  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-05",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "184127",
      "product_res_number": "Z-0322-2021",
      "event_date_initiated": "2020-09-24",
      "event_date_posted": "2020-11-02",
      "recall_status": "Open, Classified",
      "res_event_number": "86595",
      "product_code": "OLO",
      "k_numbers": [
        "K200095"
      ],
      "product_description": "Streamline OCT Navigated Instruments, Catalog numbers 26-DRILL-MDN\n26-FTAP-30-MDN\n26-FTAP-35-MDN\n26-FTAP-40-MDN\n26-FTAP-40-MDN\n26-FTAP-45-MDN\n26-S-FINDER-MDN   \n Components of a surgical navigation device.",
      "code_info": "Lots  370855 373847 373848 373849 377629 373850 373846",
      "firm_fei_number": "1000115331",
      "recalling_firm": "RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.)",
      "address_1": "375 River Park Cir",
      "city": "Marquette",
      "state": "MI",
      "postal_code": "49855-1781",
      "additional_info_contact": "Christine Thomas\n906-226-9909",
      "reason_for_recall": "System does not include a drill guide.",
      "root_cause_description": "Device Design",
      "action": "On September 24, 2020, the firm sent an Urgent Medical Devices Recall letter to customers to inform them of the product issue. \n\nCustomers were instructed to remove affected product from all user sites and inventory locations. Customers are to return these instrument systems to RTI Surgical using RGA number 63000897 along with a completed copy of the Acknowledgement and Receipt Form. If the product has been further distributed, please immediately notify your customers by providing them a copy of the recall letter and the response form for completion.",
      "product_quantity": "76 units",
      "distribution_pattern": "Distributed to one distributor in Michigan.",
      "openfda": {
        "k_number": [
          "K252088",
          "K130748",
          "K230064",
          "K173946",
          "K162195",
          "K243958",
          "K250086",
          "K162341",
          "K153237",
          "K172219",
          "K252989",
          "K182077",
          "K252742",
          "K083872",
          "K240647",
          "K243326",
          "K220958",
          "K162567",
          "K252873",
          "K240062",
          "K261205",
          "K240568",
          "K190556",
          "K183644",
          "K213033",
          "K201966",
          "K191100",
          "K232800",
          "K232413",
          "K201006",
          "K243671",
          "K120867",
          "K192140",
          "K242373",
          "K223927",
          "K261113",
          "K203005",
          "K223070",
          "K110240",
          "K213380",
          "K220146",
          "K130887",
          "K220038",
          "K110021",
          "K230202",
          "K213708",
          "K202749",
          "K251282",
          "K233255",
          "K200935",
          "K172518",
          "K191039",
          "K213877",
          "K173372",
          "K243659",
          "K222510",
          "K260571",
          "K093206",
          "K193135",
          "K153396",
          "K251314",
          "K233299",
          "K212245",
          "K203751",
          "K171267",
          "K241517",
          "K261024",
          "K183630",
          "K212023",
          "K221180",
          "K252880",
          "K243073",
          "K090953",
          "K202184",
          "K182104",
          "K210457",
          "K201255",
          "K122742",
          "K191369",
          "K200095",
          "K232521",
          "K233227",
          "K242890",
          "K231172",
          "K250599",
          "K133759",
          "K241625",
          "K252058",
          "K221224",
          "K191285",
          "K170977",
          "K182731",
          "K251315",
          "K093998",
          "K201661",
          "K093118",
          "K220973",
          "K240625",
          "K223765",
          "K083644",
          "K152574",
          "K223808",
          "K213546",
          "K230195",
          "K172517",
          "K203573",
          "K262121",
          "K190881",
          "K252841",
          "K241130",
          "K170937",
          "K191998",
          "K172623",
          "K231886",
          "K130387",
          "K170360",
          "K240701",
          "K251971",
          "K193128",
          "K203561",
          "K180352",
          "K093425",
          "K252037",
          "K233256",
          "K083483",
          "K192663",
          "K253623",
          "K192911",
          "K140596",
          "K172115",
          "K254157",
          "K192653",
          "K171525",
          "K242121",
          "K092497",
          "K083310",
          "K220241",
          "K222383",
          "K143628",
          "K242264",
          "K083009",
          "K200845",
          "K162937",
          "K100742",
          "K253444",
