{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-05",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "183091",
      "product_res_number": "Z-2948-2020",
      "event_date_initiated": "2020-07-31",
      "event_date_posted": "2020-09-09",
      "recall_status": "Open, Classified",
      "res_event_number": "86211",
      "product_code": "LWJ",
      "k_numbers": [
        "K003016",
        "K091423",
        "K100866"
      ],
      "product_description": "PROFEMUR Titanium and Cobalt Chrome modular necks, Item Numbers and descriptions:\nItem Number\tDescription\tDescription 2\nPHAC1202\tPROFEMUR NECK NEUTRAL\tSHORT COBALT CHROME\nPHAC1204\tPROFEMUR NECK NEUTRAL\tLONG COBALT CHROME\nPHAC1212\tPROFEMUR NECK A/R VAR/VAL 2\tSHORT COBALT CHROME\nPHAC1214\tPROFEMUR NECK A/R VAR/VAL 2\tLONG COBALT CHROME\nPHAC1222\tPROFEMUR NECK A/R VAR/VAL 1\tSHORT COBALT CHROME\nPHAC1224\tPROFEMUR NECK A/R VAR/VAL 1\tLONG COBALT CHROME\nPHAC1232\tPROFEMUR NECK A/R 8DG\tSHORT COBALT CHROME\nPHAC1234\tPROFEMUR NECK A/R 8DG\tLONG COBALT CHROME\nPHAC1242\tPROFEMUR NECK A/R 15DG\tSHORT COBALT CHROME\nPHAC1244\tPROFEMUR NECK A/R 15DG\tLONG COBALT CHROME\nPHAC1252\tPROFEMUR NECK VAR/VAL 8DG\tSHORT COBALT CHROME\nPHA01202\tPROFEMUR NECK NEUTRALSHORT\tSHORT\nPHA01204\tPROFEMUR NECK NEUTRAL\tLONG\nPHA01206\tPROFEMUR NECK NEUTRAL\tX LONG\nPHA01212\tPROFEMUR NECK A/R VAR/VAL 2\tSHORT\nPHA01214\tPROFEMUR NECK A/R VAR/VAL 2\tLONG\nPHA01222\tPROFEMUR NECK A/R VAR/VAL 1\tSHORT\nPHA01224\tPROFEMUR NECK A/R VAR/VAL 1\tLONG\nPHA01232\tPROFEMUR NECK 8DG A/R\tSHORT\nPHA01234\tPROFEMUR NECK 8DG A/R\tLONG\nPHA01236\tPROFEMUR NECK 8DG A/R\tX LONG\nPHA01242\tPROFEMUR NECK 15DG A/R\tSHORT\nPHA01244\tPROFEMUR NECK 15DG A/R\tLONG\nPHA01252\tPROFEMUR NECK 8DG VAR/VAL\tSHORT\nPHA01254\tPROFEMUR NECK 8DG VAR/VAL\tLONG\nPHA01256\tPROFEMUR NECK 8DG VAR/VAL\tX LONG\nPHA01262\tPROFEMUR NECK 15DG VAR/VAL\tSHORT\nPHA01264\tPROFEMUR NECK 15DG VAR/VAL\tLONG\n\nIntended Use: MPO Total Hip Systems are indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients.",
      "code_info": "All lots 1431860 1435705 1430154 1431065 1431308 1436522 1437003 0298131491437686 06113732261437685 1286531651437687 1430775 1431243 1431309 1431622 1431825 1431864 1432439 1432440 1433425 1433430 1433432 1433433 1434152 1435449 1435602 1435699 1435700 1436524 1436526 1436556 1437004 W123831741436428 1433424 1434969 14080361437684 1433431 1434993 1436525 1436527 1436528 1438068 1438130 1435702 1437055 1438069 1438131 1438455 X013918551437840 14277201440118 1434151 1434965 1435450 1435701 1435727 1435728 1436529 1436557 1436713 1438245 1438609 1438456 1438775 1438776 0398262151439704 1435603 1436712 1438872 1439145 V072510101439739 W0989675B1439738 1437581 1437877 1438664 1438873 1439144 1439379 1439380 1439387 1439582 U06811981440149 U06811981440151 V062154321440171 V092442911440169 1432436 1434964 1436365 1439944 1439945 1439946 W063349591440289 W073506871440084 1437451 1437582 1438953 1439581 1439947 U11852411440150 V062154311440165 V072329521440166 W093590691440288 W09919661440428 W103727321440287 1435852 1436366 1440186 1436367 1433438 1434153 1436555 1437050 1437580 1440654 1436364 1436711 1440187 1438954 1440655 1437057 1437876 1438297 1438454 1439325 1439512 1439513 1440239 1439323 1439750 1439752 1440238 1440575 1440576 1441171 1439514 1441170 1441535 1439753 1442404 01010095631444294 01112878341444296 0197579261444305 0197610691444306 0197842261444320 0285403251444307 0285523101444308 0285562241444309 02895142011444310 0297967521444312 0485730671444289 06113570991444291 0686252461444293 0798931781444290 1441359 1441718 1442049 03010220831444429 03113107711444440 03113335641444441 03113336361444442 0374147011444444 0397842251444450 0397913511444451 0398054701444453 07113789351444150 08011853901444155 1439324 1441410 1441716 1441961 1441962 1442050 1442313 1442407 1426645 1441492 1442201 1442405 1442406 1443050 1443053 1444193 1443262 1443489 1443490 U04705491444408 V112725371443806 W103767491443808 X064183591443810 03010819061444439 0363229471444443 0385562251444446 0385670021444447 0385730661444449 05113531101443811 1444194 1444342 W053277871443804 1440184 1440185 1443263 1432950 1444631 1444902 1440189 1444632 1445653 1445921 1445654 1446056 03113263491447082 04113343381447081 10012317451447124 1444903 1446966 W113767461447122 W113767531447123 W113857491447121 1447399 1447599 1446055 1446287 1446288 1446968 01112894471448087 1440127 1445938 1446020 1446939 1447397 1447600 W113767521447939 0586023501448096 1446176 1447786 1447787 1448190 V112725451447945 1446052 1447215 1448058 1448059 1448060 1448061 1448063 1448064 1448191 1448192 1448927 1448928 01112886231449440 05113539081449442 05113565991449443 0598437561449701 06011499591449438 08011897241449439 0986935661449700 1074719851449436 1074766231449711 11012435971449434 12012613881449435 1446021 1446053 1446177 1446938 1446940 V122874111449452 14154801449569 1444905 1447216 1448382 01010147751449539 02112969061449756 02113041371449534 03113144021449766 03113338081449531 0398310121449537 05113533911449792 05113652461449791 06011555561449765 07011838171450002 11460621449632 1299867951449762 14064991449764 1444904 1447213 1448381 1449428 1449429 U03667141449793 U06779811449794 U07764831449795 U07764931449798 U09852421449799 U10852431449797 U111161981449836 U111162101449819 U121437931449820 V062017171449800 V082154331449451 V082154331449801 W023009161449812 W023087681449805 W023087711449821 W033145651449822 W043212111449802 W043212131449807 W053212141449813 W063277911449823 W063349591449803 W063349661449824 W073349611449808 W073349621449814 W073441571449815 W073506901449816 W093590661449817 W09888841449809 W0989685111449818 W103590651449810 W103727301449804 W103767501449811 W103767561449825 0786716701450491 1448969 1448923 1448924 1448970 05113533871450501 1447214 1450118 1450206 1450208 W023143321450794 W033145591450795 0197842241451697 0398146081451827 04010969711451830 04011005261451832 04011041821451834 04113390251451835 0498281021451844 0498281031451845 0498383671451847 14456531449473 1447234 1449727 1450207 U06716531450796 U121355681451032 W013009121451017 W023090201450790 W023090201450792 W023090201451018 W033145561451020 W033145591450981 W073441541451026 W073443721451038 W0989693A1451044 W09919661451033 W103727291451039 W113767431451027 W123831761451040 1087165731450953 1449728 U08814851451031 W023143311450980 W073349581451023 W073441541450978 W093590691451029 W09896821451043 01112878341451696 02010398131451698 03010598641451700 03113234941451701 0398164731451828 04011041811451833 0497983621451840 0498281011451842 0498305411451846 0498383681451848 0498489881451849 05011398691451850 05113652481451851 0598352541451854 06011629991451855 06113570991451856 0698693431451866 0698876431451867 0698887841451868 1448769 1448971 1450791 1450793 1451169 1451170 01010209681452418 01010348151452419 01010356271451695 01010356271452380 0197801011452421 0197842261452381 02010398131452382 02010526571452424 02112949751452425 0297938031452427 0298024131452428 0298107921452431 03010819061452384 03010882901452610 03010950931452611 03113336361452385 0398274481452616 0398305431452617 04010767531452619 04011033941452620 04011041821452387 04011064471452621 0498054691452388 0498281001451841 0498489881452389 05011413401452624 0698717001452668 07011650861452671 07011692901452672 0798751101452673 0798760101452116 0798832121452119 0798850801452674 0798931791452393 0999258591452675 0999266771452676 1086935791452313 1099266791452317 1099266801452318 1099329651452678 1099486041452347 1099552971452348 11466421449734 11693481449761 1199795781452357 1199795791452358 1199858691452359 12012719041452360 1299890961452376 13986161452379 1448789 1449751 1451214 U06716521451015 U11888781451016 W093727261451024 0197842191452422 02010483451452423 0298156491452606 03010584321452607 03010634371452608 03010869181452609 0398164731452386 0398336321452618 06011450281452628 06011596281452630 0698887851452112 0798693441452115 0798760131452117 0798760141452118 0798931791452120 0999266811452127 1086935771452133 1086935781452310 1087182671452315 1099574741452350 1199741481452356 1299952611452377 1299952621452378 1450110 1451212 10838881449759 1449753 1451213 1451215 1451349 1451357 1451823 1451824 1451825 V122874091451022 W093727261451025 1450116 1451216 1451826 U121374451451042 W013009131451041 W023090181450977 W113767451451021 W113767451452511 1450996 1451217 1451354 V031690401452846 W073349631452853 W093590691452850 W113767431452844 X033925031450518 0285403271452834 0463345721452871 0474161341452868 0474237731452862 0485730671452861 0563444871452835 0585942351452860 0663564271452833 0686252461452836 0686252471452832 0686375171452870 0686375191452830 0686375201452865 0686375211452872 0686375221452831 1450993 1451218 U02667041452917 W113857491452914 0685953181452829 1449430 1449752 1450113 1451168 1451350 1452533 1453089 1453090 0285366341453638 0285515951453640 0385612011453641 0385749931453635 0485942471453637 0585991031453649 0752110021453670 0874618721453668 0874641721453671 0886576391453632 0886576391453663 0974730831453681 0986862821453664 1052603271453667 1074888181453666 1174899351453636 1174905181453673 1275063741453646 1449431 1450552 1453319 1453320 1453366 0185167121453653 0285523101453682 0385698381453644 0385776411453647 0686375191453683 0686375201453633 0686375221453634 0986862841453674 1063797651453669 1074803001453630 1074803001453665 1074803011453672 1074815181453631 1074888181453639 1450112 1452534 1453367 1453703 W123831761453512 01010179871454266 01010304091454267 01010368331454254 0197557121454255 0197664621454268 0197721881454250 0197801031454269 0197814401454270 0197825161454271 0197825161454337 02010526571454272 0297938021454273 03010584311454275 03010634371454256 03010634381454338 03010643981454339 0398262231454278 04010991651454279 04011005191454280 04011020201454281 04011033951454282 04011086271452622 04011086271454283 04011179011454284 05011363531454343 05011363541454285 05011363541454344 0899266761454263 0999329641454289 0999393811454290 0999410401454291 0999428241454292 1086935701454293 1086935711454294 1087126991454346 1087164691454295 1099311331454264 1099574701454296 1099605211454265 1199795751454298 1287520481454334 12910062841454335 12910095571454359 12910095591454249 12910095591454336 129833630011454253 14512121454405 03010784111454340 0498352531454341 05011326361454342 0798778051454345 1099428251454347 1099626711454348 1199731831454297 1453933 01112893861454744 1451167 1452462 1453321 1453704 1451166 1452786 1453932 1454437 1454643 1453927 1454644 0385776371455240 0685807981455241 0686197241455244 0686531711455238 0786630421455245 0885807901455242 0885807911455243 1074848451455239 11012546091454997 1275063791455246 1453928 1453946 1454436 1450994 1451964 1454770 1455008 1455009 1454771 1454882 1454883 10011713661455611 1450114 1450789 1451965 1452460 1454645 1455595 1456061 0986831451456404 1449432 1451951 1452461 1452465 1452787 1452790 1453696 1453697 1455418 1456045 R11976081456079 U05737651456077 U09811991456078 0197864881456655 04011098971456741 1454772 1456046 1456121 1456361 U09811991456399 V102725421456400 05113646761456876 V031690401456867 V041858521456869 V092438591456870 0598519051454286 1452789 1453443 14539331457036 14544371457034 1456062 1456122 1456783 V052153671456868 W013009111456865 W093590691456866 1456123 1456362 1456784 V041690411456987 W013009131457210 W023090201457211 1450470 1456124 1456781 1456782 V021374691457389 V041858521457390 V10100772A1457212 V041859401458073 V052153671458092 V092442911458078 V122874091458081 W043277831458056 W073441541458059 W09896771458076 W123831751458089 W123831811458071 06113573301458410 09012185781458428 14157521458424 W013009121458087 1457484 1457485 1457737 1457971 1457973 1458653 W0189693A1458069 W09896761458060 1451352 1456119 1457483 1457486 1457736 1457738 1457739 1457970 1457974 1457975 1458654 1450995 1452464 1456120 1458868 1458870 01010023091459188 01112872081459191 06113573301459186 1087164961459193 1451966 1451967 1452788 1453699 1454012 1454781 1457972 1458869 1458871 02010712471460419 05011265561460421 089771930011460424 089815779011460422 12910020351460417 1459305 W023090201459860 W033145571459822 W033145571459857 W053277851459846 W073349581459827 W073349581459862 W073349631459837 W073443721459841 1454008 1459306 1459379 W013009121459825 