{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-17",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "197576",
      "product_res_number": "Z-1018-2023",
      "event_date_initiated": "2022-12-15",
      "event_date_posted": "2023-01-26",
      "recall_status": "Open, Classified",
      "res_event_number": "91365",
      "product_code": "HRY",
      "k_numbers": [
        "K133940"
      ],
      "product_description": "Oxford Partial Knee System, Fixed Lateral Tibial Construct- Intended to replace part of a knee joint when used in conjunction with a femoral and tibial component\nItem Number: 154341",
      "code_info": "UDI: (01) 05019279515349 (17) 241123 (10) 077830\nLot Number: 077830",
      "firm_fei_number": "1825034",
      "recalling_firm": "Biomet, Inc.",
      "address_1": "56 E Bell Dr",
      "city": "Warsaw",
      "state": "IN",
      "postal_code": "46582-6989",
      "additional_info_contact": "411 Technical Services\n574-371-3071",
      "reason_for_recall": "Packaged in the incorrect shelf carton. \nIncorrect shelf carton; outer carton is smaller than the required outer carton, which forces the Tyvex pouch\nto be folded and packaged in the carton causing compound creases and compromise the sterile barrier",
      "root_cause_description": "Package design/selection",
      "action": "Zimmer Biomet issued Urgent Medical Device Recall Letter on 12/15/22 to Distributors and Risk Managers/Surgeons. Letter states reason for recall, health risk and action to take:\nReview this notification and ensure that affected personnel are aware of the contents.\n2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility.\n3. If the product has been further distributed, provide your customers with the notice for hospitals and ensure documentation.\n4. Complete Attachment 1   Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form shall be returned even if you do not have affected products at your facility.\n5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation.\n6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com.",
      "product_quantity": "7 units",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of IN, KY and the countries of Canada, New Zealand, Japan, Netherlands.",
      "openfda": {
        "k_number": [
          "K040268",
          "K042093",
          "K022437",
          "K050764",
          "K982859",
          "K810230",
          "K821341",
          "K193549",
          "K834298",
          "K072704",
          "K933935",
          "K001451",
          "K790348",
          "K983657",
          "K190633",
          "K020741",
          "K862125",
          "K030193",
          "K012591",
          "K792619",
          "K122277",
          "K020264",
          "K010810",
          "K940389",
          "K152631",
          "K790350",
          "K961900",
          "K033351",
          "K955778",
          "K831143",
          "K760245",
          "K884837",
          "K061681",
          "K033810",
          "K011509",
          "K810060",
          "K864070",
          "K030446",
          "K081293",
          "K063515",
          "K902538",
          "K071881",
          "K842367",
          "K820030",
          "K780299",
          "K761278",
          "K974434",
          "K133940",
          "K820002",
          "K032052",
          "K011138",
          "K810616",
          "K883969",
          "K060412",
          "K911373",
          "K810540",
          "K012762",
          "K082567",
          "K960617",
          "K812012",
          "K053299",
          "K890044",
          "K051721",
          "K790762",
          "K040889",
          "K971938",
          "K760744",
          "K021621",
          "K014171",
          "K923011",
          "K000172",
          "K792325",
          "K070267",
          "K860133",
          "K882241",
          "K131167",
          "K012800",
          "K944538",
          "K110415",
          "K873601",
          "K910968",
          "K150410",
          "K011795",
          "K760334",
          "K802446",
          "K954481",
          "K760842",
          "K781608",
          "K980665",
          "K790349",
          "K100973"
        ],
        "pma_number": [
          "P840049"
        ],
        "registration_number": [
          "3004641308",
          "3009888740",
          "9616680",
          "1822565",
          "3010041693",
          "1020279",
          "3009973505",
          "3009882462",
          "9610726",
          "1649518",
          "1833506",
          "1721676",
          "3010047402",
          "2133928",
          "1047843",
          "3012725451",
          "9681465",
          "3009973336",
          "3003477135",
          "1828464",
          "3008021110",
          "3017397663",
          "9614209",
          "3017410889",
          "1219602",
          "3006015326",
          "3003998208",
          "3016090213",
          "3008395353",
          "3036756245",
          "3011683674",
          "2134285",
          "3026771806",
          "1219655",
          "3007923096",
          "3009848551",
          "3010097171",
          "1827096",
          "8010916",