          "K223351",
          "K162962",
          "K202769",
          "K202320",
          "K254158",
          "K192133",
          "K220255",
          "K222355",
          "K171651",
          "K200047",
          "K243419",
          "K203816",
          "K214011",
          "K210574",
          "K242864",
          "K241893",
          "K213918",
          "K233725",
          "K191507",
          "K201569",
          "K180796",
          "K200719",
          "K210121",
          "K162375",
          "K193618",
          "K222966",
          "K172301",
          "K182524",
          "K122011",
          "K251943",
          "K253921",
          "K200888",
          "K211442",
          "K251073",
          "K223301",
          "K103829",
          "K140331",
          "K170011",
          "K162921",
          "K210668",
          "K233228",
          "K233542",
          "K091411",
          "K200376",
          "K240951",
          "K252597",
          "K212007",
          "K253381",
          "K231494",
          "K170581",
          "K223511",
          "K193094",
          "K260011",
          "K141989",
          "K180271",
          "K152041",
          "K210555",
          "K102967",
          "K170891",
          "K160537",
          "K242806",
          "K250520",
          "K203110",
          "K183381",
          "K251316",
          "K211596",
          "K212825",
          "K180690",
          "K181449",
          "K242009",
          "K110054",
          "K233595",
          "K121765",
          "K181890",
          "K201189",
          "K102990",
          "K220348",
          "K200892",
          "K172268",
          "K250182",
          "K230243",
          "K253604",
          "K170593",
          "K212088",
          "K232116",
          "K100599",
          "K193120",
          "K243751",
          "K200642",
          "K192526",
          "K242464",
          "K260222",
          "K233199",
          "K212537",
          "K250167",
          "K203040",
          "K180038",
          "K121064",
          "K153700",
          "K202552",
          "K191943",
          "K170584",
          "K242933",
          "K232802",
          "K260369",
          "K202956",
          "K221928",
          "K170430",
          "K241171",
          "K230880",
          "K163338",
          "K181606",
          "K252755",
          "K181848",
          "K243814",
          "K143752",
          "K140167",
          "K220522",
          "K262506",
          "K243560",
          "K090375",
          "K170280",
          "K232044",
          "K230783",
          "K203291",
          "K131425",
          "K260480",
          "K171321",
          "K220774",
          "K112507",
          "K230700",
          "K211441",
          "K220291",
          "K231184",
          "K182121",
          "K182721",
          "K151511",
          "K163581",
          "K143597",
          "K233706",
          "K171556",
          "K241031",
          "K141694",
          "K222291",
          "K192495",
          "K172522",
          "K251780",
          "K110204",
          "K090313",
          "K240721",
          "K223767",
          "K251060",
          "K141941",
          "K180277",
          "K241644",
          "K124004",
          "K260329",
          "K121936",
          "K232533",
          "K250608",
          "K122326",
          "K102251",
          "K252562",
          "K210472",
          "K191854",
          "K233254",
          "K220862",
          "K160385",
          "K180523",
          "K221028",
          "K230087",
          "K201022",
          "K212028",
          "K232345",
          "K191722",
          "K213876",
          "K180307",
          "K252087",
          "K243285",
          "K212756",
          "K182964",
          "K201251",
          "K153603",
          "K223649",
          "K172034",
          "K141601",
          "K140927",
          "K221618",
          "K163379",
          "K243387",
          "K162364",
          "K223108",
          "K203679",
          "K252544",
          "K222698",
          "K231527",
          "K220593",
          "K261665",
          "K211616",
          "K230716",
          "K090420",
          "K182848",
          "K223553",
          "K193515",
          "K153203",
          "K111843",
          "K230369",
          "K191012",
          "K260240",
          "K142530",
          "K190653",
          "K100107",
          "K150231",
          "K223350",
          "K140689",
          "K173338",
          "K242914",
          "K192080",
          "K220024",
          "K252962",
          "K201327",
          "K240211",
          "K182345",
          "K254148",
          "K261852",
          "K192336",
          "K131130",
          "K252950",
          "K243514",
          "K183196",
          "K151364",
          "K171780",
          "K231229",
          "K243481",
          "K183605",
