  "results": [
    {
      "cfres_id": "198492",
      "product_res_number": "Z-1249-2023",
      "event_date_initiated": "2023-02-01",
      "event_date_posted": "2023-03-13",
      "recall_status": "Open, Classified",
      "res_event_number": "91676",
      "product_code": "DQY",
      "k_numbers": [
        "K210601"
      ],
      "product_description": "Plato 17 Microcatheter: Straight Tip, PL17-160-000 ; 45 degree Tip, PL17-160-045; 90 degree Tip, PL17-160-090",
      "code_info": "All lots, REF/UDI-DI/Lot(Expiration): \nPL17-160-000/00818075010455/ 900164(8/31/2022), 900165(8/31/2022), 900172(8/31/2022), 900177(9/30/2022), 900180(9/30/2022), 900186(10/31/2022), 900187(10/31/2022), 900188(10/31/2022), 900193(10/31/2022), 900196(12/31/2022), 900199(12/31/2022), 900202(12/31/2022), 900221(2/28/2023), 900222(2/28/2023), 900225(2/28/2023), 900226(3/31/2023), 20441(11/30/2023), 20476(11/30/2023), 20650(12/31/2023);\n\nPL17-160-045/00818075010462/900167(8/31/2022), 900168(8/31/2022), 900169(8/31/2022), 900174(8/31/2022), 900175(8/31/2022), 900179(9/30/2022), 900181(9/30/2022), 900183(10/31/2022), 900185(10/31/2022),  900205(1/31/2023), 900206(1/31/2023), 900210(1/31/2023), 900211(1/31/2023), 900215(1/31/2023), 900220(1/31/2023), 900219(1/31/2023), 900217(2/28/2023), 900223(2/28/2023), 900224(2/28/2023);\n\nPL17-160-090/00818075010479/900170(8/31/2022), 900171(8/31/2022), 900178(9/30/2022), 900182(9/30/2022), 900184(10/31/2022), 900192(11/30/2022), 900195(11/30/2022), 900209(1/31/2023), 900228(3/31/2023)",
      "firm_fei_number": "3009546466",
      "recalling_firm": "Scientia Vascular, Inc.",
      "address_1": "2460 S 3270 W",
      "city": "West Valley City",
      "state": "UT",
      "postal_code": "84119-1116",
      "additional_info_contact": "888-385-9016",
      "reason_for_recall": "Microcatheters for embolic coil and diagnostic agent introduction to peripheral/neuro vasculature, due to manufacturing non-conformance, and due to flush, guidewire insertion and manipulation prior to placement, may have loose particles within catheter lumens that may be introduced to patients, which could cause embolic event, small vessel occlusion, and subacute to chronic inflammatory response.",
      "root_cause_description": "Device Design",
      "action": "On 02/01/23, recall notices were mailed and emailed to customers who were told the following:\n\nActions for Customer/User: Identify and segregate affected devices.\nActions for Recalling Firm: Sales representatives will visit and assist with completing the acknowledgement and receipt form and the collection of affected devices. If devices have been transferred from one facility to another the additional consignees will be contacted upon identification.  \n\nSales reps received the recall notice and a response form they would take to customers to obtain additional information and the affected devices.\n\nCustomers with questions can contact customer support at 1-888-385-9016 9016 between the hours of 9:30 AM and 6:30 PM ET (7:30 AM and 4:30 PM MST), Monday through Friday, or via email to regulatory@scientiavascular.com",
      "product_quantity": "353",
      "distribution_pattern": "US Nationwide distribution in the states of IL, AL, FL, NC, WI, GA, SC, IN, KS, CA, OH, NY, NH, CO, MA, NV, UT, TX.",
      "openfda": {
        "k_number": [
          "K832784",
          "K180088",
          "K042910",
          "K992622",
          "K082337",
          "K220331",
          "K021899",
          "K895753",
          "K101542",
          "K122581",
          "K013517",
          "K141354",
          "K003410",
          "K152456",
          "K143459",
          "K203521",
          "K023402",
          "K840240",
          "K900978",
          "K130257",
          "K813514",
          "K853714",
          "K152447",
          "K862975",
          "K030779",
          "K262240",
          "K863684",
          "K974683",
          "K092298",
          "K083422",
          "K192135",
          "K212738",
          "K163259",
          "K955651",
          "K934685",
          "K952185",
          "K250219",
          "K894084",
          "K884491",
          "K961280",
          "K900998",
          "K152249",
          "K053116",
          "K090822",
          "K022552",
          "K083125",
          "K092299",
          "K032869",
          "K121051",
          "K223633",
          "K072750",
          "K965247",
          "K083127",
          "K935292",
          "K093476",
          "K111850",
          "K895932",
          "K120707",
          "K110891",
          "K063252",
          "K100826",
          "K902587",
          "K913830",
          "K091527",
          "K052257",
          "K973371",
          "K183305",
          "K122772",
          "K003880",
          "K931403",
          "K052004",
          "K193322",
          "K080757",
          "K192774",
          "K131204",
          "K842084",
          "K010831",
          "K042538",
          "K962362",
          "K133026",
          "K820009",
          "K024032",
