  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-17",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "194648",
      "product_res_number": "Z-1513-2022",
      "event_date_initiated": "2022-05-31",
      "event_date_created": "2022-08-11",
      "recall_status": "Open, Classified",
      "res_event_number": "90529",
      "product_code": "MHX",
      "k_numbers": [
        "K141156"
      ],
      "product_description": "Xhibit Telemetry Receiver, Model: 96280",
      "code_info": "UDI-DI: 10841522100246, Software version 1.4.0.\nSerial Numbers:\n6280-001538,\n6280-000590,\n6280-000589,\n6280-000339,\n9280-000711,\n6280-001103,\n6280-001102,\n6280-001469,\n6280-001467,\n6280-001686,\n6280-001399,\n6280-001323,\n6280-000457,\n6280-000456,\n6280-000455,\n6280-000440,\n6280-000439,\n6280-000438,\n6280-000429,\n6280-000476,\n6280-000475,\n6280-001693,\n6280-000703,\n6280-000702,\n6280-000701,\n6280-000700,\n6280-000314,\n6280-000258,\n6280-000157,\n6280-001223,\n6280-001222,\n6280-001221,\n6280-001220,\n6280-001219,\n6280-001218,\n6280-001217,\n6280-001216,\n6280-001215,\n6280-001214,\n6280-001207,\n6280-001127,\n6280-001125,\n6280-001124,\n6280-001123,\n6280-001056,\n6280-001055,\n6280-001054,\n6280-001053,\n6280-001052,\n6280-001051,\n6280-001050,\n6280-001288,\n6280-001287,\n6280-001490,\n6280-001692,\n6280-000588,\n6280-000517,\n6280-000516,\n6280-000515,\n6280-001668,\n6280-000206,\n6280-000162,\n6280-000210,\n6280-000663,\n6280-000377,\n6280-001589,\n6280-001385,\n6280-001384,\n6280-001382,\n6280-001405,\n6280-000594,\n6280-000593,\n6280-000592,\n6280-000596,\n6280-000274,\n6280-001844,\n6280-001843,\n6280-001721,\n6280-001720,\n6280-000136,\n6280-001784,\n6280-001783,\n6280-000359,\n6280-000340,\n6280-000336,\n6280-001718,\n6280-001706,\n6280-001705,\n6280-001704,\n6280-001703,\n6280-001702,\n6280-001699,\n6280-001698,\n6280-001672,\n6280-001669,\n6280-000622,\n6280-001797,\n6280-000097,\n6280-001558,\n6280-001551,\n6280-001649,\n6280-001648,\n6280-001647,\n6280-001646,\n6280-001645,\n6280-001594,\n6280-000093,\n6280-001168,\n6280-001531,\n6280-001670,\n6280-000219,\n6280-000550,\n6280-000878,\n6280-000877,\n6280-000872,\n6280-000862,\n6280-000858,\n6280-000857,\n6280-000221,\n6280-000089,\n6280-000204,\n6280-000212,\n6280-000510,\n6280-000509,\n6280-000209,\n6280-000320,\n6280-000538,\n6280-000528,\n6280-000205,\n6280-000361,\n6280-000360,\n6280-000223,\n6280-000963,\n6280-000529,\n6280-001575,\n6280-001574,\n6280-001573,\n6280-001572,\n6280-001571,\n6280-001570,\n6280-000467,\n6280-000466,\n6280-000465,\n6280-000464,\n6280-000463,\n6280-000462,\n6280-000461,\n6280-000460,\n6280-000459,\n6280-000458,\n6280-000222,\n6280-000264,\n6280-000275,\n6280-000354,\n6280-001677,\n6280-001048,\n6280-001043,\n6280-001041,\n6280-001038,\n6280-001037,\n6280-001036,\n6280-001023,\n6280-001022,\n6280-001016,\n6280-000787,\n6280-001418,\n6280-001419,\n6280-001416,\n6280-000894,\n6280-000866,\n6280-000865,\n6280-000863,\n6280-000902,\n6280-000898,\n6280-000724,\n6280-000420,\n6280-000419,\n6280-000798,\n6280-000325,\n6280-000532,\n6280-000521,\n6280-000213,\n6280-000254,\n6280-000856,\n6280-000855,\n6280-000854,\n6280-000853,\n6280-000852,\n6280-000851,\n6280-001143,\n6280-001142,\n6280-001141,\n6280-000472,\n6280-000375,\n6280-000368,\n6280-000421,\n6280-000519,\n6280-001213,\n6280-000836,\n6280-000835,\n6280-000834,\n6280-000833,\n6280-000832,\n6280-000831,\n6280-000586,\n6280-001473,\n6280-000319,\n6280-000263,\n6280-000312,\n6280-000579,\n6280-000092,\n6280-000090,\n6280-000084,\n6280-000083,\n6280-000079,\n6280-001380,\n6280-001466,\n6280-001458,\n6280-001611,\n6280-001592,\n6280-001591,\n6280-000388,\n6280-000161,\n6280-000160,\n6280-000159,\n6280-000158,\n6280-000066,\n6280-000053,\n6280-000174,\n6280-000173,\n6280-000172,\n6280-000170,\n6280-000168,\n6280-000167,\n6280-000077,\n6280-000598,\n6280-000355,\n6280-000344,\n6280-000341,\n6280-000338,\n6280-000468,\n6280-000423,\n6280-000478,\n6280-000477,\n6280-000454,\n6280-000428,\n6280-000261,\n6280-001614,\n6280-001642,\n6280-001641,\n6280-001640,\n6280-001639,\n6280-001638,\n6280-001637,\n6280-001636,\n6280-001635,\n6280-001