039927, NPL3039928, NPL3039935, NPL3039936, NPL3039941, NPL3039943, NPL3039944, NPL3039945, NPL3039950, NPL3039960, NPL3039961, NPL3039963, NPL3039964, NPL3039966, NPL3039967, NPL3039968, NPL3039969, NPL3039970, NPL3039974, NPL3039977, NPL3039980, NPL3039981, NPL3039983, NPL3039984, NPL3039987, NPL3039988, NPL3039992, NPL3040006, NPL3040007, NPL3040008, NPL3040013, NPL3040017, NPL3040018, NPL3040019, NPL3040021, NPL3040022, NPL3040023, NPL3040026, NPL3040027, NPL3040028, NPL3040029, NPL3040030, NPL3040035, NPL3040036, NPL3040037, NPL3040038, NPL3040039, NPL3040041, NPL3040042, NPL3040043, NPL3040045, NPL3040046, NPL3040047, NPL3040049, NPL3040050, NPL3040051, NPL3040052, NPL3040053, NPL3040054, NPL3040055, NPL3040056, NPL3040057, NPL3040058, NPL3040060, NPL3040062, NPL3040063, NPL3040064, NPL3040065, NPL3040068, NPL3040069, NPL3040074, NPL3040076, NPL3040078, NPL3040085, NPL3040086, NPL3040087, NPL3040096, NPL3040097, NPL3040108, NPL3040113, NPL3040114, NPL3040118, NPL3040119, NPL3040120, NPL3040121, NPL3040123, NPL3040124, NPL3040125, NPL3040127, NPL3040129, NPL3040130, NPL3040131, NPL3040132, NPL3040133, NPL3040134, NPL3040135, NPL3040136, NPL3040139, NPL3040140, NPL3040145, NPL3040169, NPL3040173, NPL3040183, NPL3040184, NPL3040186, NPL3040187, NPL3040188, NPL3040195, NPL3040196, NPL3040200, NPL3040201, NPL3040202, NPL3040208, NPL3040215, NPL3040226, NPL3040230, NPL3040236, NPL3040249, NPL3040250, NPL3040264, NPL3040265, NPL3040267, NPL3040269, NPL3040279, NPL3040287, NPL3040288, NPL3040299, NPL3040302, NPL3040303, NPL3040316, NPL3040317, NPL3040320, NPL3040321, NPL3040325, NPL3040333, NPL3040334, NPL3040343, NPL3040347, NPL3040348, NPL3040350, NPL3040352, NPL3040356, NPL3040360, NPL3040372, NPL3040385, NPL3040388, NPL3040389, NPL3040391, NPL3040392, NPL3040407, NPL3040418, NPL3040426, NPL3040473, NPL3040488, NPL3040490, NPL3040496, NPL3040515, NPL3040569, NPL3040575, NPL3040606, NPL3040637, NPL3040645, NPL3040663, NPL3040692, NPL3040749, NPL3040752, NPL3040766, NPL3040787, NPL3040833, NPL3040836, NPL3040837, NPL3040839, NPL3040840, NPL3040841, NPL3040842, NPL3040843, NPL3040905, NPL3040958, NPL3044979, NPL3045061, NPL3045088, NPL3045127, NPL3045136, NPL3045141, NPL3045144, NPL3045162, NPL3045173, NPL3045179, NPL3045229, NPL3045236, NPL3045262; \n\n6. Model Number CT900F, (1) GTIN 00763000926144, Serial Numbers:  NPL3041229, NPL3041230, NPL3041231, NPL3041234, NPL3041235, NPL3041236, NPL3041237, NPL3041239, NPL3041240, NPL3041241, NPL3041244, NPL3041246, NPL3041248, NPL3041261, NPL3041264, NPL3041265, NPL3041266, NPL3041277, NPL3041278, NPL3041281, NPL3041282, NPL3041286, NPL3041288, NPL3041298, NPL3041301, NPL3041307, NPL3041311, NPL3041312, NPL3041324, NPL3041326, NPL3041335, NPL3041336, NPL3041337, NPL3041338, NPL3041339, NPL3041340, NPL3041352, NPL3041353, NPL3041356, NPL3041357, NPL3041359, NPL3041360, NPL3041457, NPL3041481, NPL3041486, NPL3041489, NPL3041491, NPL3041493, NPL3041505, NPL3041506, NPL3041527, NPL3041528, NPL3041532, NPL3041537, NPL3041545, NPL3041546, NPL3041551, NPL3041565, NPL3041566, NPL3041569, NPL3041585, NPL3041586, NPL3041589, NPL3041590, NPL3041614, NPL3041622, NPL3041632, NPL3041633, NPL3041640, NPL3041641, NPL3041651, NPL3041655, NPL3041656, NPL3041657, NPL3041661, NPL3041663, NPL3041682, NPL3041709, NPL3041713, NPL3041725, NPL3041728, NPL3041737, NPL3041744, NPL3041754, NPL3041763, NPL3041764, NPL3041788, NPL3041790, NPL3041793, NPL3041794, NPL3041795, NPL3041801, NPL3041807, NPL3041825, NPL3041868, NPL3041878, NPL3041884, NPL3041890, NPL3041891, NPL3041892, NPL3041893, NPL3041896, NPL3041897, NPL3041903, NPL3041910, NPL3041911, NPL3041942, NPL3041948, NPL3041959, NPL3041960, NPL3041966, NPL3041970, NPL3041975, NPL3041976, NPL3042017, NPL3042076, NPL3042095, NPL3042098, NPL3042261, NPL3042272, NPL3042283, NPL3042306, NPL3042308, NPL3042317, NPL3042326, NPL3042327, NPL3042329, NPL3042330, NPL3042331, NPL3042332, NPL3042354, NPL3042355, NPL3042361, NPL3042363, NPL3042366, NPL3042367, NPL3042368, NPL3042371, NPL3042376, NPL3042378, NPL3042416, NPL3042417, NPL3042419, NPL3042429, NPL3042432, NPL3042436, NPL3042449, NPL3042462, NPL3042476, NPL3042477, NPL3042479, NPL3042499, NPL3042509, NPL3042510, NPL3042514, NPL3042529, NPL3042532, NPL3042534, NPL3042537, NPL3042541, NPL3042553, NPL3042576, NPL3042621, NPL3042639, NPL3042644, NPL3042646, NPL3042661, NPL3042663, NPL3042665, NPL3042682, NPL3042713, NPL3042718, NPL3042730, NPL3043028, NPL3043034, NPL3043036, NPL3043037, NPL3043038, NPL3043044, NPL3043047, NPL3043049, NPL3043050, NPL3043051, NPL3043066, NPL3043075, NPL3043076, NPL3043078, NPL3043079, NPL3043080, NPL3043082, NPL3043083, NPL3043091, NPL3043092, NPL3043094, NPL3043095, NPL3043097, NPL3043100, NPL3043101, NPL3043113, NPL3043165, NPL3043167, NPL3043213, NPL3043217, NPL3043223, NPL3043224, NPL3043225, NPL3043226, NPL3043227, NPL3043229, NPL3043231, NPL3043233, NPL3043244, NPL3043249, NPL3043250, NPL3043251, NPL3043252, NPL3043253, NPL3043255, NPL3043257, NPL3043258, NPL3043259, NPL3043260, NPL3043261, NPL3043262, NPL3043263, NPL3043264, NPL3043265, NPL3043266, NPL3043267, NPL3043268, NPL3043271, NPL3043273, NPL3043274, NPL3043282, NPL3043288, NPL3043290, NPL3043321, NPL3043332, NPL3043339, NPL3043341, NPL3043342, NPL3043343, NPL3043344, NPL3043358, NPL3043360, NPL3043361, NPL3043363, NPL3043364, NPL3043366, NPL3043373, NPL3043381, NPL3043399, NPL3043400, NPL3043440, NPL3043441, NPL3043442, NPL3043444, NPL3043474, NPL3043475, NPL3043477, NPL3043488, NPL3043491, NPL3043503, NPL3043504, NPL3043505, NPL3043506, NPL3043507, NPL3043513, NPL3043517, NPL3043518, NPL3043519, NPL3043520, NPL3043526, NPL3043527, NPL3043529, NPL3043533, NPL3043535, NPL3043537, NPL3043538, NPL3043540, NPL3043541, NPL3043542, NPL3043543, NPL3043544, NPL3043574, NPL3043575, NPL3043580, NPL3043581, NPL3043582, NPL3043583, NPL3043590, NPL3043591, NPL3043597, NPL3043632, NPL3043660, NPL3043667, NPL3043668, NPL3043673, NPL3043698, NPL3043701, NPL3043719, NPL3043749, NPL3043751, NPL3043810, NPL3043811, NPL3043813, NPL3043816, NPL3043839, NPL3043840, NPL3043852, NPL3043857, NPL3043867, NPL3043873, NPL3043888, NPL3043894, NPL3043899, NPL3043903, NPL3043904, NPL3043921, NPL3043928, NPL3043961, NPL3043962, NPL3044007, NPL3044022, NPL3044024, NPL3044026, NPL3044029, NPL3044032, NPL3044033, NPL3044034, NPL3044035, NPL3044038, NPL3044039, NPL3044041, NPL3044042, NPL3044045, NPL3044053, NPL3044055, NPL3044057, NPL3044069, NPL3044073, NPL3044076, NPL3044090, NPL3044127, NPL3044128, NPL3044129, NPL3044137, NPL3044151, NPL3044153, NPL3044156, NPL3044169, NPL3044180, NPL3044181, NPL3044183, NPL3044208, NPL3044217, NPL3044239, NPL3044240, NPL3044241, NPL3044242, NPL3044243, NPL3044244, NPL3044253, NPL3044258, NPL3044260, NPL3044280, NPL3044290, NPL3044292, NPL3044300, NPL3044301, NPL3044345, NPL3044346, NPL3044347, NPL3044351, NPL3044352, NPL3044357, NPL3044358, NPL3044359, NPL3044360, NPL3044363, NPL3044364, NPL3044367, NPL3044368, NPL3044372, NPL3044373, NPL3044374, NPL3044375, NPL3044376, NPL3044378, NPL3044379, NPL3044380, NPL3044381, NPL3044382, NPL3044383, NPL3044384, NPL3044385, NPL3044386, NPL3044387, NPL3044388, NPL3044389, NPL3044390, NPL3044391, NPL3044392, NPL3044393, NPL3044394, NPL3044395, NPL3044396, NPL3044397, NPL3044398, NPL3044399, NPL3044400, NPL3044401, NPL3044402, NPL3044403, NPL3044404, NPL3044405, NPL3044406, NPL3044407, NPL3044408, NPL3044409, NPL3044410, NPL3044412, NPL3044415, NPL3044425, NPL3044431, NPL3044437, NPL3044438, NPL3044439, NPL3044467, NPL3044490, NPL3044493, NPL3044495, NPL3044505, NPL3044506, NPL3044512, NPL3044517, NPL3044518, NPL3044519, NPL3044520, NPL3044531, NPL3044533, NPL3044555, NPL3044557, NPL3044558, NPL3044559, NPL3044560, NPL3044562, NPL3044563, NPL3044564, NPL3044569, NPL3044571, NPL3044572, NPL3045323, NPL3045335, NPL3045359, NPL3045370, NPL3045404, NPL3045530, NPL3045549, NPL3045600, NPL3045696, NPL3045697, NPL3045698, NPL3045705, NPL3045706, NPL3045707, NPL3045708, NPL3045709, NPL3045710, NPL3045741, NPL3045744, NPL3045745, NPL3045746, NPL3045747, NPL3045763, NPL3045764, NPL3045770, NPL3045772, NPL3045773, NPL3045779, NPL3045783, NPL3045804, NPL3045805, NPL3045807, NPL3045808, NPL3045809, NPL3045810, NPL3045811, NPL3045812, NPL3045813, NPL3045814, NPL3045839, NPL3045840, NPL3045842, NPL3045849, NPL3045851, NPL3045852, NPL3045870, NPL3045878, NPL3045879, NPL3045881, NPL3045882, NPL3045883, NPL3045886, NPL3045889, NPL3045890, NPL3045891, NPL3045892, NPL3045897, NPL3045898, NPL3045911, NPL3045912, NPL3045924, NPL3045930, NPL3045933, NPL3045934, NPL3045937, NPL3045939, NPL3045940, NPL3045957, NPL3045958, NPL3045960, NPL3045965, NPL3045966, NPL3045973, NPL3045977, NPL3045984, NPL3045985, NPL3046034, NPL3046041, NPL3046042, NPL3046043, NPL3046045, NPL3046066, NPL3046078, NPL3046090, NPL3046091, NPL3046094, NPL3046095, NPL3046097, NPL3046099, NPL3046101, NPL3046103, NPL3046106, NPL3046107, NPL3046109, NPL3046110, NPL3046111, NPL3046112, NPL3046114, NPL3046115, NPL3046116, NPL3046