W043145601459819 W043212101459859 W053277851459861 W063349641459858 1459380 1459381 1459382 1459676 1459678 01010398051460415 01010398061460418 04011116161460420 1199892391460416 12910020351460423 1459679 1459745 1459746 1460086 01010245651460383 1459677 1459680 1460087 1460088 1460617 0385669731461388 0485879521461383 08011913901461386 0874714591461162 1459828 1460095 1460096 1460618 1460624 1460876 1460089 1460619 1460620 1460625 1461026 1461029 W0889678B1461614 1461030 1461031 0398274501461970 0498336281461962 0874664451461892 0886766761461960 10012350331461968 1086607521461969 1086862881461955 10897610011461967 1163839531461893 1460877 1461027 1461028 1461032 04113321771462439 05113539141462450 1461033 1461923 W123831751462177 01010355941463447 01010433461463472 0197693791462423 0297967511462701 03010768381463397 03010796731463474 03010829301463400 04010981721462800 04011198381463434 089772182011463476 089815787011463477 1086576411462799 1086862881462661 1199813321463418 1199813321463421 1286531651462798 14381301462703 1461034 1461924 1462132 W063349641462650 W073349631462652 W113767531462653 W123831761462649 01010017461463457 01010209461463393 01010209461463408 01010231841463429 01010271781463458 01010284621463459 01010306751463440 01010306761463441 01010411221463416 01010433451463430 01010433451463439 01010471251463431 02010526311463442 02010538661463460 02010598201463395 02010598211463443 02010608581463396 03010768401463398 03010783871463399 03010829311463401 03010838901463433 03010896181463444 06011515671463435 06011555561463411 07011756031463402 07011779511463403 089772180011463417 089772181011463436 089815780011463404 089815780011463409 089815782011463414 089815786011463437 089815787011463446 089902014011463405 1199813321463415 12910114151463406 12910147411463407 12910186001463438 12910198351463419 0398274501462938 1462131 1462133 1462135 1462455 W103727271462651 1453444 1462134 1462136 1462458 1462459 1462456 1462457 1462460 01010411191464085 1462987 1462988 1462989 1462990 1462991 1462992 1463549 1463550 1463850 1463851 11062351462474 07011756031463412 14621341465808 1463973 1463974 1463976 1463981 1464364 1464368 02113022721465393 07113799281465391 14154861465394 1463849 1463977 1463979 1463980 1464366 1464369 1463975 1464365 1464367 1465033 1465035 1465037 1465243 1465246 1465509 1465034 1465036 1465038 1465244 1465245 1465510 1465511 1465573 1465574 1465575 1465831 1465835 0197913531467425 0297842281467426 0398024171467427 04010931061467428 0498373751467429 1465832 1465833 1465834 1465836 1466171 1466172 1466173 1466275 1466338 1466339 1466276 1466277 1466463 1466464 1466465 1466462 1466797 1466798 1466891 1466893 1466894 1466898 1466896 1467060 1467163 1467164 1467167 1467168 1467291 1467292 0586057091467897 1287431811467925 1466899 1467392 01010185531467440 01010231951467441 01010231951467840 01010245691467442 01010245691467841 01010262141467443 01010262141467842 0109952641467444 0197579521467447 0197700431467448 0197700441467449 0197913531467450 0197913531467847 0285494821467911 0285520141467910 0285521491467909 0285547641467922 0285604581467923 0297842291467859 0297842291467957 0297913541467837 0298054641467860 03010698521467456 0398024171467457 0398054631467458 04010931061467862 0485775371467967 0498373751467459 0498373751467864 0498383681467042 05011422071467460 05011422071467867 0586057091467908 0598610251467869 0674388601467968 0685953181467969 0686279371467898 0698611191467432 0698611191467935 0798931801467552 0886784701467921 0886862911467920 1087078511467556 109770043011467944 1098703061467435 1098703061467557 1099329681467046 1099627701467946 1199685201467436 1199685201467973 1199741491467437 1199741491467558 1199741491467948 129802419011467838 1299952651467835 1456420 1467061 1467165 1467166 1467274 1467293 1467296 T091285441467043 X013831821467914 X023884811467917 X023964141467912 X023964141467913 X034012381467916 X034012381467924 X044108751467919 X054183621467918 X074201641467915 1467632 0463336721468744 0586085061468739 0686279391468720 0763562501468721 0863651571468745 0886784701468722 089815779011468323 1263900731468749 1467294 V041858571468298 V051858561468296 W033145611468297 W043277831468293 W053277851468294 01010185531467839 0197579521467845 0197700441467846 0285403301468727 0285494821468728 0285520141468730 0285521491468732 0285547641468733 0285565811468735 0285596081468737 0285604581468738 0297842281467858 04010926281467861 0498347221467863 06011505641467932 0686279371468740 0686279381468741 0698611191467960 0698899061467936 0798931811467938 0799008061469424 0886862911468723 0886862911468742 0963762441468724 0974717421468747 1063853541468748 1086576441467943 1099516761467945 1099627701467965 1199795801467949 1275063961468743 1287254031467833 1287441811467952 12910062861467834 12910062861467953 1299952651467955 1467631 1468046 1468194 R121008071467445 T121346971467865 U01632491467941 U03663431467956 U03667211467930 U06665751467931 U06665751467959 V011507561468301 V021355651468291 V021567311468295 V042122891468300 V112767211468292 W043212131468299 X034012381468725 0285403291468726 1467393 1467815 S011019101467446 T021217411467947 V031690461468302 0298024191467987 0398281101467992 0698899061468005 1199741491468012 1467390 0109952641467976 0197579521467977 0297913541467986 0398024171467975 0398024171467988 04010931061467993 0498373751467994 0598610251467999 0598611171468000 0598611181468001 06011505641468002 06011505641468020 0786607501469421 0798931811468006 1086576441468009 1087078511468010 1087078521468011 1287254041468014 1287431811468015 1287441811468016 12910062861468017 1299952651468018 W113857501469390 01010186041468869 01010209731468870 01010304131468871 01010348171468872 01010386611468873 0109988671468874 0109988681468875 0197800691468954 02010471511468959 0285476291468961 0297968001468866 03010643991468969 03010656581468970 0385716731468973 0398209701468976 04011020171468985 04011041851468986 04011041861468987 05011318261468990 0574307511468993 0585953281468994 0698702291469129 0698769251469130 0698769271469131 07011681461468836 0785808671469132 0786576421469133 0798730361469134 0798931821469135 0885808661469136 0885808701469137 0886619971469138 0886831431469139 0899220781469140 0974770421469141 0999369081469142 0999398161469143 0999398171469144 0999398191469145 1074804311469146 1074808061469147 1074878451469148 1074882911469149 1086935751469151 1087091501469152 108709151101469153 1087091531469154 1087098451469156 1087161721469157 1468189 1468190 1468191 1468195 01010147771468867 0197721911468876 0197721921468953 0197800701468955 0197814441468956 02010170811468957 02010411841469572 02010538901468960 0285516721468962 0297950661468963 0297967541468964 0297968001468965 0297968021468966 0298041721468968 03010784161468971 0385716721468972 039820969146897570ZM22C5 0398221201468977 0398310121468978 0398310131468979 04010846971468980 04010906851468981 04010926291468838 04010926291468982 04010926301468983 04011020181468839 04011041861468840 0485942301468831 0498398041468988 0498429911468849 05010931071468989 05011386361468991 05011400231468841 05011400231468992 0586140551468995 0586140561468996 0598433241468997 0598577671468998 06011481391468833 06011481391469000 06011485081469003 06011606161468834 07011681461468844 0899220781468845 0899258701468854 1187420181468857 1287536661469564 1299476681468859 1299476681469565 1299941941469566 1299941951469567 1467391 1468192 1468193 1468198 1468896 01010135791469569 0109988681469570 0185413211468847 0297967991468848 0298005691468967 0398209701468846 04010931081468984 04011041851468835 05011318261468832 05011318261468861 0585942321468850 0585942321468862 0598475261468851 0598576901468842 06011481391468863 0698760191468843 0798832111468837 0798832111468864 0798909971468852 0974754651468855 0999266821468856 1199698501469560 1199753071469561 1199837801469562 1199868011468858 1199868011469563 1299941951468860 1299962141469568 1467394 1468897 1469212 1469513 1469548 1086619961469150 1468196 1469213 1453401 1453402 1453408 1468197 1469240 1469547 1469964 1469966 01010306761470480 01010398081470267 01010433451470475 01010433451470478 01010458651470472 02010518961470457 02010538671470474 02010598201470396 02010598211470481 02010695601470482 02010726541470154 0274057861470569 028101641021470587 0285403291470570 0285520141470572 0285547641470573 0285565811470574 0285596081470575 0285604581470576 03010768381470268 03010783871470269 03010783871470271 03010783901470483 03010829301470397 03010829311470398 03010896191470479 03010979501470473 04011086361470156 04011087141470157 04011107021470155 04011107021470158 04011116161470159 0574321961470577 0586085061470578 0586085071470579 0663544901470580 0686279371470581 0874613771470582 0886784701470583 0886862911470584 089772179011470279 089815779011470160 089815780011470394 089815780011470399 089815782011470458 089815786011470477 0974717421470585 1199813321470459 1199813321470466 1199885971470460 1199885971470467 1199892401470468 1199950091470395 1274719791470586 12910114151470400 12910120691470401 12910198351470461 12910198351470469 1469934 1469963 1469965 1470299 1470300 1470301 X034012381470588 1453398 1453399 1470297 1470302 1470560 1470561 1471187 03010896191470881 089772182011470882 1470562 0185255571470938 0297938021470952 04011086271470956 0598519051470953 0999258591470944 0999329641470945 1470298 1470745 1471188 0197721881470939 03010584321470940 0498417101470941 0674481131470958 0798676891470942 0798778051470943 0999393811470946 0999428231470947 1099525921470948 1187334871470949 1199809871470950 1275063831470957 1453405 1469237 1471189 1471517 1471518 1435710 1468941 1469929 01010356111470463 01010411221470464 089772180011470465 089772181011470476 1468942 1469239 1469512 1470747 1470748 1471519 1471520 W123831811472320 1470564 1470746 1469543 1471773 1471775 1471856 1472657 T031217431468974 U02663541468865 1470769 1471866 1470770 1471858 1472636 1472687 1469545 1469546 1472633 1472658 1472659 1472882 1473029 1472686 1473030 1473031 1473032 0485841601473346 1472637 1473332 1473333 1473268 02113041311473318 0297967511473532 0297967511473534 04113380721473307 06113696621473530 06113764131473312 09012042731473371 1086607521473319 1086607521473528 1287441801473533 1287520451473529 13976741473308 14071071473317 14341521473531 1469238 1472883 1473334 1471774 1471776 1472471 1473018 1473019 1473266 1473269 1472642 1473267 1473335 1473652 1473653 1472643 1473264 1474008 1474010 1471860 1474009 1453400 1473265 1474007 1474214 1474215 1474546 1474547 1474549 1453397 1453406 1453407 1473654 1474548 1474992 1474993 1474012 1474256 1474962 1474994 1475532 1475533 1474257 1474545 1474995 1475498 1475499 1453395 1475534 0197576031475992 07011732221475990 1087165731475991 14401861475994 1453404 1472430 1473416 1474213 1475791 1476035 X033964091475528 1474140 1474483 1474544 1474990 1474991 1475535 1473118 1453403 1474011 1474141 1476654 1453409 1453410 1458172 1476268 1476272 1476485 1453396 1458164 1474212 1475122 1476182 1476270 1476656 1477106 1477107 1477108 1477648 05113426951475510 11515581475508 1477006 1477007 1477649 1472634 1474963 1475120 1475790 1477650 1458184 1476269 1477677 1475584 1476891 1477009 1477651 1477675 1458173 1475585 1477008 1477011 1477242 1477247 1477676 1477678 1477679 1477680 1478295 1478296 1478297 1478298 1478299 1478301 1458174 1458175 1458176 1458177 1458185 1474284 1474997 1474999 1475000 1475121 1475789 1476487 1476488 1477243 1477466 1477472 1477949 1478034 1478300 1478793 1479063 1479198 1479199 1479200 1478294 1474996 1475001 1476029 1476030 1476658 1476659 1477467 1477793 1479062 1477244 1477792 1477794 1477970 1478036 1479204 1479206 1477245 1477246 1478320 1479205 1480407 1474998 1477468 1478321 1478634 1479822 1479823 1479826 1479827 1480409 1480734 1480735 1479059 1480267 1480268 1480397 1480408 1480410 1480733 1480982 1477469 1477973 1481022 1478318 1479202 1481764 1453411 1458166 1476032 1478792 1479201 1481023 1481024 1481025 1481258 1481765 1458167 1458182 1480654 1481026 1481027 0197611181482715 0285403251482716 0298146101482598 0397842251482645 0397842251482717 0398164731482646 0398164741482600 0485879501482647 0498383681482648 0585942331482719 0598352541482601 0598437491482720 0652328641482721 0685953181482602 0686375191482606 0698876441482723 0798783031482637 0798931781482638 0798931791482639 089637518011482640 0986862841482682 0986862841482724 