{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-17",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "214768",
      "product_res_number": "Z-2272-2025",
      "event_date_initiated": "2025-07-07",
      "event_date_posted": "2025-08-05",
      "recall_status": "Open, Classified",
      "res_event_number": "97180",
      "product_code": "JGS",
      "k_numbers": [
        "K200107"
      ],
      "product_description": "epoc BGEM BUN Test Card [25pk]. Material Number: 10736515.",
      "code_info": "Material Number: 10736515. GTIN: 00809708121860. Lot Numbers: 02-25095-30, 02-25096-30, 02-25098-30, 02-25099-30, 02-25100-30, 02-25101-30, 02-25102-30, 02-25103-30, 02-25105-30, 02-25111-30, 02-25112-30, 02-25114-30, 02-25115-30, 02-25116-30, 02-25117-30, 02-25119-30, 02-25121-30, 02-25122-30, 02-25124-30, 02-25096-40, 02-25097-40, 02-25098-40, 02-25099-40, 02-25101-40, 02-25102-40, 02-25105-40, 02-25106-40, 02-25121-40, 02-25122-40, 02-25124-40, 02-25125-40, 02-25096-50, 02-25098-50, 02-25099-50, 02-25100-50, 02-25102-50, 02-25103-50, 02-25105-50, 02-25107-50, 02-25108-50, 02-25111-50, 02-25112-50, 02-25113-50, 02-25114-50, 02-25115-50, 02-25119-50, 02-25120-50, 12-25096-60, 12-25097-60, 12-25098-60, 12-25100-60, 12-25101-60, 12-25102-60, 12-25103-60, 12-25111-60, 02-25112-60, 02-25113-60, 02-25114-60, 02-25115-60, 02-25116-60, 02-25117-60, 02-25118-60, 02-25120-60, 02-25121-60, 02-25122-60, 02-25123-60, 02-25124-60, 02-25099-70, 02-25101-70, 02-25104-70, 02-25105-70, 02-25107-70, 02-25108-70, 02-25110-70, 02-25111-70, 02-25112-70, 02-25113-70, 02-25115-70, 02-25116-70, 02-25118-70, 02-25120-70.",
      "firm_fei_number": "3002637618",
      "recalling_firm": "Siemens Healthcare Diagnostics Inc",
      "address_1": "2 Edgewater Dr",
      "city": "Norwood",
      "state": "MA",
      "postal_code": "02062-4637",
      "additional_info_contact": "Customer Service USA\n877-229-3711",
      "reason_for_recall": "Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in some epoc test card lots. All other analytes are performing as intended. The\r\nobserved average bias for sodium was -4.4 mmol/L. The maximum bias observed was -14 mmol/L, which occurred at a higher sodium concentration around 150 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing.",
      "root_cause_description": "Under Investigation by firm",
      "action": "Siemens Healthcare Diagnostics notified consignees via letter on about 07/07/2025. Consignees were instructed to review the letter with the Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable, discard and request a replacement of any affected lots or if your institution does not need to report sodium, you can turn off sodium following the instruction of the owner s manual. Consignees were also instructed to complete and return the Field Correction Effectiveness Check Form included with letter, notify customers if products were further distributed, and to retain this letter with your laboratory records and forward this letter to those who may have received this product at your site.",
      "product_quantity": "37,554 units",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Bahrain, Bolivia, Brazil, Canada, Canary Islands, Chile, Croatia, Denmark, Ecuador, Estonia, France, Fren. Polynesia, Germany, Greece, Guatemala, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Jordan, Kazakhstan, Lithuania, Malaysia, Martinique, Mexico, Morocco, Netherlands, New Caledonia, Paraguay, Philippines, Poland, Portugal, Qatar, Rep. Korea (S), Romania, Russian Fed., San Marino, Slovakia, Slovenia, South Africa, Spain, Srilanka, Sweden, Switzerland, Taiwan, Thailand, Turkey, Uganda, Ukraine, Unit.Arab Emir., United Kingdom, Uruguay, Japan.",
      "openfda": {
        "k_number": [
          "K852595",
          "K811289",
          "K161954",
          "K882104",
          "K844587",
          "K801795",
          "K061597",
          "K844559",
          "K894218",
          "K832070",
          "K895672",
          "K172232",
          "K830588",
          "K901042",
          "K923806",
          "K932179",
          "K962334",
          "K863907",
          "K925612",
          "K800584",
          "K924547",
          "K191365",
          "K811896",
          "K963627",
          "K874832",
          "K812762",
          "K873126",
          "K052724",
          "K020364",
          "K222438",
          "K940194",
          "K872190",
          "K874185",
          "K843530",
          "K905600",
          "K023049",
          "K822039",
          "K840557",
          "K180428",
          "K800676",
          "K963763",
          "K100673",
          "K842873",
          "K875263",
          "K123726",
          "K102959",
          "K800216",
          "K992842",
          "K904335",
          "K961099",
          "K213486",
          "K841417",
          "K053165",
          "K183549",
          "K895936",
          "K011213",
          "K001154",
          "K870183",
          "K833072",
          "K200403",
          "K882801",
          "K801896",
          "K861696",
          "K160370",
          "K002855",
          "K830589",
          "K082462",
          "K002407",
          "K933862",
          "K081411",
          "K802155",
          "K843453",
          "K924471",
          "K771839",
          "K993211",
          "K894914",
          "K820395",
          "K852473",
          "K941400",
          "K140373",
          "K863617",
          "K833712",
          "K252475",
          "K921718",
          "K912016",
          "K070104",
          "K871028",
          "K153357",
          "K841418",
          "K903564",
          "K132418",
          "K955375",
          "K881255",