          "K043489",
          "K915521",
          "K091254",
          "K042568",
          "K080583",
          "K902999",
          "K112035",
          "K023223",
          "K950635",
          "K120585",
          "K083657",
          "K141981",
          "K833303",
          "K062420",
          "K990707",
          "K833433",
          "K231311",
          "K974684",
          "K251277",
          "K031766",
          "K870675",
          "K173662",
          "K914856",
          "K190393",
          "K950047",
          "K033320",
          "K142065",
          "K261825",
          "K222037",
          "K082932",
          "K061843",
          "K243691",
          "K062594",
          "K032737",
          "K112032",
          "K020161",
          "K161126",
          "K182487",
          "K110540",
          "K001577",
          "K963663",
          "K820736",
          "K170914",
          "K140307",
          "K094019",
          "K001856",
          "K884209",
          "K062084",
          "K121077",
          "K121352",
          "K080994",
          "K150201",
          "K905547",
          "K243224",
          "K833801",
          "K912199",
          "K201271",
          "K082290",
          "K971572",
          "K040700",
          "K133501",
          "K120120",
          "K141547",
          "K151409",
          "K231608",
          "K212725",
          "K070238",
          "K081130",
          "K100135",
          "K162253",
          "K261766",
          "K832064",
          "K862117",
          "K261631",
          "K853997",
          "K883564",
          "K152876",
          "K914984",
          "K051580",
          "K150865",
          "K213466",
          "K041151",
          "K090335",
          "K790083",
          "K173709",
          "K922611",
          "K221750",
          "K994229",
          "K981359",
          "K132197",
          "K140910",
          "K926421",
          "K092075",
          "K954251",
          "K062885",
          "K111252",
          "K042388",
          "K250492",
          "K072283",
          "K193607",
          "K853860",
          "K911747",
          "K982770",
          "K192996",
          "K122940",
          "K120178",
          "K890152",
          "K043534",
          "K033875",
          "K180169",
          "K102979",
          "K852099",
          "K872182",
          "K150970",
          "K903158",
          "K962830",
          "K191560",
          "K861827",
          "K950179",
          "K981191",
          "K021587",
          "K080988",
          "K903216",
          "K111556",
          "K111334",
          "K102439",
          "K970401",
          "K982883",
          "K121763",
          "K120964",
          "K231246",
          "K103250",
          "K875269",
          "K240885",
          "K992673",
          "K943622",
          "K161362",
          "K233350",
          "K130525",
          "K192625",
          "K162384",
          "K231600",
          "K090918",
          "K161901",
          "K130378",
          "K050599",
          "K142449",
          "K864118",
          "K093303",
          "K082933",
          "K132556",
          "K212026",
          "K900657",
          "K253409",
          "K221846",
          "K110483",
          "K131592",
          "K220691",
          "K170041",
          "K113112",
          "K053491",
          "K250962",
          "K962717",
          "K171612",
          "K132914",
          "K160389",
          "K042541",
          "K865074",
          "K130649",
          "K190475",
          "K232647",
          "K112482",
          "K140034",
          "K834109",
          "K131063",
          "K112917",
          "K010954",
          "K022138",
          "K053118",
          "K861165",
          "K151396",
          "K011889",
          "K843263",
          "K912564",
          "K260289",
          "K870075",
          "K130252",
          "K003920",
          "K210230",
          "K852791",
          "K083313",
          "K903330",
          "K262392",
          "K860699",
          "K161986",
          "K905249",
          "K130858",
          "K212211",
          "K152202",
          "K123452",
          "K152401",
          "K101117",
          "K112886",
          "K231044",
          "K080360",
          "K933033",
          "K250410",
          "K842176",
          "K151082",
          "K021088",
          "K912877",
          "K063638",
          "K840761",
          "K022061",
          "K203275",
          "K092396",
          "K103386",
          "K010880",
          "K841828",
          "K031692",
          "K031505",
          "K943783",
          "K924591",
          "K170996",
          "K091793",
          "K932619",
          "K162467",
          "K120229",
          "K854984",
          "K182937",
          "K945705",
          "K060927",
          "K052791",
          "K051838",
          "K162359",
          "K980278",
          "K863791",
          "K230156",
          "K810962",
          "K042850",
          "K260993",
