{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-10",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "190691",
      "product_res_number": "Z-0415-2022",
      "event_date_initiated": "2021-11-12",
      "event_date_created": "2021-12-22",
      "recall_status": "Open, Classified",
      "res_event_number": "89170",
      "product_code": "PGW",
      "k_numbers": [
        "K180948"
      ],
      "product_description": "TruDi NAV Suction Instruments: 0-Degree, REF: TDNS000Z; 70-Degree, REF: TDNS070Z; and 90-Degree, REF: TDNS090Z",
      "code_info": "0-Degree Lots: 1903120, 1903250, 1904060, 1904180, 1904240, 1905030, 1905070, 1906260, 1907120, 1907170, 1907220, 1907240, 1907260, 2003040, 2009110, 2011240, 2102080, 2103090, 2104260, 2105040, 2105260, 2106100, 2107070.\n70-Degree Lots: 1903137, 1903257, 1904067, 1905017, 1905037, 1905077, 1906067, 1906267, 1907127, 1908137, 1908157, 2003127.\n90-Degree: 1903139, 1903259, 1904069, 1905019, 1905139, 1906079, 1908269, 1909279, 1909309, 1910299, 2003029, 2004029",
      "firm_fei_number": "3005172759",
      "recalling_firm": "Acclarent, Inc.",
      "address_1": "31 Technology Dr  Suite 200 31 Technology Dr",
      "address_2": "Suite 200",
      "city": "Irvine",
      "state": "CA",
      "postal_code": "92618-2346",
      "additional_info_contact": "949-784-9350",
      "reason_for_recall": "Suction Instruments were incorrectly calibrated so they may not meet the system accuracy specification for visual verification of device location in the patient anatomy. If the surgeon does not detect the accuracy error, this could lead to an intraprocedural delay, a potential for reversible tissue damage or damage to the carotid artery, CSF leak, or damage to optic nerve.",
      "root_cause_description": "Process control",
      "action": "On 11/12/21, recall notices were sent to OR Directors, Risk Managers, or Material Managers.  Customers were asked to do the following:\n1) Examine your inventory immediately to determine if you have product subject to this recall on hand and quarantine such product(s). 2) Remove the product subject to this recall and communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. 3) If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return. Please consider including a copy of this recall letter when communicating. 4) Complete the Business Reply Form and return to acclarent8930@sedgwick.com 5) After the Business reply Form is received  a return kit will be sent. 6) Keep this notice visibly posted for awareness until all product subject to this recall has been returned. Customers with additional questions are encouraged to call the recalling firm at 1-877-775-2789, 4:00am-5:00pm PST.",
      "product_quantity": "1543",
      "distribution_pattern": "US: AZ, MI, IA, LA, IL, VA, SC, NY, IN, OR, SD, TX, PA, NE, FL, CT, MN, NV, CA, NJ, UT, AL, MA, WI, TN, WV, OH, NC, KY, HI¿, OK, GA, CO, MO, WY, MS, NH, HI, MD, ID",
      "openfda": {
        "k_number": [
          "K153555",
          "K201174",
          "K200723",
          "K243053",
          "K162176",
          "K150473",
          "K133665",
          "K180948",
          "K141456",
          "K193118",
          "K161701",
          "K253395",
          "K141551",
          "K223734",
          "K262235",
          "K173628",
          "K231862",
          "K221037",
          "K161555",
          "K163416",
          "K160369",
          "K170018",
          "K133573",
          "K202609",
          "K193502",
          "K161697",
          "K221098",
          "K190532",
          "K211291",
          "K261540",
          "K160479",
          "K171661",
          "K200041",
          "K193453",
          "K161491",
          "K153247",
          "K192397",
          "K161940"
        ],
        "device_name": "Ear, Nose, And Throat Stereotaxic Instrument",
        "medical_specialty_description": "Neurology",
        "regulation_number": "882.4560",
        "device_class": "2",
        "registration_number": [
          "3017060084",
          "3007738736",
          "1221485",
          "3015131155",
          "3010202439",
          "3011874573",
          "3010707607",
          "3008837339",
          "2183744",
          "9610905",
          "3005677016",
          "1058084",
          "3011543740",
          "3008087383",
          "3014342096",
          "3018094310",
          "3006697241",
          "8040510",
          "3010273872",
          "2020550",
          "3010313588",
          "1928237",
          "3003418325",
          "3018565642",
          "3005423519",
          "9610617",
          "1824206",
          "9680721",
          "1225492",
          "3009217531",
          "3005172759",
          "3009018440",
          "1723170",
          "1811755",
          "3007421149",
          "8010177",
          "3008203003",
          "3003500271",
          "8043933",
          "3015462848",
          "3007361893",
          "3015967359",
          "3010009632",
          "3003678543",
          "3016245399"
        ],
        "fei_number": [
          "1000517638",
          "3017060084",
          "3007738736",
          "1221485",
          "3015131155",
          "3011874573",
          "3010707607",
          "3008837339",
          "2183744",
          "3005677016",
          "3002821504",
          "3007475226",
          "3006165545",
          "3011543740",
          "3008087383",
          "3001237338",
          "3014342096",
          "1000221981",
          "3018094310",
          "3006697241",
          "3002808268",
          "3010273872",
          "2020550",
          "3010313588",
          "3001319476",
          "3003418325",
          "3016245399",
          "3018565642",
          "1928237",
          "1824206",
          "3009217531",
          "3005172759",
          "3009018440",
          "1811755",
          "3007421149",
          "3008203003",
          "3003500271",
          "3015462848",
          "3002807464",
          "3007361893",
          "3015967359",
          "3010009632",
          "3003678543",
          "1000244733",
          "3002619595"
        ]
      }
    }
  ]
}