{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-20",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "199077",
      "product_res_number": "Z-1281-2023",
      "event_date_initiated": "2023-02-17",
      "event_date_posted": "2023-03-22",
      "recall_status": "Open, Classified",
      "res_event_number": "91867",
      "product_code": "MUJ",
      "k_numbers": [
        "K190387"
      ],
      "product_description": "Raystation treatment planning system for radiation therapy, Model Numbers 9.0.0, 9.0.1, 9.1.0, 9.2.0, 10.0.0, 10.0.1, 10.0.2, 10.1.0, 10.1.1, 11.0.0, 11.0.1, 11.0.3, 11.0.4, 12.0.0, 12.1.0, 12.1.1, 12.0.3, 12.1.2, 12.0.4, 12.1.3, 13.0.0, 13.1.0.  Software Version Numbers 9A, 9B, 10A, 10B, 11A, 11B and 12A including service packs",
      "code_info": "UDI-DI: 0735000201067920221007;",
      "firm_fei_number": "3007774465",
      "recalling_firm": "RAYSEARCH LABORATORIES AB",
      "address_1": "Eugeniavagen 18c 18c eugeniavagen",
      "city": "Stockholm",
      "country": "Sweden",
      "reason_for_recall": "It is possible to set a non-zero collimator angle in the Virtual Simulation module. The angle will be correct in the user interface but if the plan is DICOM exported by using the VSIM Export button, the angle will not be correct in the exported RT plan. In the exported RT plan, the DICOM attribute Beam Limiting Device Angle (300A,0120) is set to zero, regardless of the selected collimator angle.",
      "root_cause_description": "Under Investigation by firm",
      "action": "A Field Safety Notice, Medical Device Correction letter dated 2/16/23 was sent to customers.\n\nActions to be taken by the user\n\" Be aware that an RT plan exported by using the  VSIM Export  button will always have the DICOM attribute Beam Limiting Device Angle (300A,0120) set to zero.\n\" Be aware that an RT plan exported by using the  VSIM Export  button will always have the DICOM attribute Number of Wedges (300A,00D0) set to zero and no wedge information will be exported.\n\" Be aware that an RT plan exported by using the  VSIM Export  button will never include the DICOM attribute Applicator Sequence (300A,0107) and no cone information will be exported.\n\" Do not use or export Virtual Simulation plans with non-zero collimator angle, wedges or cones using the  VSIM Export  button.\n\" Be aware that a plan imported as a Virtual Simulation plan may not contain all intended information and that some attributes affecting dose may be different from the original plan.\n\" Always perform dose computation for imported Virtual Simulation plans before export. Do not use the  VSIM Export  for export of such plans.\n\" Educate planning staff and all users about this workaround.\n\" Inspect your product and identify all installed units with the above software version number(s).\n\" Confirm you have read and understood this notice by replying to the notification email.\n\nSolution\nThis issue will be resolved in the next version of RayStation, scheduled for market release in June 2023 (subject to market clearance in some markets). If customers wish to continue using versions of RayStation affected by this notice, all users must maintain awareness of this notice. Alternatively, customers can choose to upgrade to the new version once it becomes available for clinical use.\n\nTransmission of this Notice\nThis notice needs to be passed on to all those who need to be aware within your organization. Maintain awareness of this notice as long as any affected version is in use.",
      "product_quantity": "10770 systems",
      "distribution_pattern": "US Nationwide",
      "openfda": {
        "k_number": [
          "K102216",
          "K013535",
          "K090531",
          "K093588",
          "K151380",
          "K220582",
          "K181145",
          "K102011",
          "K171885",
          "K083672",
          "K091146",
          "K103246",
          "K121448",
          "K040688",
          "K991815",
          "K100056",
          "K063482",
          "K112723",
          "K130617",
          "K013112",
          "K242378",
          "K984532",
          "K202789",
          "K190178",
          "K011246",
          "K220783",
          "K180379",
          "K030821",
          "K162370",
          "K222312",
          "K032456",
          "K090808",
          "K082606",
          "K133655",
          "K092928",
          "K082005",
          "K093391",
          "K103146",
          "K141860",
          "K203681",
          "K992762",
          "K193640",
          "K994413",
          "K032345",
          "K180308",
          "K170086",
          "K172163",
          "K031349",
          "K151560",
          "K252784",
          "K213438",
          "K212274",
          "K160093",
          "K182665",
          "K200569",
          "K002240",
          "K974770",
          "K063230",
          "K240398",
          "K080601",
          "K133331",
          "K120387",
          "K253078",
          "K240671",
          "K042861",
          "K010682",
          "K253091",
          "K091145",
          "K000922",
          "K984196",
          "K082642",
          "K241876",
          "K013644",
          "K091492",
          "K173635",
          "K061893",
          "K083502",
          "K171536",
          "K241318",
          "K022411",
          "K041403",
          "K100417",
          "K250064",
          "K022961",
          "K983343",
          "K231825",
          "K151887",
          "K990475",
          "K992303",
          "K041971",
          "K193109",
          "K211511",
          "K012575",
          "K250440",
          "K101076",
          "K242748",
          "K032209",
          "K090182",
          "K111311",
          "K192554",
          "K013661",
          "K992864",
          "K001295",
          "K993895",
          "K182888",
          "K020971",
          "K150547",
          "K152488",
          "K260207",
          "K242463",
          "K143269",
          "K142108",
          "K040474",
          "K232400",