121, NPL3046123, NPL3046125, NPL3046126, NPL3046134, NPL3046135, NPL3046136, NPL3046137, NPL3046138, NPL3046139, NPL3046140, NPL3046141, NPL3046142, NPL3046143, NPL3046144, NPL3046145, NPL3046146, NPL3046147, NPL3046148, NPL3046149, NPL3046150, NPL3046151, NPL3046153, NPL3046154, NPL3046156, NPL3046157, NPL3046159, NPL3046160, NPL3046161, NPL3046162, NPL3046163, NPL3046164, NPL3046165, NPL3046166, NPL3046168, NPL3046169, NPL3046170, NPL3046171, NPL3046172, NPL3046173, NPL3046174, NPL3046176, NPL3046177, NPL3046178, NPL3046179, NPL3046180, NPL3046181, NPL3046182, NPL3046183, NPL3046184, NPL3046185, NPL3046186, NPL3046187, NPL3046188, NPL3046189, NPL3046190, NPL3046191, NPL3046192, NPL3046193, NPL3046194, NPL3046196, NPL3046197, NPL3046198, NPL3046199, NPL3046201, NPL3046203, NPL3046204, NPL3046205, NPL3046206, NPL3046207, NPL3046208, NPL3046209, NPL3046210, NPL3046211, NPL3046212, NPL3046213, NPL3046214, NPL3046215, NPL3046216, NPL3046217, NPL3046218, NPL3046219, NPL3046220, NPL3046221, NPL3046223, NPL3046225, NPL3046226, NPL3046227, NPL3046228, NPL3046229, NPL3046230, NPL3046231, NPL3046232, NPL3046234, NPL3046236, NPL3046238, NPL3046239, NPL3046240, NPL3046241, NPL3046243, NPL3046245, NPL3046246, NPL3046247, NPL3046248, NPL3046249, NPL3046251, NPL3046252, NPL3046253, NPL3046254, NPL3046255, NPL3046256, NPL3046257, NPL3046258, NPL3046259, NPL3046260, NPL3046261, NPL3046262, NPL3046263, NPL3046264, NPL3046265, NPL3046266, NPL3046268, NPL3046269, NPL3046270, NPL3046271, NPL3046272, NPL3046274, NPL3046275, NPL3046276, NPL3046279, NPL3046280, NPL3046281, NPL3046282, NPL3046283, NPL3046284, NPL3046285, NPL3046286, NPL3046287, NPL3046288, NPL3046289, NPL3046290, NPL3046291, NPL3046292, NPL3046293, NPL3046294, NPL3046296, NPL3046297, NPL3046298, NPL3046299, NPL3046300, NPL3046302, NPL3046303, NPL3046304, NPL3046307, NPL3046310, NPL3046311, NPL3046313, NPL3046314, NPL3046316, NPL3046317, NPL3046318, NPL3046319, NPL3046320, NPL3046321, NPL3046323, NPL3046324, NPL3046325, NPL3046326, NPL3046327, NPL3046328, NPL3046329, NPL3046330, NPL3046331, NPL3046332, NPL3046334, NPL3046335, NPL3046338, NPL3046339, NPL3046341, NPL3046342, NPL3046343, NPL3046344, NPL3046345, NPL3046346, NPL3046347, NPL3046348, NPL3046349, NPL3046350, NPL3046351, NPL3046352, NPL3046353, NPL3046354, NPL3046355, NPL3046356, NPL3046357, NPL3046358, NPL3046359, NPL3046360, NPL3046361, NPL3046362, NPL3046363, NPL3046364, NPL3046365, NPL3046366, NPL3046367, NPL3046368, NPL3046369, NPL3046370, NPL3046371, NPL3046372, NPL3046373, NPL3046374, NPL3046376, NPL3046377, NPL3046378, NPL3046379, NPL3046380, NPL3046381, NPL3046382, NPL3046383, NPL3046384, NPL3046386, NPL3046389, NPL3046391, NPL3046393, NPL3046394, NPL3046396, NPL3046397, NPL3046398, NPL3046400, NPL3046401, NPL3046402, NPL3046403, NPL3046404, NPL3046406, NPL3046407, NPL3046408, NPL3046409, NPL3046410, NPL3046412, NPL3046413, NPL3046414, NPL3046415, NPL3046416, NPL3046417, NPL3046418, NPL3046419, NPL3046420, NPL3046422, NPL3046423, NPL3046424, NPL3046425, NPL3046427, NPL3046428, NPL3046430, NPL3046432, NPL3046433, NPL3046434, NPL3046435, NPL3046436, NPL3046437, NPL3046438, NPL3046439, NPL3046440, NPL3046441, NPL3046442, NPL3046443, NPL3046446, NPL3046447, NPL3046448, NPL3046449, NPL3046451, NPL3046452, NPL3046453, NPL3046454, NPL3046455, NPL3046457, NPL3046458, NPL3046459, NPL3046460, NPL3046461, NPL3046462, NPL3046463, NPL3046464, NPL3046465, NPL3046466, NPL3046467, NPL3046468, NPL3046469, NPL3046471, NPL3046472, NPL3046473, NPL3046474, NPL3046476, NPL3046477, NPL3046478, NPL3046479, NPL3046480, NPL3046481, NPL3046482, NPL3046483, NPL3046484, NPL3046485, NPL3046486, NPL3046492, NPL3046493, NPL3046494, NPL3046497, NPL3046498, NPL3046499, NPL3046500, NPL3046501, NPL3046502, NPL3046503, NPL3046504, NPL3046505, NPL3046517, NPL3046518, NPL3046519, NPL3046520, NPL3046521, NPL3046522, NPL3046523, NPL3046532, NPL3046533, NPL3046535, NPL3046537, NPL3046538, NPL3046540, NPL3046542, NPL3046543, NPL3046544, NPL3046545, NPL3046546, NPL3046547, NPL3046548, NPL3046549, NPL3046550, NPL3046551, NPL3046552, NPL3046553, NPL3046554, NPL3046560, NPL3046563, NPL3046564, NPL3046565, NPL3046566, NPL3046567, NPL3046568, NPL3046569, NPL3046570, NPL3046571, NPL3046572, NPL3046573, NPL3046574, NPL3046575, NPL3046577, NPL3046579, NPL3046580, NPL3046581, NPL3046582, NPL3046583, NPL3046584, NPL3046587, NPL3046588, NPL3046589, NPL3046590, NPL3046591, NPL3046592, NPL3046593, NPL3046595, NPL3046597, NPL3046599, NPL3046601, NPL3046603, NPL3046604, NPL3046605, NPL3046606, NPL3046607, NPL3046630, NPL3046631, NPL3046632, NPL3046633, NPL3046634, NPL3046635, NPL3046636, NPL3046637, NPL3046638, NPL3046639, NPL3046640, NPL3046641, NPL3046642, NPL3046643, NPL3046644, NPL3046645, NPL3046646, NPL3046647, NPL3046648, NPL3046649, NPL3046650, NPL3046651, NPL3046652, NPL3046653, NPL3046654, NPL3046655, NPL3046656, NPL3046657, NPL3046659, NPL3046660, NPL3046661, NPL3046662, NPL3046663, NPL3046664, NPL3046665, NPL3046667, NPL3046668, NPL3046669, NPL3046670, NPL3046671, NPL3046672, NPL3046674, NPL3046675, NPL3046676, NPL3046677, NPL3046678, NPL3046679, NPL3046680, NPL3046682, NPL3046684, NPL3046686, NPL3046689, NPL3046705, NPL3046709, NPL3046711, NPL3046712, NPL3046713, NPL3046714, NPL3046715, NPL3046716, NPL3046717, NPL3046719, NPL3046720, NPL3046721, NPL3046723, NPL3046724, NPL3046725, NPL3046726, NPL3046727, NPL3046728, NPL3046729, NPL3046748, NPL3046749, NPL3046750, NPL3046751, NPL3046752, NPL3046753, NPL3046754, NPL3046755, NPL3046756, NPL3046757, NPL3046758, NPL3046759, NPL3046760, NPL3046761, NPL3046762, NPL3046763, NPL3046764, NPL3046766, NPL3046767, NPL3046768, NPL3046771, NPL3046773, NPL3046774, NPL3046775, NPL3046776, NPL3046777, NPL3046779, NPL3046780, NPL3046781, NPL3046782, NPL3046783, NPL3046784, NPL3046785, NPL3046787, NPL3046788, NPL3046793, NPL3046795, NPL3046797, NPL3046800, NPL3046802, NPL3046803, NPL3046804, NPL3046805, NPL3046806, NPL3046807, NPL3046808, NPL3046809, NPL3046810, NPL3046811, NPL3046812, NPL3046813, NPL3046814, NPL3046815, NPL3046816, NPL3046817, NPL3046818, NPL3046819, NPL3046820, NPL3046821, NPL3046822, NPL3046824, NPL3046825, NPL3046826, NPL3046843, NPL3046844, NPL3046845, NPL3046846, NPL3046847, NPL3046848, NPL3046866, NPL3046870, NPL3046871, NPL3046872, NPL3046873, NPL3046874, NPL3046875, NPL3046876, NPL3046877, NPL3046878, NPL3046879, NPL3046880, NPL3046881, NPL3046882, NPL3046883, NPL3046884, NPL3046885, NPL3046886, NPL3046887, NPL3046888, NPL3046889, NPL3046890, NPL3046891, NPL3046892, NPL3046893, NPL3046894, NPL3046895, NPL3046896, NPL3046899, NPL3046900, NPL3046901, NPL3046902, NPL3046903, NPL3046904, NPL3046907, NPL3046908, NPL3046909, NPL3046910, NPL3046911, NPL3046912, NPL3046914, NPL3046915, NPL3046916, NPL3046917, NPL3046919, NPL3046920, NPL3046921, NPL3046922, NPL3046923, NPL3046925, NPL3046926, NPL3046927, NPL3046928, NPL3046929, NPL3046930, NPL3046931, NPL3046932, NPL3046933, NPL3046934, NPL3046935, NPL3046936, NPL3046937, NPL3046938, NPL3046939, NPL3046940, NPL3046941, NPL3046942, NPL3046948, NPL3046949, NPL3046950, NPL3046951, NPL3046952, NPL3046953, NPL3046954, NPL3046955, NPL3046956, NPL3046957, NPL3046958, NPL3046959, NPL3046960, NPL3046961, NPL3046962, NPL3046963, NPL3046964, NPL3046980, NPL3046981, NPL3046992, NPL3046993, NPL3046995, NPL3046997, NPL3046998, NPL3047003, NPL3047005, NPL3047006, NPL3047007, NPL3047010, NPL3047015, NPL3047016, NPL3047019, NPL3047020, NPL3047022, NPL3047023, NPL3047031, NPL3047032, NPL3047034, NPL3047036, NPL3047038, NPL3047041, NPL3047044, NPL3047045, NPL3047047, NPL3047048, NPL3047052, NPL3047053, NPL3047054, NPL3047055, NPL3047065, NPL3047087, NPL3047088, NPL3047090, NPL3047091, NPL3047092, NPL3047095, NPL3047096, NPL3047098, NPL3047099, NPL3047103, NPL3047104, NPL3047105, NPL3047106, NPL3047107, NPL3047108, NPL3047111, NPL3047113, NPL3047114, NPL3047116, NPL3047119, NPL3047123, NPL3047137, NPL3047138, NPL3047139, NPL3047140, NPL3047141, NPL3047142, NPL3047143, NPL3047144, NPL3047145, NPL3047146, NPL3047147, NPL3047148, NPL3047149, NPL3047150, NPL3047151, NPL3047152, NPL3047153, NPL3047154, NPL3047155, NPL3047156, NPL3047157, NPL3047158, NPL3047161, NPL3047162, NPL3047163, NPL3047164, NPL3047165, NPL3047166, NPL3047167, NPL3047168, NPL3047169, NPL3047170, NPL3047177, NPL3047179, NPL3047180, NPL3047182, NPL3047183, NPL3047184, NPL3047187, NPL3047189, NPL3047195, NPL3047203, NPL3047204, NPL3047225, NPL3047226, NPL3047227, NPL3047228, NPL3047231, NPL3047233, NPL3047234, NPL3047235, NPL3047237, NPL3047238, NPL3047239, NPL3047240, NPL3047241, NPL3047242, NPL3047243, NPL3047244, NPL3047245, NPL3047246, NPL3047247, NPL3047248, NPL3047249, NPL3047250, NPL3047253, NPL3047254, NPL3047255, NPL3047256, NPL3047258, NPL3047259, NPL3047260, NPL3047261, NPL3047262, NPL3047263, NPL3047264, NPL3047265, NPL3047266, NPL3047270, NPL3047271, NPL3047272, NPL3047273, NPL3047274, NPL3047275, NPL3047277, NPL3047278, NPL3047279, NPL3047280, NPL3047281, NPL3047282, NPL3047283, NPL3047284, NPL3047285