  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-05",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "220847",
      "product_res_number": "Z-2280-2026",
      "event_date_initiated": "2026-04-15",
      "event_date_posted": "2026-06-11",
      "recall_status": "Open, Classified",
      "res_event_number": "99111",
      "product_code": "RDW",
      "product_description": "Fresh Roast Systems ColorTrack.\nModel Name: ColorTrack. Model Number: BENCH R-100",
      "code_info": "Only Model Number provide. Model Number: BENCH R-100",
      "firm_fei_number": "3008254664",
      "recalling_firm": "Fresh Roast Systems Inc",
      "address_1": "200 Ocean Lane Dr Apt 806 200 Ocean Lane Dr",
      "address_2": "Apt 806",
      "city": "Key Biscayne",
      "state": "FL",
      "postal_code": "33149-1447",
      "reason_for_recall": "A defect has been identified in the manufacture of the Fresh Roast Systems ColorTrack instrument model number BENCH R-100 which could under very rare specific misuse circumstances potentially allow human access to infrared laser light levels in excess of the safety Classification of the product.",
      "root_cause_description": "Radiation Control for Health and Safety Act",
      "action": "Fresh Roast Safety, Inc. sent Safety Notice to owners and users of the ColorTrack roasting sensor Model Number 'BENCH R-100' manufactured by Fresh Roast Systems, Inc.",
      "product_quantity": "38",
      "distribution_pattern": "U.S.",
      "openfda": {
        "device_name": "General Purpose Laser Products",
        "medical_specialty_description": "Unknown",
        "regulation_number": "",
        "device_class": "N"
      }
    }
  ]
}