{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-10",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 6
    }
  },
  "results": [
    {
      "cfres_id": "217955",
      "product_res_number": "Z-1158-2026",
      "event_date_initiated": "2025-12-19",
      "event_date_posted": "2026-02-05",
      "recall_status": "Open, Classified",
      "res_event_number": "98237",
      "product_code": "KNS",
      "k_numbers": [
        "K233318"
      ],
      "product_description": "AXIOS Stent and Electrocautery-Enhanced Delivery System - 20mm x 10mm;\nMaterial Number (UPN): M00553660;",
      "code_info": "Material Number (UPN): M00553680;\nUDI-DI: 8714729951179;\nLot Numbers: 35961006\n\n 35961007\n\n 35962524\n\n 35967379\n\n 35968520, 35968521, 35968597, 35968598, 35971077, 35971078, 35971079, 35972100, 35982272, 35982273, 35982274, 35982275, 35982276, 35982277, 35982278, 35982279, 35982420, 35982424, 35982425, 35985313, 35985314, 35993619, 35993720, 35993721, 36109761, 36109763, 36109764, 36109765, 36109766, 36111314, 36111315, 36118993, 36118994, 36122066, 36122067, 36122075, 36132801, 36132802, 36132803, 36132804, 36134972, 36144058, 36144059, 36144120, 36144121, 36174475, 36174476, 36178005, 36187588, 36187589, 36187687, 36187688, 36188887, 36188888, 36188889, 36194446, 36194447, 36197239, 36198580, 36217768, 36217769, 36217770, 36227341, 36227342, 36227343, 36228603, 36228604, 36228605, 36235934, 36236151, 36240589, 36240590, 36241715, 36241716, 36241717, 36251363, 36251364, 36253000, 36253008, 36253041, 36260860, 36260861, 36260862, 36261599, 36261780, 36261781, 36261834, 36261835, 36280395, 36280396, 36280397, 36281912, 36281913, 36281914, 36283860, 36292682, 36292683, 36292684, 36292689, 36292690, 36292691, 36294495, 36301621, 36301622, 36301623, 36301624, 36302290, 36302291, 36302292, 36319024, 36319025, 36319026, 36320016, 36320017, 36323789, 36323790, 36324720, 36325159, 36325560, 36325568, 36325569, 36332101, 36332102, 36341577, 36342114, 36342418, 36352008, 36352009, 36355049, 36355050, 36355752, 36359567, 36362263, 36362264, 36362265, 36363150\n\n 36363151\n\n 36364647\n\n 36372361\n\n 36372362, 36377757, 36378548, 36380549, 36388531, 36390612, 36396985, 36399492, 36403441, 36408983, 36408984, 36417820, 36417821, 36418721, 36418722, 36424650, 36424651, 36424652, 36425889, 36425890, 36425891, 36425892, 36462323, 36462324, 36469427, 36469428, 36469812, 36472250, 36472253, 36472254, 36475370, 36478478, 36478479, 36572740, 36572741, 36573582, 36573583, 36573587, 36573588, 36573589, 36573590, 36591073, 36591074, 36591075, 36603384, 36603385, 36603386, 36604188, 36604189, 36604190, 36604191, 36604748, 36604749, 36607249, 36607250, 36607251, 36607252, 36607253, 36610818, 36610819, 36611060, 36615703, 36615704, 36621589, 36621590, 36623324, 36623325, 36623326, 36693849, 36693850, 36693851, 36693852, 36694699, 36695020, 36695021, 36695023, 36698247, 36698249, 36698251, 36698527, 36698528, 36698529, 36698530, 36698531, 36698532, 36706712, 36706713, 36706714, 36706715, 36709766, 36709767, 36709768, 36709769, 36713255, 36713256, 36713257, 36713258, 36717717, 36717718, 36717719, 36717800, 36717801, 36717802, 36718966, 36718967, 36723909, 36723910, 36728154, 36764237, 36764238, 36767449, 36767450, 36771884, 36771885, 36771886, 36785593, 36788793, 36788794, 36788795, 36795369, 36795370, 36796430, 36796431, 36804013, 36804375, 36839665, 36895362, 36895363, 36895364, 36958216, 36958217, 36962750, 36970218, 36970219, 36980309, 36980310\n\n 36981855\n\n 36981856\n\n 36982694\n\n 36983909, 36985933, 36988630, 36988631, 36988632, 36990382, 36990383, 36991168, 36991170, 36991683, 36992721, 37014102, 37014103, 37014104, 37027339, 37028180, 37028181, 37030756, 37032104, 37036344, 37036345, 37037505, 37037506, 37040169, 37041012, 37041435, 37041436, 37046449, 37046450, 37046928, 37046929, 37048665, 37049464, 37049465, 37049874, 37050321, 37050322, 37050323, 37055465, 37055466, 37055467, 37056013, 37056014, 37057032, 37059006, 37059010, 37094844, 37106164, 37106583, 37131143, 37131144, 37131145, 37131146, 37177253, 37177271, 37177272, 37179955, 37179956, 37180902, 37180903, 37182323, 37183490, 37188257, 37188258, 37197563, 37197578, 37198908, 37199019, 37199360, 37199361, 37199972, 37199973, 37209169, 37209170, 37209171, 37211282, 37211285, 37212243, 37212244, 37212245, 37212246, 37213246, 37213247, 37213248, 37213249, 37240012, 37240013, 37240236, 37240237, 37240249, 37240250, 37240252, 37240254, 37242247, 37242248, 37242249, 37243754, 37243755, 37243756, 37247304, 37247305, 37247306, 37283604, 37283605, 37291851, 37291852, 37292951, 37292952, 