, NPL3047286, NPL3047287, NPL3047288, NPL3047289, NPL3047290, NPL3047291, NPL3047292, NPL3047293, NPL3047294, NPL3047296, NPL3047297, NPL3047298, NPL3047301, NPL3047302, NPL3047303, NPL3047305, NPL3047314, NPL3047318, NPL3047322, NPL3047323, NPL3047325, NPL3047326, NPL3047327, NPL3047328, NPL3047331, NPL3047332, NPL3047333, NPL3047334, NPL3047335, NPL3047336, NPL3047337, NPL3047338, NPL3047339, NPL3047340, NPL3047341, NPL3047342, NPL3047343, NPL3047344, NPL3047345, NPL3047346, NPL3047347, NPL3047348, NPL3047349, NPL3047352, NPL3047353, NPL3047354, NPL3047355, NPL3047356, NPL3047357, NPL3047358, NPL3047359, NPL3047360, NPL3047361, NPL3047362, NPL3047368, NPL3047369, NPL3047370, NPL3047371, NPL3047372, NPL3047373, NPL3047374, NPL3047375, NPL3047376, NPL3047377, NPL3047378, NPL3047379, NPL3047380, NPL3047382, NPL3047383, NPL3047384, NPL3047386, NPL3047387, NPL3047388, NPL3047389, NPL3047390, NPL3047391, NPL3047392, NPL3047393, NPL3047395, NPL3047398, NPL3047399, NPL3047400, NPL3047402, NPL3047403, NPL3047406, NPL3047408, NPL3047409, NPL3047410, NPL3047411, NPL3047412, NPL3047413, NPL3047414, NPL3047415, NPL3047416, NPL3047417, NPL3047418, NPL3047420, NPL3047421, NPL3047422, NPL3047423, NPL3047424, NPL3047425, NPL3047426, NPL3047427, NPL3047428, NPL3047429, NPL3047430, NPL3047432, NPL3047433, NPL3047434, NPL3047435, NPL3047436, NPL3047437, NPL3047438, NPL3047439, NPL3047440, NPL3047442, NPL3047443, NPL3047446, NPL3047452, NPL3047453, NPL3047454, NPL3047455, NPL3047456, NPL3047457, NPL3047462, NPL3047463, NPL3047464, NPL3047466, NPL3047467, NPL3047468, NPL3047469, NPL3047470, NPL3047471, NPL3047472, NPL3047475, NPL3047476, NPL3047479, NPL3047480, NPL3047482, NPL3047483, NPL3047484, NPL3047486, NPL3047487, NPL3047488, NPL3047489, NPL3047491, NPL3047492, NPL3047494, NPL3047495, NPL3047499, NPL3047500, NPL3047501, NPL3047502, NPL3047503, NPL3047504, NPL3047505, NPL3047506, NPL3047507, NPL3047508, NPL3047509, NPL3047510, NPL3047511, NPL3047512, NPL3047513, NPL3047516, NPL3047517, NPL3047518, NPL3047521, NPL3047525, NPL3047526, NPL3047527, NPL3047528, NPL3047529, NPL3047532, NPL3047548, NPL3047549, NPL3047550, NPL3047551, NPL3047559, NPL3047560, NPL3047561, NPL3047562, NPL3047563, NPL3047565, NPL3047566, NPL3047567, NPL3047568, NPL3047569, NPL3047570, NPL3047572, NPL3047573, NPL3047575, NPL3047576, NPL3047577, NPL3047579, NPL3047611, NPL3047614, NPL3047615, NPL3047616, NPL3047617, NPL3047618, NPL3047619, NPL3047622, NPL3047623, NPL3047625, NPL3047626, NPL3047627, NPL3047628, NPL3047629, NPL3047635, NPL3047636, NPL3047638, NPL3047639, NPL3047640, NPL3047641, NPL3047642, NPL3047643, NPL3047644, NPL3047645, NPL3047646, NPL3047647, NPL3047649, NPL3047650, NPL3047651, NPL3047652, NPL3047654, NPL3047656, NPL3047657, NPL3047659, NPL3047662, NPL3047664, NPL3047665, NPL3047668, NPL3047671, NPL3047672, NPL3047673, NPL3047674, NPL3047675, NPL3047676, NPL3047677, NPL3047678, NPL3047679, NPL3047680, NPL3047681, NPL3047682, NPL3047683, NPL3047684, NPL3047685, NPL3047686, NPL3047687, NPL3047689, NPL3047691, NPL3047692, NPL3047693, NPL3047694, NPL3047696, NPL3047697, NPL3047698, NPL3047699, NPL3047700, NPL3047701, NPL3047702, NPL3047703, NPL3047704, NPL3047709, NPL3047710, NPL3047711, NPL3047712, NPL3047713, NPL3047714, NPL3047715, NPL3047716, NPL3047717, NPL3047718, NPL3047719, NPL3047720, NPL3047721, NPL3047722, NPL3047723, NPL3047724, NPL3047725, NPL3047726, NPL3047729, NPL3047731, NPL3047732, NPL3047733, NPL3047734, NPL3047735, NPL3047736, NPL3047738, NPL3047739, NPL3047740, NPL3047741, NPL3047743, NPL3047744, NPL3047745, NPL3047746, NPL3047747, NPL3047748, NPL3047752, NPL3047753, NPL3047754, NPL3047755, NPL3047756, NPL3047757, NPL3047758, NPL3047759, NPL3047760, NPL3047761, NPL3047762, NPL3047763, NPL3047764, NPL3047765, NPL3047769, NPL3047772, NPL3047773, NPL3047774, NPL3047775, NPL3047776, NPL3047777, NPL3047778, NPL3047779, NPL3047781, NPL3047782, NPL3047783, NPL3047784, NPL3047785, NPL3047786, NPL3047787, NPL3047788, NPL3047789, NPL3047790, NPL3047791, NPL3047792, NPL3047793, NPL3047794, NPL3047795, NPL3047796, NPL3047797, NPL3047798, NPL3047799, NPL3047800, NPL3047801, NPL3047802, NPL3047803, NPL3047804, NPL3047805, NPL3047806, NPL3047807, NPL3047808, NPL3047809, NPL3047810, NPL3047811, NPL3047814, NPL3047815, NPL3047816, NPL3047817, NPL3047818, NPL3047820, NPL3047822, NPL3047823, NPL3047824, NPL3047826, NPL3047829, NPL3047830, NPL3047833, NPL3047834, NPL3047835, NPL3047836, NPL3047837, NPL3047838, NPL3047839, NPL3047841, NPL3047842, NPL3047843, NPL3047848, NPL3047849, NPL3047853, NPL3047854, NPL3047855, NPL3047856, NPL3047857, NPL3047858, NPL3047859, NPL3047860, NPL3047861, NPL3047862, NPL3047863, NPL3047864, NPL3047866, NPL3047872, NPL3047875, NPL3047878, NPL3047879, NPL3047880, NPL3047882, NPL3047883, NPL3047884, NPL3047885, NPL3047886, NPL3047887, NPL3047908, NPL3047909, NPL3047910, NPL3047912, NPL3047914, NPL3047915, NPL3047917, NPL3047920, NPL3047921, NPL3047922, NPL3047924, NPL3047925, NPL3047926, NPL3047927, NPL3047928, NPL3047929, NPL3047930, NPL3047935, NPL3047936, NPL3047937, NPL3047938, NPL3047939, NPL3047941, NPL3047944, NPL3047945, NPL3047946, NPL3047947, NPL3047948, NPL3047949, NPL3047950, NPL3047951, NPL3047952, NPL3047954, NPL3047955, NPL3047956, NPL3047957, NPL3047958, NPL3047959, NPL3047960, NPL3047961, NPL3047962, NPL3047963, NPL3047964, NPL3047965, NPL3047966, NPL3047967, NPL3047968, NPL3047969, NPL3047970, NPL3047971, NPL3047972, NPL3047973, NPL3047975, NPL3047976, NPL3047977, NPL3047978, NPL3047979, NPL3047980, NPL3047981, NPL3047982, NPL3047983, NPL3047984, NPL3047987, NPL3047989, NPL3047990, NPL3047991, NPL3047992, NPL3047993, NPL3047994, NPL3047995, NPL3047996, NPL3047997, NPL3047998, NPL3047999, NPL3048000, NPL3048001, NPL3048002, NPL3048003, NPL3048004, NPL3048005, NPL3048011, NPL3048013, NPL3048014, NPL3048015, NPL3048016, NPL3048017, NPL3048019, NPL3048020, NPL3048021, NPL3048022, NPL3048023, NPL3048024, NPL3048025, NPL3048032, NPL3048033, NPL3048034, NPL3048035, NPL3048036, NPL3048038, NPL3048039, NPL3048041, NPL3048046, NPL3048047, NPL3048048, NPL3048052, NPL3048053, NPL3048054, NPL3048055, NPL3048062, NPL3048067, NPL3048068, NPL3048069, NPL3048071, NPL3048075, NPL3048082, NPL3048088, NPL3048089, NPL3048090, NPL3048091, NPL3048092, NPL3048094, NPL3048095, NPL3048096, NPL3048097, NPL3048098, NPL3048099, NPL3048100, NPL3048101, NPL3048102, NPL3048103, NPL3048104, NPL3048105, NPL3048106, NPL3048107, NPL3048109, NPL3048110, NPL3048111, NPL3048112, NPL3048113, NPL3048114, NPL3048115, NPL3048116, NPL3048117, NPL3048118, NPL3048119, NPL3048120, NPL3048121, NPL3048122, NPL3048123, NPL3048124, NPL3048125, NPL3048129, NPL3048130, NPL3048135, NPL3048136, NPL3048137, NPL3048138, NPL3048139, NPL3048142, NPL3048143, NPL3048148, NPL3048149, NPL3048150, NPL3048151, NPL3048152, NPL3048153, NPL3048154, NPL3048160, NPL3048161, NPL3048163, NPL3048164, NPL3048165, NPL3048166, NPL3048168, NPL3048169, NPL3048171, NPL3048173, NPL3048174, NPL3048175, NPL3048176, NPL3048177, NPL3048178, NPL3048180, NPL3048183, NPL3048185, NPL3048186, NPL3048187, NPL3048188, NPL3048193, NPL3048194, NPL3048197, NPL3048200, NPL3048206, NPL3048207, NPL3048208, NPL3048209, NPL3048210, NPL3048214, NPL3048217, NPL3048218, NPL3048219, NPL3048220, NPL3048221, NPL3048222, NPL3048223, NPL3048224, NPL3048225, NPL3048226, NPL3048227, NPL3048229, NPL3048230, NPL3048231, NPL3048232, NPL3048233, NPL3048236, NPL3048237, NPL3048239, NPL3048241, NPL3048243, NPL3048244, NPL3048245, NPL3048246, NPL3048248, NPL3048249, NPL3048250, NPL3048251, NPL3048252, NPL3048254, NPL3048255, NPL3048256, NPL3048257, NPL3048258, NPL3048259, NPL3048261, NPL3048262, NPL3048264, NPL3048265, NPL3048267, NPL3048268, NPL3048269, NPL3048270, NPL3048271, NPL3048272, NPL3048273, NPL3048274, NPL3048277, NPL3048278, NPL3048279, NPL3048283, NPL3048284, NPL3048285, NPL3048286, NPL3048287, NPL3048288, NPL3048289, NPL3048290, NPL3048291, NPL3048292, NPL3048294, NPL3048295, NPL3048296, NPL3048297, NPL3048298, NPL3048300, NPL3048301, NPL3048302, NPL3048303, NPL3048304, NPL3048305, NPL3048306, NPL3048307, NPL3048308, NPL3048310, NPL3048311, NPL3048313, NPL3048314, NPL3048315, NPL3048316, NPL3048318, NPL3048320, NPL3048321, NPL3048330, NPL3048331, NPL3048332, NPL3048333, NPL3048335, NPL3048336, NPL3048337, NPL3048338, NPL3048339, NPL3048340, NPL3048342, NPL3048343, NPL3048345, NPL3048347, NPL3048348, NPL3048350, NPL3048351, NPL3048352, NPL3048353, NPL3048354, NPL3048356, NPL3048357, NPL3048358, NPL3048359, NPL3048360, NPL3048363, NPL3048364, NPL3048365, NPL3048366, NPL3048368, NPL3048369, NPL3048370, NPL3048371, NPL3048372, NPL3048373, NPL3048375, NPL3048376, NPL3048377, NPL3048378, NPL3048379, NPL3048381, NPL3048382, NPL3048383, NPL3048384, NPL3048385, NPL3048386, NPL3048387, NPL3048389, NPL3048393, NPL3048394, NPL3048395, NPL3048396, NPL3048397, NPL3048398, NPL3048399, NPL3048400, NPL3048401, NPL3048402, NPL3048403, NPL3048404, NPL3048405, NPL3048406, NPL3048407, NPL3048409, NPL3048410, NPL3048411