  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-05",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "220364",
      "product_res_number": "Z-2398-2026",
      "event_date_initiated": "2026-05-07",
      "event_date_posted": "2026-06-10",
      "recall_status": "Open, Classified",
      "res_event_number": "98985",
      "product_code": "MLM",
      "k_numbers": [
        "K150168"
      ],
      "product_description": "Siemens Dimension Tacrolimus Flex Reagent Cartridge (TAC), REF DF207, SMN 10700795; enzyme immunoassay",
      "code_info": "Manufacturing Lot/Serial Numbers:  FA6254, FA6260, FA6286; \nUDIs:  (01)00842768035425(10)FA6254(17)20260911; \n(01)00842768035425(10)FA6260(17)20260917; \n(01)00842768035425(10)FA6286(17)20261013.",
      "firm_fei_number": "2517506",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "500 Gbc Dr",
      "city": "Newark",
      "state": "DE",
      "postal_code": "19702-2466",
      "additional_info_contact": "Customer Care Center\n800-275-7795",
      "reason_for_recall": "A potential positive bias in quality control (QC) and patient sample results in three lots of the Dimension Tacrolimus (TAC) Flex reagent cartridge.",
      "root_cause_description": "Under Investigation by firm",
      "action": "Siemens Healthineers issued an Urgent Medical Device Correction notice to its consignees on 5/12/2026 via FedEx.  The notice described the issue, its impact to QC and patient results, and requested the following:  \n\"Customer Actions . Immediately discontinue use of and discard the affected kit lot numbers.  \n. Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable.  \n. Review your inventory of these products to determine your laboratory s replacement needs and to provide information to Siemens Healthineers for required regulatory reporting.  \n. Complete and return the Field Correction Effectiveness Check Form attached to this letter within (30) days.  \n. Please retain this letter with your laboratory records and forward it to anyone who may have received or used this product.\"  \nFor questions, contact your Siemens Healthineers Customer Care Center at (312) 275-7795 or your local Siemens Healthineers technical support representative.",
      "product_quantity": "2878 units",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Brazil, Croatia, Germany, Hungary, India, Italy, Japan, Korea, Lithuania, Pakistan, Poland, Saudi Arabia, Spain, South Africa, Sri Lanka, Turkey, United Arab Emirates.",
      "openfda": {
        "k_number": [
          "K050206",
          "K060502",
          "K060385",
          "K070820",
          "K123343",
          "K063868",
          "K173857",
          "K203833",
          "K150168"
        ],
        "pma_number": [
          "P970025",
          "P970007"
        ],
        "registration_number": [
          "1415939",
          "9610126",
          "2432235",
          "2521625",
          "3005333358",
          "2517506",
          "1218959",
          "3010939897",
          "3003951180",
          "9616673",
          "1643621",
          "3008344661",
          "3026848661"
        ],
        "fei_number": [
          "1415939",
          "3002806559",
          "2432235",
          "2521625",
          "3005333358",
          "2517506",
          "1218959",
          "3010939897",
          "3003951180",
          "1643621",
          "3008344661",
          "3026848661",
          "3003179344"
        ],
        "device_name": "Enzyme Immunoassay, Tracrolimus",
        "medical_specialty_description": "Clinical Chemistry",
        "regulation_number": "862.1678",
        "device_class": "2"
      }
    }
  ]
}