          "3002907620",
          "1644408",
          "3005718816",
          "1054811",
          "2249697",
          "2183744",
          "3003387384",
          "2242737",
          "3003639920",
          "1000550978",
          "1000517406",
          "3002806470",
          "3010288357",
          "3015542154",
          "3008658204",
          "3025603301",
          "3009555440",
          "3006395932",
          "1643264",
          "1423662",
          "3005706667",
          "3033509898",
          "1529009",
          "3014662844",
          "3021008900",
          "3009241418",
          "1064858",
          "1834331",
          "3006356043",
          "3011061242",
          "1833920",
          "3004748528",
          "8010379",
          "2249615",
          "9610612",
          "3035366890",
          "3010039400",
          "3007311878",
          "3016438694",
          "3007366790",
          "1220477",
          "3014315669",
          "1226544",
          "3004371426",
          "3038503932",
          "1825034",
          "3010536692",
          "1057425",
          "1526534",
          "3002806603",
          "1319660",
          "1450662",
          "3009555447",
          "3010173425",
          "3009475821",
          "3008868758",
          "3020703621",
          "3007200272",
          "3015916578",
          "1421101",
          "2245304",
          "3027484613",
          "3011295718",
          "3010220595",
          "3005061536",
          "9616671",
          "3013014058",
          "3013055499",
          "1038671",
          "3015516266",
          "3015207155",
          "3011274144",
          "3006017180",
          "1818910",
          "2244478",
          "3002808022",
          "3007539489",
          "3002807310",
          "3014302784",
          "3009546990",
          "3020745236",
          "3013176080",
          "3006801265",
          "1061927",
          "3013791180"
        ],
        "fei_number": [
          "1000116912",
          "3004641308",
          "3009888740",
          "3000203000",
          "3010041693",
          "1020279",
          "3009973505",
          "1832656",
          "3009882462",
          "1721676",
          "3003569173",
          "3010047402",
          "3003070421",
          "1047843",
          "3008397795",
          "3012725451",
          "3009973336",
          "3003477135",
          "1828464",
          "3008021110",
          "3017397663",
          "1219602",
          "3006015326",
          "3003998208",
          "3016090213",
          "1000115656",
          "3008395353",
          "3036756245",
          "3011683674",
          "3026771806",
          "1219655",
          "3007923096",
          "3009848551",
          "3010097171",
          "1827096",
          "3001236277",
          "3002907620",
          "3005718816",
          "2183744",
          "3003387384",
          "3003639920",
          "1000550978",
          "1000517406",
          "3002806470",
          "3010288357",
          "3001548076",
          "3015542154",
          "3008658204",
          "3025603301",
          "3009555440",
          "3006395932",
          "1643264",
          "3002854251",
          "3002807113",
          "1000199976",
          "3005706667",
          "3001049282",
          "3000215273",
          "3033509898",
          "3002806725",
          "3002287318",
          "3002806315",
          "3014662844",
          "3021008900",
          "3009241418",
          "3007282753",
          "1000221916",
          "3006356043",
          "3011061242",
          "3004748528",
          "3003386935",
          "3035366890",
          "3010039400",
          "1000136478",
          "3007311878",
          "3016438694",
          "1220477",
          "3014315669",
          "3002807172",
          "3038503932",
          "1000306340",
          "1825034",
          "3010536692",
          "1526534",
          "3002806603",
          "1000114185",
          "1450662",
          "3009555447",
          "1000220733",
          "3009475821",
          "3008868758",
          "3020703621",
          "1000149660",
          "3002459190",
          "3007200272",
          "3015916578",
          "3002807036",
          "1421101",
          "2245304",
          "3027484613",
          "3011295718",
          "3010220595",
          "3005061536",
          "3013014058",
          "3013055499",
          "1038671",
          "3015516266",
          "3015207155",
          "3005985723",
          "3011274144",
          "1000136269",
          "3006017180",
          "1818910",
          "3000717902",
          "3002808022",
          "3007539489",
          "3002807310",
          "3014302784",
          "3009546990",
          "3020745236",
          "3003128516",
          "3013176080",
          "3006801265",
          "3013791180"
        ],
        "device_name": "Prosthesis, Knee, Femorotibial, Semi-Constrained, Cemented, Metal/Polymer",
        "medical_specialty_description": "Orthopedic",
        "regulation_number": "888.3530",
        "device_class": "2"
      }
    }
  ]
}