          "K220652",
          "K182982",
          "K241011",
          "K253661",
          "K140454",
          "K203115",
          "K261060",
          "K233638",
          "K230861",
          "K153647",
          "K213491",
          "K222753",
          "K233443",
          "K230989",
          "K181596",
          "K214059",
          "K163221",
          "K252871",
          "K240160",
          "K260582",
          "K233513",
          "K153240",
          "K172166",
          "K172462",
          "K143668",
          "K200791",
          "K223200",
          "K172849",
          "K243258",
          "K202836",
          "K240662",
          "K181111",
          "K241082",
          "K232572",
          "K160713",
          "K220459",
          "K213873",
          "K103400",
          "K193307",
          "K223558",
          "K111443",
          "K230876",
          "K203205",
          "K250877",
          "K213720",
          "K191223",
          "K230180",
          "K260601",
          "K180816",
          "K111811",
          "K091998",
          "K192173",
          "K231963",
          "K261410",
          "K234047",
          "K240139",
          "K232140",
          "K260546",
          "K260010",
          "K250879"
        ],
        "registration_number": [
          "3037407500",
          "3015536273",
          "3010041430",
          "3006846753",
          "3012639025",
          "1057425",
          "8031020",
          "3009971621",
          "3012429289",
          "3014725904",
          "1030489",
          "3007014520",
          "1835251",
          "1834331",
          "3021060474",
          "3015869121",
          "2024311",
          "2183946",
          "3011713200",
          "9617297",
          "1835444",
          "3043520111",
          "3005706667",
          "9610622",
          "3012911444",
          "2031910",
          "2244478",
          "3021948057",
          "3016261131",
          "1424434",
          "3003500271",
          "3002907620",
          "3004641308",
          "3013422238",
          "3006776201",
          "3034676720",
          "1018470",
          "3003935342",
          "3004976965",
          "3016090213",
          "9616696",
          "1833920",
          "3005751028",
          "3033509898",
          "3010129044",
          "1625507",
          "9611390",
          "2531477",
          "3025603301",
          "3009973699",
          "3003541440",
          "3010047402",
          "1818910",
          "3010370554",
          "1833986",
          "3014498598",
          "3013404959",
          "1721964",
          "3009433026",
          "3013552417",
          "9611112",
          "9610612",
          "3006984710",
          "3008868758",
          "3003139373",
          "3003604053",
          "1835296",
          "3016050940",
          "3009380063",
          "3010097171",
          "3003639920",
          "2133772",
          "3005075696",
          "3015516266",
          "3011656326",
          "1221485",
          "1935627",
          "3007993775",
          "3014170143",
          "3012447612",
          "8043792",
          "2183449",
          "1649518",
          "3003617485",
          "2133928",
          "9612277",
          "1721676",
          "3023852420",
          "3030412764",
          "3009882675",
          "3010986197",
          "3024788684",
          "3011782009",
          "8043933",
          "3005739886",
          "1319639",
          "3007515069",
          "3006654680",
          "1834379",
          "1220477",
          "2133786",
          "2528981",
          "1225492",
          "3011796723",
          "3000270450",
          "9617840",
          "3010313588",
          "3011795235",
          "3007700286",
          "3008455034",
          "3005677016",
          "3003637761",
          "3033536255",
          "3027153110",
          "1000548922",
          "3011683674",
          "3015876273",
          "3010120326",
          "1720929",
          "3013485140",
          "3012764194",
          "3006791286",
          "3009522766",
          "3026776",
          "3019807891",
          "3007344102",
          "8010468",
          "3013917867",
          "1824199",
          "3003761012",
          "1424263",
          "3007922509",
          "2090010",
          "1047843",
          "3021976927",
          "3011015572",
          "9617544",
          "1064858",
          "3027305794",
          "3005061536",
          "3015907115",
          "1526534",
          "3005819474",
          "3014252644",
          "3008114965",
          "3010155648",
          "3017410889",
          "3010462278",
          "3004209178",
          "3016084569",
          "1421101",
          "3010162973",
          "3015179588",
          "3002743211",
          "3009050824",
          "2953359",
          "3014655115",
          "3006563559",
          "3011577940",