634,\n6280-001633,\n6280-001632,\n6280-001631,\n6280-001630,\n6280-001629,\n6280-001913,\n6280-000138,\n6280-000437,\n6280-000436,\n6280-000177,\n6280-000176,\n6280-000337,\n6280-001857,\n6280-001853,\n6280-001749,\n6280-001359,\n6280-001160,\n6280-001159,\n6280-001158,\n6280-001157,\n6280-001156,\n6280-001154,\n6280-001132,\n6280-001112,\n6280-001109,\n6280-000991,\n6280-000983,\n6280-001189,\n6280-001188,\n6280-001205,\n6280-001204,\n6280-001203,\n6280-001202,\n6280-001200,\n6280-001194,\n6280-001193,\n6280-001184,\n6280-001166,\n6280-000358,\n6280-000357,\n6280-000345,\n6280-000432,\n6280-000431,\n6280-000811,\n6280-000693,\n6280-000690,\n6280-000929,\n6280-001063,\n6280-001062,\n6280-001061,\n6280-001060,\n6280-000391,\n6280-001273,\n6280-001554,\n6280-001553,\n6280-001261,\n6280-000426,\n6280-001525,\n6280-000229,\n6280-000876,\n6280-000875,\n6280-000745,\n6280-000208,\n6280-000574,\n6280-000218,\n6280-001087,\n6280-000045,\n6280-000041,\n6280-000374,\n6280-000373,\n6280-001257,\n6280-001256,\n6280-001255,\n6280-001254,\n6280-001253,\n6280-001252,\n6280-001251,\n6280-001250,\n6280-001249,\n6280-000840,\n6280-000486,\n6280-001825,\n6280-000543,\n6280-000542,\n6280-000839,\n6280-000879,\n6280-001325,\n6280-001071,\n6280-001070,\n6280-001069,\n6280-000900,\n6280-000897,\n6280-001822,\n6280-001874,\n6280-001872,\n6280-001871,\n6280-000695,\n6280-000540,\n6280-000541,\n6280-000539,\n6280-000166,\n6280-000624,\n6280-000614,\n6280-000613,\n6280-000612,\n6280-000611,\n6280-000610,\n6280-000609,\n6280-000608,\n6280-000607,\n6280-000630,\n6280-000629,\n6280-000628,\n6280-000627,\n6280-000626,\n6280-001644,\n6280-001619,\n6280-001617,\n6280-000587,\n6280-000490,\n6280-000489,\n6280-000488,\n6280-000487,\n6280-000974,\n6280-000973,\n6280-000972,\n6280-000971,\n6280-000970,\n6280-000969,\n6280-000968,\n6280-000967,\n6280-000966,\n6280-000965,\n6280-000329,\n6280-000326,\n6280-000318,\n6280-001621,\n6280-001609,\n6280-001596,\n6280-001536,\n6280-001381,\n6280-001340,\n6280-001339,\n6280-001338,\n6280-001337,\n6280-001336,\n6280-001335,\n6280-001334,\n6280-001315,\n6280-001312,\n6280-001311,\n6280-000547,\n6280-000546,\n6280-000545,\n6280-000544,\n6280-000518,\n6280-001764,\n6280-001752,\n6280-001751,\n6280-001746,\n6280-001548,\n6280-001408,\n6280-001431,\n6280-001729,\n6280-001046,\n6280-000601,\n6280-001350,\n6280-000689,\n6280-001487,\n6280-001475,\n6280-000928,\n6280-000927,\n6280-000926,\n6280-000925,\n6280-000901,\n6280-001329,\n6280-001331,\n6280-001330,\n6280-001342,\n6280-001341,\n6280-000514,\n6280-000513,\n6280-000512,\n6280-000511,\n6280-000530,\n6280-000108,\n6280-000107,\n6280-000106,\n6280-000105,\n6280-000104,\n6280-000103,\n6280-000102,\n6280-000101,\n6280-000100,\n6280-000600,\n6280-000276,\n6280-001374,\n6280-000195,\n6280-000192,\n6280-000189,\n6280-000187,\n6280-000178,\n6280-000194,\n6280-001820,\n6280-001805,\n6280-001444,\n6280-001225,\n6280-000850,\n6280-000725,\n6280-000694,\n6280-000553,\n6280-000549,\n6280-000548,\n6280-000330,\n6280-000317,\n6280-000188,\n6280-000182,\n6280-000290,\n6280-001818,\n6280-001765,\n6280-001602,\n6280-001601,\n6280-001450,\n6280-001449,\n6280-001377,\n6280-001376,\n6280-001178,\n6280-001049,\n6280-000796,\n6280-000743,\n6280-000491,\n6280-000356,\n6280-000295,\n6280-000293,\n6280-000292,\n6280-000291,\n6280-000231,\n6280-000196,\n6280-000061,\n6280-000060,\n6280-000473,\n6280-000232,\n6280-000062,\n6280-001319,\n6280-001313,\n6280-000384,\n6280-000199,\n6280-000198,\n6280-000065,\n6280-000064,\n6280-000237,\n6280-000406,\n6280-000401,\n6280-000362,\n6280-000270,\n6280-000400,\n6280-001245,\n6280-001101,\n6280-000873,\n6280-000871,\n6280-000870,\n6280-000861,\n6280-000860,\n6280-000815,\n6280-000554,\n6280-000527,\n6280-000499,\n6280-000416,\n6280-000371,\n6280-000197,\n6280-000523,\n6280-000063",
      "firm_fei_number": "3010157426",
      "recalling_firm": "Spacelabs Healthcare, Inc.",
      "address_1": "35301 Se Center St",
      "city": "Snoqualmie",
      "state": "WA",
      "postal_code": "98065-9216",
      "additional_info_contact": "Wendy Machleid\n425-396-3300",
