  "results": [
    {
      "cfres_id": "194993",
      "product_res_number": "Z-1442-2022",
      "event_date_initiated": "2022-04-07",
      "event_date_created": "2022-08-01",
      "recall_status": "Open, Classified",
      "res_event_number": "90597",
      "product_code": "REN",
      "product_description": "NanoSight NS300",
      "code_info": "CDRH accession #0510530-003",
      "firm_fei_number": "3014373630",
      "recalling_firm": "MALVERN PANALYTICAL LTD",
      "address_1": "Grovewood Road",
      "city": "Malvern",
      "country": "United Kingdom",
      "reason_for_recall": "Product found to not comply to Class 1 Laser safety requirement.",
      "root_cause_description": "Radiation Control for Health and Safety Act",
      "action": "Customers are being made aware of this issue. All affected customers will be provided with a remediation kit with Class 3R labels and \nmanual addenda with revised safety instructions.",
      "product_quantity": "507 units",
      "distribution_pattern": "US Nationwide - Worldwide Distribution",
      "openfda": {
        "device_name": "Particle-Size Measuring Instrument",
        "medical_specialty_description": "Unknown",
        "regulation_number": "",
        "device_class": "N"
      }
    }
  ]
}