          "3019767615",
          "9615128",
          "1219602",
          "3004638600",
          "3017528621",
          "3003525643",
          "3016678045",
          "2530808",
          "3021902355",
          "3010197239",
          "1045834",
          "1828288",
          "9612420",
          "3007485169",
          "3004126422",
          "3004608878",
          "3012495575",
          "3009185973",
          "3004181214",
          "3015542154",
          "3010822102",
          "3010120104",
          "3025783933",
          "3004142400",
          "1226544",
          "3010363503",
          "9611813",
          "3009051471",
          "3041276509",
          "3005641619",
          "3017492965",
          "3004024955",
          "3009144915",
          "3014273424",
          "1058584",
          "3009337690",
          "3007125392",
          "1020279",
          "3015967359",
          "2133641",
          "3009879967",
          "3010863450",
          "3004717358",
          "3010057495",
          "3009887475",
          "3012263546",
          "3004188066",
          "3010287737",
          "3009554293",
          "3007997036",
          "3013194153",
          "3009959868",
          "2133810",
          "3009973336",
          "3017406274",
          "3019878714",
          "3015399803",
          "3010232390",
          "3007539489",
          "3004893332",
          "3003387384",
          "3003477135",
          "3006460162",
          "3009417901",
          "3012098629",
          "3007521480",
          "3013302242",
          "3017565094",
          "3003596442",
          "3018405359",
          "2031093",
          "2031966",
          "3005497913",
          "3006493760",
          "3015207155",
          "3010173425",
          "1833824",
          "1644408",
          "3001239363",
          "3010120339",
          "3025110439",
          "1218017",
          "2135156",
          "1643264",
          "3005874553",
          "3009266751",
          "3010120135",
          "3008027912",
          "3030154790",
          "3010663372",
          "8010177",
          "3008812560",
          "1928237",
          "3008992889",
          "9611579",
          "3008837339",
          "3015877656",
          "3015941638",
          "3010531060",
          "3004635447",
          "1723170",
          "2246552",
          "3015537048",
          "3010032903",
          "3025343017",
          "3010123206",
          "1226188",
          "9680515",
          "9614209",
          "1833506",
          "3004719693",
          "3012227716",
          "8010916",
          "3016619943",
          "3003120897",
          "3005180920",
          "3040335671",
          "3016075969",
          "1123169",
          "3010049501",
          "3007366790",
          "3009293341",
          "3016438694",
          "1000517406",
          "3009381600",
          "3026576110",
          "3005216725",
          "3006125668",
          "3012966183",
          "3015831087",
          "3005012805",
          "1056553",
          "3016496998",
          "3020375573",
          "3010047454",
          "3017980495",
          "2134947",
          "3005083075",
          "3036756245",
          "1835572",
          "2648623",
          "2134470",
          "3014680795",
          "3013422236",
          "3017123055",
          "3004774118",
          "3043543260",
          "3014782362",
          "1221053",
          "3012401682",
          "3009394448",
          "3003678543",
          "3040312029",
          "3024999441",
          "3009493875",
          "3009049161",
          "3009165919",
          "1319660",
          "3004513872",
          "3014479313",
          "3010141593",
          "3010009632",
          "3010039400",
          "3007728266",
          "3009888740",
          "3012131184",
          "3007441485",
          "3012428435",
          "3004464325",
          "3039520898",
          "3010408671",
          "3004727093",
          "3004060780",
          "3001284227",
          "3010120148",
          "3016443334",
          "3018565653",
          "3014411622",
          "2647346",
          "3013682146",
          "2081135",
          "3021632375",
          "3008110533",
          "1450662",
          "3038503932",
          "3030400539",
          "1221763",
          "3000719653",
          "3008583793",
          "1423662",
          "9611827",
          "3010041499",
          "1314417",
          "2031917",
          "3004153896",
          "2087234",
          "2011171",
          "3006783837",
          "1836357",
          "3002807090",
          "2027487",
          "1043653",
          "2030624",
          "3011674645",
          "3033536319",
          "3017823294",
          "2249615",
          "2027467",
          "3006639944",
          "3003661942",
          "3021008900",
          "3018094310",