          "K151767",
          "K091861",
          "K183688",
          "K864862",
          "K910305",
          "K915176",
          "K892500",
          "K833036",
          "K800628",
          "K896526",
          "K220977",
          "K863521",
          "K990346",
          "K952179",
          "K880917",
          "K873074",
          "K861967",
          "K791691",
          "K935120",
          "K051646",
          "K990036",
          "K925611",
          "K861087",
          "K810135",
          "K963542",
          "K992844",
          "K901040",
          "K910554",
          "K912560",
          "K853832",
          "K160225",
          "K832415",
          "K181201",
          "K102647",
          "K921937",
          "K880750",
          "K770601",
          "K833992",
          "K933373",
          "K841424",
          "K843452",
          "K905264",
          "K894035",
          "K832744",
          "K000895",
          "K001761",
          "K801011",
          "K100829",
          "K955374",
          "K053110",
          "K801007",
          "K040971",
          "K914810",
          "K161837",
          "K922215",
          "K801882",
          "K864872",
          "K862819",
          "K171476",
          "K872716",
          "K230275",
          "K850209",
          "K040977",
          "K903188",
          "K912648",
          "K842025",
          "K912387",
          "K121140",
          "K912559"
        ],
        "registration_number": [
          "3003637681",
          "1000455123",
          "2050012",
          "2030861",
          "3005146313",
          "3023162659",
          "9680746",
          "3023268435",
          "8020888",
          "3002807877",
          "3037000637",
          "1319681",
          "3015997711",
          "3003795116",
          "3005509212",
          "3004493545",
          "3000148879",
          "3012963943",
          "1220649",
          "3007603826",
          "3014506106",
          "3013513377",
          "3012817228",
          "8020316",
          "1217183",
          "3007490412",
          "3003593973",
          "9610529",
          "2939693",
          "2431530",
          "3008244372",
          "9614698",
          "3002807968",
          "3004102403",
          "3002642396",
          "2250051",
          "2050010",
          "3012104670",
          "3003971136",
          "2221819",
          "3004438098",
          "2247288",
          "1319436",
          "3004959793",
          "3011989923",
          "3000308930",
          "1000256526",
          "3002637618",
          "3009156722",
          "3009473156",
          "1317202",
          "8030478",
          "9612296",
          "3002809144",
          "3016244760",
          "1219913",
          "1222035",
          "1000414342",
          "2245578",
          "9610816",
          "3006198300",
          "3016438761",
          "3008629785",
          "1616487",
          "1319809",
          "3003030793",
          "1036223",
          "9610633",
          "3014607865",
          "3004378326",
          "3005333358",
          "2517506",
          "1717966",
          "2085064",
          "3010324880",
          "9610126",
          "1827821",
          "3000126629",
          "1181121",
          "1036362",
          "3007662974",
          "3010825766",
          "1220972",
          "1219029",
          "9614673",
          "2432235",
          "3005643513"
        ],
        "fei_number": [
          "1000120319",
          "3003637681",
          "1000455123",
          "2050012",
          "3005146313",
          "3023162659",
          "1000135102",
          "3023268435",
          "3002807877",
          "3037000637",
          "3015997711",
          "3003795116",
          "3005509212",
          "3004493545",
          "3000148879",
          "3002807167",
          "3007019572",
          "3012963943",
          "3002807663",
          "1220649",
          "3007603826",
          "1000122044",
          "3014506106",
          "1000305840",
          "3002808382",
          "3013513377",
          "3012817228",
          "3002806560",
          "3007490412",
          "3003593973",
          "1217183",
          "2939693",
          "2431530",
          "3008244372",
          "3002807968",
          "3004102403",
          "3002642396",
          "2250051",
          "2050010",
          "3012104670",
          "3003971136",
          "2221819",
          "3004438098",
          "3001576181",
          "3000215862",
          "3004959793",
          "3011989923",
          "3000308930",
          "1000256526",
          "3002637618",
          "3009156722",
          "3003559576",
          "3009473156",
          "1317202",
          "3002809144",
          "3016244760",
          "1219913",
          "1222035",
          "1000136573",
          "1000414342",
          "2245578",
          "1000206808",
          "3002806559",
          "3006198300",
          "3016438761",
          "3008629785",
          "1616487",
          "3003030793",
          "1036223",
          "3014607865",
          "3004378326",
          "3005333358",
          "3002806809",
          "2517506",
          "1717966",
          "3010324880",
          "1827821",
          "3000126629",
          "1181121",
          "1036362",
          "3007662974",
          "3010825766",
          "1220972",
          "1219029",
          "2432235",
          "3002806693",
          "3005643513"
        ],
        "device_name": "Electrode, Ion Specific, Sodium",
        "medical_specialty_description": "Clinical Chemistry",
        "regulation_number": "862.1665",
        "device_class": "2"
      }
    }
  ]
}