  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-10",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 2
    }
  },
  "results": [
    {
      "cfres_id": "195984",
      "product_res_number": "Z-1761-2022",
      "event_date_initiated": "2022-02-28",
      "event_date_posted": "2022-09-26",
      "recall_status": "Open, Classified",
      "res_event_number": "90835",
      "product_code": "REN",
      "product_description": "NanoSight LM10 fitted with LM12 module.  A laboratory instrument for determination of particle characteristics (e.g. size).",
      "code_info": "CDRH accession # 0510530-001",
      "firm_fei_number": "3014373630",
      "recalling_firm": "MALVERN PANALYTICAL LTD",
      "address_1": "Grovewood Road",
      "city": "Malvern",
      "country": "United Kingdom",
      "reason_for_recall": "Product found to not comply to Class 1 Laser safety requirement.",
      "root_cause_description": "Radiation Control for Health and Safety Act",
      "action": "Customers are being made aware of this issue. Malvern Panalytical will retrofit modified microscope plates and cell tops with a new safety cover on the LM10 to maintain the original Class 1 laser product classification at no charge to the purchaser.",
      "product_quantity": "111 units",
      "distribution_pattern": "US Nationwide Distribution",
      "openfda": {
        "device_name": "Particle-Size Measuring Instrument",
        "medical_specialty_description": "Unknown",
        "regulation_number": "",
        "device_class": "N"
      }
    }
  ]
}