          "3012120772",
          "3009973505",
          "3001592626",
          "3003955307",
          "3009871135",
          "3020703621",
          "3006128100",
          "1828464",
          "3004499989",
          "2219920",
          "3011187779",
          "3004016520",
          "1526439",
          "3009962553",
          "3010009693",
          "3010197224",
          "3005823819",
          "3014680149"
        ],
        "fei_number": [
          "3037407500",
          "3001551180",
          "3003164255",
          "3000210784",
          "3003640911",
          "3015536273",
          "1000115656",
          "3010041430",
          "3006846753",
          "3012639025",
          "3009971621",
          "3012429289",
          "3014725904",
          "1000522098",
          "3007014520",
          "1000221916",
          "3021060474",
          "3015869121",
          "1000139987",
          "2024311",
          "3011713200",
          "3043520111",
          "3001503333",
          "3005706667",
          "3012911444",
          "3002807156",
          "3021948057",
          "3016261131",
          "3000204692",
          "3003500271",
          "3002907620",
          "3002821504",
          "3004641308",
          "3013422238",
          "3006776201",
          "3000717687",
          "3034676720",
          "1018470",
          "1000138054",
          "3003935342",
          "3004976965",
          "1000477302",
          "3016090213",
          "3005751028",
          "3033509898",
          "3010129044",
          "1625507",
          "3025603301",
          "3003112432",
          "3009973699",
          "3003541440",
          "3010047402",
          "1818910",
          "3002904345",
          "3010370554",
          "3014498598",
          "3002806577",
          "3002807421",
          "3009433026",
          "3013552417",
          "3006984710",
          "3008868758",
          "3003139373",
          "3003604053",
          "3016050940",
          "1000517638",
          "3009380063",
          "3010097171",
          "3003639920",
          "3001237338",
          "3005075696",
          "3015516266",
          "3011656326",
          "1221485",
          "1935627",
          "1832656",
          "3007993775",
          "3014170143",
          "3012447612",
          "3008599177",
          "2183449",
          "3003617485",
          "3009256700",
          "3002808269",
          "1721676",
          "3023852420",
          "3030412764",
          "3010986197",
          "3002806773",
          "3024788684",
          "3005739886",
          "3007515069",
          "1000123143",
          "1000306340",
          "3006654680",
          "1220477",
          "2528981",
          "3011796723",
          "3000270450",
          "3010313588",
          "3011795235",
          "3000203066",
          "3007700286",
          "3008455034",
          "3005677016",
          "3003637761",
          "3033536255",
          "3027153110",
          "1000116912",
          "1000548922",
          "3011683674",
          "3015876273",
          "3010120326",
          "3002795782",
          "1720929",
          "3013485140",
          "3012764194",
          "3006791286",
          "3009522766",
          "3019807891",
          "3007344102",
          "3013917867",
          "1824199",
          "3003761012",
          "3000206435",
          "3007922509",
          "1047843",
          "3021976927",
          "3011015572",
          "3027305794",
          "3009322142",
          "3005061536",
          "1000219637",
          "3015907115",
          "1526534",
          "3005819474",
          "3014252644",
          "3008114965",
          "3010155648",
          "3010462278",
          "3001452082",
          "3016084569",
          "3001236549",
          "1421101",
          "3003128516",
          "3015179588",
          "3002743211",
          "3009050824",
          "3002980729",
          "3014655115",
          "3001642380",
          "3006563559",
          "3011577940",
          "3002807041",
          "3019767615",
          "1219602",
          "3004638600",
          "3008820440",
          "3017528621",
          "3003179719",
          "3003525643",
          "3016678045",
          "3002806315",
          "3021902355",
          "3010197239",
          "1045834",
          "1828288",
          "3007485169",
          "3004126422",
          "3004608878",
          "3002806685",
          "3002926812",
          "1000136478",
          "1000115331",
          "3012495575",
          "3009185973",
          "3004181214",
          "3015542154",
          "3010822102",
          "3010120104",
          "3025783933",
          "3004142400",
          "3010363503",
          "3009051471",
          "3041276509",
          "3005641619",
          "3017492965",
          "3004024955",
          "3009144915",
          "3014273424",