1099486041482641 1099486041482726 1199574731482643 1199795781482644 1199795791482727 1263942411482728 1480405 1480653 1480981 1481766 0498383671482718 0686375191482722 1479203 1481021 1481028 1482023 0197664641482918 0385670021482599 0398221201482919 0686200271482920 0686314751482921 0686314761482922 0999398181482923 1087164941482912 1199698511482914 1478791 1479824 1481020 1482474 1482475 1482476 U03667181482642 X013831791482684 X023884781482686 X023964111482688 X033964121482689 X044012351482690 X044012361482693 X064183581482696 X064183591482713 X074201621482714 1458168 1480402 1482025 1483284 1483285 1479060 1458169 1479061 1479825 1480403 1482027 1482029 1482477 1483576 1086935731483650 1086935751482909 1087009661482910 1087124661483651 1087164891483652 1087164931482911 1099329681482725 1099486061483653 1099574761483654 1099664391483655 1163908421483656 1187310011483657 1187334731482913 1199664371483658 1458165 1478794 1480247 1482478 1483982 1480265 1480406 1482479 1484105 1484186 1483878 1478795 1480404 1480471 1481761 1484101 1484102 1484106 1484107 1484108 1484109 1484110 1484187 1484188 X013884851482915 X044012431482904 X044012441482916 X054183661482905 X064183671482917 X074201671482906 1481454 1481762 1481763 1483574 1483932 1483934 1484103 1484826 1484827 1485008 0109988671483672 0197619521483673 0297950671483767 0498429911483675 0598475261483676 0598577671483677 0798832111483680 0999220771483681 0999266821483682 0999398161483683 1087164931483684 1187310011483685 1187334731483686 1287403561483687 1299962141483688 1484104 0197842201485392 0298107921485393 0298200441485394 0398216021485454 0498352611483667 0498417101485455 0586140571483668 0686066141485456 0686200241483669 0686375141485458 0798751101485459 0798898921483670 0798909971483671 0885807901485461 0999266751485391 12910062841485390 1453394 1458178 1483575 1483936 1483981 1484481 1484802 1484803 1484828 1485007 1485059 1485144 1485272 1485275 1485290 1485561 1485575 1480398 1485267 1485276 1485291 1485562 0899266761485471 0999258591485473 0999258601485474 0999393811485475 1063833491485476 1099671161485477 1187370341485478 1199731811485479 1485273 1485534 T071269971483679 1458162 1485274 1485563 1485576 1479300 1485277 1485535 1486030 1486041 1485060 1486463 10686301482234 1482289 1482290 1483933 1484480 1485564 1485811 1486042 1486464 1486465 1486467 1486869 1486872 1486875 1486876 1480399 1486039 1486049 1486468 1486873 1481018 1481019 1486014 1486199 1486874 1487506 1487507 1487508 1487510 1458179 1480128 1485143 1486050 1487943 1482768 1483980 1486051 1488214 1488215 1489052 1482769 1485536 1486040 1486870 14779731489146 1485810 1486200 1486878 1487511 1487944 1488148 1488785 T111347771485472 1486203 1487509 14776491485245 1488198 1489345 1486052 1487022 1487722 1488502 1488503 1488504 1488505 1488781 1488782 1489987 1489988 1489989 1488783 1489990 1489999 1490000 1476489 1476493 1486201 1488200 1488204 1489053 1489346 1489991 1489995 1490001 1490002 1490859 1490860 1458163 1458170 1476491 1476712 1476716 1486202 1486871 1486877 1487945 1488205 1488206 1489054 1490796 1490854 1490866 1491776 1487515 1488199 1476713 1488211 1488212 1488498 1489993 1489992 1476490\n1476492 1476714 1488499 1488500 1488501 1488784 1489703 1489704 1489994 1489996 1489997 1489998 1490867 1490872 1492182 1492268 1492269 1476715 1487721 1490873 1490873 1491777 1492106 1492107 1492271 1486204 1488786 1490862 1490870 1492104 1492105 1492400 1492401 1492538 1487723 1490871 1492270 1492422 1492423 1492272 1489287 1489532 1492536 1492537 1493446 1493447 1493448 1489705 1490861 1490868 1493002 1493449 1493682 1493683 1493686 1494068 1494069 1494724 1494722 1486881 1490857 1490869 1493003 1493539 1493769 1494631 1494720 1494726 1487348 1488652 1490856 1490858 1490864 1493550 1493768 1494073 1494189 1494728 11249251495803 14154861495806 14261181495804 1490863 1494732 1493688 1494404 1494405 1494734 1495334 1490865 1493540 1494187 1494188 1494902 1495335 1495732 1492108 1494852 1494903 1495731 1495733 1495734 1496004 1496005 1496948 1497046 065209594101498430 07113899921497041 13982671497040 13982671497042 1495337 1495730 1496006 1496340 1496341 1496406 1496407 1496949 1497045 01010271781496271 01010284621496272 01010355971496274 01010356111496250 01010398081496236 01010411221496252 01010458651496267 01010471281496302 02010608581496237 02010726551496232 02010726561496254 02010743021496268 03010768321496263 03010838911496264 03010979501496265 05011398781496233 089771930011496234 089772180011496255 089772181011496269 089815779011496235 089815786011496266 089815787011496304 1199813321496256 12910147411496249 12910163101496257 12910198351496258 12910209591496259 1299961801496270 045203692101498593 1492427 1493684 1494070 1494632 1494633 1494685 1497047 1497048 1497049 W123831801497032 X033964071496475 X054183631498826 X064183541496477 X074183651498751 1492273 1494071 1498636 X011112371496474 X013831771497027 X013884801497033 X013884801497039 X0193876A1497034 X023964131497035 X033964101497028 X044012311496476 X044012371497036 X044183601497037 X064183571497029 X064201631497038 X074396921497030 X084288451497031 X104201591496478 X104288421496479 1492428 1493685 1494072 1494684 1495972 1496189 1496381 1496382 1497738 1498635 1498637 1498974 1498975 1492274 1494399 1497101 1497739 1498223 1498976 1498977 1498978 1498979 1496946 1500008 1500009 1492109 1492336 1494932 1495339 1499211 1499298 1500222 1500223 1500669 1500767 1496947 1500224 1500225 1500666 1500668 1500747 1500750 1500768 1500787 1500789 1501085 1501086 1501427 1501084 1501087 1501428 1501587 1500214 1500215 1500667 1492402 1494730 1494934 1501403 1501404 1501586 1501588 1501589 1501638 1502264 1493083 1493770 1494401 1495970 1495971 1501848 1501637 1492403 1501076 1492404 1492405 1494935 1495336 1496832 1501077 1501847 1502480 1501078 1501079 1500221 W053277851503394 1494403 1496188 1497102 1498224 1501374 1502265 1502510 1502511 1500744 1498980 1500226 1501083 1501375 1502589 1502596 1502597 1502758 1502760 1502912 1503305 1497732 1497733 1497740 1501376 1501377 1502491 1502590 1503308 1504058 1504524 1492406 1492407 1492408 1492409 1492411 1492412 1497741 1500054 1502481 1502759 1503094 1503306 1504056 1504528 1504529 1505180 1505190 1505395 1505961 1504525 1492410 1494933 1496732 1496835 1497744 1500216 1500220 1503093 1503310 1504236 1504237 1504526 1504527 1505189 1505191 1505385 1505646 1506866 1492413 1497745 1497947 1500217 1500227 1502323 1503311 1504057 1505647 1505951 1505962 1505970 1506033 1506186 1506613 1506867 1501846 1502324 1502913 1503095 1503907 1506032 1506190 1506614 1506616 1507171 1506612 1501082 1502914 1506187 1506188 1506723 1506762 1507170 1504239 1506615 1506761 1507166 1507169 1507172 1507250 1505396 1507252 1507253 1507802 1507804 1507165 1507251 1507259 1507803 1507806 1508653 1505163 1505383 1505384 1506180 1507805 1507807 1507810 1508652 1509196 1509201 0798611161510105 0799008081510106 14742151510101 14900001510103 1503309 1504238 1504240 1505952 1506181 1506189 1508382 1509207 1509208 1507256 1507258 1507811 1508383 1508938 1508976 1509370 1507257 1509197 1509278 1509371 1510289 1510290 1510292 1510305 1510309 1510790 1497948 1497962 1502593 1507264 1508937 1509279 1510295 1510791 1510792 1510793 1510910 1497949 1497952 1497953 1497954 1510313 1510316 1511915 1511916 1511917 1497963 1497965 1511923 1497956 1497955 1497968 1507255 1507260 1507262 1507263 1510911 1510913 1511924 1511925 1512326 1512327 1512406 1512677 X013884721513380 X023964051513381 X044012301513382 X054183521513379 02010743021512943 03010838911512945 08011913541512950 089772179011512938 089772181011512948 089772182011512951 1199813321512942 1497961 1501843 1501844 1502492 1502757 1505398 1507265 1508974 1509280 1509281 1509375 1510795 1511918 1511919 1512405 1512408 1512474 1513162 0752302471513697 0852452851513696 1507812 1510794 1510796 1512473 1506030 1507813 1509374 1510914 1513163 1513165 1513561 1506031 1510797 1497966 1501845 1503096 1503307 1507261 1509373 1510319 1510320 1510322 1510325 1510912 1510915 1510917 1511920 1511928 1511929 1511931 1512328 1512407 1512614 1512615 1512678 1512680 1512862 1512864 1513565 1514129 1507254 1512329 1512863 1513164 1513533 1513562 1513563 1497967 1512861 1513168 1513170 1514128 1514154 1513171 1513971 1513973 1515458 1513974 1514131 1514483 1514484 1515457 1514130 1514733 1515460 1515610 1515611 1514155 1514157 1515711 1511930 1513572 1513573 1514735 1515612 1515868 1516674 1516675 1515871 1510916 1513169 1514737 1514738 1515459 1515613 1515867 1515870 1515875 1515877 1516670 1516671 1516676 1516677 1516924 1517247 1517758 1509202 1514734 1517158 1517248 1513166 1514739 1514740 1515709 1515725 1515874 1516921 1517157 1517760 1497964 1505397 1509372 1513972 0686200241483669 0686375141485458 0798751101485459 0798898921483670 0798909971483671 0885807901485461 0999266751485391 12910062841485390 1453394 1458178 1483575 1483936 1483981 1484481 1484802 1484803 1484828 1485007 1485059 1485144 1485272 1485275 1485290 1485561 1485575 1480398 1485267 1485276 1485291 1485562 0899266761485471 0999258591485473 0999258601485474 0999393811485475 1063833491485476 1099671161485477 1187370341485478 1199731811485479 1485273 1485534 T071269971483679 1458162 1485274 1485563 1485576 1479300 1485277 1485535 1486030 1486041 1485060 1486463 10686301482234 1482289 1482290 1483933 1484480 1485564 1485811 1486042 1486464 1486465 1486467 1486869 1486872 1486875 1486876 1480399 1486039 1486049 1486468 1486873 1481018 1481019 1486014 1486199 1486874 1487506 1487507 1487508 1487510 1458179 1480128 1485143 1486050 1487943 1482768 1483980 1486051 1488214 1488215 1489052 1482769 1485536 1486040 1486870 14779731489146 1485810 1486200 1486878 1487511 1487944 1488148 1488785 T111347771485472 1486203 1487509 14776491485245 1488198 1489345 1486052 1487022 1487722 1488502 1488503 1488504 1488505 1488781 1488782 1489987 1489988 1489989 1488783 1489990 1489999 1490000 1476489 1476493 1486201 1488200 1488204 1489053 1489346 1489991 1489995 1490001 1490002 1490859 1490860 1458163 1458170 1476491 1476712 1476716 1486202 1486871 1486877 1487945 1488205 1488206 1489054 1490796 1490854 1490866 1491776 1487515 1488199 1476713 1488211 1488212 1488498 1489993 1489992 1476490 1476492 1476714 1488499 1488500 1488501 1488784 1489703 1489704 1489994 1489996 1489997 1489998 1490867 1490872 1492182 1492268 1492269 1476715 1487721 1490873 1490873 1491777 1492106 1492107 1492271 1486204 1488786 1490862 1490870 1492104 1492105 1492400 1492401 1492538 1487723 1490871 1492270 1492422 1492423 1492272 1489287 1489532 1492536 1492537 1493446 1493447 1493448 1489705 1490861 1490868 1493002 1493449 1493682 1493683 1493686 1494068 1494069 1494724 1494722 1486881 1490857 1490869 1493003 1493539 1493769 1494631 1494720 1494726 1487348 1488652 1490856 1490858 1490864 1493550 1493768 1494073 1494189 1494728 11249251495803 14154861495806 14261181495804 1490863 1494732 1493688 1494404 1494405 1494734 1495334 1490865 1493540 1494187 1494188 1494902 1495335 1495732 1492108 1494852 1494903 1495731 1495733 1495734 1496004 1496005 1496948 1497046 065209594101498430 07113899921497041 13982671497040 13982671497042 1495337 1495730 1496006 1496340 1496341 1496406 1496407 1496949 1497045 01010271781496271 01010284621496272 01010355971496274 01010356111496250 01010398081496236 01010411221496252 01010458651496267 01010471281496302 02010608581496237 02010726551496232 02010726561496254 02010743021496268 03010768321496263 03010838911496264 03010979501496265 05011398781496233 089771930011496234 089772180011496255 089772181011496269 089815779011496235 089815786011496266 089815787011496304 1199813321496256 12910147411496249 12910163101496257 12910198351496258 12910209591496259 1299961801496270 045203692101498593 1492427 1493684 1494070 1494632 1494633 1494685 1497047 1497048 1497049 W123831801497032 X033964071496475 X054183631498826 X064183541496477 X074183651498751 1492273 1494071 1498636 X011112371496474 X013831771497027 X013884801497033 X013884801497039 X0193876A1497034 X023964131497035 X033964101497028 X044012311496476 X044012371497036 X044183601497037 X064183571497029 X064201631497038 X074396921497030 X084288451497031 X104201591496478 X104288421496479 1492428 1493685 1494072 1494684 1495972 1496189 1496381 1496382 1497738 1498635 1498637 1498974 1498975 1492274 1494399 1497101 1497739 1498223 1498976 1498977 1498978 1498979 1496946 