          "K182097",
          "K230363",
          "K014117",
          "K871425",
          "K042381",
          "K091345",
          "K121219",
          "K021284",
          "K042624",
          "K060745",
          "K885237",
          "K123531",
          "K151667",
          "K182360",
          "K223728",
          "K934988",
          "K032312",
          "K820648",
          "K062609",
          "K091624",
          "K180201",
          "K190581",
          "K032069",
          "K012002",
          "K172468",
          "K012913",
          "K191551",
          "K831807",
          "K101885",
          "K111839",
          "K873485",
          "K211990",
          "K950490",
          "K062809",
          "K210322",
          "K171176",
          "K873035",
          "K220632",
          "K921311",
          "K102657",
          "K161546",
          "K193034",
          "K173560",
          "K101282",
          "K121859",
          "K180864",
          "K013789",
          "K013581",
          "K193119",
          "K820516",
          "K852388",
          "K101877",
          "K260499",
          "K211251",
          "K061416",
          "K090913",
          "K212228",
          "K010531",
          "K082579",
          "K091299",
          "K122684",
          "K201811",
          "K182512",
          "K222115",
          "K001969",
          "K042117",
          "K082646",
          "K832178",
          "K894131",
          "K870067",
          "K123153",
          "K113778",
          "K052236",
          "K072313",
          "K210110",
          "K831729",
          "K122394",
          "K021721",
          "K851241",
          "K882969",
          "K822393",
          "K160681",
          "K060807",
          "K180959",
          "K841797",
          "K050073",
          "K014118",
          "K941137",
          "K170403",
          "K062868",
          "K112281",
          "K210188",
          "K872163",
          "K965211",
          "K013963",
          "K952345",
          "K183493",
          "K252417",
          "K030476",
          "K210009",
          "K181563",
          "K193420",
          "K250147",
          "K020028",
          "K953909",
          "K904404",
          "K862056",
          "K112335",
          "K082868",
          "K193012",
          "K250828",
          "K083596",
          "K210431",
          "K093977",
          "K870411",
          "K220647",
          "K922609",
          "K031211",
          "K861008",
          "K042553",
          "K132659",
          "K141529",
          "K173548",
          "K953265",
          "K915777",
          "K040922",
          "K833573",
          "K214107",
          "K021589",
          "K233637",
          "K112262",
          "K163000",
          "K190401",
          "K021256",
          "K082670",
          "K013638",
          "K040481",
          "K982632",
          "K132701",
          "K840336",
          "K112579",
          "K131978",
          "K150107",
          "K101659",
          "K231629",
          "K101591",
          "K121839",
          "K123765",
          "K201726",
          "K192712",
          "K010853",
          "K854610",
          "K903217",
          "K821480",
          "K091841",
          "K232437",
          "K222023",
          "K011464",
          "K142565",
          "K874751",
          "K092174",
          "K021455",
          "K840762",
          "K113589",
          "K920331",
          "K991601",
          "K052258",
          "K173799",
          "K131225",
          "K182570",
          "K103534",
          "K861829",
          "K042230",
          "K001136",
          "K870412",
          "K022722",
          "K242644",
          "K151981",
          "K951159",
          "K241611",
          "K183265",
          "K032931",
          "K232690",
          "K821189",
          "K991059",
          "K102725",
          "K070051",
          "K040243",
          "K905720",
          "K171146",
          "K080849",
          "K192296",
          "K214024",
          "K240018",
          "K911869",
          "K032298",
          "K042489",
          "K232190",
          "K162563",
          "K100184",
          "K934009",
          "K200721",
          "K852815",
          "K061601",
          "K031459",
          "K980453",
          "K897082",
          "K033540",
          "K142014",
          "K003085",
          "K083833",
          "K171913",
          "K862724",
          "K925439",
          "K952205",
          "K052149",
          "K032587",
          "K852914",
          "K081878",
          "K243436",
          "K904884",
          "K220716",
          "K183353",
          "K150894",
          "K152957",
          "K242276",
          "K072329",
          "K863322",
          "K895440",
          "K033622",
          "K871573",
          "K953987",
          "K043387",
          "K153703",
          "K122917",
          "K132542",
          "K913517",
          "K871080",
          "K060327",
          "K030630",
          "K033633",
          "K030589",