      "reason_for_recall": "Software related alarm escalation defect occurs after approximately 25 days of continuous use. Technical alarm escalation for the following alarms, signal loss, all leads off, low battery, SpO2 sensor off, and signal interference, does not occur in the telemetry receiver as specified. Caregiver may not respond to the initial technical alarm and may not be aware of the continuing technical failure.",
      "root_cause_description": "Software Design Change",
      "action": "On 05/31/22, correction notices were mailed to customers who were asked to take the following actions that can be taken to minimize risk to patients or users:\nWeigh the benefits versus the risks when deciding whether to continue to use the alarm escalation feature within XTR 1.4. \n\nSuggestions from Spacelabs to mitigate the risk:\n- Set the priority to the same level as the  escalation max  on the product. This will ensure that the highest priority alarm is triggered for the technical alarm.\n- Do not run the XTR 1.4 for more than 24 days consecutively. Restarting the XTR after 24 days will mitigate the issue for another 24 days.\n- The recalling firm is working on a 1.4.1 release of XTR which will address this defect.\n\nFor additional information or technical assistance, please contact: Technical Support, 1-800-522-7025 and select 2",
      "product_quantity": "529",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of NY, AR, CO, FL, NJ, AL, KY, TN, SC, WA, OK, MN, PA, GA, VA, CA, OH, IA, KS, MI, IN, TX, MT, MO, ME, LA, IL, SD, MA, OR, ID, MD, AZ, AK and the countries of MB, ON, FR, QC, BC, IS, CT, RM.",
      "openfda": {
        "k_number": [
          "K040458",
          "K090556",
          "K070791",
          "K123074",
          "K071426",
          "K242962",
          "K201746",
          "K992595",
          "K962424",
          "K031340",
          "K091594",
          "K062271",
          "K112190",
          "K191323",
          "K221879",
          "K082785",
          "K071164",
          "K042904",
          "K031320",
          "K242562",
          "K101692",
          "K120598",
          "K062095",
          "K191106",
          "K150329",
          "K030931",
          "K950264",
          "K042656",
          "K231477",
          "K002209",
          "K060281",
          "K060039",
          "K031403",
          "K252250",
          "K120173",
          "K011000",
          "K053207",
          "K123472",
          "K180017",
          "K100217",
          "K213234",
          "K043276",
          "K030738",
          "K040246",
          "K020290",
          "K092970",
          "K041741",
          "K192137",
          "K911051",
          "K131929",
          "K030812",
          "K141963",
          "K250179",
          "K210906",
          "K080495",
          "K082583",
          "K120631",
          "K151080",
          "K933624",
          "K103134",
          "K033600",
          "K180482",
          "K993974",
          "K143676",
          "K030657",
          "K250854",
          "K040955",
          "K013311",
          "K041607",
          "K173765",
          "K253277",
          "K063690",
          "K223531",
          "K082280",
          "K071020",
          "K132533",
          "K992929",
          "K090702",
          "K061379",
          "K123556",
          "K003243",
          "K260098",
          "K202336",
          "K001268",
          "K000854",
          "K990476",
          "K091834",
          "K002478",
          "K990068",
          "K973474",
          "K213363",
          "K061610",
          "K090172",
          "K103646",
          "K013624",
          "K033365",
          "K021422",
          "K024108",
          "K150352",
          "K213336",
          "K130208",
          "K121480",
          "K100046",
          "K232962",
          "K001348",
          "K072346",
          "K161531",
          "K213799",
          "K211619",
          "K000815",
          "K970445",
          "K061052",
          "K043551",
          "K090443",
          "K051217",
          "K011540",
          "K110953",
          "K221056",
          "K083271",
          "K133222",
          "K041790",