1500008 1500009 1492109 1492336 1494932 1495339 1499211 1499298 1500222 1500223 1500669 1500767 1496947 1500224 1500225 1500666 1500668 1500747 1500750 1500768 1500787 1500789 1501085 1501086 1501427 1501084 1501087 1501428 1501587 1500214 1500215 1500667 1492402 1494730 1494934 1501403 1501404 1501586 1501588 1501589 1501638 1502264 1493083 1493770 1494401 1495970 1495971 1501848 1501637 1492403 1501076 1492404 1492405 1494935 1495336 1496832 1501077 1501847 1502480 1501078 1501079 1500221 W053277851503394 1494403 1496188 1497102 1498224 1501374 1502265 1502510 1502511 1500744 1498980 1500226 1501083 1501375 1502589 1502596 1502597 1502758 1502760 1502912 1503305 1497732 1497733 1497740 1501376 1501377 1502491 1502590 1503308 1504058 1504524 1492406 1492407 1492408 1492409 1492411 1492412 1497741 1500054 1502481 1502759 1503094 1503306 1504056 1504528 1504529 1505180 1505190 1505395 1505961 1504525 1492410 1494933 1496732 1496835 1497744 1500216 1500220 1503093 1503310 1504236 1504237 1504526 1504527 1505189 1505191 1505385 1505646 1506866 1492413 1497745 1497947 1500217 1500227 1502323 1503311 1504057 1505647 1505951 1505962 1505970 1506033 1506186 1506613 1506867 1501846 1502324 1502913 1503095 1503907 1506032 1506190 1506614 1506616 1507171 1506612 1501082 1502914 1506187 1506188 1506723 1506762 1507170 1504239 1506615 1506761 1507166 1507169 1507172 1507250 1505396 1507252 1507253 1507802 1507804 1507165 1507251 1507259 1507803 1507806 1508653 1505163 1505383 1505384 1506180 1507805 1507807 1507810 1508652 1509196 1509201 0798611161510105 0799008081510106 14742151510101 14900001510103 1503309 1504238 1504240 1505952 1506181 1506189 1508382 1509207 1509208 1507256 1507258 1507811 1508383 1508938 1508976 1509370 1507257 1509197 1509278 1509371 1510289 1510290 1510292 1510305 1510309 1510790 1497948 1497962 1502593 1507264 1508937 1509279 1510295 1510791 1510792 1510793 1510910 1497949 1497952 1497953 1497954 1510313 1510316 1511915 1511916 1511917 1497963 1497965 1511923 1497956 1497955 1497968 1507255 1507260 1507262 1507263 1510911 1510913 1511924 1511925 1512326 1512327 1512406 1512677 X013884721513380 X023964051513381 X044012301513382 X054183521513379 02010743021512943 03010838911512945 08011913541512950 089772179011512938 089772181011512948 089772182011512951 1199813321512942 1497961 1501843 1501844 1502492 1502757 1505398 1507265 1508974 1509280 1509281 1509375 1510795 1511918 1511919 1512405 1512408 1512474 1513162 0752302471513697 0852452851513696 1507812 1510794 1510796 1512473 1506030 1507813 1509374 1510914 1513163 1513165 1513561 1506031 1510797 1497966 1501845 1503096 1503307 1507261 1509373 1510319 1510320 1510322 1510325 1510912 1510915 1510917 1511920 1511928 1511929 1511931 1512328 1512407 1512614 1512615 1512678 1512680 1512862 1512864 1513565 1514129 1507254 1512329 1512863 1513164 1513533 1513562 1513563 1497967 1512861 1513168 1513170 1514128 1514154 1513171 1513971 1513973 1515458 1513974 1514131 1514483 1514484 1515457 1514130 1514733 1515460 1515610 1515611 1514155 1514157 1515711 1511930 1513572 1513573 1514735 1515612 1515868 1516674 1516675 1515871 1510916 1513169 1514737 1514738 1515459 1515613 1515867 1515870 1515875 1515877 1516670 1516671 1516676 1516677 1516924 1517247 1517758 1509202 1514734 1517158 1517248 1513166 1514739 1514740 1515709 1515725 1515874 1516921 1517157 1517760 1497964 1505397 1509372 1513972 1514156 1515710 1515712 1515849 1515864 1515865 1515866 1515869 1516672 1516673 1517761 X054183551517080 X104201601517085 1512679 1519215 1511926 1511927 1514736 1515878 1517763 1517764 1518338 1518339 1497970 1513167 1516923 1517759 1497971 1497973 1497974 1497987 1497988 1497990 1497990 1497991 1497991 1515880 1515882 1517762 1517982 1497989 1515726 1497986 1515729 1497999 1497999 1497992 1497994 1497995 1497998 1498000 1515730 1517765 1517980 1519211 01010368331516895 02010198511517086 02010411821517013 0297842211517841 0297842221516899 0297842221517842 0297842231517016 0298107921516896 0398146141516900 0398146141517017 0398229591517087 0498347171517091 0597991621516902 0798618861517082 0798931761517084 0798947551516905 0798947551517076 1287397021516897 1287431851517007 1287441801517008 1299988631517009 T101327051517003 U05705531517005 1517981 01010209611521128 01010306761521134 01010471281521135 02010496941521129 02010598201521123 03010783871521124 07011756031521125 089771930011521120 089772180011521131 089815779011521121 089815780011521127 089815786011521133 089815787011521137 1199892391521119 12910087781521130 12910120691521126 1299961801521132 1497993 1517983 1519212 1519790 1519875 1520521 1521011 1497996 1517978 1519788 1521748 1517979 1519789 1520745 1521549 1521550 1521787 X013884801517887 X023964051517884 X044012311517885 X044012371517888 X064183581517886 1497997 1498006 1521803 1521804 1521805 1522351 1521551 1516922 1519213 1520520 1522355 1520518 1520519 1521014 1521976 1521978 1521979 1522560 1522561 1522563 1522564 1519874 1520742 1523638 1521548 1522122 1522353 1522562 1523908 1524405 1520740 1523909 1522123 1524404 15093731527014 15217871527015 14673921527030 14783001527007 14807351526994 14868731527003 14875091527029 14922741527016 14979971527070 15024801527006 15033111527017 15078111526982 1525564 07113810151527002 14772451527000 14936851527068 14969471526999 14979651527067 14979741527097 14979951527095 15002241527064 1523440 1523469 1523637 1523639 14798221526998 14804711527042 14899921527040 14924271526986 15027601526978 15089741528079 15089741528079 15093741528088 15103201527057 15135731528066 15158671528034 15218031527039 1522350 1522356 1524401 1524407 1525757 1525759 14663381528077 14695451527032 14770071528080 14850071527001 14852761528076 14868771526993 14940711527072 14979631527065 14979711527073 14979861527012 14979871527058 14979921527050 15002251527038 15002251527038 15006671526996 15015871526971 15025101526983 15025961526979 15025971526989 15040571527044 15042401527074 15053851527041 15072611527062 15083831526985 15147371528033 15147391527036 15147401527052 15158661527061 1520743 1521977 1522121 1522124 1523467 1523468 1523906 1524403 1525368 1525375 1525376 1525761 1526274 1526275 1526362 1526363 1526658 1526946 1526948 1526949 1526950 1527135 1527136 1527139 1527644 1526951 14822901528526 14827691528529 14839801528528 14908541528511 14946841527071 14979541528513 14979701528347 14979731528348 14979911528349 14980001527096 15007891528336 15042391528512 15053971528520 15072601528514 15078131528515 15103191528517 15119271528522 15119311528025 1523470 1523691 1525062 1525063 1525760 1525762 1526364 1526374 1526375 1526553 1526555 1526947 1527128 1527138 1527605 1527606 1527645 1527646 1527999 0986862901529034 14969461528752 1498007 1521988 1522520 1522809 1523005 1523258 1524949 1525064 1525758 1526378 1526945 1527137 1528172 1528219 0297842221528749 03010882931528755 03010950911528738 0999258591528969 1299988631528747 14848271528731 14868711528769 14899901528745 14924051528753 14979621528758 1498002 15059521528757 15072621528760 15072631528759 15078121528754 15131671528768 15171581528737 1523004 1523473 1526608 1526660 1527814 1527818 1527997 1528218 1528252 1528253 1528434 1528449 1528450 14581691528946 1526554 0298107921529461 03010882951529005 0398024171529030 0398221201529038 0686375191529024 0698876441529028 0798931761529015 0885808141529014 0986862841529022 0999266751529464 0999266821529054 0999398181529045 1063833491528968 1099664391529061 1099671161528970 14629891528936 14792061528942 14802651529455 14899891529452 15072591526980 15103191530221 1524951 1524952 1525065 1525565 1526944 1527648 1528251 1528627 1528933 R11976081528993 T111347771529456 01010348161530685 01112837221530450 0297991631530672 0398054661529494 0485841601529477 05113521691529443 0698876411529507 1287297951529486 14921081530435 14922701529961 14924071529957 14934471529963 14946321530438 1526379 1526380 1527816 1527817 1528451 1529185 0274106021529492 0297842221529543 0298054651529493 0298054651529495 03010828401529476 0385530171529496 0397842251529922 0398146141529529 0574334061529485 0585879531529505 0798618861529537 0898618871530679 1099525931530770 1099574701530423 1187334731530035 1199795771529515 14790621529449 14810281529891 14887861529947 14908571530430 14924061529964 14924081530018 14934461529960 14936881530788 15014041530419 15109161530017 1524950 1526371 1528018 1529189 0285565811530004 0285612261530094 0598437491529896 07113821111530523 07113821111530524 10012247441530533 1099486061530032 11012317441530591 11012341471530594 13795111530746 14979491529967 1498003 1526370 1529190 1529583 0109988671531036 01112894491531082 0197693781531112 02010198511531109 0274103961530077 0398146141530570 04011041851531071 04113450461531092 04113450461531094 0698899061530966 0752456651530081 079889898101530526 079893176101530577 0798931811530968 1099486041529912 11012317441530593 1199574731529927 1199685201530971 1199685201530972 1199795771531110 1199868011531111 1275045211530072 12910062861530974 1299809901530774 13939211531103 14011151530548 14776801530909 14860411530415 14908631531066 1492539 14946331530979 1498028 15089381530915 1528000 1528019 1528173 1528290 1528452 1528935 1529187 1529188 1529191 1529192 01010262141530923 0197721921531037 0285403241529302 0285612261529293 0297968021531045 0298005691531046 04011041851531052 0498373751530951 05010931071531054 05011317991531095 06011505641530965 0786619881529282 1098703061530969 11012247341531100 1299962141531065 13795111530747 13983821530545 14980021531523 14980071531524 15053981530793 15053981530794 1526372 1526381 1528289 1528939 1530814 1531072 1531260 R0489693A1529392 039829635101530571 0485730641529396 05011422071530952 0563476601529390 0652095941529391 0785808191529308 0798931801530967 1527647 1527998 1528285 1528435 1528453 1530047 1530604 1530605 1531181 1531182 1531262 1531556 1529953 1530642 02010712691531084 1523907 1527857 1528286 1530797 1531073 1528287 1528625 1530600 1530601 1531557 X013831871517889 X013858441517890 X044012431517891 1498008 1498010 1526376 1528628 1530044 1530638 1530810 14489711533609 1528938 1529186 1530640 0485942381530120 0886766761530118 1526377 1527800 1528013 1528015 1528454 1530606 1530639 1530812 1530816 1531256 1532904 1530815 15033051526992 15089371527066 1530813 1531906 1532903 1530045 1530046 1530648 1530817 1533496 1517637 1519122 1527856 1528883 1529952 1530608 1530645 1531074 1531075 1531257 1531902 1531903 1532148 1532150 1532151 1532254 1532256 1532259 1533009 1533012 1533225 1533497 1528016 1528882 1530611 1530612 1531258 1531259 1531904 1531905 1531907 1532146 1532147 1532149 1532255 1532258 1532267 1532268 1532269 1532430 1532431 1533010 1533011 1533393 1533498 1533500 1533519 1534170 1534171 1534408 1534409 1534496 1534497 1534634 1530607 1533518 1530610 0185403171529419 0285561851529421 1528626 1532257 1532261 1533391 1533499 1528291 1531784 1531786 1531908 1533014 1533224 1534407 1528292 1530609 1531781 1531783 1532260 1532852 1533392 1533394 1533501 1533521 1534028 1534029 1534635 1535625 1535853 1530303 1530603 1533502 1533520 1535866 1532262 1532270 1532271 1533016 1534175 1533397 1536038 01010209461535570 01010397961535562 01010433461535575 02010598211535576 03010783901535577 03010829311535564 03010968361535574 089815779011535559 089815779011535563 1199813321535571 1199885971535572 1199950091535565 12910198351535573 1526373 1531909 1532264 1532265 1532434 1533015 1533017 1533975 1534632 1535668 1536034 1536035 1536039 1528014 1532266 1534127 1534136 1534173 1534174 1534178 1534180 1534412 1534633 1535626 1536627 1536628 1537674 1537675 1537676 1532112 1532263 1532432 1532577 1533503 1533524 1533525 1534128 1534172 1534406 1536273 1536608 1537671 1537677 03010829301537537 1534168 1534182 1536042 1536049 1536271 1536629 1537470 1537471 1535251 1536187 1537476 1529249 1533522 1533523 1535252 1535665 1535666 1535667 1535867 1536045 1536048 1536272 1536609 1537305 1537475 1537669 1537672 14446321536078 14772451539312 14791991539499 14908571539519 14922741539522 14946321539501 14979651539314 14979741539338 14979921539318 14980071539315 1498009 15006671539311 15007871539500 15024801539310 15024811539313 15027571539524 15072601539305 15072611539296 1534410 1535864 1536265 1536266 1536606 1537307 1537673 14908581539587 14921081539567 14979961539326 14980061539317 15078121539339 15103201539307 1532114 1536607 1536630 1537013 1537507 8200441539594 