37296017, 37296018, 37296019, 37297941, 37300544, 37300545, 37300546, 37302814, 37308038, 37308039, 37325355, 37328835, 37328836, 37328838, 37331112, 37332012, 37333358, 37333362, 37337364, 37340128, 37340129, 37340130, 37340131, 37340270, 37340271, 37341078, 37341079, 37344296\n\n 37344297\n\n 37346443\n\n 37346445\n\n 37348330, 37348331, 37352140, 37352141, 37354943, 37354944, 37357958, 37357959, 37358120, 37359884, 37359885, 37365502, 37365503, 37365504, 37365505, 37375861, 37388591, 37388592, 37388593, 37400608, 37402464, 37402466, 37410374, 37411886, 37411887, 37415923, 37415924, 37415925, 37427250, 37429329, 37429330, 37438134, 37438135, 37439609, 37439610, 37439611, 37451308, 37451309, 37454250, 37454251, 37456595, 37456596, 37459573, 37567407, 37573089, 37573090, 37573091, 37575640, 37578876, 37578877, 37578879, 37579368, 37579369, 37579370, 37582027, 37585524, 37588798, 37588799, 37589260, 37592382, 37592383, 37592384, 37596205, 37596206, 37596207, 37596208, 37601263, 37602176, 37602177, 37602178, 37602179, 37603348, 37607457, 37607458, 37607459, 37607540, 37625485, 37625486, 37626455, 37626456, 37631100, 37631101, 37633190, 37633191, 37638120, 37638121, 37638122, 37639137, 37639138, 37644407, 37644408, 37644409, 37649491, 37649492, 37686664, 37686665, 37686666, 37700937, 37700938, 37706532, 37706533, 37706534, 37708203, 37708204, 37716332, 37716333, 37718549, 37720553, 37720554, 37720555, 37727831, 37730055, 37754383, 37758168, 37758169, 37758170, 37759554, 37761362, 37761363, 37785682, 37785683, 37787089, 37787090, 37789164, 37789168, 37789169, 37820578, 37823439, 37826531, 37826532, 37832473, 37832474, 37833962, 37833963, 37834751, 37844023;",
      "firm_fei_number": "3005099803",
      "recalling_firm": "Boston Scientific Corporation",
      "address_1": "100 Boston Scientific Way",
      "city": "Marlborough",
      "state": "MA",
      "postal_code": "01752-1234",
      "additional_info_contact": "Renee Archie\n508-683-4523",
      "reason_for_recall": "Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one.\nThe most serious reasonably foreseeable outcome in cases where the first flange is unable to deploy or expand would be a cascade of events resulting in additional intervention, either endoscopic or surgical, to remove the stent and close the puncture site.",
      "root_cause_description": "Process control",
      "action": "On December 19, 2025, Urgent Medical Device Removal - Immediate Action Required letters were sent to customers. \nActions to be taken: \n- Immediately stop further distribution or use of any remaining affected devices.\nRemove them from your facility s inventory and segregate them in a secure location until they can be returned to Boston Scientific.\n\" Immediately post this information in a visible location near the affected products to ensure this information is readily accessible to all handlers and users of the device.\n\" Share this notice with any health care professional within your organization and with any organization where the affected products may have been transferred.\n\" If you are a distributor, this notification must be forwarded to your customers to ensure notification of this product removal is carried out to the end-user level.\n\" If you are a facility that has sent products to another hospital or a facility within your network, ensure this notification is forwarded to them.\n\" Boston Scientific reminds healthcare professionals that AXIOS Stents should only be used as indicated in Instructions For Use.",
      "product_quantity": "557 units",
      "distribution_pattern": "Worldwide distribution. US Nationwide including Puerto Rico;  Algeria, Argentina, Austria, Belgium, Brazil, Canada, CHILE, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, JAPAN, Jordan, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Morocco, Netherlands, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, PERU, Poland, Portugal, Romania, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, UAE, and United Kingdom.",
      "openfda": {
        "k_number": [
          "K002236",
          "K800975",
          "K884306",
          "K173559",
          "K896426",
          "K970945",
          "K960231",
          "K920089",
          "K845015",
          "K861904",
          "K002802",
          "K200404",
          "K002583",
          "K911648",
          "K232825",
          "K896612",
          "K990376",
          "K150142",
          "K961349",
          "K992402",
          "K913881",
          "K150786",
          "K250351",
          "K223256",
          "K251104",
          "K982557",