          "K073700",
          "K024035",
          "K923245",
          "K110055",
          "K950094",
          "K251014",
          "K040771",
          "K190431",
          "K923518",
          "K914422",
          "K080987",
          "K001435",
          "K000677",
          "K080428",
          "K153729",
          "K110813",
          "K033217",
          "K021120",
          "K874874",
          "K192525",
          "K915374",
          "K001694",
          "K911527",
          "K853998",
          "K234115",
          "K082385",
          "K023463",
          "K180919",
          "K010579",
          "K861828",
          "K151821",
          "K950634",
          "K151896",
          "K180797",
          "K111738",
          "K152840",
          "K172014",
          "K810461",
          "K161262",
          "K010169",
          "K120296",
          "K972978",
          "K834232",
          "K913052",
          "K132409",
          "K061480",
          "K902905",
          "K250558",
          "K142239",
          "K912777",
          "K083214",
          "K120129",
          "K940545",
          "K101015",
          "K894105",
          "K840175",
          "K222603",
          "K252390",
          "K190749",
          "K030671",
          "K030715",
          "K182420",
          "K143608",
          "K863564",
          "K014212",
          "K833873",
          "K012169",
          "K180889",
          "K143744",
          "K183463",
          "K881504",
          "K150366",
          "K902256",
          "K944537",
          "K972484",
          "K093969",
          "K250972",
          "K071665",
          "K944668",
          "K000715",
          "K090613",
          "K110461",
          "K171672",
          "K243184",
          "K190833",
          "K001992",
          "K180986",
          "K180715",
          "K241177",
          "K023548",
          "K260559",
          "K111704",
          "K083919",
          "K123324",
          "K943777",
          "K021802",
          "K831563",
          "K163314",
          "K222873",
          "K930476",
          "K120781",
          "K003495",
          "K211240",
          "K133539",
          "K841587",
          "K032784",
          "K042537",
          "K010838",
          "K182829",
          "K935738",
          "K013601",
          "K160561",
          "K083139",
          "K133177",
          "K072618",
          "K241538",
          "K940559",
          "K024327",
          "K191237",
          "K940603",
          "K121611",
          "K162565",
          "K191717",
          "K082339",
          "K984482",
          "K863790",
          "K231176",
          "K013284",
          "K953515",
          "K883916",
          "K112404",
          "K852491",
          "K171653",
          "K192782",
          "K131493",
          "K940187",
          "K010874",
          "K911333",
          "K012050",
          "K041093",
          "K963962",
          "K112828",
          "K012130",
          "K200324",
          "K884129",
          "K061710",
          "K182918",
          "K946272",
          "K082561",
          "K850353",
          "K101647",
          "K051846",
          "K140557",
          "K090659",
          "K050389",
          "K070126",
          "K142321",
          "K210647",
          "K132673",
          "K171268",
          "K910116",
          "K850829",
          "K842316",
          "K894682",
          "K091750",
          "K863916",
          "K112629",
          "K061119",
          "K111380",
          "K933019",
          "K183000",
          "K882803",
          "K093184",
          "K946204",
          "K132420",
          "K200094",
          "K053217",
          "K051343",
          "K012083",
          "K131492",
          "K111750",
          "K102022",
          "K172090",
          "K083343",
          "K000598",
          "K073378",
          "K111154",
          "K170664",
          "K241158",
          "K212588",
          "K082559",
          "K123311",
          "K143412",
          "K221886",
          "K081799",
          "K193596",
          "K243757",
          "K810460",
          "K191163",
          "K112100",
          "K163233",
          "K191664",
          "K172297",
          "K093160",
          "K233858",
          "K935425",
          "K904608",
          "K061680",
          "K081680",
          "K073289",
          "K101479",
          "K052036",
          "K122826",
          "K140080",
          "K242588",
          "K060655",
          "K970823",
          "K171819",
          "K023320",
          "K080154",
          "K833825",
          "K160848",
          "K021593",
          "K032591",
          "K970668",
          "K181000",
          "K972065",
          "K950026",
          "K082524",
          "K895972",
          "K022738",
          "K981413",
          "K912191",
          "K200317",
          "K831636",
          "K860339",
          "K033441",
          "K053431",
          "K881793",
          "K042194",
          "K003643",