          "K130711",
          "K172966",
          "K102040",
          "K063725",
          "K972535",
          "K160685",
          "K072070",
          "K120366",
          "K170567",
          "K252726",
          "K152973",
          "K201764",
          "K041772",
          "K980411",
          "K070588",
          "K000168",
          "K922974",
          "K010923",
          "K132075",
          "K030431",
          "K921014",
          "K030038",
          "K191249",
          "K220308",
          "K191149",
          "K914838",
          "K942700",
          "K982279",
          "K134046",
          "K131032",
          "K030313",
          "K192514",
          "K132026",
          "K050141",
          "K033492",
          "K062283",
          "K102495",
          "K171801",
          "K183387",
          "K032290",
          "K131223",
          "K260343",
          "K053121",
          "K983996",
          "K121367",
          "K063315",
          "K122439",
          "K153275",
          "K113623",
          "K071073",
          "K203088",
          "K945134",
          "K110015",
          "K062324",
          "K171580",
          "K152902",
          "K093788",
          "K081793",
          "K013199",
          "K945385",
          "K911646",
          "K092449",
          "K053522",
          "K980619",
          "K231671",
          "K120616",
          "K051367",
          "K023674",
          "K052801",
          "K923054",
          "K131829",
          "K060135",
          "K993608",
          "K191769",
          "K060223",
          "K092680",
          "K113657",
          "K043439",
          "K033694",
          "K031760",
          "K031124",
          "K032369",
          "K881643",
          "K062194",
          "K131414",
          "K113054",
          "K950436",
          "K033711",
          "K922668",
          "K971840",
          "K913263",
          "K083750",
          "K113187",
          "K101449",
          "K030575",
          "K080251",
          "K153160",
          "K222586",
          "K240320",
          "K042642",
          "K112637",
          "K170514",
          "K053484",
          "K190939",
          "K151063",
          "K143195",
          "K945099",
          "K190008",
          "K033444",
          "K122253",
          "K152552",
          "K150975",
          "K120727",
          "K922210",
          "K041087",
          "K151027",
          "K102004",
          "K081140",
          "K200494",
          "K192972",
          "K093268",
          "K110622",
          "K173768",
          "K172510",
          "K163584",
          "K031481",
          "K171901",
          "K073203",
          "K030722",
          "K052707",
          "K260579",
          "K912627",
          "K030613",
          "K062576",
          "K023569",
          "K113653",
          "K082045",
          "K150030",
          "K250349",
          "K141015",
          "K131971",
          "K070566",
          "K981576",
          "K130401",
          "K022485",
          "K012915",
          "K142601",
          "K133576",
          "K171621",
          "K180472",
          "K250453",
          "K001693",
          "K041235",
          "K113045",
          "K153707",
          "K163459",
          "K920445",
          "K030105",
          "K170876",
          "K023698",
          "K203435",
          "K182075",
          "K173454",
          "K102239",
          "K181831",
          "K200859",
          "K062144",
          "K113798",
          "K170155",
          "K110922",
          "K101521",
          "K251508",
          "K080157",
          "K012016",
          "K050742",
          "K213181",
          "K982091",
          "K031849",
          "K014213",
          "K101067",
          "K112962",
          "K131876",
          "K083697",
          "K011093",
          "K200015",
          "K944860",
          "K092369",
          "K123048",
          "K132662",
          "K160981",
          "K012226",
          "K021279",
          "K252840",
          "K152407",
          "K213490",
          "K033964",
          "K014261",
          "K101619",