          "K973366",
          "K011787",
          "K895855",
          "K242061",
          "K810336",
          "K860366",
          "K133933",
          "K182408",
          "K851964",
          "K991847",
          "K082464",
          "K851652",
          "K212999",
          "K800915",
          "K141991",
          "K911600",
          "K092309",
          "K251226",
          "K160417",
          "K232633",
          "K093079",
          "K012587",
          "K201121",
          "K894861",
          "K962083",
          "K944955",
          "K912129",
          "K171993",
          "K232476",
          "K943629",
          "K800973",
          "K853680",
          "K800965",
          "K895842",
          "K250584",
          "K152771",
          "K860593",
          "K885005",
          "K842810",
          "K222421",
          "K942301",
          "K050578",
          "K171983",
          "K191505",
          "K955247",
          "K013153",
          "K262460",
          "K896507",
          "K030315",
          "K172288",
          "K990855",
          "K884254",
          "K932293",
          "K781353",
          "K110887",
          "K800974",
          "K820240",
          "K172665",
          "K031026",
          "K003569",
          "K183595",
          "K915408",
          "K903323",
          "K241271",
          "K072383",
          "K230328",
          "K083292",
          "K222542",
          "K800321",
          "K260962",
          "K870705",
          "K022595",
          "K010568",
          "K033135",
          "K160241",
          "K192905",
          "K820405",
          "K912601",
          "K862060",
          "K241855",
          "K092571",
          "K193039",
          "K023706",
          "K211282",
          "K820430",
          "K884252",
          "K180165",
          "K250945",
          "K000508",
          "K760178",
          "K211909",
          "K973826",
          "K961762",
          "K130623",
          "K123319",
          "K232162",
          "K171158",
          "K970278",
          "K812213",
          "K972674",
          "K974168",
          "K990850",
          "K780774",
          "K080946",
          "K083275",
          "K934315",
          "K946109",
          "K061684",
          "K252388",
          "K243568",
          "K233113",
          "K122203",
          "K192342",
          "K252889",
          "K942068",
          "K901443",
          "K963484",
          "K021266",
          "K880748",
          "K231533",
          "K052413",
          "K950836",
          "K990145",
          "K962346",
          "K851958",
          "K770280",
          "K181758",
          "K820196",
          "K052051",
          "K921029",
          "K122505",
          "K242192",
          "K180237",
          "K101139",
          "K190093",
          "K945764",
          "K242774",
          "K813434",
          "K802444",
          "K190621",
          "K864413",
          "K102273",
          "K040981",
          "K990628",
          "K950166",
          "K851096",
          "K994166",
          "K941728",
          "K990852",
          "K970053",
          "K911700",
          "K192339",
          "K210934",
          "K982289",
          "K030194",
          "K033203",
          "K882048",
          "K892895",
          "K213627",
          "K884025",
          "K233318",
          "K082114",
          "K800964"
        ],
        "registration_number": [
          "",
          "2183613",
          "1825146",
          "8010047",
          "3010707607",
          "3007593903",
          "1219602",
          "9610773",
          "3007960282",
          "1724474",
          "3004215117",
          "1223925",
          "2246552",
          "2133936",
          "9617070",
          "3004111573",
          "3007305485",
          "3006621103",
          "3025311469",
          "3008386005",
          "3008720584",
          "1225687",
          "3024021261",
          "1056129",
          "3002807310",
          "3031233199",
          "3014925413",
          "1651252",
          "9610612",
          "2183744",
          "9681260",
          "3002807314",
          "3012494290",
          "3012536737",
          "3015895045",
          "1057358",
          "1222140",
          "9616684",
          "1721676",
          "3006696607",
          "3030446844",
          "3007695959",
          "1220592",
          "1649518",
          "1222168",
          "3014579161",
          "3013440400",
          "1320894",
          "1000393132",
          "9611283",
          "3013247477",