0586023471539732 0586023481539733 14072751539386 14639731539586 14791981539566 14899871539521 14899921539593 14908561539591 14940701539576 14946851539520 14969471539340 15018441539533 15024921539570 15053971539321 15078051539387 1533013 1535253 1535254 1537505 1538277 1538278 14908541539530 14908581539532 14792031539771 14804061539568 14804711539612 14921081539613 14922731539528 14940711539518 14979541539325 14979901539324 14979971539308 14980021539364 14980031539363 15002251539531 15025101540112 15166761540107 1538386 1539208 1539209 10062841539595 10368331539726 14792001539525 14792021539687 14899911539590 14936851539772 14979931539376 14979941539316 14980051539367 15007891539569 15027581539523 15042401539493 15078131539375 15103191539469 15346341539497 1537477 1537950 1538279 1539210 9258601539747 14787931539526 14802471539673 14924021539306 14979951539352 14980081539365 15002241539589 15042391539370 15072591539787 15078131539346 15119261539369 15119271539368 15223531540119 15223541540117 15269471540306 1537508 1537510 10981711540510 14370031536082 14804071539351 14899881539753 14899891539748 14924041539350 14924051539372 14924101536083 14979681539361 14979731539358 14979911539359 14980071539348 15006671539344 15083831539782 15119281539366 15147401540335 15263631540278 15271391540027 1537509 1538280 1538719 1539001 02010726561537538 089815786011541314 089815787011541315 14868761540470 14868771540476 14885031539789 14979661539360 14979671539354 14979711539362 14989801540459 15025961540331 15027601540038 15067621539918 15071661539798 15078101539921 15078111539794 15092011539808 15119311540462 15158661540464 15171571539917 15171581539793 15218031539920 15218051539922 15223551540307 15234401539792 15236371539919 15236381539797 15236911539931 15253761540114 15262741539788 15262751539795 15281721539929 15282511539796 15286281539928 15308141540321 15321121540333 15321511540332 15332241540505 15332251540461 15333911540334 1534185 15360381539790 1536043 1536051 1538274 1539211 10950911540024 12559021541340 14804711539729 14841071540325 14848271539930 14979971539345 15018481540320 15025961540328 15033051540457 15033101539496 15033111539353 15053981539470 15093721539371 15131681539688 15147371540463 15147391540329 15158671540465 15172471540292 15218041539791 15234701541200 15236381540297 15236391540043 15236921540305 15244071541325 15250621541313 15253681540111 15263741541201 15263761541159 15269451540285 15269511539916 15271281540113 15271361539933 15271371540118 15276051540286 15276471541346 15278181540021 15280141541181 15282191540309 15289391541348 15291921540460 15303031541152 15306481540287 15319051540295 15322571540120 15324341541150 15330151541147 15334971540521 15334981541334 15334991541335 15335001541339 15339751541309 15341731541175 1534630 15356681541319 1537296 1537298 1538101 1538102 1538103 1538104 1538273 1539002 1539783 1539785 1539786 1540746 14925391541911 14979611541912 14979641541901 14979881541904 14979991541903 14980281541896 15109151541897 15250631541318 15263751541310 15269441540091 15312581541905 1534413 1537611 1537613 1539003 1539213 1537612 1540267 14261221541638 14979871541894 14980001541895 15059521541902 1532103 1538275 1538276 1538387 1539582 1540266 1541004 1542166 10896391541913 14804101540086 14817631541935 14827681541952 14827691541923 14835761539932 14839801541946 14979621541915 14979631541906 14979631541908 14979891541898 14979981541900 15018431539689 15053851541907 15072621541917 15072631541916 15278571542036 1532100 15321461541909 15322541541918 1536047 1537949 1539004 1539212 1539214 1540747 1541790 14798221541963 14822901541934 15051911541910 15109141541899 1540268 1541003 1542167 1542168 14764911540108 14809821540092 14812581540109 14850071540308 14855751540296 15157101540290 15265551540074 15278171540115 15282521540095 15322561540090 1534169 1534403 15346351539466 1535865 15360341540025 1536040 1536431 1539583 1539584 1541543 14885021540110 14804091540304 14810221540293 14924061539356 14924071539347 15072601539342 15223561540298 15236371540322 15278161540326 15284501540301 15322601540283 15330141540299 1534405 1534631 1536270 1537472 1537951 1537952 1539215 1539585 1539957 1540544 1541005 1541791 0485942391529415 13573141540458 14868731540289 15000091540324 15002221540327 15007501540310 15042361540302 15078101540291 15131651540288 15166751540281 15257611540280 15276471540277 15280001540279 15282181540284 15284491540311 15286271540294 15291851540303 15300471540300 15330111540282 15333971540330 1541793 14022171539592 14979471539341 14979911539476 15024801539349 15278141540323 15282521540312 1529242 1532116 15344961539455 15344961539473 15358531539374 1539958 1541569 1541006 1541716 1542575 14295011541914 14946321539632 1537415 0298107921539529 0363242351542035 0586051751529292 07113821111536080 0885807901539690 0999393811539527 14695451541891 14776491539696 14802651539597 14817621541933 14980031535010 15071651541626 15089371541920 15358531539343 1537309 1537311 1537414 1537512 1537608 1538384 1538385 1538718 1540146 1540999 1541000 1541568 1541572 1541574 1541717 1541794 1541795 1542162 1542163 1542571 1542574 1542576 1542708 0297864871541630 14899901539674 1542191 1542572 0899266761539715 1287297951541631 1532105 1532109 1537473 1537474 1537513 1542165 1542192 1542300 1542573 1542827 03010950931541627 12710271539492 15014271540313 1537511 1542164 1542188 1542190 1542703 1542707 1542826 0896375181542032 14868711541892 1540545 1542570 1542704 1542706 02010608581541312 14381301539304 14639811541919 14810281541965 14924091541636 15060321541635 15068661541637 1533018 1536185 1536186 1541575 1541718 1542169 1542301 1542675 1542705 1542788 1542828 1542829 1543230 1543231 1543391 15053981539472 1528017 1541719 1542626 1542709 1543232 1543233 1543234 1543235 1543383 1543384 1543385 1543386 1543388 1543392 1543393 1543394 1543395 1543403 1543972 1543975 1544312 1544314 1544353 1544376 1502594 1537012 1543389 1543390 1543396 1543850 1543854 1543855 1544153 1544316 1544415 1537077 1543852 1543853 1535624 1536354 1536844 1537014 1541573 1543387 1543851 1545288 1536733 1541001 1542303 1543404 1543973 1544151 1544319 1544606 1545234 1287431851541625 14980281535020 1537410 1541002 1541968 1542302 1542627 1543971 1543978 1544315 1544399 1544400 1544605 1544607 1544698 1544846 1544847 05011277131541628 1543856 1543969 1543970 1544396 1544697 1545079 1498011 1532118 1541969 1543405 1543857 1544092 1544152 0298146101542027 0385670021542034 0598352541541966 1087120521541629 1287441801541624 14357091539309 14563621539357 14887841541633 14924001541634 14969471539373 14969471539454 14979481541632 14979491539468 14979681539478 14979701539355 14979711539460 14979731539479 14980081541893 15344081539467 15344091539495 15358661539491 1541972 1543397 1543398 1543406 1543976 1543977 1544318 1544416 1544608 1544609 1544798 1544848 1545082 1545198 1545232 1545233 1545360 1545361 1545362 1545385 1545400 1545401 U03667181542030 X023884781541967 X033964121541964 1498001 1541974 1498012 1541970 1541971 1541973 1544317 1545608 1544397 1546621 1541975 1544398 1544602 1545386 1545507 1545552 1545603 1545604 1545605 1545607 1546176 1545599 1545601 1545602 1545606 1546170 1546174 1546401 1546402 1546552 1546553 1546655 1546673 01010471251547379 02010526301547376 02010726561547265 03010768351547380 089772180011547269 089772181011547381 089815779011547264 089815782011547266 089815787011547385 089815787011547386 12910087781547267 1546173 1546860 1547014 1547016 1547154 1547157 07011812811547270 089815786011547382 1546169 1547012 1547156 1547189 1547333 1547143 1545600 1546651 1546652 1546653 1546654 1546922 1547018 1547144 1547155 1547332 1547334 1547749 1547841 1547842 1547847 1548143 01010397961547263 1547336 02010695601547383 089815780011547378 12910198351547268 1547790 1547335 1547752 1547789 1547845 1547848 1548488 1548669 1549178 03010950141547384 03010768401547377 1541976 1541977 1548491 1541980 1541981 1548146 1541978 1541979 1544814 1548445 1549541 1549800 15441511551525 14980051551528 15410021551614 15421661552015 15446051551922 1544603 1547572 1547751 1547846 1548672 1548836 1548974 1550091 1550294 1550295 1550453 1550454 1550881 1547788 1547839 1549494 1549495 1549496 1549542 1549543 1549801 1551132 15372981551283 1547780 1547838 1548084 1549485 1550455 1550672 1551134 1551533 1551606 1550673 1547840 1548081 1548774 1549415 1550841 1551420 1551534 1551607 1551608 1552008 1552173 1552174 1548629 1548630 1549414 1552009 1552144 1552332 1552539 1547750 1548835 1549540 1551604 1551834 1552175 1552333 1552540 1552541 0752456651555188 14839801555202 14979871555052 14980071554819 15329031555197 15341711555195 15360451555196 15363541554702 15419771555090 15421691554701 15461731555092 1548447 1548490 1548671 1550670 1550671 1552696 1552969 1548446 1549802 1551605 1552330 1552697 1553250 15315561555485 15341711555443 15356651555484 15477491555445 15477501555557 15471571554459 1548145 15486721555225 15504531555226 1553249 1549799 1551047 1551419 1551833 1552328 1552970 1553370 04113265991554455 15477521556098 15486711556284 1541982 1548082 1553251 1553369 1553588 1553589 14980011556041 15025941556042 15025941556761 15356261556751 15363541556039 15370141556038 15370141556760 15410001556762 15419771556000 15443981556749 15477501556002 15477891556657 15478381556653 15484471556001 15484471556763 12910198351556682 01010209461556679 01010262871556684 01010284621556685 02010608581556680 03010768321556683 1199813321556681 14255231557325 15495431557326 15516081557327 15532511557324 1541983 1541985 1548083 1551131 1552007 1552968 1553366 1553368 1553587 1553999 1554003 1554144 1554145 1554284 1554285 1554286 1554843 15484881557622 15506731557626 15515341557624 066353694101551730 075210935101551731 1541984 1552172 1552331 1552653 1553364 1554001 1554002 1554142 1554143 1554488 1555446 14979991557805 15027581557797 15360491557809 15448141557807 15484461557803 1551133 1552698 1555641 1554490 1554845 15470121558714 1554643 1554842 1555287 1555640 1555642 1556323 14980121558600 15315571558599 15356261558601 15360491558603 15370131558596 15375121558597 15392111558595 15419841558594 15478401558593 15484451558591 15510471558598 15526531558602 1552967 1552971 1554000 1554140 15542851558908 1555448 1556107 1556618 1556872 14980011559177 15025941559175 15329031559171 15356241559179 15358651559172 15360511559181 15370131559173 15392111559170 15511311559169 15514191559174 15539991559180 1555790 1556322 1556324 1556619 14980121559520 15176371559518 15346311559517 15356241559521 15360401559523 15374701559516 15410051559522 15480831559515 15510471559519 02010598211558739 02010726551558736 02010743021558738 089815782011558737 15477501559896 15488351559897 15540011559932 1555789 14979891560043 15360401560042 15370131560098 15374731560061 15392091560095 15523281560044 15529671560099 1553367 15544881560096 1557541 1557542 15548451560463 01112837221560715 0398054661560716 1287297951560717 14908581560713 14979981560721 15319061560720 15335001560714 15335181560723 15341711560718 15506711560712 15523281560719 15529671560722 1553248 1555643 1556620 1556871 1556873 1557094 1557096 1557097 1557368 1557370 1557766 1557963 1558050\n1558224 1555286 1555447 1556108 1557367 1558223 1558340 1558341 1558565 1558592 1559250 01010209461561233 01010306751561244 01010356111561237 01010398081561234 01010458651561241 01010471251561242 02010518961561239 02010538661561243 02010598201561235 03010768381561236 03010796731561245 04011087141561232 089815787011561246 12910087781561240 1549798 1551048 1551917 1554139 1554489 1555298 1557043 1557761 1557765 1558225 1558342 1558343 1558566 1558567 1558802 1558951 1558952 1558953 1558954 1559251 1559548 14822901561598 14940711561593 15072621561599 15334991561594 15341691561595 15360401561602 15362711561596 15363541561600 15375121561597 15511311561592 15541401561601 1548775 1552006 1552654 1553586 1554652 1557763 1559252 1559605 1559606 1560143 15541431562025 1554487 15583411562062 1559550 14980121562142 15315561562141 15356261562143 15360451562145 15392101562137 15510481562139 15529681562136 15541391562144 1556617 15577611562138 1559823 15525411562480 15548451562479 1559549 14357121551282 14817631562713 14979641562714 15360451562717 15421911562710 1549177 15510481562712 1551130 1552171 15526541562709 1552816 15535861562716 1556236 15577611562711 1559249 1560413 T021217411562715 15526541562890 15532481562914 