, NPL3048412, NPL3048413, NPL3048414, NPL3048415, NPL3048416, NPL3048417, NPL3048418, NPL3048419, NPL3048420, NPL3048421, NPL3048422, NPL3048423, NPL3048424, NPL3048425, NPL3048426, NPL3048427, NPL3048428, NPL3048429, NPL3048430, NPL3048431, NPL3048432, NPL3048433, NPL3048434, NPL3048435, NPL3048436, NPL3048437, NPL3048438, NPL3048439, NPL3048440, NPL3048441, NPL3048442, NPL3048443, NPL3048444, NPL3048445, NPL3048446, NPL3048447, NPL3048448, NPL3048450, NPL3048452, NPL3048453, NPL3048465, NPL3048466, NPL3048467, NPL3048468, NPL3048469, NPL3048470, NPL3048471, NPL3048472, NPL3048473, NPL3048474, NPL3048475, NPL3048476, NPL3048477, NPL3048478, NPL3048479, NPL3048480, NPL3048481, NPL3048482, NPL3048483, NPL3048484, NPL3048485, NPL3048486, NPL3048487, NPL3048488, NPL3048489, NPL3048490, NPL3048499, NPL3048504, NPL3048513, NPL3048514, NPL3048521, NPL3048523, NPL3048530, NPL3048533, NPL3048536, NPL3048544, NPL3048548, NPL3048549, NPL3048552, NPL3048561, NPL3048562, NPL3048563, NPL3048565, NPL3048571, NPL3048581, NPL3048587, NPL3048588, NPL3048594, NPL3048595, NPL3048598, NPL3048599, NPL3048600, NPL3048601, NPL3048602, NPL3048604, NPL3048612, NPL3048614, NPL3048615, NPL3048619, NPL3048621, NPL3048622, NPL3048623, NPL3048625, NPL3048634, NPL3048651, NPL3048652, NPL3048667, NPL3048668, NPL3048671, NPL3048672, NPL3048673, NPL3048674, NPL3048675, NPL3048677, NPL3048678, NPL3048679, NPL3048684, NPL3048685, NPL3048687, NPL3048690, NPL3048691, NPL3048694, NPL3048695, NPL3048706, NPL3048710, NPL3048715, NPL3048716, NPL3048717, NPL3048731, NPL3048732, NPL3048734, NPL3048737, NPL3048738, NPL3048739, NPL3048741, NPL3048742, NPL3048757, NPL3048766, NPL3048771, NPL3048775, NPL3048777, NPL3048783, NPL3048784, NPL3048787, NPL3048797, NPL3048799, NPL3048801, NPL3048806, NPL3048810, NPL3048814, NPL3048818, NPL3048821, NPL3048827, NPL3048829, NPL3048835, NPL3048836, NPL3048837, NPL3048838, NPL3048839, NPL3048840, NPL3048842, NPL3048843, NPL3048844, NPL3048846, NPL3048848, NPL3048849, NPL3048850, NPL3048851, NPL3048852, NPL3048853, NPL3048854, NPL3048855, NPL3048856, NPL3048857, NPL3048858, NPL3048859, NPL3048860, NPL3048861, NPL3048862, NPL3048863, NPL3048864, NPL3048868, NPL3048870, NPL3048871, NPL3048872, NPL3048873, NPL3048874, NPL3048875, NPL3048876, NPL3048877, NPL3048878, NPL3048879, NPL3048880, NPL3048881, NPL3048882, NPL3048883, NPL3048884, NPL3048885, NPL3048886, NPL3048887, NPL3048888, NPL3048889, NPL3048890, NPL3048891, NPL3048892, NPL3048893, NPL3048894, NPL3048895, NPL3048896, NPL3048897, NPL3048898, NPL3048899, NPL3048900, NPL3048901, NPL3048902, NPL3048903, NPL3048904, NPL3048905, NPL3048906, NPL3048907, NPL3048908, NPL3048909, NPL3048910, NPL3048911, NPL3048912, NPL3048913, NPL3048914, NPL3048915, NPL3048916, NPL3048917, NPL3048918, NPL3048919, NPL3048920, NPL3048921, NPL3048927, NPL3048928, NPL3048929, NPL3048930, NPL3048931, NPL3048933, NPL3048934, NPL3048935, NPL3048936, NPL3048938, NPL3048939, NPL3048940, NPL3048941, NPL3048942, NPL3048943, NPL3048944, NPL3048945, NPL3048946, NPL3048947, NPL3048948, NPL3048949, NPL3048951, NPL3048952, NPL3048953, NPL3048954, NPL3048955, NPL3048956, NPL3048957, NPL3048958, NPL3048959, NPL3048960, NPL3048964, NPL3048965, NPL3048966, NPL3048967, NPL3048971, NPL3048974, NPL3048975, NPL3048989, NPL3048990, NPL3049001, NPL3049004, NPL3049005, NPL3049006, NPL3049007, NPL3049008, NPL3049009, NPL3049010, NPL3049011, NPL3049013, NPL3049014, NPL3049015, NPL3049016, NPL3049017, NPL3049018, NPL3049039, NPL3049045, NPL3049046, NPL3049052, NPL3049053, NPL3049054, NPL3049058, NPL3049059, NPL3049064, NPL3049065, NPL3049099, NPL3049106, NPL3049107, NPL3049108, NPL3049109, NPL3049110, NPL3049117, NPL3049119, NPL3049120, NPL3049121, NPL3049122, NPL3049123, NPL3049124, NPL3049125, NPL3049126, NPL3049127, NPL3049128, NPL3049129, NPL3049130, NPL3049132, NPL3049133, NPL3049134, NPL3049135, NPL3049136, NPL3049137, NPL3049140, NPL3049142, NPL3049145, NPL3049146, NPL3049147, NPL3049148, NPL3049149, NPL3049150, NPL3049151, NPL3049153, NPL3049154, NPL3049156, NPL3049157, NPL3049158, NPL3049159, NPL3049160, NPL3049161, NPL3049163, NPL3049164, NPL3049165, NPL3049166, NPL3049167, NPL3049168, NPL3049169, NPL3049170, NPL3049171, NPL3049172, NPL3049173, NPL3049175, NPL3049176, NPL3049177, NPL3049178, NPL3049179, NPL3049180, NPL3049182, NPL3049183, NPL3049184, NPL3049185, NPL3049187, NPL3049188, NPL3049189, NPL3049192, NPL3049194, NPL3049199, NPL3049200, NPL3049202, NPL3049203, NPL3049204, NPL3049206, NPL3049217, NPL3049218, NPL3049219, NPL3049220, NPL3049221, NPL3049222, NPL3049223, NPL3049231, NPL3049234, NPL3049235, NPL3049243, NPL3049254, NPL3049255, NPL3049256, NPL3049257, NPL3049262, NPL3049263, NPL3049264, NPL3049265, NPL3049267, NPL3049269, NPL3049284, NPL3049289, NPL3049290, NPL3049291, NPL3049300, NPL3049303, NPL3049309, NPL3049310, NPL3049312, NPL3049320, NPL3049328, NPL3049329, NPL3049330, NPL3049332, NPL3049334, NPL3049336, NPL3049337, NPL3049346, NPL3049347, NPL3049359, NPL3049360, NPL3049361, NPL3049362, NPL3049364, NPL3049365, NPL3049377, NPL3049380, NPL3049381, NPL3049382, NPL3049383, NPL3049384, NPL3049398, NPL3049404, NPL3049409, NPL3049421, NPL3049429, NPL3049433, NPL3049434, NPL3049440, NPL3049450, NPL3049456, NPL3049457, NPL3049458, NPL3049459, NPL3049460, NPL3049461, NPL3049463, NPL3049464, NPL3049466, NPL3049467, NPL3049470, NPL3049471, NPL3049474, NPL3049475, NPL3049477, NPL3049479, NPL3049481, NPL3049482, NPL3049483, NPL3049484, NPL3049485, NPL3049486, NPL3049487, NPL3049489, NPL3049505, NPL3049508, NPL3049509, NPL3049510, NPL3049511, NPL3049512, NPL3049513, NPL3049514, NPL3049515, NPL3049516, NPL3049517, NPL3049518, NPL3049519, NPL3049520, NPL3049521, NPL3049522, NPL3049523, NPL3049524, NPL3049525, NPL3049526, NPL3049527, NPL3049528, NPL3049529, NPL3049530, NPL3049531, NPL3049532, NPL3049533, NPL3049534, NPL3049535, NPL3049536, NPL3049537, NPL3049538, NPL3049539, NPL3049540, NPL3049543, NPL3049550, NPL3049551, NPL3049552, NPL3049554, NPL3049555, NPL3049556, NPL3049557, NPL3049558, NPL3049559, NPL3049563, NPL3049564, NPL3049565, NPL3049566, NPL3049567, NPL3049568, NPL3049569, NPL3049571, NPL3049572, NPL3049576, NPL3049582, NPL3049587, NPL3049588, NPL3049589, NPL3049590, NPL3049591, NPL3049592, NPL3049593, NPL3049594, NPL3049595, NPL3049596, NPL3049597, NPL3049598, NPL3049599, NPL3049600, NPL3049601, NPL3049602, NPL3049604, NPL3049606, NPL3049607, NPL3049609, NPL3049611, NPL3049612, NPL3049613, NPL3049614, NPL3049615, NPL3049616, NPL3049617, NPL3049618, NPL3049619, NPL3049620, NPL3049621, NPL3049622, NPL3049623, NPL3049624, NPL3049625, NPL3049626, NPL3049627, NPL3049628, NPL3049629, NPL3049630, NPL3049631, NPL3049632, NPL3049633, NPL3049634, NPL3049635, NPL3049636, NPL3049637, NPL3049638, NPL3049639, NPL3049640, NPL3049641, NPL3049642, NPL3049643, NPL3049644, NPL3049645, NPL3049646, NPL3049647, NPL3049648, NPL3049649, NPL3049650, NPL3049651, NPL3049652, NPL3049653, NPL3049654, NPL3049655, NPL3049656, NPL3049657, NPL3049658, NPL3049659, NPL3049660, NPL3049661, NPL3049662, NPL3049663, NPL3049664, NPL3049665, NPL3049666, NPL3049667, NPL3049668, NPL3049669, NPL3049670, NPL3049671, NPL3049672, NPL3049673, NPL3049674, NPL3049675, NPL3049676, NPL3049677, NPL3049678, NPL3049679, NPL3049680, NPL3049681, NPL3049682, NPL3049683, NPL3049684, NPL3049685, NPL3049686, NPL3049687, NPL3049688, NPL3049690, NPL3049691, NPL3049692, NPL3049693, NPL3049694, NPL3049695, NPL3049696, NPL3049697, NPL3049698, NPL3049699, NPL3049700, NPL3049701, NPL3049702, NPL3049703, NPL3049704, NPL3049705, NPL3049706, NPL3049707, NPL3049708, NPL3049709, NPL3049710, NPL3049711, NPL3049712, NPL3049713, NPL3049714, NPL3049715, NPL3049716, NPL3049717, NPL3049718, NPL3049719, NPL3049720, NPL3049721, NPL3049722, NPL3049723, NPL3049724, NPL3049725, NPL3049726, NPL3049727, NPL3049728, NPL3049729, NPL3049730, NPL3049731, NPL3049732, NPL3049733, NPL3049734, NPL3049735, NPL3049737, NPL3049738, NPL3049739, NPL3049740, NPL3049741, NPL3049742, NPL3049743, NPL3049744, NPL3049745, NPL3049746, NPL3049747, NPL3049748, NPL3049749, NPL3049750, NPL3049751, NPL3049752, NPL3049753, NPL3049754, NPL3049755, NPL3049756, NPL3049757, NPL3049758, NPL3049759, NPL3049761, NPL3049763, NPL3049765, NPL3049770, NPL3049772, NPL3049774, NPL3049775, NPL3049777, NPL3049778, NPL3049779, NPL3049783, NPL3049786, NPL3049787, NPL3049788, NPL3049789, NPL3049793, NPL3049794, NPL3049795, NPL3049797, NPL3049798, NPL3049806, NPL3049819, NPL3049820, NPL3049821, NPL3049822, NPL3049823, NPL3049825, NPL3049826, NPL3049829, NPL3049830, NPL3049832, NPL3049833, NPL3049835, NPL3049837, NPL3049838, NPL3049839, NPL3049840, NPL3049843, NPL3049857, NPL3049876, NPL3050007, NPL3050117, NPL3050122, NPL3050146, NPL3050205, NPL3050212, NPL3050217, NPL3050275, NPL3050286, NPL3050310, NPL3050333, NPL3050349, NPL3050364, NPL3050392, NPL3050414, NPL3050424, NPL3050426, NPL3050443, NPL3050449, NPL3050497, NPL3050499, NPL3050508, NPL3050518, NPL3050538, NPL3050562, NPL3050575, NPL3050581, NPL3050583, NPL3050592