  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-17",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "215838",
      "product_res_number": "Z-2478-2025",
      "event_date_initiated": "2025-08-04",
      "event_date_posted": "2025-09-11",
      "recall_status": "Open, Classified",
      "res_event_number": "97520",
      "product_code": "REM",
      "product_description": "Wasatch Photonics 785/830-SR and 785-ER-ILC Raman Spectrometers, \nProduct Models:\n785-SR-ILP, 785-SR-ILC, 830-SR-ILP, 830-SR-ILC,785-ER-ILC.",
      "code_info": "all codes included in the model numbers.",
      "firm_fei_number": "3014813917",
      "recalling_firm": "WASATCH PHOTONICS",
      "address_1": "808 Aviation Pkwy Ste 1400 808 Aviation Pkwy",
      "address_2": "Ste 1400",
      "city": "Morrisville",
      "state": "NC",
      "postal_code": "27560-6600",
      "reason_for_recall": "Two new potential laser safety failure modes have been identified in the product design. Overcurrent of laser diode, and a fail-unsafe mode.\nThe failure modes have never been found to occur in these models, whether in the field or factory, but could potentially occur under excessive force or extreme, continuous vibration.",
      "root_cause_description": "Radiation Control for Health and Safety Act",
      "action": "Notifications will be sent out to all purchasers of affected models. Notification will include an offer of free RMA (return to factory for repair, then shipment back to customer). All repairs will be performed at the site of original manufacture in Morrisville, NC.\n\nCustomer notifications to be delivered by 31-Aug-2025.",
      "product_quantity": "192",
      "distribution_pattern": "U.S.",
      "openfda": {
        "device_name": "Spectroscopy Instrument",
        "medical_specialty_description": "Unknown",
        "regulation_number": "",
        "device_class": "N"
      }
    }
  ]
}