15535861562891 15544901562913 14393241563374 15362711563523 15487751563522 15510481563524 15535881563535 15541401563525 15575421563373 15583411563375 15419771563737 1552329 1552815 1553365 1554141 1554644 1554651 1554653 15548451563701 1555299 1555636 1555637 1555639 1556615 1557363 1557364 15577611563736 15585661563702 1559475 1559603 1559705 1559825 1560142 1560319 1560955 1560956 1560957 1561169 15356661563960 15368441564053 15375131564052 1550783 1555296 1556235 1556616 1560028 1561163 1561633 15518341564172 15544871564187 14860411559862 15362711564297 15497991564296 1555638 1556614 1557365 1558066 1558563 1558564 1558801 1559248 1559474 1559547 1559701 1560026 1560141 1560320 1560414 1560588 1560741 1560743 1560902 1560903 1560958 1561202 U06665751564298 14235841564565 15282911564836 15324311564561 15529711564566 15548431564564 15582241564563 15583401564562 15583411564835 1561634 1562198 076356251101560473 14658351562012 14702991561704 14798221565073 14817631565071 14827681565072 14848021559859 14887831559861 14980001565074 15176371565070 15344071565069 15360451565077 15370121565075 15421911565067 15446021565076 15540001565066 15546521565065 15577611565068 1558800 1560029 1561711 1561714 1561716 1563035 14524601565775 15362711565530 15510481565531 15577611565529 15466531566051 15521751566069 15521751566138 15563241566068 15563241566148 02010566171564759 03010896211564762 05113426951566218 089815780011564760 089815782011564761 14334381566216 15542841566214 15556421566217 15561081566213 15577661566215 14265781566639 15083831566635 15291871566528 15563241566537 15577611566527 15579631566638 01112894491563451 0463303831557473 14638491561703 14681941562003 14899891562004 14899951562002 1550788 1555297 1558065 1558950 1559704 1560411 1560959 1561161 1561204 1561715 1561905 1561948 1561951 1561952 1561953 1561954 1562412 1562777 1563034 1563037 14979961566887 14980001566891 14980061566888 15315561566889 15315571566890 15346321566886 15360511566893 15374711566885 15390021566892 15466511566884 1563815 0663536941557481 10355971565263 13971341565262 14193231567156 14563621566931 14924051566932 14924061566933 14924071566934 14979491566938 14979651566935 14979701566936 14979901566937 14980011566939 15027571566929 15321161566941 15335181566943 15344121565260 15356651566940 15375111566930 15466521566928 1550784 1550786 1550792 1557762 1558063 1560027 1560431 1561712 1561717 1561718 1561955 1562776 1562779 1562780 1563304 1563432 1563814 1564109 1564111 1564476 1550782 15570961567409 1561512 1563306 1563813 1550787 1561205 1561950 1562410 1563431 1563434 1564107 05011277131567848 1497981567850 15269511567846 15271391567847 15341681567999 15346331568001 15356671568008 15362711567849 15410001568006 15410061568010 15423021567995 15516081568195 15526531567851 15556431568193 15561081567822 15577611567997 15589511567993 15592501568194 1561652 X0339641568004 14658351568406 14968351565691 15072551565690 15507821568271 1550791 1558067 1558068 1562413 14250801568526 14791991568670 14822901568678 14979871568679 14979941568675 14979971568676 14980061568673 15089371568680 15344091568677 15352521568672 15358661568674 15360491568687 15443971568686 15455071568668 15455521568669 15466521568667 15516071568527 15525411568528 15535861568682 15539991568681 15541391568683 15541401568684 15544871568685 15577611568671 1561164 1563601 1565300 1550789 15511301569009 1563307 1563602 1563812 1564246 0363242351569338 14979631569145 15292681569339 15321161569146 15344961569700 15346321569144 15358531569699 15419741569335 15419751569143 15433831569334 15464011569333 1550793 1550794 1565546 1565549 1566036 1566044 1569143 14112241566503 14138431569655 14318631566498 15346331569637 1550781 1563038 1564472 1564799 1564801 1566032 14125261570079 14217991570078 15376131569962 15415681570077 15427081569963 15438541570064 15439771569961 15484901570076 1086576441570202 1099486041570197 1287431811570201 1287441811570200 14011151570195 14791991570193 14810281570198 14979991570199 15322601570184 15344121569971 15346331570196 15360431570204 15421901570185 15438511570181 15439731570186 15443151570182 15443961570203 15455521570183 15456021570187 15471431570175 15477881570188 15477901570176 15480831570178 15480841570173 15494151570179 15495401570174 15497991570189 15506701570192 15514201570190 1552608 15529681570177 15533671570180 1557042 15577611570194 1558064 1564112 1564247 1564617 1565302 1566337 0197693781570457 03010950931570456 14776491570455 14980001570458 15360451570459 15507841570361 14011011571104 14261221571108 14318631571110 14357121570927 15029131571105 15352521571182 15356261571185 15360451571188 15360481571071 15361851571070 15374701571180 15375121571183 15409991571184 15425741571106 15438551571186 15438561571187 15487751571181 15491771570899 15507891571126 1550795 15521711571048 15552961571127 1559703 1564108 1564245 1564798 1566042 15072631571087 15525401571323 15579631571241 1563036 1564474 1564800 1565547 1564616 14817621572186 15346321572016 15446051572389 15507831572174 15507941572215 15533651572216 1561162 15617161572396 1561946 15619531572185 1564110 1565051 1565548 1566465 1566466 1566467 1567002 1568075 1568314 1569042 14899951572173 14979871572637 14980051572175 15312581572655 15446051572941 15507811572679 15507951572683 15526081572671 15600271572685 15604311572684 15616331572672 15507811573023 15507831572990 15546441572942 15562351572991 1564475 1564615 1568463 0598437491573273 0686375191573276 14822901573274 15356671573282 15358651573278 15360491573270 15374701573279 15507871573423 15580641573424 15627771573460 15664661573461 1565297 1568462 1569359 0598352541573493 15053851573272 15507861573532 15511301573480 1552605 1564477 1565049 15655481573533 1566336 1566463 1566703 1567498 1567499 1567653 1568757 1569043 0385670021573275 0398146141573790 0698876441573791 0896375181573793 1199574731573792 14908581573825 15334991573824 15360481573795 15361861573794 15363541573827 15375121573823 15409991573822 1566702 15580631574194 15609571574192 15611621574195 15611631574191 15611691574190 15507881574468 15566141574478 15619511574476 15644751574475 0185403171574708 14791991574709 15010871562000 15335201574706 15507811574621 1552606 1566029 1567655 1568313 1568465 1564797 1566039 1566764 1566765 1567654 1569921 1571831 04113450461574748 14269261574838 15487751574837 15650511575036 01010348161575642 01112894491565566 0985808401565264 10012247441575641 14899881575640 15282181565261 15360511575644 15361861575643 15407471575718 1567865 1568756 1570569 1571832 15310751575885 15346311575887 15360511575892 15375121575886 15421911575888 15407471576093 1550796 1559706 1563033 1563430 1563811 1564473 1567001 1567268 1567497 1567864 1568312 1569040 1570248 1570772 1570773 1571382 1571383 1571386 1571827 1571835 1572267 1572270 1573231 1561709 1566700 1569038 1571834 1572269 14791991576331 15335231576326 15341691576329 15356651576342 1570570 1571830 1573235 1573236 15604131576672 15607431576673 1565288 1566462 1571387 1572560 14275641576868 15507931576674 1560430 1561203 1566335 1571381 1572112 1572114 1573229 1574306 1567652 1568461 1569045 1569923 1569924 1572111 1573232 12910147411576225 0886766761577669 14791991577672 15356671577674 15360421577668 15360481577879 15526061577835 1573871 14259581577947 14259581578118 1566464 1567003 1568755 1570572 1571829 1572559 1573869 1573872 1574061 1574308 1574309 02113033801574108 04113380731574109 15507881578157 1550798 15526051578166 15573641578158 15600261578159 15612021578160 15634301578171 15644731578170 1567651 1572556 1574542 1563600 1566701 1570249 1572266 1574543 1574545 1574760 1574763 15360511578655 15363541578654 15375121578653 14240731578829 14269261578743 15341691578652 15423031578651 1550797 1565050 1570418 1570568 1570769 1571828 1572110 1573868 1574546 1574762 1575575 1575752 1575927 14259581579070 14899921578957 15341691578953 15356241578959 15360401578950 1576163 1561713 1563032 1565545 1566333 1570419 1573867 1574186 1574759 1575577 1575925 1576280 1576281 1576713 14261201579292 14269261579248 15528161579301 15507911579358 15619461579360 03111887311576979 03113192221577002 07113937291577001 1575574 1575755 1575756 1576075 1576130 1576900 14181031580043 14275641579816 15432321580045 15511331580044 1561710 1575923 1576165 1576472 1576898 1577263 1577510 14016251576999 15585641580632 15597061580634 15641071580638 15644731580631 1574060 1577262 15335211580588 15356671580587 15722661580637 1568759 1575753 1576076 1576129 1575573 1576279 1576282 1576475 1576903 1577512 0863722581574114 14080861581134 14261201581259 14946321581131 1576162 1576471 1578085 15335231581130 15356671581129 1576715 1577763 1577765 14638491581491 15507971581358 15507981581800 15630351581489 15650511581492 15655461581488 15684611581490 08011976211576993 1575751 1577592 1578082 1578204 1578389 1578885 14269261581732 15585671581926 0297842221582105 0297991631582104 1199795771582103 14240731582175 14247171582010 15360481582100 15690421582274 1573230 03010896211582005 1570771 1577924 14240731582788 14271891582857 14271891582986 15573631582495 15638121582930 15646161582958 15683121582963 15725561582961 1575924 1577926 1578391 1578392 15552961583094 15721101583100 0863722571574113 14318631583052 15660391583053 1566334 15747621583054 1577766 1578385 1579082 1579418 15552961583328 15566161583326 15650501583325 15718321583522 15721101583327 15566151583839 15597061583838 15647981583837 0898618871583772 15341691583770 15360481583774 15363541583773 15638111583974 15638121583975 1578691 1579231 1579417 1579420 14275641584231 15556391584397 15562361584396 15619531584402 15641071584300 15667021584401 15670011584398 15674981584399 15676531584400 1579696 14791991584530 14979861584537 15059521584536 15089371584539 15335231584544 15335241584542 15341691584532 15358651584531 15361851584540 15361861584541 15375111584535 15375131584534 14213711584523 15007871584968 1578887 1580500 15421911584896 15507871584787 15721121585116 089815787011584958 1576278 1576470 1576896 1577261 1579233 1579915 1581763 1581020 1581762 1582115 1576137 1577591 1578083 1578205 1578884 1579413 1579421 1579697 1579913 1579917 1580151 1580498 1580658 1580660 1581019 1581626 1581878 1581879 1582324 1579695 1580656 1580818 1581022 1582497 1582850 1583305 15312581587235 15507931587798 15663341587541 15667641587808 15687551587545 15690431587561 15690451587564 15722691587786 15732351587563 15757531587562 15775121587560 15775911587543 1578086 15782041587753 15782051587729 1579084 1579698 15799151587542 15799151587746 1580155 15806561587544 1580657 1580821 1580823 1581628 1582114 1582499 1582500 1583568 1583572 1583806 1584062 15507961587856 15664651587875 15704181587860 15732351587876 15740601587743 15757561587877 15761651587874 15775121587785 15775121587873 15775911587858 15783911587776 15792331587726 15799151587857 15808181587859 01010263791586842 02010598201586836 03010829301586837 089771930011586834 089815780011586838 089815786011586841 12910095511586839 12910209591586840 1559826 1579081 1580145 1580154 1580661 1580819 1582325 1583531 1583569 1584196 1584708 1585074 15745451588734 15805001588725 1583570 1584562 1584700 1585399 1585639 12910209591587762 15507831589625 15507911589626 1578690 1579230 1583160 1583351 1583949 1584701 1585388 1586024 1586026 1586371 1586502 15732291589641 15747591589642 15794181589640 1586369 15526061589763 15552971589762 15609561589766 15616341589765 15644741589761 1586025 1567866 1584865 1585391 1586734 1587441 1587454 1587188 1587640 14979871590339 14979871590344 15109151590498 15312581590338 15053851590559 15322551590499 15423021590494 15089371590646 15407471590647 15322541590630 15409991590919 1552614 1583948 1584561 1585640 1586735 10896391591271 14295011591272 14925391591247 14979611591261 14979631591258 14979861591267 14979881591251 14979881591252 14979881591253 14979991591269 14980001591249 14980111591254 14980281591248 14980281591260 15051911591250 15072631591259 15072631591265 15089381591264 15109141591266 15310741591256 15312581591255 15321461591268 15322541591262 15356671591276 15410001591257 15808181591270 W123831801591263 06011400241589571 0798931771589570 089771930011589562 14908571591275 15407471591718 T101347631589568 W103727331591393 14316221583858 14512161591538 15526141591648 1587459 1587835 1588245 15072621591762 15790811591650 1584565 1584689 1584705 1585078 1585080 1585389 1585397 1585642 1586370 1586503 