, NPL3050607, NPL3050620, NPL3050674, NPL3050675, NPL3050677, NPL3050678, NPL3050679, NPL3050680, NPL3050682, NPL3050683, NPL3050685, NPL3050687, NPL3050689, NPL3050690, NPL3050692, NPL3050693, NPL3050694, NPL3050695, NPL3050697, NPL3050700, NPL3050701, NPL3050702, NPL3050703, NPL3050704, NPL3050705, NPL3050706, NPL3050707, NPL3050708, NPL3050709, NPL3050711, NPL3050712, NPL3050714, NPL3050715, NPL3050717, NPL3050721, NPL3050724, NPL3050725, NPL3050726, NPL3050735, NPL3050738, NPL3050739, NPL3050740, NPL3050742, NPL3050743, NPL3050746, NPL3050750, NPL3050751, NPL3050752, NPL3050753, NPL3050755, NPL3050756, NPL3050757, NPL3050758, NPL3050759, NPL3050760, NPL3050761, NPL3050762, NPL3050763, NPL3050764, NPL3050765, NPL3050766, NPL3050767, NPL3050768, NPL3050769, NPL3050770, NPL3050771, NPL3050772, NPL3050773, NPL3050774, NPL3050775, NPL3050776, NPL3050777, NPL3050778, NPL3050779, NPL3050780, NPL3050781, NPL3050782, NPL3050783, NPL3050784, NPL3050785, NPL3050786, NPL3050788, NPL3050789, NPL3050790, NPL3050792, NPL3050793, NPL3050795, NPL3050797, NPL3050799, NPL3050800, NPL3050801, NPL3050802, NPL3050804, NPL3050805, NPL3050806, NPL3050807, NPL3050808, NPL3050809, NPL3050810, NPL3050811, NPL3050812, NPL3050817, NPL3050820, NPL3050821, NPL3050823, NPL3050824, NPL3050830, NPL3050831, NPL3050832, NPL3050833, NPL3050836, NPL3050837, NPL3050839, NPL3050842, NPL3050843, NPL3050844, NPL3050845, NPL3050847, NPL3050848, NPL3050850, NPL3050851, NPL3050854, NPL3050856, NPL3050861, NPL3050862, NPL3050863, NPL3050864, NPL3050865, NPL3050867, NPL3050868, NPL3050873, NPL3050878, NPL3050882, NPL3050883, NPL3050885, NPL3050886, NPL3050887, NPL3050888, NPL3050889, NPL3050890, NPL3050892, NPL3050893, NPL3050895, NPL3050896, NPL3050897, NPL3050898, NPL3050899, NPL3050900, NPL3050902, NPL3050903, NPL3050904, NPL3050907, NPL3050909, NPL3050911, NPL3050912, NPL3050918, NPL3050928, NPL3050930, NPL3050931, NPL3050933, NPL3050934, NPL3050935, NPL3050937, NPL3050938, NPL3050950, NPL3050952, NPL3050953, NPL3050954, NPL3050955, NPL3050956, NPL3050957, NPL3050958, NPL3050962, NPL3050963, NPL3050964, NPL3050965, NPL3050967, NPL3050968, NPL3050969, NPL3050970, NPL3050971, NPL3050973, NPL3050974, NPL3050976, NPL3050979, NPL3050984, NPL3050992, NPL3050993, NPL3050995, NPL3051000, NPL3051005, NPL3051013, NPL3051015, NPL3051018, NPL3051019, NPL3051020, NPL3051021, NPL3051022, NPL3051026, NPL3051027, NPL3051028, NPL3051031, NPL3051032, NPL3051035, NPL3051036, NPL3051037, NPL3051038, NPL3051043, NPL3051046, NPL3051047, NPL3051050, NPL3051056, NPL3051057, NPL3051059, NPL3051060, NPL3051063, NPL3051066, NPL3051067, NPL3051068, NPL3051069, NPL3051071, NPL3051073, NPL3051074, NPL3051075, NPL3051083, NPL3051084, NPL3051087, NPL3051088, NPL3051089, NPL3051090, NPL3051091, NPL3051092, NPL3051095, NPL3051107, NPL3051109, NPL3051112, NPL3051114, NPL3051115, NPL3051117, NPL3051119, NPL3051120, NPL3051121, NPL3051123, NPL3051129, NPL3051130, NPL3051133, NPL3051134, NPL3051135, NPL3051136, NPL3051137, NPL3051138, NPL3051140, NPL3051142, NPL3051143, NPL3051144, NPL3051150, NPL3051155, NPL3051156, NPL3051157, NPL3051160, NPL3051161, NPL3051169, NPL3051173, NPL3051183, NPL3051184, NPL3051187, NPL3051192, NPL3051204, NPL3051208, NPL3051213, NPL3051222, NPL3051227, NPL3051235, NPL3051237, NPL3051239, NPL3051250, NPL3051256, NPL3051272, NPL3051274, NPL3051284, NPL3051286, NPL3051292, NPL3051295, NPL3051300, NPL3051306, NPL3051312, NPL3051313, NPL3051314, NPL3051324, NPL3051326, NPL3051332, NPL3051336, NPL3051338, NPL3051341, NPL3051343, NPL3051348, NPL3051361, NPL3051362, NPL3051368, NPL3051388, NPL3051392, NPL3051393, NPL3051394, NPL3051400, NPL3051401, NPL3051402, NPL3051403, NPL3051405, NPL3051406, NPL3051410, NPL3051411, NPL3051412, NPL3051413, NPL3051414, NPL3051415, NPL3051416, NPL3051417, NPL3051418, NPL3051419, NPL3051421, NPL3051422, NPL3051423, NPL3051424, NPL3051425, NPL3051427, NPL3051428, NPL3051429, NPL3051431, NPL3051432, NPL3051433, NPL3051442, NPL3051450, NPL3051451, NPL3051461, NPL3051462, NPL3051480, NPL3051486, NPL3051490, NPL3051518, NPL3051520, NPL3051521, NPL3051522, NPL3051523, NPL3051525, NPL3051526, NPL3051527, NPL3051532, NPL3051535, NPL3051536, NPL3051543, NPL3051545, NPL3051546, NPL3051547, NPL3051548, NPL3051549, NPL3051550, NPL3051551, NPL3051552, NPL3051556, NPL3051557, NPL3051562, NPL3051564, NPL3051569, NPL3051571, NPL3051573, NPL3051574, NPL3051575, NPL3051577, NPL3051578, NPL3051579, NPL3051580, NPL3051581, NPL3051582, NPL3051583, NPL3051584, NPL3051585, NPL3051586, NPL3051587, NPL3051588, NPL3051589, NPL3051590, NPL3051591, NPL3051592, NPL3051593, NPL3051595, NPL3051596, NPL3051597, NPL3051598, NPL3051599, NPL3051600, NPL3051601, NPL3051602, NPL3051603, NPL3051604, NPL3051605, NPL3051608, NPL3051609, NPL3051610, NPL3051611, NPL3051612, NPL3051613, NPL3051614, NPL3051615, NPL3051616, NPL3051617, NPL3051618, NPL3051619, NPL3051620, NPL3051622, NPL3051623, NPL3051624, NPL3051625, NPL3051626, NPL3051627, NPL3051628, NPL3051630, NPL3051631, NPL3051632, NPL3051633, NPL3051634, NPL3051635, NPL3051636, NPL3051637, NPL3051638, NPL3051639, NPL3051640, NPL3051641, NPL3051642, NPL3051643, NPL3051644, NPL3051645, NPL3051646, NPL3051647, NPL3051649, NPL3051650, NPL3051651, NPL3051652, NPL3051653, NPL3051654, NPL3051655, NPL3051656, NPL3051657, NPL3051658, NPL3051659, NPL3051660, NPL3051661, NPL3051662, NPL3051663, NPL3051664, NPL3051665, NPL3051666, NPL3051667, NPL3051670, NPL3051671, NPL3051673, NPL3051674, NPL3051677, NPL3051678, NPL3051679, NPL3051680, NPL3051681, NPL3051682, NPL3051683, NPL3051685, NPL3051687, NPL3051688, NPL3051690, NPL3051691, NPL3051693, NPL3051694, NPL3051695, NPL3051696, NPL3051697, NPL3051698, NPL3051699, NPL3051700, NPL3051701, NPL3051704, NPL3051706, NPL3051707, NPL3051709, NPL3051710, NPL3051711, NPL3051712, NPL3051714, NPL3051717, NPL3051718, NPL3051720, NPL3051721, NPL3051722, NPL3051723, NPL3051724, NPL3051725, NPL3051726, NPL3051727, NPL3051728, NPL3051730, NPL3051733, NPL3051734, NPL3051736, NPL3051737, NPL3051738, NPL3051739, NPL3051740, NPL3051743, NPL3051744, NPL3051745, NPL3051746, NPL3051747, NPL3051748, NPL3051750, NPL3051752, NPL3051753, NPL3051755, NPL3051757, NPL3051758, NPL3051759, NPL3051760, NPL3051761, NPL3051762, NPL3051763, NPL3051764, NPL3051767, NPL3051768, NPL3051769, NPL3051770, NPL3051771, NPL3051775, NPL3051776, NPL3051777, NPL3051778, NPL3051779, NPL3051780, NPL3051785, NPL3051799, NPL3051804, NPL3051805, NPL3051806, NPL3051810, NPL3051812, NPL3051813, NPL3051817, NPL3051820, NPL3051824, NPL3051826, NPL3051827, NPL3051828, NPL3051830, NPL3051832, NPL3051834, NPL3051835, NPL3051836, NPL3051837, NPL3051840, NPL3051842, NPL3051843, NPL3051844, NPL3051846, NPL3051847, NPL3051848, NPL3051850, NPL3051851, NPL3051852, NPL3051857, NPL3051858, NPL3051861, NPL3051866, NPL3051868, NPL3051869, NPL3051870, NPL3051871, NPL3051872, NPL3051874, NPL3051875, NPL3051877, NPL3051878, NPL3051879, NPL3051881, NPL3051883, NPL3051884, NPL3051885, NPL3051886, NPL3051887, NPL3051889, NPL3051890, NPL3051891, NPL3051892, NPL3051893, NPL3051896, NPL3051897, NPL3051898, NPL3051899, NPL3051900, NPL3051901, NPL3051908, NPL3051910, NPL3051911, NPL3051913, NPL3051919, NPL3051920, NPL3051921, NPL3051922, NPL3051923, NPL3051924, NPL3051925, NPL3051926, NPL3051927, NPL3051928, NPL3051929, NPL3051930, NPL3051931, NPL3051932, NPL3051933, NPL3051935, NPL3051936, NPL3051937, NPL3051938, NPL3051940, NPL3051941, NPL3051943, NPL3051944, NPL3051946, NPL3051947, NPL3051949, NPL3051952, NPL3051956, NPL3051957, NPL3051959, NPL3051975, NPL3051977, NPL3051978, NPL3051979, NPL3051980, NPL3051981, NPL3051982, NPL3051983, NPL3051984, NPL3051985, NPL3051986, NPL3051987, NPL3051991, NPL3051992, NPL3051993, NPL3051994, NPL3051995, NPL3051996, NPL3051997, NPL3051998, NPL3051999, NPL3052000, NPL3052001, NPL3052002, NPL3052003, NPL3052004, NPL3052005, NPL3052006, NPL3052007, NPL3052008, NPL3052009, NPL3052010, NPL3052011, NPL3052013, NPL3052014, NPL3052015, NPL3052016, NPL3052017, NPL3052018, NPL3052019, NPL3052020, NPL3052021, NPL3052022, NPL3052023, NPL3052024, NPL3052026, NPL3052030, NPL3052031, NPL3052032, NPL3052033, NPL3052034, NPL3052038, NPL3052039, NPL3052042, NPL3052046, NPL3052047, NPL3052048, NPL3052049, NPL3052050, NPL3052051, NPL3052052, NPL3052054, NPL3052055, NPL3052056, NPL3052057, NPL3052058, NPL3052059, NPL3052060, NPL3052062, NPL3052063, NPL3052064, NPL3052070, NPL3052074, NPL3052077, NPL3052079, NPL3052080, NPL3052090, NPL3052092, NPL3052093, NPL3052095, NPL3052096, NPL3052098, NPL3052102, NPL3052103, NPL3052104, NPL3052105, NPL3052107, NPL3052108, NPL3052110, NPL3052111, NPL3052112, NPL3052114, NPL3052115, NPL3052120, NPL3052123, NPL3052130, NPL3052132, NPL3052133, NPL3052134, NPL3052137, NPL3052142, NPL3052143, NPL3052145, NPL3052151, NPL3052153, NPL3052172, NPL3052174, NPL3052177, NPL3052183, NPL3052186, NPL3052200, NPL3052207, NPL3052226, NPL3052241, NPL3052256, NPL3052265, NPL3052267, NPL3052269, NPL3052270, NPL3052277, NPL3052286, NPL3052290, NPL3052291, NPL3052292, NPL3052293, NPL3052294, NPL3052295, NPL3052297, NPL3052299, NPL3052301, NPL3052302, NPL3052305, NPL3052306, NPL3052317, NPL3052318, NPL3052320, NPL3052322, NPL3052323, NPL3052324, NPL3052325, NPL3052326, NPL3052327, NPL3052328, NPL3052329, NPL3052330, NPL3052331, NPL3052332, NPL3052333, NPL3052334, NPL3052335, NPL3052336, NPL3052337, NPL3052338, NPL3052339, NPL3052340, NPL3052341, NPL3052342, NPL3052343, NPL3052344, NPL3052345, NPL3052346, NPL3052347, NPL3052348, NPL3052349, NPL3052350, NPL3052351, NPL3052352, NPL3052353, NPL3052354, NPL3052355, NPL3052356, NPL3052357, NPL3052358, NPL3052359, NPL3052360, NPL3052361, NPL3052364, NPL3052365, NPL3052366, NPL3052367, NPL3052368, NPL3052374, NPL3052375, NPL3052377, NPL3052378, NPL3052381, NPL3052382, NPL3052383, NPL3052384, NPL3052385, NPL3052386, NPL3052389, NPL3052390, NPL3052391, NPL3052392, NPL3052394, NPL3052395, NPL3052396, NPL3052397, NPL3052398, NPL3052399, NPL3052400, NPL3052403, NPL3052404, NPL3052405, NPL3052407, NPL3052408, NPL3052415, NPL3052416, NPL3052418, NPL3052432, NPL3052436, NPL3052438, NPL3052442, NPL3052447, NPL3052453, NPL3052459, NPL3052461, NPL3052462, NPL3052479, NPL3052483, NPL3052484, NPL3052485, NPL3052491, NPL3052492, NPL3052495, NPL3052503, NPL3052504, NPL3052519, NPL3052521, NPL3052525, NPL3052527, NPL3052529, NPL3052530, NPL3052531, NPL3052532, NPL3052533, NPL3052536, NPL3052537, NPL3052546, NPL3052547, NPL3052552, NPL3052553, NPL3052554, NPL3052555, NPL3052557, NPL3052558, NPL3052567, NPL3052568, NPL3052569, NPL3052572, NPL3052576, NPL3052577, NPL3052579, NPL3052580, NPL3052581, NPL3052582, NPL3052583, NPL3052584, NPL3052585, NPL3052588, NPL3052590, NPL3052593, NPL3052594, NPL3052595, NPL3052596, NPL3052597, NPL3052598, NPL3052599, NPL3052600, NPL3052603, NPL3052604, NPL3052607, NPL3052608, NPL3052609, NPL3052610, NPL3052611, NPL3052612, NPL3052613, NPL3052614, NPL3052615, NPL3052616, NPL3052618, NPL3052625, NPL3052626, NPL3052627, NPL3052644, NPL3052646, NPL3052649, NPL3052653, NPL3052654, NPL3052683, NPL3052684, NPL3052708, NPL3052709, NPL3052710, NPL3052711, NPL3052712, NPL3052713, NPL3052715, NPL3052716, NPL3052717, NPL3052718, NPL3052719, NPL3052720, NPL3052721, NPL3052722, NPL3052723, NPL3052724, NPL3052725, NPL3052726, NPL3052727, NPL3052728, NPL3052729, NPL3052732, NPL3052733, NPL3052734, NPL3052737, NPL3052738, NPL3052739, NPL3052740, NPL3052742, NPL3052743, NPL3052745, NPL3052753, NPL3052766, NPL3052767, NPL3052769, NPL3052770, NPL3052772, NPL3052773, NPL3052774, NPL3052775, NPL3052776, NPL3052777, NPL3052778, NPL3052780, NPL3052781, NPL3052782, NPL3052783, NPL3052785, NPL3052786, NPL3052787, NPL3052788, NPL3052789, NPL3052790, NPL3052792, NPL3052793, NPL3052794, NPL3052796, NPL3052797, NPL3052800, NPL3052801, NPL3052802, NPL3052804, NPL3052808, NPL3052811, NPL3052812, NPL3052813, NPL3052814, NPL3052822, NPL3052824, NPL3052825, NPL3052828, NPL3052830, NPL3052831, NPL3052832, NPL3052833, NPL3052834, NPL3052837, NPL3052838, NPL3052840, NPL3052842, NPL3052846, NPL3052848, NPL3052849, NPL3052850, NPL3052851, NPL3052852, NPL3052856, NPL3052858, NPL3052861, NPL3052862, NPL3052864, NPL3052865, NPL3052868, NPL3052869, NPL3052870, NPL3052875, NPL3052877, NPL3052881, NPL3052883, NPL3052884, NPL3052885, NPL3052886, NPL3052887, NPL3052889, NPL3052890, NPL3052892, NPL3052898, NPL3052899, NPL3052901, NPL3052902, NPL3052903, NPL3052904, NPL3052908, NPL3052916, NPL3052928, NPL3052930, NPL3052931, NPL3052932, NPL3052933, NPL3052934, NPL3052935, NPL3052936, NPL3052937, NPL3052939, NPL3052940, NPL3052941, NPL3052946, NPL3052947, NPL3052949, NPL3052950, NPL3052951, NPL3052952, NPL3052953, NPL3052955, NPL3052956, NPL3052957, NPL3052958, NPL3052961, NPL3052963, NPL3052964, NPL3052965, NPL3052967, NPL3052968, NPL3052969, NPL3052970, NPL3052972, NPL3052973, NPL3052974, NPL3052975, NPL3052976, NPL3052977, NPL3052978, NPL3052980, NPL3052981, NPL3052982, NPL3052984, NPL3052985, NPL3052987, NPL3052988, NPL3052989, NPL3052990, NPL3052991, NPL3052992, NPL3052993, NPL3052994, NPL3052995, NPL3052996, NPL3052997, NPL3052998, NPL3052999, NPL3053000, NPL3053002, NPL3053003, NPL3053004, NPL3053005, NPL3053006, NPL3053007, NPL3053009, NPL3053010, NPL3053011, NPL3053012, NPL3053013, NPL3053014, NPL3053015, NPL3053016, NPL3053017, NPL3053018, NPL3053019, NPL3053020, NPL3053021, NPL3053022, NPL3053023, NPL3053024, NPL3053025, NPL3053026, NPL3053027, NPL3053028, NPL3053030, NPL3053031, NPL3053032, NPL3053035, NPL3053036, NPL3053037, NPL3053038, NPL3053043, NPL3053044, NPL3053045, NPL3053046, NPL3053047, NPL3053048, NPL3053049, NPL3053050, NPL3053051, NPL3053052, NPL3053053, NPL3053054, NPL3053056, NPL3053057, NPL3053058, NPL3053059, NPL3053060, NPL3053061, NPL3053062, NPL3053063, NPL3053064, NPL3053065, NPL3053067, NPL3053068, NPL3053069, NPL3053070, NPL3053071, NPL3053072, NPL3053073, NPL3053074, NPL3053075, NPL3053076, NPL3053077, NPL3053078, NPL3053079, NPL3053080, NPL3053081, NPL3053082, NPL3053084, NPL3053085, NPL3053086, NPL3053087, NPL3053092, NPL3053093, NPL3053095, NPL3053096, NPL3053099, NPL3053100, NPL3053103, NPL3053104, NPL3053111, NPL3053112, NPL3053113, NPL3053115, NPL3053117, NPL3053118, NPL3053119, NPL3053120, NPL3053121, NPL3053122, NPL3053123, NPL3053124, NPL3053125, NPL3053126, NPL3053134, NPL3053137, NPL3053141, NPL3053142, NPL3053143, NPL3053144, NPL3053145, NPL3053147, NPL3053149, NPL3053150, NPL3053152, NPL3053153, NPL3053154, NPL3053155, NPL3053157, NPL3053158, NPL3053159, NPL3053160, NPL3053161, NPL3053163, NPL3053166, NPL3053167, NPL3053172, NPL3053173, NPL3053175, NPL3053176, NPL3053177, NPL3053178, NPL3053179, NPL3053185, NPL3053187, NPL3053193, NPL3053194, NPL3053196, NPL3053197, NPL3053198, NPL3053204, NPL3053206, NPL3053207, NPL3053211, NPL3053212, NPL3053213, NPL3053215, NPL3053216, NPL3053217, NPL3053218, NPL3053219, NPL3053220, NPL3053221, NPL3053222, NPL3053223, NPL3053224, NPL3053225, NPL3053226, NPL3053227, NPL3053229, NPL3053230, NPL3053231, NPL3053232, NPL3053233, NPL3053234, NPL3053235, NPL3053238, NPL3053239, NPL3053240, NPL3053241, NPL3053246, NPL3053251, NPL3053252, NPL3053253, NPL3053254, NPL3053255, NPL3053257, NPL3053259, NPL3053260, NPL3053261, NPL3053262, NPL3053263, NPL3053264, NPL3053265, NPL3053266, NPL3053267, NPL3053268, NPL3053270, NPL3053271, NPL3053273, NPL3053274, NPL3053280, NPL3053287, NPL3053288, NPL3053289, NPL3053293, NPL3053294, NPL3053295, NPL3053296, NPL3053299, NPL3053300, NPL3053304, NPL3053305, NPL3053308, NPL3053310, NPL3053314, NPL3053316, NPL3053317, NPL3053318, NPL3053319, NPL3053320, NPL3053322, NPL3053323, NPL3053325, NPL3053326, NPL3053327, NPL3053329, NPL3053330, NPL3053331, NPL3053332, NPL3053333, NPL3053334, NPL3053335, NPL3053336, NPL3053340, NPL3053341, NPL3053342, NPL3053343, NPL3053344, NPL3053345, NPL3053346, NPL3053347, NPL3053348, NPL3053349, NPL3053350, NPL3053351, NPL3053352, NPL3053353, NPL3053359, NPL3053360, NPL3053364, NPL3053365, NPL3053366, NPL3053367, NPL3053368, NPL3053369, NPL3053370, NPL3053371, NPL3053373, NPL3053374, NPL3053375, NPL3053376, NPL3053377, NPL3053378, NPL3053379, NPL3053380, NPL3053381, NPL3053382, NPL3053383, NPL3053384, NPL3053386, NPL3053387, NPL3053388, NPL3053390, NPL3053391, NPL3053392, NPL3053393, NPL3053395, NPL3053396, NPL3053397, NPL3053399, NPL3053400, NPL3053401, NPL3053402, NPL3053403, NPL3053404, NPL3053405, NPL3053406, NPL3053407, NPL3053409, NPL3053411, NPL3053412, NPL3053413, NPL3053414, NPL3053415, NPL3053416, NPL3053417, NPL3053418, NPL3053419, NPL3053420, NPL3053421, NPL3053422, NPL3053425, NPL3053426, NPL3053427, NPL3053428, NPL3053429, NPL3053431, NPL3053432, NPL3053433, NPL3053434, NPL3053435, NPL3053436, NPL3053437, NPL3053438, NPL3053439, NPL3053441, NPL3053443, NPL3053444, NPL3053445, NPL3053446, NPL3053447, NPL3053448, NPL3053449, NPL3053450, NPL3053451, NPL3053452, NPL3053453, NPL3053454, NPL3053456, NPL3053457, NPL3053459, NPL3053460, NPL3053461, NPL3053462, NPL3053463, NPL3053465, NPL3053466, NPL3053470, NPL3053473, NPL3053474, NPL3053475, NPL3053477, NPL3053478, NPL3053479, NPL3053480, NPL3053481, NPL3053484, NPL3053485, NPL3053486, NPL3053487, NPL3053491, NPL3053494, NPL3053495, NPL3053496, NPL3053497, NPL3053502, NPL3053505, NPL3053509, NPL3053510, NPL3053517, NPL3053518, NPL3053521, NPL3053522, NPL3053528, NPL3053530, NPL3053531, NPL3053534, NPL3053535, NPL3053536, NPL3053538, NPL3053540, NPL3053542, NPL3053543, NPL3053545, NPL3053546, NPL3053547, NPL3053548, NPL3053560, NPL3053561, NPL3053563, NPL3053567, NPL3053568, NPL3053571, NPL3053575, NPL3053576, NPL3053577, NPL3053579, NPL3053581, NPL3053583, NPL3053584, NPL3053585, NPL3053586, NPL3053587, NPL3053588, NPL3053589, NPL3053590, NPL3053591, NPL3053592, NPL3053593, NPL3053594, NPL3053595, NPL3053596, NPL3053597, NPL3053599, NPL3053601, NPL3053602, NPL3053603, NPL3053605, NPL3053606, NPL3053607, NPL3053609, NPL3053611, NPL3053613, NPL3053614, NPL3053615, NPL3053616, NPL3053617, NPL3053618, NPL3053619, NPL3053620, NPL3053621, NPL3053622, NPL3053626, NPL3053627, NPL3053628, NPL3053631, NPL3053633, NPL3053634, NPL3053635, NPL3053636, NPL3053638, NPL3053640, NPL3053641, NPL3053642, NPL3053646, NPL3053648, NPL3053649, NPL3053651, NPL3053652, NPL3053653, NPL3053656, NPL3053658, NPL3053660, NPL3053661, NPL3053662, NPL3053664, NPL3053666, NPL3053667, NPL3053673, NPL3053674, NPL3053675, NPL3053676, NPL3053677, NPL3053678, NPL3053681, NPL3053684, NPL3053685, NPL3053686, NPL3053688, NPL3053689, NPL3053690, NPL3053691, NPL3053693, NPL3053694, NPL3053696, NPL3053697, NPL3053698, NPL3053701, NPL3053706, NPL3053712, NPL3053713, NPL3053716, NPL3053718, NPL3053719, NPL3053723, NPL3053725, NPL3053726, NPL3053734, NPL3053735, NPL3053736, NPL3053738, NPL3053741, NPL3053742, NPL3053745, NPL3053759, NPL3053760, NPL3053761, NPL3053762, NPL3053763, NPL3053764, NPL3053770, NPL3053771, NPL3053772, NPL3053773, NPL3053774, NPL3053775, NPL3053776, NPL3053777, NPL3053778, NPL3053779, NPL3053780, NPL3053781, NPL3053782, NPL3053783, NPL3053784, NPL3053785, NPL3053786, NPL3053787, NPL3053788, NPL3053789, NPL3053790, NPL3053791, NPL3053794, NPL3053795, NPL3053796, NPL3053797, NPL3053798, NPL3053799, NPL3053800, NPL3053801, NPL3053802, NPL3053803, NPL3053804, NPL3053805, NPL3053806, NPL3053809, NPL3053810, NPL3053815, NPL3053816, NPL3053817, NPL3053818, NPL3053819, NPL3053820, NPL3053821, NPL3053824, NPL3053825, NPL3053826, NPL3053829, NPL3053830, NPL3053832, NPL3053834, NPL3053835, NPL3053836, NPL3053837, NPL3053841, NPL3053844, NPL3053845, NPL3053848, NPL3053851, NPL3053853, NPL3053854, NPL3053855, NPL3053856, NPL3053857, NPL3053858, NPL3053859, NPL3053860, NPL3053861, NPL3053862, NPL3053863, NPL3053865, NPL3053868, NPL3053871, NPL3053872, NPL3053873, NPL3053877, NPL3053879, NPL3053881, NPL3053882, NPL3053884, NPL3053886, NPL3053888, NPL3053889, NPL3053892, NPL3053894, NPL3053895, NPL3053896, NPL3053897, NPL3053898, NPL3053899, NPL3053901, NPL3053902, NPL3053903, NPL3053905, NPL3053906, NPL3053907, NPL3053908, NPL3053910, NPL3053912, NPL3053913, NPL3053915, NPL3053916, NPL3053917, NPL3053919, NPL3053920, NPL3053921, NPL3053922, NPL3053923, NPL3053924, NPL3053925, NPL3053926, NPL3053927, NPL3053928, NPL3053929, NPL3053930, NPL3053931, NPL3053932, NPL3053934, NPL3053936, NPL3053937, NPL3053938, NPL3053939, NPL3053940, NPL3053945, NPL3053946, NPL3053948, NPL3053949, NPL3053951, NPL3053952, NPL3053953, NPL3053954, NPL3053955, NPL3053956, NPL3053957, NPL3053958, NPL3053959, NPL3053960, NPL3053961, NPL3053963, NPL3053964, NPL3053965, NPL3053966, NPL3053967, NPL3053968, NPL3053969, NPL3053971, NPL3053972, NPL3053973, NPL3053975, NPL3053976, NPL3053983, NPL3053984, NPL3053985, NPL3053988, NPL3053989, NPL3053990, NPL3053991, NPL3053992, NPL3053994, NPL3053996, NPL3054010, NPL3054012, NPL3054013, NPL3054016, NPL3054017, NPL3054019, NPL3054020, NPL3054021, NPL3054022, NPL3054023, NPL3054024, NPL3054025, NPL3054026, NPL3054027, NPL3054028, NPL3054032, NPL3054033, NPL3054034, NPL3054035, NPL3054036, NPL3054037, NPL3054038, NPL3054043, NPL3054046, NPL3054047, NPL3054048, NPL3054049, NPL3054050, NPL3054051, NPL3054053, NPL3054054, NPL3054055, NPL3054056, NPL3054057, NPL3054058, NPL3054059, NPL3054060, NPL3054061, NPL3054070, NPL3054076, NPL3054077, NPL3054078, NPL3054079, NPL3054080, NPL3054084, NPL3054085, NPL3054087, NPL3054090, NPL3054092, NPL3054093, NPL3054096, NPL3054098, NPL3054102, NPL3054103, NPL3054104, NPL3054138, NPL3054147, NPL3054150, NPL3054159, NPL3054167, NPL3054175, NPL3054186, NPL3054191, NPL3054194, NPL3054199, NPL3054200, NPL3054202, NPL3054203, NPL3054204, NPL3054205, NPL3054206, NPL3054217, NPL3054223, NPL3054225, NPL3054229, NPL3054231, NPL3054235, NPL3054238, NPL3054239, NPL3054245, NPL3054257, NPL3054260, NPL3054262, NPL3054266, NPL3054269, NPL3054276, NPL3054277, NPL3054283, NPL3054284, NPL3054287, NPL3054290, NPL3054292, NPL3054293, NPL3054294, NPL3054295, NPL3054296, NPL3054297, NPL3054298, NPL3054299, NPL3054300, NPL3054301, NPL3054302, NPL3054303, NPL3054307, NPL3054309, NPL3054311, NPL3054313, NPL3054316, NPL3054318, NPL3054321, NPL3054322, NPL3054325, NPL3054326, NPL3054327, NPL3054328, NPL3054329, NPL3054330, NPL3054331, NPL3054332, NPL3054333, NPL3054341, NPL3054357, NPL3054361, NPL3054362, NPL3054363, NPL3054364, NPL3054365, NPL3054366, NPL3054377, NPL3054378, NPL3054379, NPL3054381, NPL3054385, NPL3054386, NPL3054387, NPL3054388, NPL3054391, NPL3054393, NPL3054396, NPL3054397, NPL3054398, NPL3054399, NPL3054402, NPL3054403, NPL3054404, NPL3054405, NPL3054406, NPL3054407, NPL3054408, NPL3054409, NPL3054410, NPL3054411, NPL3054413, NPL3054414, NPL3054415, NPL3054425, NPL3054426, NPL3054427, NPL3054459, NPL3054463, NPL3054464, NPL3054465, NPL3054467, NPL3054469, NPL3054470, NPL3054472, NPL3054473, NPL3054479, NPL3054480, NPL3054481, NPL3054483, NPL3054484, NPL3054485, NPL3054486, NPL3054491, NPL3054496, NPL3054497, NPL3054498, NPL3054500, NPL3054501, NPL3054502, NPL3054504, NPL3054505, NPL3054507, NPL3054508, NPL3054509, NPL3054510, NPL3054511, NPL3054512, NPL3054514, NPL3054515, NPL3054525, NPL3054527, NPL3054539, NPL3054543, NPL3054545, NPL3054546, NPL3054547, NPL3054548, NPL3054549, NPL3054550, NPL3054571, NPL3054572, NPL3054574, NPL3054577, NPL3054578, NPL3054579, NPL3054583, NPL3054584, NPL3054585, NPL3054587, NPL3054588, NPL3054589, NPL3054591, NPL3054592, NPL3054595, NPL3054605, NPL3054609, NPL3054610, NPL3054611, NPL3054612, NPL3054621, NPL3054622, NPL3054623, NPL3054624, NPL3054626, NPL3054627, NPL3054630, NPL3054632, NPL3054634, NPL3054636, NPL3054640, NPL3054642, NPL3054643, NPL3054644, NPL3054714, NPL3054715, NPL3054724, NPL3054726, NPL3054744, NPL3054746, NPL3054747, NPL3054753, NPL3054754, NPL3054761, NPL3054762, NPL3054763, NPL3054764, NPL3054767, NPL3054768, NPL3054778, NPL3054779, NPL3054788, NPL3054789, NPL3054790, NPL3054793, NPL3054794, NPL3054796, NPL3054797, NPL3054799, NPL3054800, NPL3054801, NPL3054804, NPL3054805, NPL3054806, NPL3054807, NPL3054810, NPL3054814, NPL3054815, NPL3054820, NPL3054821, NPL3054822, NPL3054823, NPL3054832, NPL3054845, NPL3054858, NPL3054859, NPL3054860, NPL3054866, NPL3054867, NPL3054868, NPL3054869, NPL3054871, NPL3054879, NPL3054880, NPL3054881, NPL3054882, NPL3054884, NPL3054885, NPL3054886, NPL3054887, NPL3054908, NPL3054909, NPL3054910, NPL3054911, NPL3054912, NPL3054913, NPL3054938, NPL3054988, NPL3054989, NPL3055007, NPL3055008, NPL3055011, NPL3055012, NPL3055013, NPL3055014, NPL3055015, NPL3055023, NPL3055025, NPL3055026, NPL3055027, NPL3055032, NPL3055034, NPL3055035, NPL3055037, NPL3055039, NPL3055040, NPL3055041, NPL3055042, NPL3055046, NPL3055050, NPL3055051, NPL3055054, NPL3055057, NPL3055059, NPL3055061, NPL3055063, NPL3055064, NPL3055066, NPL3055067, NPL3055068, NPL3055069, NPL3055071, NPL3055072, NPL3055073, NPL3055074, NPL3055078, NPL3055079, NPL3055080, NPL3055081, NPL3055082, NPL3055083, NPL3055084, NPL3055085, NPL3055086, NPL3055087, NPL3055088, NPL3055090, NPL3055091, NPL3055092, NPL3055093, NPL3055094, NPL3055096, NPL3055097, NPL3055098, NPL3055114, NPL3055122, NPL3055123, NPL3055130, NPL3055131, NPL3055132, NPL3055135, NPL3055140, NPL3055141, NPL3055142, NPL3055143, NPL3055144, NPL3055145, NPL3055146, NPL3055147, NPL3055151, NPL3055152, NPL3055156, NPL3055157, NPL3055163, NPL3055165, NPL3055166, NPL3055167, NPL3055168, NPL3055169, NPL3055170, NPL3055171, NPL3055172, NPL3055173, NPL3055175, NPL3055176, NPL3055177, NPL3055178, NPL3055186, NPL3055187, NPL3055190, NPL3055193, NPL3055194, NPL3055197, NPL3055198, NPL3055199, NPL3055202, NPL3055203, NPL3055204, NPL3055205, NPL3055206, NPL3055207, NPL3055208, NPL3055209, NPL3055210, NPL3055212, NPL3055214, NPL3055215, NPL3055216, NPL3055217, NPL3055218, NPL3055219, NPL3055225, NPL3055226, NPL3055227, NPL3055229, NPL3055232, NPL3055233, NPL3055234, NPL3055235, NPL3055236, NPL3055237, NPL3055238, NPL3055239, NPL3055240, NPL3055241, NPL3055242, NPL3055243, NPL3055244, NPL3055245, NPL3055246, NPL3055253, NPL3055254, NPL3055255, NPL3055257, NPL3055258, NPL3055259, NPL3055260, NPL3055261, NPL3055262, NPL3055263, NPL3055265, NPL3055267, NPL3055268, NPL3055271, NPL3055273, NPL3055275, NPL3055276, NPL3055277, NPL3055278, NPL3055279, NPL3055281, NPL3055282, NPL3055283, NPL3055284, NPL3055287, NPL3055288, NPL3058210, NPL3058213, NPL3058214, NPL3058215, NPL3058216, NPL3058224, NPL3058229, NPL3058230, NPL3058231, NPL3058232, NPL3058233, NPL3058234, NPL3058236, NPL3058238, NPL3058240, NPL3058244, NPL3058247, NPL3058248, NPL3058250, NPL3058252, NPL3058253, NPL3058254, NPL3058255, NPL3058256, NPL3058257, NPL3058258, NPL3058260, NPL3058261, NPL3058263, NPL3058264, NPL3058267, NPL3058268, NPL3058269, NPL3058270, NPL3058271, NPL3058273, NPL3058274, NPL3058275, NPL3058276, NPL3058277, NPL3058278, NPL3058280, NPL3058282, NPL3058283, NPL3058284, NPL3058287, NPL3058288, NPL3058289, NPL3058291, NPL3058292, NPL3058293, NPL3058294, NPL3058295, NPL3058296, NPL3058297, NPL3058298, NPL3058299, NPL3058300, NPL3058301, NPL3058302, NPL3058304, NPL3058305, NPL3058306, NPL3058307, NPL3058309, NPL3058310, NPL3058311, NPL3058312, NPL3058313, NPL3058314, NPL3058315, NPL3058316, NPL3058318, NPL3058321, NPL3058322, NPL3058323, NPL3058324, NPL3058328, NPL3058333, NPL3058335, NPL3058342, NPL3058343, NPL3058344, NPL3058346, NPL3058347, NPL3058348, NPL3058349, NPL3058355, NPL3058356, NPL3058357, NPL3058359, NPL3058360, NPL3058361, NPL3058364, NPL3058366, NPL3058469, NPL3058470, NPL3058471, NPL3058472, NPL3058477, NPL3058481, NPL3058482, NPL3058486, NPL3058631, NPL3058632, NPL3058634, NPL3058635, NPL3058636, NPL3058637, NPL3058655, NPL3058657, NPL3058658, NPL3058659, NPL3058660, NPL3058662, NPL3058663, NPL3058666, NPL3058667, NPL3058670, NPL3058677, NPL3058681, NPL3058682, NPL3058683, NPL3058684, NPL3058685, NPL3058686, NPL3058687, NPL3058688, NPL3058689, NPL3058690, NPL3058691, NPL3058692, NPL3058693, NPL3058694, NPL3058695, NPL3058698, NPL3058699, NPL3058701, NPL3058709, NPL3058710, NPL3058713, NPL3058714, NPL3058717, NPL3058719, NPL3058721, NPL3058723, NPL3058727, NPL3058728, NPL3058729, NPL3058730, NPL3058739, NPL3058741, NPL3058742, NPL3058748, NPL3058749, NPL3058764, NPL3058765, NPL3058766, NPL3058767, NPL3058768, NPL3058774, NPL3058775, NPL3058897, NPL3058903, NPL3058917, NPL3058921, NPL3058925, NPL3058928, NPL3058929, NPL3058931, NPL3058936, NPL3058938, NPL3058944, NPL3058945, NPL3058948, NPL3058950, NPL3058951, NPL3058964, NPL3058965, NPL3058967, NPL3058969, NPL3058970, NPL3058971, NPL3058976, NPL3058977, NPL3058980, NPL3058984, NPL3058985, NPL3058986, NPL3058987, NPL3058999, NPL3059006, NPL3059007, NPL3059027, NPL3059030, NPL3059031, NPL3059033, NPL3059036, NPL3059042, NPL3059050, NPL3059051, NPL3059053, NPL3059061, NPL3059066, NPL3059074, NPL3059075, NPL3059076, NPL3059077, NPL3059078, NPL3059087, NPL3059093, NPL3059094, NPL3059095, NPL3059105, NPL3059106, NPL3059110, NPL3059111, NPL3059112, NPL3059113, NPL3059114, NPL3059115, NPL3059116, NPL3059119, NPL3059120, NPL3059123, NPL3059124, NPL3059137, NPL3059151, NPL3059154, NPL3059155, NPL3059157, NPL3059158, NPL3059163, NPL3059174, NPL3059177, NPL3059178, NPL3059183, NPL3059184, NPL3059185, NPL3059186, NPL3059187, NPL3059188, NPL3059189, NPL3059190, NPL3059192, NPL3059195, NPL3059203, NPL3059208, NPL3059232, NPL3059293, NPL3059297, NPL3059389, NPL3059440, NPL3059516, NPL3059553, NPL3059711, NPL3059717, NPL3059748, NPL3059777, NPL3059806, NPL3059814, NPL3059837, NPL3059840, NPL3059867, NPL3059911, NPL3059920, NPL3059925, NPL3059928, NPL3059929, NPL3059933, NPL3059937, NPL3059945, NPL3059947, NPL3059949, NPL3059954, NPL3059957, NPL3059962, NPL3059972, NPL3059987, NPL3059990, NPL3059994, NPL3060001, NPL3060031, NPL3060033, NPL3060041, NPL3060043, NPL3060060, NPL3060063, NPL3060072, NPL3060073, NPL3060075, NPL3060081, NPL3060083, NPL3060091, NPL3060112, NPL3060125, NPL3060126, NPL3060127, NPL3060130, NPL3060136, NPL3060140, NPL3060141, NPL3060143, NPL3060144, NPL3060145, NPL3060174, NPL3060176, NPL3060178, NPL3060179, NPL3060180, NPL3060191, NPL3060194, NPL3060201, NPL3060202, NPL3060205, NPL3060208, NPL3060213, NPL3060217, NPL3060218, NPL3060245, NPL3060275, NPL3060617, NPL3060638, NPL3060755, NPL3060814, NPL3060815; \n\n(2) GTIN 00763000984649, Serial Numbers:  NPL3049809, NPL3049810, NPL3049811, NPL3049812, NPL3049814, NPL3049815, NPL3049816, NPL3049817, NPL3049818, NPL3049862, NPL3049898, NPL3049900, NPL3049906, NPL3049931, NPL3049939, NPL3049940, NPL3049969, NPL3050484, NPL3050493, NPL3051007, NPL3051010, NPL3051054, NPL3051096, NPL3051097, NPL3051099, NPL3051100, NPL3051104, NPL3051170, NPL3051171, NPL3051172, NPL3051178, NPL3051179, NPL3051180, NPL3051181, NPL3051195, NPL3051196, NPL3051200, NPL3051202, NPL3051209, NPL3051211, NPL3051212, NPL3051214, NPL3051219, NPL3051224, NPL3051229, NPL3051230, NPL3051231, NPL3051233, NPL3051234, NPL3051238, NPL3051241, NPL3051244, NPL3051245, NPL3051246, NPL3051247, NPL3051248, NPL3051249, NPL3051251, NPL3051253, NPL3051254, NPL3051257, NPL3051260, NPL3051261, NPL3051262, NPL3051263, NPL3051265, NPL3051266, NPL3051267, NPL3051269, NPL3051270, NPL3051275, NPL3051276, NPL3051277, NPL3051278, NPL3051279, NPL3051281, NPL3051282, NPL3051287, NPL3051288, NPL3051289, NPL3051290, NPL3051291, NPL3051296, NPL3051297, NPL3051298, NPL3051299, NPL3051301, NPL3051303, NPL3051304, NPL3051305, NPL3051307, NPL3051310, NPL3051311, NPL3051316, NPL3051317, NPL3051318, NPL3051319, NPL3051322, NPL3051325, NPL3051327, NPL3051328, NPL3051329, NPL3051333, NPL3051334, NPL3051337, NPL3051339, NPL3051340, NPL3051342, NPL3051345, NPL3051347, NPL3051349, NPL3051350, NPL3051351, NPL3051352, NPL3051353, NPL3051354, NPL3051355, NPL3051358, NPL3051359, NPL3051360, NPL3051363, NPL3051365, NPL3051366, NPL3051369, NPL3051370, NPL3051371, NPL3051381, NPL3051382, NPL3051385, NPL3051386, NPL3051435, NPL3051437, NPL3051438, NPL3051439, NPL3051440, NPL3051441, NPL3051443, NPL3051444, NPL3051446, NPL3051447, NPL3051452, NPL3051453, NPL3051456, NPL3051457, NPL3051458, NPL3051459, NPL3051460, NPL3051464, NPL3051465, NPL3051466, NPL3051467, NPL3051469, NPL3051495, NPL3051498, NPL3051505, NPL3051506, NPL3051507, NPL3051509, NPL3051510, NPL3051511, NPL3051512, NPL3051513, NPL3051516, NPL3051519, NPL3059593.",
      "firm_fei_number": "2182207",
      "recalling_firm": "Medtronic Neuromodulation",
      "address_1": "7000 Central Ave Ne",
      "city": "Minneapolis",
      "state": "MN",
      "postal_code": "55432-3568",
      "additional_info_contact": "Naomi Rodiles\n763-514-2000",
      "reason_for_recall": "Software issue:  When programming the SynchroMed Pump with the A810 CP App in Flex Infusion Mode, the infusion steps should automatically be arranged in consecutive order, from earliest to latest. However, in rare cases, the steps may be displayed out of order. If this occurs, the programmed Flex Infusion schedule will not be delivered at the intended time, if at all.",
      "root_cause_description": "Under Investigation by firm",
      "action": "Medtronic issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignees on 4/8/2026 via email and mail.  The notice explained the issue with the device and provided the following:  \n\"Recommendations:  \nReview the order of Flex Infusion steps for patients previously programmed in Flex Infusion Mode:  \n Review the Session Report from the prior interrogation. Access the report via the A810 application by communicating with the pump, then navigating to Settings > Reports > Session Long Report, and review the patient s previous programming session. Within Current Settings, confirm that the steps and durations are listed in the correct chronological order. If steps are not in chronological time order, either change the timing of the steps so they appear in order or reprogram while adding steps in chronological time order.  \nWhen programming in Flex Infusion mode:  \n Program Flex Infusion steps in chronological order.  \n Before completing programming, confirm on the Finish screen that all steps appear in chronological time order (earliest to latest).  \n If steps are not in chronological time order, reorder.  \nActions:  \n Share this notice with all those who need to be aware of this issue within your organization and maintain a copy of this notice in your records.  \n Please complete and return the Customer Confirmation Form enclosed with this letter, acknowledging that you have received this information. The form may be emailed to neuro.quality@medtronic.com.\"  \nFor questions regarding this communication, please contact Medtronic Technical Services at 1-800-707-0933 or your local Medtronic representative.",
      "product_quantity": "10165 units",
      "distribution_pattern": "Worldwide distribution.",
      "openfda": {
        "pma_number": [
          "P140032",
          "P890055",
          "P860004",
          "P990034",
          "P800036",
          "P080012"
        ],
        "registration_number": [
          "2135338",
          "2246552",
          "2032227",
          "1722746",
          "1422634",
          "3004209178",
          "3003174284",
          "2183319",
          "3003120897",
          "2124215",
          "1526611",
          "3030198397",
          "3007829657",
          "3004593495",
          "2523676",
          "3015537318",
          "3018094310",
          "2182207",
          "3009183442"
        ],
        "fei_number": [
          "2246552",
          "3003166194",
          "3003174284",
          "2183319",
          "3003120897",
          "2124215",
          "1526611",
          "3005700659",
          "3004369318",
          "3030198397",
          "3007829657",
          "1000514656",
          "1000306216",
          "3004593495",
          "2523676",
          "3015537318",
          "3018094310",
          "2182207",
          "3009183442"
        ],
        "device_name": "Pump, Infusion, Implanted, Programmable",
        "medical_specialty_description": "Unknown",
        "regulation_number": "",
        "device_class": "3"
      }
    }
  ]
}