          "3009337690",
          "3000215205",
          "3004369318",
          "1020279",
          "3007125392",
          "3015967359",
          "3002807328",
          "3002807036",
          "3009879967",
          "3010863450",
          "3004717358",
          "3010057495",
          "3009887475",
          "3012263546",
          "3004188066",
          "3010287737",
          "3009554293",
          "3008782585",
          "3007997036",
          "3013194153",
          "3009959868",
          "3009973336",
          "3017406274",
          "3003015552",
          "3019878714",
          "3015399803",
          "3010232390",
          "3007539489",
          "3002807231",
          "3004893332",
          "3003387384",
          "3003477135",
          "3006460162",
          "3009417901",
          "3012098629",
          "3007521480",
          "3013302242",
          "3008647423",
          "3017565094",
          "3003596442",
          "3018405359",
          "1000117423",
          "3005497913",
          "3006493760",
          "3010938185",
          "3006003617",
          "3015207155",
          "3008342029",
          "3001239363",
          "3010120339",
          "3025110439",
          "1218017",
          "1643264",
          "3005874553",
          "3009266751",
          "3010120135",
          "3002462840",
          "1000136269",
          "3008027912",
          "3030154790",
          "3010663372",
          "3001236277",
          "3010666513",
          "3008812560",
          "1928237",
          "3008992889",
          "3005985723",
          "3005532635",
          "3008837339",
          "3015877656",
          "3010531060",
          "3004635447",
          "3008397795",
          "2246552",
          "3015537048",
          "3010032903",
          "3025343017",
          "3010123206",
          "3000203466",
          "3011764740",
          "3004719693",
          "3012227716",
          "3016619943",
          "3003120897",
          "3005180920",
          "3000205591",
          "3016075969",
          "3040335671",
          "3010049501",
          "3009293341",
          "3016438694",
          "1000517406",
          "3007282753",
          "3009381600",
          "3026576110",
          "3005216725",
          "3006125668",
          "3012966183",
          "3015831087",
          "3002806725",
          "3005012805",
          "3016496998",
          "3020375573",
          "3010047454",
          "3017980495",
          "3005083075",
          "3036756245",
          "2648623",
          "3014680795",
          "3013422236",
          "3003064967",
          "3017123055",
          "3043543260",
          "3014782362",
          "1221053",
          "3012401682",
          "3009394448",
          "3003678543",
          "3040312029",
          "3003569173",
          "3024999441",
          "1000132370",
          "3009493875",
          "3009049161",
          "3009165919",
          "3004513872",
          "3014479313",
          "3010141593",
          "3010009632",
          "3010039400",
          "3007728266",
          "3009888740",
          "3012131184",
          "3007441485",
          "3012428435",
          "3004464325",
          "3039520898",
          "3002807830",
          "3010408671",
          "3005969827",
          "3004727093",
          "3004060780",
          "3001284227",
          "3010120148",
          "3016443334",
          "3018565653",
          "3014411622",
          "2647346",
          "3013682146",
          "2081135",
          "3021632375",
          "3002929183",
          "3002619595",
          "3008110533",
          "1000306303",
          "1450662",
          "3038503932",
          "3030400539",
          "3001049282",
          "1221763",
          "3000719653",
          "1000116116",
          "3008583793",
          "3000209794",
          "1314417",
          "3004153896",
          "2011171",
          "3000717902",
          "2027487",
          "1043653",
          "3003200961",
          "3011674645",
          "3033536319",
          "3017823294",
          "2027467",
          "3006639944",
          "3003661942",
          "1000135253",
          "3021008900",
          "3018094310",
          "3012120772",
          "3000215273",
          "3009973505",
          "3001592626",
          "3003955307",
          "3009871135",
          "3020703621",
          "3006128100",
          "1828464",
          "3004499989",
          "2219920",
          "3011187779",
          "3004016520",
          "1526439",
          "3009962553",
          "3010009693",
          "3010197224",
          "3014680149",
          "3003714383"
        ],
        "device_name": "Orthopedic Stereotaxic Instrument",
        "medical_specialty_description": "Neurology",
        "regulation_number": "882.4560",
        "device_class": "2"
      }
    }
  ]
}