          "K891069",
          "K913783",
          "K090509",
          "K033989",
          "K090849",
          "K033670",
          "K072316",
          "K090977",
          "K253361",
          "K053277",
          "K060418",
          "K242873",
          "K220340",
          "K022764",
          "K855201",
          "K130714",
          "K191229",
          "K131482",
          "K151638",
          "K162228",
          "K032622",
          "K990726",
          "K925131",
          "K831717",
          "K072156",
          "K011562",
          "K935597",
          "K854137",
          "K883174",
          "K193474",
          "K062202",
          "K082854",
          "K240384",
          "K182602",
          "K802287",
          "K192558",
          "K021735",
          "K223178",
          "K173661",
          "K112082",
          "K112303",
          "K051772",
          "K171335",
          "K040745",
          "K250426",
          "K093050",
          "K142235",
          "K911519",
          "K070372",
          "K160448",
          "K161152",
          "K210601",
          "K932933",
          "K093431",
          "K032660",
          "K103367",
          "K070970",
          "K101776",
          "K981198",
          "K051060",
          "K030984",
          "K102418",
          "K202610",
          "K973298",
          "K132652",
          "K854000",
          "K151880",
          "K243180",
          "K240787",
          "K842645",
          "K033678",
          "K060572",
          "K043580",
          "K151103",
          "K171452",
          "K120381",
          "K091188",
          "K242337",
          "K130850",
          "K070502",
          "K811971",
          "K031417",
          "K182844",
          "K214106",
          "K141649",
          "K050038",
          "K951603",
          "K111963",
          "K202252"
        ],
        "registration_number": [
          "3010882065",
          "3006131984",
          "3006451981",
          "3039020025",
          "3005255160",
          "3016678045",
          "3014687026",
          "3012931345",
          "3027319104",
          "8023054",
          "2518902",
          "1721504",
          "3008814558",
          "3009018440",
          "3026788751",
          "3024801436",
          "2182208",
          "1820334",
          "1625425",
          "3012494290",
          "3016948450",
          "3013556777",
          "3015053858",
          "9710641",
          "2032112",
          "3008853977",
          "1721676",
          "3003955307",
          "1047843",
          "2030598",
          "2024168",
          "3008423561",
          "3005070406",
          "3013858269",
          "3015970743",
          "1721686",
          "2135147",
          "3003910212",
          "2027111",
          "3000308637",
          "3010155661",
          "3004718255",
          "3014590708",
          "1028232",
          "3016119728",
          "3032916632",
          "3012448339",
          "3009756153",
          "3030236028",
          "3003915875",
          "3004415014",
          "2024311",
          "3008881809",
          "3012279212",
          "1048735",
          "3015002281",
          "9680794",
          "3006082230",
          "3008264254",
          "3022392708",
          "3016522967",
          "3005168196",
          "3008114965",
          "2521402",
          "1220452",
          "2939561",
          "3008110587",
          "3009905888",
          "3004063494",
          "3016540856",
          "2017233",
          "3013758550",
          "9617601",
          "3003610571",
          "3030733800",
          "3011706110",
          "3032277090",
          "3005580113",
          "3012159165",
          "3047894544",
          "1318694",
          "3000247873",
          "2126666",
          "3044156918",
          "1016427",
          "3019807891",
          "3011088743",
          "3010273872",
          "2020394",
          "3012179728",
          "3031311812",
          "2124215",
          "3027058844",
          "1225687",
          "9681260",
          "3020347218",
          "2183744",
          "1118880",
          "3005635992",
          "3008573567",
          "3015615738",
          "3014579161",
          "1061124",
          "1531050",
          "3004859241",
          "3033589330",
          "1646831",
          "3017708755",
          "2029214",
          "3007740583",
          "3007882731",
          "2183787",
          "3020677060",
          "3004784537",
          "3009756327",
          "3020954126",
          "2953200",
          "1724474",
          "8010026",
          "3014779787",
          "2528981",
          "3010665433",
          "3012543881",
          "1313046",
          "3010421104",
          "3007695959",
          "3008361782",
          "3010909876",
          "1722746",
          "3008789872",
          "3012154226",
          "3017210488",
          "3006942602",
          "3046434506",
          "3008700817",
          "3008824097",
          "9681477",