1586505 1586736 1586737 1587190 1587631 1588247 1588730 1588828 1578692 1581018 1584564 1586086 1586739 1587192 1587632 1587836 1587896 1587899 1588704 1588830 1588832 1588834 1588857 1586507 1586847 1588517 08011975941589735 08012000841589728 14204611589730 15053851592240 15322541592245 1587837 1587897 1588501 1588852 03010981711589732 14370031587778 14581771587775 14726861587777 14736541589729 14745441589733 15059521592015 15059521592429 15059521592430 15072621592423 15390031584640 15455991584645 1584087 1587189 03010538881592801 06011500001592802 0886784711574986 11012404851592800 14188941589734 14776781589731 15223561584647 1582323 1587895 1589110 14921081592671 15335231592668 15363541592672 1588829 1588836 1589111 1589196 1552611 15761291593335 15786901593340 15790811593259 15823231593338 15856401593339 1588851 1589534 0563424911574112 14232061589174 14330181589171 14330191589172 14629891589173 14676311589249 14742141589177 14776491589254 14782991589256 1552609 15526111594024 1552612 15808191594019 1589531 14749941589175 15360511594195 15423031594198 15526141594023 15874541594020 1589533 1591450 1552613 14946321594796 15335221594349 15335221594795 15361861594353 15374101594797 15526091595014 15660421595022 15407471595270 15526121595114 15847051595116 15867371595043 15868471595104 03010828401569205 1063826961591564 1163891071591565 14791991594351 14899871595577 15158681575581 1588854 1589199 1590899 1591449 1592255 027405657101574139 086372256101591563 106382397101591592 1552610 15619051595870 15619481595840 15634341595828 15704181595827 15718351595868 15780861595851 15799171595991 15801511596000 15804981596001 15845641596057 15847001595847 15856421595842 15876321595831 1589528 15914501595932 1592254 1592374 1592549 1592614 1592914 1593046 1561944 1584864 1587630 1587898 1587900 1588243 1588831 1593209 1593566 1594010 15526131596247 15533651596251 15630351596249 15782051595999 15806581596002 15895311596250 15507881596802 15895311596799 15335221597190 15360421597191 15361861597192 15650491597411 1585071 1585072 1587455 1590684 1593567 1593847 1594005 1594011 1595012 1590682 1594006 1594705 1594747 1594837 1595236 15407461597708 15526131597605 15653001597614 15760761597908 15833051597613 15835721597611 15856401597575 15865031597576 15865031597576 15865051597604 15867371597904 15871901597602 15878371597601 15926141597586 1594610 15630381597907 15876301597906 15407461598605 15619441598821 15768961598812 15772611598813 1587641 1588242 15906821598715 15922541598814 15935671598815 1594748 1594838 1595085 1595237 1595238 1595583 1595844 1587191 1588856 1589197 1589200 1590683 1591196 1595348 1596318 01010271781593762 01010356111593757 01010398061593754 01010433461593764 02010655801593758 03010768381593755 03010783901593763 03010838901593760 06011528341592794 089815782011593759 12910123891593756 12910186001593761 14868711599250 15407461599247 15407471599243 12910186001596311 1590686 15911961599591 15950851599590 15953481599589 1595940 1596492 1595086 1595845 1596147 1596494 15498021599908 1588518 1595846 1596491 1596760 1597017 1597251 1597427 1586506 1593265 1595667 1596493 1596495 1596758 1597019 1597250 027405657101600310 14776781600309 14639731600661 14787931600664 14791981600662 14791991600658 14804711600657 14899891600668 14921081600663 15335251600654 15392081600665 15421681600659 0197615991595969 12012592161595970 15360511600851 15370771600656 1590900 1596759 1597018 1597550 1597640 1597806 1597945 15487751600823 1593047 1597426 1598290 1598578 1594609 1597249 1597392 1597428 1597638 1597639 1598287 1598481 1599140 15810221601349 15952381601350 15956671601360 15974271601351 15526101601468 15950861601600 15329041601660 15334971601657 15487751601656 1552615 1580144 1597391 1598883 1598884 1598980 15853971601899 15964911601898 01010263791596313 01010356111596304 01010398081596300 01010411191596301 01010411221596305 03010796731596314 06011515671596309 089771930011596298 089772180011596306 089815779011596299 089815780011596302 089815782011596307 089815786011596310 089815787011596315 1199950091596303 12910163101596308 1299961801596312 1590685 1597946 1598576 1599138 1599367 1599660 01010263791599780 01010398071599776 06011528341597463 089815782011599777 089815786011599779 089815787011599781 15361871602512 1599922 1600056 1600057 0574251481600217 1263927411599635 1591198 15970171603855 15974281603854 15982901603856 1600235 V10100772A1589737 15972501604239 15952381604299 15964951604326 15984811604327 16000571604325 14787951595966 15946091604557 15948381604554 1597248 1599366 1599368 1599369 1599659 1599921 1600679 1600865 1600999 0574309471602570 0574360801602571 0574371131602572 14122331602566 14271891604651 1599365 1599923 1600058 1600462 14138061602567 15379501602568 1600461 1600867 14271891605606 1599919 1600236 1600460 1600680 1600681 1600868 1601145 1601146 1601147 1601461 1601654 1602665 1602999 1597948 1598575 1599658 1600234 1601252 1601653 1601967 1602447 1600678 1600998 1602282 1603569 V062017171601668 14266461606315 15952381606360 15963181606355 14247171606554 15989801606639 15993681606622 15996591606623 15609571603814 1597425 1597905 1598289 1598577 1599920 1600260 1600864 1601144 1601251 1601459 1601809 1601810 1601964 1601965 1601968 1603306 1603575 1603681 1603993 089815787011606872 15507971606963 15764701606961 15769001606955 15799151606962 15824991606957 15871911606981 15891971606877 15914501606878 1598882 1599313 15996581606874 15999211606882 1603299 1603572 1603773 1603994 1604017 14173731591978 14577381595977 14600861595972 14807331591981 14887841595976 15060311591979 15135721595979 15856391607061 1595013 15963181607063 1597156 1598479 15993671607062 16004601607052 1601460 16016541607055 16029991607060 1603577 1603775 1603777 1604018 1604284 1604570 1604571 1604592 1604818 1604999 049833628101574110 0574251481599637 066349348101598742 14224281595971 14271891607377 14330181595964 14546451595975 14732651595968 14924101595978 15053981591980 15053981595974 15263751604001 15882431607317 16019651607316 1604019 W033145561601666 14266461607575 1603000 0174005091605542 13942391607816 14112181607811 15382781604003 15487741605014 15523301603847 0274025431602863 14181031605243 14217991607997 14231751608013 14266461607996 1597947 1602998 1603581 1604021 1604283 1604285 1604591 14213711608403 15850781608256 16009981608259 16011441608260 16012511608257 16019671608261 16042851608258 14000161606176 14009461606169 14122341606177 14334311606171 14341511606172 15516051603846 15611621606173 1603770 1603778 1604287 1604820 1605263 1605264 1605413 1605641 1605642 1605787 1605789 1605931 0274125631602869 0574251541603815 1074832841605543 116385243101603816 14268391606170 1590687 1605266 1606191 1606193 15979471609791 15996601609792 16008641609782 16030001609794 16035721609790 1604286 1605643 1606192 1600259 1604020 1605265 1606324 1606399 1606621 1606674 1607157 14301541606178 15690451608865 1591200 1604819 1605788 1606398 1607048 1607211 1607212 1607213 1607214 15767151609175 1591201 15979471610973 16042841610964 16054131611002 16063241611012 16066211610965 16070481611001 1607859 1608058 1608310 1608484 1608737 16057871611179 16061911611174 16072111611178 15874411611488 16032991611505 1606194 16072131611484 1607433 1607434 1607624 1608736 16052631611980 16056421611981 16061931611982 16063981611983 01010471251607042 15912001612182 1606675 1606910 1607625 1607860 1608285 1608309 03010783871609344 05113426951609345 05113433571609346 1607215 1609194 1609478 1607861 1607862 1608483 1608734 1608885 1608886 1609193 1609837 16037731613108 1610185 16087361613110 1608312 1609015 1611053 0698898931610834 16110531614086 0274102001611588 037413276101611587 0398262141611865 047415871101609209 047417160101611586 05113646921610462 0574302021611592 057430344101611591 057434345101611585 0874714031610578 106382240101611584 1163865541611589 12910114151609883 15670011609651 16040201612537 1608486 1608735 1608739 1609171 1609192 1609477 1610186 1611051 1611052 1611054 1611308 1611569 16039941614177 16076251614190 16090151614178 1598288 16040191612536 16074341612542 1608662 1609663 1609835 1609836 1610018 1610183 1610184 1612733 1612838 14349691611560 15511341611559 16035771612532 16037771612535 16078611612543 1609191 1610398 1610594 1610808 1611042 1611567 1612320 1612855 1612894 16042861612538 16052651612541 1610589 1611040 03113263451613243 07011842601613244 15282871613754 1610592 1610807 1611041 1611343 1611568 1612456 1613168 16078611615860 16101851615876 16110531615858 1612319 01010411221610521 01010433451610523 03010896211610524 15360481616324 16048191612539 1609474 1611043 1611882 1612452 1612892 1613398 1613631 1614134 1614277 1614342 15801511613911 1611973 1612453 1612454 1612776 1613792 1613928 15982881616542 1609030 1609475 1609832 1609834 1611881 1612780 1614121 1614517 1614659 1614801 1615004 1615035 1613793 1614636 1614664 1615005 1615223 1608887 1609195 1613169 15335211617108 15335211617125 15335231617109 15335251617123 15360451617124 15360481617122 15360491617119 15368441617126 1614663 1615221 1615406 16048201617400 16052661617404 1610399 16110531617401 16128941617407 16142771617405 1614637 1614660 1615003 1615220 1615460 04113331971615532 1609479 1609662 1610182 1611044 1612895 1612898 1614662 1615409 1615969 15368441617582 04113331971618238 11012450811618286 15370771618466 15375111618464 1615459 1615724 1615966 01112864131610461 0274066401609105 0474189551611036 14003301610463 1610019 1613170 1615034 1615407 1615594 1615597 1615931 1615932 1615933 1615936 1616042 1616222 1616223 01112837191610459 10012211561609088 1163881481609912 15269441609652 1610017 1610181 1612893 1614136 1614343 1615593 1615596 1615726 1616043 1616373 1616655 1616657 15376081618285 04113441801619418 08012007931619417 16157261619497 15374101619171 15375121619148 0174003981613246 0274127181613247 0374130731611864 057429443101611863 0574345651615259 0874662221615260 1608738 1611306 1616870 1617014 1617170 1610397 1610588 1611342 1614661 1615595 1616656 1616769 1616770 1616977 1617015 15344051621031 15367331621032 16037771620711 16087391620762 16091931620767 16131701620707 16146371620704 16154071620710 16159361620709 1616044 15360451621392 1611972 1614518 1617171 1617172 1618084 057434345101613248 087469979101615261 0974742301617677 16098321621791 16100171621793 1614137 16150351621789 1615725 16159661621790 1616976 1617016 1617173 X064201511616738 03111887311618839 1616041 1617013 1617438 1617622 1617755 16146641622403 15346311622999 15379501619879 1616863 1617435 1617437 1617621 1617756 1617780 1618064 1618324 1619227 16088871623414 16155941623415 16166551623410 0374169761619880 0374216371619884 0874699791619302 16072141612540 16110441623877 1614800 1615222 1615405 16162231623864 1617436 1617779 1618083 1618325 1618326 1618541 1618542 1620468 1621212 15498011624317 15644771619211 16100171623948 16103971623972 16113061623949 1615930 16162221623950 1619563 1619959 1620156 1620178 15370121621441 1617778 1618543 1619230 1619564 1619960 1620926 1621216 1621870 07113764191610460 1615723 1616761 1617619 1621464 16083121623011 1618321 1620928 1620929 1621215 1621465 1621871 1622105 1622451 16167711626743 01112873011625335 0173995311620348 017402091101620349 0274103961626858 027410602101620347 03010768401625284 03113136191625260 03113159791625309 03113210181625261 04113190051625262 04113380731625268 04113382631625313 05113398231625348 07113645541625349 08011921021626821 089772179011625278 10712541625324 11725471625281 13795111626856 13975431625272 14063871625251 14152231625320 14186751625275 14205641625254 14862021625357 14908581622756 14921821625253 14977441625331 15139731625332 15392111622758 15471551625334 15498021625321 1618323 1618539 1618540 1618771 1618773 1619958 1621214 1621929 1623017 1623130 1623619 017399217101620350 04113366881625336 0574345311618840 06113602401625270 08012020671625316 09012049611625360 10979491625341 11005511625326 15123271625255 15158801625344 15280171625256 15375111623207 1623131 10198341625282 W123831811626818 0285561851626859 0285612261626860 0285612261626861 0398146141626862 04113390301625354 0485942381626865 08011925151625323 09012077811625271 09012161491625351 10896211625355 11012514081625258 13795111626828 14138431625319 14141421625352 14226171625276 14393241625304 15191221625347 15470161625359 15577661625259 15583401625305 W113767531626749 057433406101622504 0974801881622981 14908571622757 1617620 