          "K061993",
          "K031781",
          "K132077",
          "K161056",
          "K221616",
          "K060221",
          "K161517",
          "K141811",
          "K062392",
          "K944495",
          "K003621",
          "K100040",
          "K230604",
          "K221904",
          "K051658",
          "K051628",
          "K060541",
          "K141156",
          "K150082",
          "K000882",
          "K113441",
          "K190011",
          "K872476",
          "K051400",
          "K102562",
          "K013427",
          "K020531",
          "K202405",
          "K240312",
          "K073550",
          "K232694",
          "K073462",
          "K102422",
          "K033957",
          "K223574",
          "K160951",
          "K912835",
          "K211900",
          "K131763",
          "K201941",
          "K011918",
          "K023475",
          "K022834",
          "K220989",
          "K250757",
          "K102376",
          "K221348",
          "K103652",
          "K172226",
          "K061185",
          "K050762",
          "K001664",
          "K112803",
          "K033304",
          "K181314",
          "K182979",
          "K081983",
          "K031376",
          "K040395",
          "K081891",
          "K130849",
          "K924575",
          "K073280",
          "K122146",
          "K251146",
          "K992273",
          "K001433",
          "K052756",
          "K091395",
          "K060833",
          "K142827",
          "K173036",
          "K912188",
          "K223498",
          "K000517",
          "K970182",
          "K001722",
          "K092027",
          "K130584",
          "K093501",
          "K000563",
          "K023871",
          "K950536",
          "K153702",
          "K992674",
          "K021325",
          "K033048",
          "K002158",
          "K003248",
          "K080342",
          "K260656",
          "K061675",
          "K083228",
          "K153448",
          "K922058",
          "K040304",
          "K043549",
          "K251702",
          "K253170",
          "K990563",
          "K022889",
          "K151812",
          "K231010",
          "K151860",
          "K131872",
          "K123900",
          "K254209",
          "K062098",
          "K100939",
          "K252589",
          "K040357"
        ],
        "registration_number": [
          "3006087789",
          "3005783425",
          "3043479346",
          "3019545235",
          "3011120183",
          "3007836437",
          "3008255748",
          "3003591740",
          "3013500228",
          "9710602",
          "3019388613",
          "3013501110",
          "3009607734",
          "3012972516",
          "3004977335",
          "3021559257",
          "1218950",
          "3009973336",
          "3009077524",
          "3015997711",
          "3012421607",
          "3008483389",
          "3016931915",
          "3007734888",
          "9611252",
          "3003832357",
          "3016701404",
          "3014848734",
          "3030446844",
          "3010552191",
          "3013896610",
          "3009673645",
          "3022123357",
          "3007603826",
          "3032109181",
          "3004526608",
          "3004035727",
          "3030516450",
          "3016852448",
          "3010157426",
          "9681377",
          "3009443693",
          "3014263632",
          "3006147705",
          "3012104670",
          "3014579161",
          "3011353843",
          "3010536822",
          "3009715978",
          "3003971136",
          "2221819",
          "3018783526",
          "2080783",
          "3012238587",
          "3007413079",
          "3004415095",
          "3003263092",
          "3001570235",
          "3005906887",
          "8030978",
          "1316463",
          "3009156722",
          "3006606827",
          "3010611950",
          "3021736577",
          "1000484830",
          "3012359877",
          "3006979678",
          "3009490946",
          "3013596024",
          "3005521864",
          "3016618143",
          "9611031",
          "1930027",
          "3008363989",