          "3008132398",
          "3004784537",
          "1828132",
          "3035434483",
          "9611503",
          "3008040402",
          "3002808148",
          "3004837686",
          "1528319",
          "3009183442",
          "3006950086",
          "3018094310",
          "3010188783",
          "3023102955",
          "3013526170",
          "3010273872",
          "2020550",
          "2650143",
          "3004083736",
          "3021171301",
          "9611102",
          "2916714",
          "3001845648",
          "1000121056",
          "3014620201",
          "3003902943",
          "3003790304",
          "8010456",
          "3014615697",
          "9612074",
          "3011050570",
          "9614641",
          "3010202439",
          "3005099803",
          "3030733800",
          "3010288205",
          "3004904811",
          "1037905",
          "1721504",
          "2431166",
          "3014342096",
          "3015309643",
          "1319639",
          "3003724334",
          "3013557681",
          "1820334",
          "9710641",
          "9610617",
          "3013557562",
          "2124215",
          "9680718",
          "3009526575",
          "8010877",
          "3006546082",
          "3009217531",
          "3033589330",
          "2183446",
          "1321175",
          "9680515",
          "2936485",
          "3005580113",
          "1220477",
          "3002894438",
          "1061927",
          "3015975397",
          "3015286855",
          "3014769442",
          "3002806337"
        ],
        "fei_number": [
          "",
          "2183613",
          "3002807091",
          "1825146",
          "3000718675",
          "3010707607",
          "1000121130",
          "3007475226",
          "3007593903",
          "3001451571",
          "1219602",
          "3001236277",
          "3007960282",
          "3002770040",
          "3004215117",
          "1000208608",
          "2246552",
          "3004111573",
          "3007305485",
          "3006621103",
          "3025311469",
          "3008386005",
          "3008720584",
          "3002807798",
          "3024021261",
          "3002807310",
          "3005969827",
          "3031233199",
          "1000122656",
          "3003995201",
          "3014925413",
          "2183744",
          "3003094851",
          "3002807314",
          "3012494290",
          "3012536737",
          "3015895045",
          "3002808268",
          "1000132436",
          "1721676",
          "3002807072",
          "3006696607",
          "3030446844",
          "3002806315",
          "3007695959",
          "3002806565",
          "1220592",
          "3002913962",
          "3014579161",
          "1222168",
          "3013440400",
          "3000280067",
          "1000393132",
          "3013247477",
          "3004784537",
          "1828132",
          "3001451463",
          "1000138054",
          "3035434483",
          "3002808335",
          "3002806816",
          "1000305720",
          "3008040402",
          "3002808148",
          "3004837686",
          "3009183442",
          "3006950086",
          "3018094310",
          "3010188783",
          "3023102955",
          "3013526170",
          "3010273872",
          "2020550",
          "3002807134",
          "3942",
          "3004083736",
          "3021171301",
          "3002807795",
          "2916714",
          "3001845648",
          "1000121056",
          "2648209",
          "3014620201",
          "3003902943",
          "3003790304",
          "3003565456",
          "3014615697",
          "3011050570",
          "3005099803",
          "3002477156",
          "3030733800",
          "3002808001",
          "3010288205",
          "3004904811",
          "3001611427",
          "1037905",
          "1721504",
          "3014342096",
          "3015309643",
          "2523865",
          "3003724334",
          "3013557681",
          "1820334",
          "3000203000",
          "3013557562",
          "2124215",
          "3009526575",
          "3006546082",
          "3009217531",
          "3033589330",
          "2183446",
          "2936485",
          "3005580113",
          "1220477",
          "3002894438",
          "3015975397",
          "3015286855",
          "3014769442",
          "3002806337"
        ],
        "device_name": "Unit, Electrosurgical, Endoscopic (With Or Without Accessories)",
        "medical_specialty_description": "Gastroenterology, Urology",
        "regulation_number": "876.4300",
        "device_class": "2"
      }
    }
  ]
}