          "1828132",
          "3025028711",
          "3007113487",
          "3004111573",
          "1319639",
          "3009121749",
          "2017865",
          "1526863",
          "3004091615",
          "3020950818",
          "3013944123",
          "3015309643",
          "3015550451",
          "2648729",
          "3013134719",
          "1058584",
          "3010291427",
          "1647149",
          "3032047719",
          "3009564766",
          "3020954746",
          "3007007790",
          "1064858",
          "3010370269",
          "2024024",
          "3009144177",
          "3009051888",
          "3006946276",
          "3005994106",
          "3010419931",
          "3009965866",
          "3000268902",
          "3007738736",
          "9616099",
          "3039062605",
          "3001374820",
          "3010079067",
          "1643817",
          "3005210579",
          "2134812",
          "3007284313",
          "9612164",
          "1220477",
          "3007284006",
          "2020735",
          "9610139",
          "3013557562",
          "3015060261",
          "3035111606",
          "3004742232",
          "3003775186",
          "9616684",
          "1018233",
          "3012050423",
          "3003775027",
          "3008627763",
          "3009546466",
          "3008261722",
          "3030088112",
          "3013162291",
          "1526534",
          "3002806603",
          "3014498720",
          "1450662",
          "2183870",
          "3010173425",
          "3016829855",
          "3003898360",
          "3016444913",
          "3038912532",
          "3013288201",
          "3007207906",
          "2027062",
          "3009766315",
          "3025341484",
          "3001124136",
          "3016727301",
          "1123137",
          "2011171",
          "3015225571",
          "3015453963",
          "3009380063",
          "3017262662",
          "2184009",
          "1000563940",
          "3005718570",
          "3031768596",
          "9617592",
          "3006260740",
          "9681834",
          "1000121056",
          "3005232773",
          "3002807314",
          "9616666",
          "3017168767",
          "3019751610",
          "3014766734",
          "3006550224",
          "3010041511",
          "3003780911",
          "3014461810",
          "3006425876",
          "2134265",
          "2523676",
          "3014211783",
          "1928237",
          "3005334138",
          "3014162263",
          "3016171942",
          "3012497308",
          "3009888344",
          "3011137372",
          "3010432890",
          "3012102437",
          "9618000",
          "3008837339",
          "3014656749",
          "3006950086",
          "2648045",
          "3000126629",
          "3005012805",
          "8043892",
          "3007628272",
          "3005941719",
          "3002807090",
          "3008307705",
          "2025870",
          "3013764800",
          "3016591327",
          "3027343398",
          "1018470",
          "2030624",
          "3021951050",
          "3012536737",
          "3016704395",
          "2649622",
          "2029275",
          "3011142981",
          "3017278832",
          "3014644549",
          "3012266672",
          "3015859709",
          "9614654",
          "1061927"
        ],
        "fei_number": [
          "3006131984",
          "3006451981",
          "3039020025",
          "3005255160",
          "3016678045",
          "3014687026",
          "3012931345",
          "1000138054",
          "3027319104",
          "1000306139",
          "3003565456",
          "2518902",
          "3002827704",
          "1721504",
          "3008814558",
          "3009018440",
          "3000203000",
          "3026788751",
          "3024801436",
          "1820334",
          "1625425",
          "3012494290",
          "3016948450",
          "3013556777",
          "2182208",
          "1000116158",
          "3015053858",
          "3008853977",
          "1721676",
          "3003955307",
          "2828",
          "3002468371",
          "3002808050",
          "1000160215",
          "1047843",
          "3008397795",
          "2024168",
          "3008423561",
          "3004635528",
          "3005070406",
          "1000306431",
          "3013858269",
          "3015970743",
          "1721686",
          "1000512168",
          "1000518731",
          "3003910212",
          "2027111",
          "3000308637",
          "3010155661",
          "3004718255",
          "3014590708",
          "1028232",
          "3016119728",
          "3032916632",
          "3012448339",
          "3009756153",
          "3030236028",
          "3003915875",
          "3004415014",
          "2024311",
          "3008881809",
          "3012279212",
          "1048735",
          "3015002281",
          "3006082230",
          "3008264254",
          "3022392708",
          "3016522967",
          "3002095335",
          "3005168196",
          "3008114965",
          "2521402",
