{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-12",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "189845",
      "product_res_number": "Z-0226-2022",
      "event_date_initiated": "2021-10-05",
      "event_date_created": "2021-11-16",
      "recall_status": "Open, Classified",
      "res_event_number": "88870",
      "product_code": "MUQ",
      "k_numbers": [
        "K150493"
      ],
      "product_description": "Sensis/ Sensis Vibe systems with software VD1X as follows:\n(1)Sensis Model 10764561\n(2) Sensis Vibe Hemo  Model 11007641\n(3)Sensis Vibe Combo  Model 11007642\n\nIntended Use:  A diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies",
      "code_info": "Serial Numbers:\n15916\n15908\n103108\n103111\n15917\n103047\n103048\n103171\n103181\n103182\n103185\n103191\n103193\n103301\n103170\n103223\n103225\n103247\n15939\n103313\n103315\n100413\n100414\n100417\n100402\n14612\n14613\n101016\n101017\n101019\n101020\n101022\n101029\n101031\n101032\n101036\n18903\n15945\n15946\n103359\n15931\n100263\n103424\n15953\n103274\n15941\n15942\n103319\n103320\n103321\n103323\n103325\n103328\n103330\n103333\n103334\n103335\n103344\n100216\n103337\n15514\n15515\n100154\n100156\n100157\n14803\n14804\n18502\n15544\n15545\n100416\n102008\n103413\n15561\n101082\n101083\n101084\n18902\n14603\n100208\n100209\n102138\n103437\n15560\n101090\n102019\n102021\n102025\n102029\n15559\n101088\n15819\n15820\n102145\n102147\n103160\n15539\n100346\n100349\n103314\n102199\n15830\n102323\n102324\n102325\n15832\n15833\n103079\n122042\n123004\n15005\n15147\n100095\n100097\n100103\n100115\n100116\n100117\n100120\n100121\n102158\n102167\n102168\n102172\n122006\n122026\n102214\n102216\n102219\n102222\n102272\n101096\n102181\n102242\n102068\n102069\n102078\n102080\n102093\n102097\n15828\n15829\n102221\n103123\n103124\n103131\n100163\n100165\n15135\n15136\n100306\n100309\n15552\n15553\n101045\n101052\n101053\n101054\n101056\n101059\n101063\n101064\n103300\n15564\n15565\n101099\n101100\n101101\n15556\n15557\n101060\n101068\n101069\n101070\n101071\n101072\n103265\n14602\n100064\n15546\n14611\n100422\n14906\n103307\n103312\n103019\n10654\n11657\n14122\n14123\n103331\n103362\n103393\n103402\n103419\n103252\n103261\n103267\n103269\n103270\n103276\n103277\n103280\n103281\n103022\n100279\n103381\n15543\n15551\n100408\n100410\n100411\n100412\n101043\n102162\n122029\n103194\n103145\n103148\n103152\n15920\n15537\n15538\n100345\n123047\n15836\n15837\n102285\n102286\n102338\n102183\n15943\n15948\n103345\n103368\n103415\n103416\n123053\n15123\n103245\n103249\n103250\n103264\n15949\n15904\n103080\n103082\n103086\n103088\n103089\n103090\n15530\n100231\n103290\n15804\n102030\n102031\n102304\n101094\n18600\n15547\n15548\n101008\n101093\n15821\n15523\n100175\n100176\n102226\n101097\n15935\n15936\n103306\n103341\n15909\n102269\n102273\n102276\n102278\n102279\n103114\n14809\n14810\n102283\n15823\n15824\n102193\n15803\n15809\n15506\n15507\n100070\n100071\n100074\n102155\n102157\n15902\n15903\n103040\n103042\n14312\n103039\n15009\n102174\n102182\n15813\n15814\n103130\n103200\n103234\n103053\n14900\n103116\n103117\n103119\n103288\n18900\n18901\n14902\n15927\n15928\n14802\n15901\n103008\n103216\n102237\n103189\n103208\n15541\n100391\n15536\n102010\n15841\n15842\n102295\n102296\n102302\n18099\n100006\n18098\n14299\n100005\n10731\n15524\n15525\n100184\n15950\n103352\n122011\n102274\n14905\n103283\n15937\n15938\n103308\n103310\n15925\n15926\n103115\n103244\n103246\n103014\n15923\n15924\n103242\n101018\n121014\n15818\n102143\n102091\n\nExpanded Recall 10/27/21 (36 units)\nSerial Numbers:\n15915\n103137\n103144\n103196\n103198\n15913\n103166\n103173\n123021\n103184\n103177\n103186\n103199\n103139\n103168\n102326\n103167\n103178\n103161\n103172\n14120\n103432\n102328\n15520\n15522\n102060\n102062\n15807\n15806\n102057\n102049\n102051\n103431\n101095\n101089\n15558",
      "firm_fei_number": "2240869",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "city": "Malvern",
      "state": "PA",
      "postal_code": "19355-1418",
      "additional_info_contact": "SAME\n610-219-4834",
      "reason_for_recall": "System operator manual states that the system should be rebooted once, every 7 days. In some cases, users are not following these instructions and the system is running for more than 7 days. Under rare circumstances, this may result in a partial freeze of the user interface which would no longer update the vital signs",
      "root_cause_description": "Software design",
      "action": "Siemens Medical Solutions USA, Inc.  initiated a Customer Safety Advisory Notice to customers via AX067/21/S dated 10/5/21.   Letter states reason for recall,health risk and action to take:\nThe software in the affected systems will be updated to display a notification message. Once the system has reached a runtime of more than 7 days, a notification message will be displayed after closing a study to remind the operator to reboot the system as it is running longer than advised.\nIn addition, the system operating manual will include a warning message that the Sensis system can have\nperformance issues when running for more than 7 days nonstop without a restart. A system running nonstop for more than 7 days may display incorrect or delayed vital signs; the values may not be saved, the report generation may be incorrect, and the system could become unavailable before or during an examination. The warning message will also contain the statement that the Sensis system should be restarted at least once a week.\nSiemens will provide a software update for all affected systems which will be distributed via Update Instruction AX066/21/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436.\n\nPlease review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein",
      "product_quantity": "403 units",
      "distribution_pattern": "Nationwide",
      "openfda": {
        "pma_number": [
          "P010003"
        ],
        "registration_number": [
          "1422634",
          "1450662",
          "1047843",
          "1526611",
          "1063481",
          "1058584"
        ],
        "fei_number": [
          "1000219637",
          "1450662",
          "1047843",
          "1526611",
          "3001451326",
          "1000514656"
        ],
        "device_name": "Glue, Surgical, Arteries",
        "medical_specialty_description": "Unknown",
        "regulation_number": "",
        "device_class": "3"
      }
    }
  ]
}