  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-10",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "218071",
      "product_res_number": "Z-1024-2026",
      "event_date_initiated": "2025-12-10",
      "event_date_posted": "2026-01-13",
      "recall_status": "Open, Classified",
      "res_event_number": "98260",
      "product_description": "MAVERICK Measurement Module Spectrometer PN 820-01367 (originally part of a kit\nPN 820-01519)\nFinished Goods PN 820-01496",
      "code_info": "PN 820-01367 (originally part of a kit\nPN 820-01519)\nFinished Goods PN 820-01496",
      "firm_fei_number": "3015685946",
      "recalling_firm": "REPLIGEN CORPORATION",
      "address_1": "111 Locke Dr",
      "city": "Marlborough",
      "state": "MA",
      "postal_code": "01752-7236",
      "reason_for_recall": "Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.",
      "root_cause_description": "Radiation Control for Health and Safety Act",
      "action": "Repligen will repair the connections in each potentially affected MAVERICK device. In\nthe meantime, all customers have been instructed to refrain from using the device.",
      "product_quantity": "33",
      "distribution_pattern": "U.S.",
      "openfda": {}
    }
  ]
}