  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-17",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "219686",
      "product_res_number": "Z-1807-2026",
      "event_date_initiated": "2026-03-06",
      "event_date_posted": "2026-04-17",
      "recall_status": "Open, Classified",
      "res_event_number": "98747",
      "product_code": "RHP",
      "product_description": "The product is a handheld ultraviolet-C germicidal wand and contains two low-pressure mercury bulbs with a peak emission wavelength at 254 nm.",
      "code_info": "None provided.",
      "firm_fei_number": "3031733296",
      "recalling_firm": "Uvlizer c/o RAIS INTERNATIONAL LLC",
      "address_1": "16192 Coastal Hwy",
      "city": "Lewes",
      "state": "DE",
      "postal_code": "19958-3608",
      "reason_for_recall": "In some foreseeable use conditions, the Products can expose nearby persons to\nUVC radiation at levels significantly above limits recommended by international safety guidelines\nfor skin and eye exposure.",
      "root_cause_description": "Radiation Control for Health and Safety Act",
      "action": "Notifications to purchasers and subsequent transferees will be transmitted through multiple communication channels: Certified Mail, Email Notifications, and Public Website Notice, describing the defect, safety considerations, and instructions to discontinue use of the Product.",
      "product_quantity": "480,000",
      "distribution_pattern": "U.S.",
      "openfda": {
        "device_name": "Uv Lamp, Germicidal",
        "medical_specialty_description": "Unknown",
        "regulation_number": "",
        "device_class": "N"
      }
    }
  ]
}