          "1220452",
          "3008110587",
          "3009905888",
          "3004063494",
          "3016540856",
          "2017233",
          "3013758550",
          "3003610571",
          "3002808904",
          "3030733800",
          "3006097687",
          "3011706110",
          "3032277090",
          "3005580113",
          "3012159165",
          "3047894544",
          "1318694",
          "3000247873",
          "2126666",
          "1000306216",
          "3044156918",
          "1016427",
          "3019807891",
          "3011088743",
          "3001452571",
          "3010273872",
          "2020394",
          "3002913962",
          "3012179728",
          "3031311812",
          "2124215",
          "3027058844",
          "3020347218",
          "2183744",
          "1118880",
          "3005635992",
          "3008573567",
          "3005747797",
          "3015615738",
          "3014579161",
          "3033589330",
          "3004859241",
          "3017708755",
          "3007740583",
          "3007882731",
          "2183787",
          "3002808327",
          "1000123180",
          "3020677060",
          "3004784537",
          "3020954126",
          "3014779787",
          "2528981",
          "3010665433",
          "3012543881",
          "1313046",
          "3010421104",
          "3007695959",
          "3008361782",
          "3010909876",
          "3001451463",
          "3002807439",
          "3008789872",
          "3012154226",
          "3017210488",
          "3006942602",
          "3046434506",
          "3008700817",
          "3003094851",
          "3008824097",
          "3002806500",
          "1828132",
          "3025028711",
          "3007113487",
          "3004111573",
          "3009121749",
          "1526863",
          "3004091615",
          "3020950818",
          "3013944123",
          "3015309643",
          "3015550451",
          "3002807111",
          "3004216185",
          "2648729",
          "3013134719",
          "3000206435",
          "3010291427",
          "3032047719",
          "3009564766",
          "3020954746",
          "3007007790",
          "3010370269",
          "2024024",
          "3009144177",
          "1000220862",
          "3009051888",
          "3006946276",
          "3005994106",
          "3010419931",
          "3003076831",
          "3009965866",
          "3000268902",
          "3007738736",
          "3039062605",
          "3001374820",
          "1000219637",
          "3010079067",
          "3002806565",
          "3005210579",
          "1643817",
          "3007284313",
          "3003245778",
          "1220477",
          "3007284006",
          "2020735",
          "3013557562",
          "3015060261",
          "3035111606",
          "3004742232",
          "3003775186",
          "1018233",
          "3012050423",
          "3003775027",
          "3008906816",
          "3008627763",
          "3009546466",
          "3008261722",
          "3030088112",
          "3013162291",
          "1526534",
          "3003737899",
          "3002806603",
          "3014498720",
          "1450662",
          "2183870",
          "3016829855",
          "3003898360",
          "3016444913",
          "3038912532",
          "3013288201",
          "3007207906",
          "2027062",
          "3009766315",
          "3002807036",
          "3025341484",
          "3001124136",
          "3016727301",
          "3015225571",
          "2011171",
          "3002534070",
          "3015453963",
          "3009380063",
          "3017262662",
          "1000563940",
          "2017865",
          "3005718570",
          "3031768596",
          "3006260740",
          "3006003617",
          "1000121056",
          "3002755653",
          "3005232773",
          "3002807314",
          "3017168767",
          "3014766734",
          "3019751610",
          "3006550224",
          "3010041511",
          "3003780911",
          "3014461810",
          "3006425876",
          "2523676",
          "3014211783",
          "1928237",
          "3005334138",
          "3014162263",
          "3016171942",
          "3012497308",
          "3009888344",
          "3012102437",
          "3010432890",
          "3008837339",
          "3014656749",
          "2648045",
          "3006950086",
          "3000126629",
          "3005012805",
          "3007628272",
          "3003528016",
          "3008307705",
          "2025870",
          "3013764800",
          "3016591327",
          "3027343398",
          "1018470",
          "3021951050",
          "3012536737",
          "3016704395",
          "3002806434",
          "2649622",
          "3002807038",
          "3011142981",
          "3017278832",
          "3014644549",
          "3012266672",
          "3015859709"
        ],
        "device_name": "Catheter, Percutaneous",
        "medical_specialty_description": "Cardiovascular",
        "regulation_number": "870.1250",
        "device_class": "2"
      }
    }
  ]
}