          "3008058135",
          "3009831876",
          "3023272766",
          "3017446092",
          "3005877899",
          "3012528160",
          "3007481893",
          "3013557562",
          "9610816",
          "8030229",
          "3013314298",
          "3031158",
          "2025816",
          "3004196561",
          "3014541700",
          "3009499478",
          "3021219789",
          "3016674095",
          "8022257",
          "3010882708",
          "1054713",
          "3010698831",
          "3004832480",
          "3020115",
          "3008642652",
          "9610105",
          "1651104",
          "3003589956",
          "3007048291",
          "1066270",
          "1000122786",
          "3008716327",
          "3010703925",
          "3014633893",
          "1219324",
          "3013679558",
          "3030597679",
          "3000162901",
          "3000126629",
          "3015181391",
          "3008729547",
          "3007305624",
          "3010390468",
          "2032233",
          "3015328411"
        ],
        "fei_number": [
          "3003757840",
          "3006087789",
          "3005783425",
          "3043479346",
          "3019545235",
          "3011120183",
          "3007836437",
          "3008255748",
          "3003591740",
          "3013500228",
          "3019388613",
          "3013501110",
          "3009607734",
          "3012972516",
          "3004977335",
          "3021559257",
          "1218950",
          "3009973336",
          "3009077524",
          "3015997711",
          "3012421607",
          "3008483389",
          "3016931915",
          "3003142870",
          "3007734888",
          "3002807167",
          "3003832357",
          "3016701404",
          "3014848734",
          "3030446844",
          "3002737156",
          "3013896610",
          "1000300598",
          "3009673645",
          "3007603826",
          "3032109181",
          "3000957797",
          "3010830833",
          "3004526608",
          "3004035727",
          "3030516450",
          "3016852448",
          "3010157426",
          "3009443693",
          "3014263632",
          "3006147705",
          "3012104670",
          "3014579161",
          "1000114111",
          "3011353843",
          "3010536822",
          "3009715978",
          "3003971136",
          "2221819",
          "3018783526",
          "2080783",
          "3002806783",
          "3012238587",
          "3007413079",
          "3004415095",
          "3000215024",
          "3003263092",
          "3001570235",
          "3005906887",
          "1316463",
          "3009156722",
          "3006606827",
          "3010611950",
          "3015191855",
          "3021736577",
          "1000484830",
          "3012359877",
          "3006979678",
          "3009490946",
          "3013596024",
          "3005521864",
          "3016618143",
          "3002808182",
          "1930027",
          "3008363989",
          "3008058135",
          "3009831876",
          "3002807739",
          "3023272766",
          "3017446092",
          "3005877899",
          "3003513129",
          "3012528160",
          "3007481893",
          "3013557562",
          "2025816",
          "3004196561",
          "3014541700",
          "3009499478",
          "3021219789",
          "3016674095",
          "3010882708",
          "3010698831",
          "3004832480",
          "3020115",
          "3008642652",
          "3003589956",
          "3007048291",
          "1000122786",
          "3002807019",
          "3008716327",
          "3010703925",
          "3014633893",
          "1219324",
          "3013679558",
          "3030597679",
          "3002807720",
          "1051088",
          "3000162901",
          "3000126629",
          "3015181391",
          "3008729547",
          "3007305624",
          "3010390468",
          "3015328411"
        ],
        "device_name": "Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)",
        "medical_specialty_description": "Cardiovascular",
        "regulation_number": "870.1025",
        "device_class": "2"
      }
    }
  ]
}