          "K151435",
          "K993923",
          "K012017",
          "K200487",
          "K042739",
          "K141283",
          "K223279",
          "K100552",
          "K051941",
          "K083264",
          "K120551",
          "K210645",
          "K170750",
          "K211867",
          "K153368",
          "K013673",
          "K211034",
          "K201350",
          "K981678",
          "K042269",
          "K260308",
          "K173940",
          "K123357",
          "K183037",
          "K211841",
          "K002132",
          "K250963",
          "K252109",
          "K203766",
          "K214081",
          "K151469",
          "K170969",
          "K180289",
          "K052325",
          "K103093",
          "K252002",
          "K091179",
          "K053584",
          "K123152",
          "K080929",
          "K193459",
          "K090893",
          "K141900",
          "K022036",
          "K041719",
          "K132816",
          "K022741",
          "K041997",
          "K110730",
          "K234068",
          "K081112",
          "K982821",
          "K062032",
          "K222803",
          "K143514",
          "K071760",
          "K002312",
          "K140187",
          "K061440",
          "K102915",
          "K220141",
          "K991237",
          "K131891",
          "K915691",
          "K061097",
          "K002147",
          "K151666",
          "K001700",
          "K211881",
          "K233236",
          "K230557",
          "K103515",
          "K020027",
          "K222728",
          "K031088",
          "K022307",
          "K082221",
          "K071873",
          "K002509",
          "K223233",
          "K252304",
          "K990071",
          "K032724",
          "K242822",
          "K071938",
          "K062967",
          "K982791",
          "K151369",
          "K132971",
          "K992753",
          "K190387",
          "K152393",
          "K202284",
          "K112446",
          "K242992",
          "K051497",
          "K203469",
          "K123263",
          "K243057",
          "K260750",
          "K133247",
          "K141248",
          "K223724",
          "K092132",
          "K083414",
          "K090972",
          "K010784",
          "K050296",
          "K232799",
          "K161136",
          "K031281",
          "K253050",
          "K151233",
          "K131428",
          "K993766",
          "K081281",
          "K201607",
          "K113102",
          "K120363",
          "K993675",
          "K200608",
          "K232479",
          "K982696",
          "K062514",
          "K992434",
          "K190664",
          "K193381",
          "K192065",
          "K101220",
          "K252863",
          "K073020",
          "K112060",
          "K211379",
          "K002237",
          "K080888",
          "K112776",
          "K101119",
          "K021268",
          "K100092",
          "K170355",
          "K213787",
          "K191384",
          "K201444"
        ],
        "registration_number": [
          "3022719374",
          "3033527281",
          "8043933",
          "3010123525",
          "3013410552",
          "3008156050",
          "3009302839",
          "3015777306",
          "3000126629",
          "3006894636",
          "3023344262",
          "3015232217",
          "3003793371",
          "1038814",
          "3014425747",
          "3006579682",
          "3010034862",
          "3030564015",
          "3004526608",
          "3009529630",
          "3016602874",
          "3012039133",
          "2030598",
          "3011716550",
          "3003418325",
          "3044735888",
          "3003094912",
          "3025765259",
          "3005136058",
          "3019834893",
          "3002808157",
          "3016686005",
          "3004363352",
          "2183553",
          "3014873236",
          "3007037519",
          "3007774465",
          "9617978",
          "3022745617",
          "3004608878",
          "3004555507",
          "1651102",
          "9611894",
          "3004561814",
          "2434141",
          "3009077524",
          "2916710",
          "3003873069",
          "3004580128",
          "3042177665",
          "3003768277",
          "3009090417",
          "3018004899",
          "3004977335",
          "3003586672",
          "3019543601",
          "2939248",
          "3007791541"
        ],
        "fei_number": [
          "3002802883",
          "3022719374",
          "3033527281",
          "3010123525",
          "3013410552",
          "3008156050",
          "3009302839",
          "3015777306",
          "3000126629",
          "3006894636",
          "3023344262",
          "3015232217",
          "3003793371",
          "3008581796",
          "1038814",
          "3014425747",
          "3006579682",
          "3010034862",
          "3030564015",
          "3004526608",
          "3009529630",
          "3016602874",
          "3012039133",
          "3011716550",
          "3003418325",
          "3044735888",
          "3003094912",
          "3025765259",
          "3005136058",
          "3019834893",
          "3002808157",
          "3002619595",
          "3016686005",
          "3004363352",
          "2183553",
          "3014873236",
          "3007037519",
          "3007774465",
          "3022745617",
          "3004608878",
          "3004555507",
          "3004561814",
          "2434141",
          "1000518731",
          "3009077524",
          "3003873069",
          "2916710",
          "3004580128",
          "3042177665",
          "3009090417",
          "3002807880",
          "3018004899",
          "3004977335",
          "3003586672",
          "3019543601",
          "2939248",
          "3002807741",
          "3007791541"
        ],
        "device_name": "System, Planning, Radiation Therapy Treatment",
        "medical_specialty_description": "Radiology",
        "regulation_number": "892.5050",
        "device_class": "2"
      }
    }
  ]
}