1619229 1619954 1619955 1623286 1623590 1624134 1624357 01010471251625342 01112863201625307 0197693781626857 03113164351625310 04113247871625264 04113325531625266 0485730641626863 05113450481625314 05113461001625340 06011462021626820 09012135521625350 11622881625356 1199885971625329 12910020351626763 14011151627519 14082921625318 14334301625353 14378761625302 15183391625346 15257591625358 15425741625338 W103727271626748 W113767531626755 W113767531626757 W113767531626758 W123831811626819 0874662221622978 1074808051622977 1174966261622980 1622453 1623525 1624498 1625983 0774591441622979 15446051622983 1620466 1623526 1623556 1624358 0698876421622982 15373111626504 1622104 1622364 1623018 1623129 1624499 1624903 1625109 1625576 1625869 1625965 1625981 1626278 16214641624278 1623527 1623994 1624500 1624902 1626329 1626332 1626720 1626331 048573065101628697 0485730651628694 15027571625839 1622452 1625706 1625870 1625966 1626951 1627213 1627234 1627239 1627923 1628473 1627924 1627942 1628474 1628709 1628712 1622691 1623132 1623528 1626721 1626609 1626723 1627922 1628276 1628278 1628705 1628706 1628711 1630815 1630816 1630823 1632523 0974717401632831 1074855961632500 1274981841632832 1624904 1625967 1628704 1629167 1630812 1630814 1631331 1631332 1631337 1631338 1632079 1632080 1632519 1633067 1633138 01010209611621234 01010356111621235 0398229591621374 06011515671621237 1199950091621233 1299952611621272 U04705521621375 02010726561621236 1628277 1628472 1631336 1632081 1632521 1632522 1632525 1633051 1633052 1633068 1633136 1633251 1628707 1629164 1629165 1629166 1632082 1632877 1633137 1633234 1633252 1633494 1633813 1633135 1633469 16098321635856 16100171635839 1625110 16259671635876 1628632 16325251635858 1633491 1633629 14154871636039 15498021635959 15498021636038 16231311636098 16231321636099 16251091636011 16272131636095 1631334 0974796131634115 0974801881633342 1174960541624655 14692401626321 15066131626320 15126781626319 1627203 16284731634116 1630821 1631333 1633433 1633588 1633632 1634013 1634061 R11979971624654 14128301636315 1626722 1631339 1633610 1633887 1615408 1632083 1634012 1634345 1634603 1634608 14148871636826 1632084 1632520 1633139 1633235 1633468 1633589 1633888 1634346 1634435 1634607 1635239 1611423 1634344 1634604 1634605 1634799 1635027 1635147 16249041637237 16272031637105 16279221637110 16282771637129 16320831637106 1632920 1634198 1635143 1635146 1635466 16128921637720 16139281637721 16141211637799 16284721637753 16287091637718 16287111637717 16287121637719 1631492 1632085 16325221637722 1633735 1634199 16346041637716 1635026 1635086 1635145 1635666 16263291637800 16291651637699 01010209611635931 02010518961635932 02010598211635935 1299952651635883 1633470 1633886 1634062 1634606 1635028 1635668 1635897 1636396 01010209461635929 01010355971635934 04011198371635933 089771930011635927 12910114151635928 12910114151635930 14016251638125 15743081638124 16308141638116 1630825 1630827 1632086 1632524 1633814 1634609 1634800 1635144 1635465 1635665 1635894 1636395 1636691 1628708 1635241 15650491638401 1635240 1635671 1636263 1636527 15604311638985 16160411638999 16320841638981 16334331638996 1635895 1635896 1635898 1636264 1636394 1636529 1636689 1636690 1636912 1637429 1637803 16308211639350 16314921639349 16335881639206 16338141639205 16351471639344 16356711639231 1636528 1636894 1637089 1637428 1637430 1637570 1637915 1637298 1637569 1638037 1638168 1637088 1637568 1637914 1638394 1638396 1638494 1638495 1638625 1635087 1637567 1637802 1638491 1638621 1638843 1638239 1638492 1638620 1638841 15374151641186 15367331641705 15374141642335 1637913 1638038 1638257 1638490 1638624 1638839 1639851 1639885 16340621639182 1637981 1638626 1638749 1638840 1639036 1639275 1639276 1639324 1639326 1639761 1639850 1639868 1640238 15346301643148 15374141643146 15375111643147 1638623 1640077 1640289 1640852 1638751 1640853 04113398251634901 0874645381639641 15319061638929 16180641638930 16287091634898 1639760 1639852 1640075 1640078 1640240 1641040 1641148 1641248 0385404031639645 0874662221639643 0974753901639647 0974796131639640 1174836031639642 1174897161639639 117500325101639644 1639119 1639325 1640076 1640239 1640379 1640856 1640883 1640953 1640954 1641149 1641451 16037751605761 1639762 1640375 1640377 1640963 1642275 15374151644332 16026651636508 1640237 1640290 1640854 1640951 1641034 1641453 1174966261639646 1639034 1640074 1640373 1640603 1640952 1641247 1641483 1642441 1642443 1642552 16398501645042 16400741645041 1642845 1640855 1640950 1641030 1641037 1641146 1641754 1642023 W073349621642784 0586057091642755 1086576441642757 1087078511642758 11012450811642762 1287254031642769 1299952651642779 1640602 1641033 1641147 1641185 1641346 1641430 1641448 1641454 1641674 1641675 1642021 1642024 1642039 1642282 1642624 1642701 1642847 1642976 15374151645733 1635893 16386201645498 0285403301642751 0586057091642753 16408551645499 16410341645496 U036672111642781 V011507561642783 W073506901642785 X034012381642786 X044108751642787 1643089 1643090 1643189 1643494 1644012 1644313 1642626 1642846 1643970 14979631646589 15568721646340 15747621646311 15805001646614 1642273 1642551 1642978 1643544 1643628 1643752 1643753 1643968 1643969 1644124 01010356411646845 03113282321646827 09012050681646846 1642844 1643625 1643626 1644285 15361861646341 15367331646335 15374141646854 15374151646677 15845641646667 15868471646666 1642274 1642625 1643190 1644126 1644476 1644695 1644772 1644922 0185312481641508 1099451591643837 15361871646344 15368441647311 15368441647327 15370121647310 15370771647326 16358981643838 1642442 1643092 1643290 1644125 1644286 1644288 1645157 1641450 1643627 1644920 1645101 1645633 1643095 1643289 1643971 1644770 1644924 15368441647838 15368441647840 15370121647837 15370771647839 1644308 1644479 1644921 1644312 1644478 1644287 1644693 1644694 1644771 1644923 1645099 1645102 1645413 1645630 1645790 1645219 1645307 1645632 1645641 1645674 1646232 1646591 16363941649322 16406021649324 16422731649320 16426241649321 16432891649332 1644290 1644477 1645221 1645222 1645414 1645415 1645636 1645673 1645799 1646092 1646920 1647062 1647063 1647064 1645100 1645789 1647315 1647551 0799022011649268 15721101649267 1645631 1646110 1647316 1647418 1647425 1647762 1645672 1646772 1647166 1647426 1648237 1648709 1645220 1646590 1647317 1647550 1647552 14203101650660 14217991650661 14275641650663 14025031651166 1646231 1647318 1647553 15805001651568 16379811651584 16441251651583 16452211651576 16465911651574 1646109 1647059 1647061 1647761 1648063 1648387 1648654 1648799 1648994 16442901651445 16446931651881 16447711651882 16457991651883 0298107921647980 0798898941651165 0885807901647982 10368331647983 1263900731647868 14791991647985 14792001647987 14899881647993 14899901647996 15356261647869 15367331647870 16435441651447 0686279391647865 0763562501647866 0863651571647867 14125231652536 14792021647989 14792031647991 14908571647997 1646771 1647067 1647165 1647303 1648388 1649667 14125231652525 15868471653065 16450991653063 16452191653064 16456301653062 16457901653061 16461101653059 16469201653060 14085521653451 14095201653450 16456321653066 16470621653067 1647066 1650468 1651203 1647065 1647760 1647419 1648236 1648386 1647164 1648385 1648264 1648533 1648796 1648993 1651204 14141411655134 16473181653068 16470591655489 16470611655490 1648655 1648797 1649051 1649214 1649554 1649633 1648798 1649212 1649215 1649555 1649634 1649977 1651533 1652264 1652638 1652639 1653336 1653434 1653890 1649976 1651448 1652424 1652881 1654079 16413461655981 16447701655982 16450991655980 16456301655979 1647313 16474181655978 16475521655988 16482361655976 16483851655977 1647549 1649213 1649972 1651446 1652422 1654151 1654543 1654947 1653889 1647060 1651532 1651653 1652262 1654078 1654544 15600271657444 16474191657296 16474251657297 16475501657295 16495551657279 16514461657294 1648060 1648062 1648795 1649975 1652263 1653520 1653684 1654080 1654268 1654270 1654761 1654936 1654950 1655293 1655296 16436281655253 16439691655252 16443131655251 16444781655250 1652423 1641673 1655294 15655471655985 16272341655986 16272391655987 16431891655983 16449201655984 1654150 1654391 1655292 1655624 1655742 1655948 1642977 16475511658626 16480631658624 16483871658625 16483881658623 1649211 16516531658642 1655138 1655285 1655287 1655288 1655949 1656350 14186751658790 1643288 1648061 1654948 1654949 1655136 1655289 1655623 1655625 1655698 1655743 1656057 03113264581657158 0385638241655809 14141411658781 16103981655807 01010355971649807 01010398071649803 06011537741649805 089815786011649806 089815787011649808 1199892401649804 1649387 1653335 1654762 1655535 1656296 1656497 0586085041655810 14940701655808 1628710 16451011659273 1651651 1656297 14095201657157 14134811657159 15361871657831 1648384 1649631 1656702 1656704 1656786 1657049 16429781658184 16456411658181 1099451591658179 15341711651164 16487091658183 1653711 1655286 1656292 1656295 1656498 1656845 1656903 1657047 1657048 1657933 16358981658180 16435441658182 15374151657832 1652261 16533351661242 1653888 1654545 1656058 1656703 1656904 1657046 1657253 1658702 16487091661490 16557431661475 1657097 1657309 16474191661603 1656899 16569031661561 1656905 019770044101659934 0285361111659480 0285361131659487 0385404031659483 06011650901659485 09012062261659479 12012613911659484 14745461659933 15617161659804 15627761659803 1654269 1654935 1654946 1655290 1656902 1657045 1657098 1657252 1657425 1657566 1657934 1657936 1657997 1658360 1658541 1658700 1658701 1659196 1659316 1660048 1660049 1655139 1655137 1658853 11818801659806 1650466 1658540 1659197 1659620 1660462 04113343381659802 15644761659807 1660055 1656705 1658852 1659622 1660054 1660438 1657640 1658396 1660463 16564981662562 1657641 1658539 1658850 1659319 1661022 14226171572638 1655135 1657255 1659320 1659504 1659769 1660050 1660052 1657938 1660461 1658851 1659315 1659767 1659770 1660464 1661021 1661822 0598577661657192 0786619911662113 15577661664630 1661206 1662279 1662280 1662311 0586085001663467 1656757 1658224 1658542 1659623 1660439 1660440 1660441 1661207 1661240 1661464 1661783 1662313 1662521 1663058 1663060 1663061 1663062 15373091665661 16585421666234 1662310 1662312 1662751 1663377 0974790341663383 15278571663771 15642471663772 16556981661034 1662750 1662937 1662938 1663375 03111887311663770 1656901 1657639 1658698 1659505 1660442 1660444 1662976 1662977 1663097 1663374 1661237 1663135 1663879 1661463 1661781 1661859 1662748 1657095 1659621 1660914 1661020 1663372 1663642 1663645 1664674 1664676 1665226 1665517 02010198511667830 15346301667829 1658699 1661855 1664675 1665225 1665514 1659318 1659768 1659819 1662417 1664080 1664624 1665228 1665518 1665797 1665973 1662282 1663876 1664626 1665974 1666217 1665976 1661782 1662089 1662749 1663913 1665028 1665975 1665977 1663581 1663644 1664967 1665531 1667760 1663166 1666219 1658543 1666423 1667759 1666424 14112241665980 14389541665978 1658544 1659494 1661019 1661857 1663878 1667217 1667218 1667287 1667360 1667563 1667761 1667763 1668023 1668079 1667562 1667762 1662975 1666220 1667215 1667219 1667361 1667765 1668024 1668025 1668080 1668171 1669205 0686375181668018 127506393101665981 15183381665979 1664078 1665027 1666010 1667764 1668452 1669206 1669471 16572551668758 1662281 1665690 1666218 1669591 1661821 1666730 1669291 1669592 1659195 1663580 1669593 1669810 1670883 1662696 1663643 1665516 1667769 1669285 1669595 1670611 1669287 1669808 1670886 038553017101671762 1665689 1669597 1664764 1665530 1669397 1669596 1669630 1669812 1670198 1670882 1671296 1665708 1665798 1670200 1670201 1670202 1670305 1670306 1670442 1672240 1663877 1664625 1665227 1670444 1670925 1671158 1667770 1670922 1670923 1671128 1671516 1672682 1665026 1668077 1671294 1671517 1672239 1672242 1665515 1667341 1672241 1672683 1672856 1669293 01111887181675022 03111887421675035 04113321771675029 04113343381675033 14016251675034 1663912 1664079 1671519 1672243 1672244 1672468 1672469 1672582 1672583 1673372 0485859711674791 0686252491674793 07113942331675025 0886707321674795 0985808391674797 1174960541674800 16375671675030 16496311675031 1666731 1667343 1670622 1670623 1671518 1672470 1673593 10011713611675023 14002981675032 15596031675024 0686085051674792 0886707301674794 0985808381674796 0985808471674798 1174908681674799 13991831675028 1671295 1672858 1673199 1673211 1674090 1674845 1674846 1675632 1667342 1671520 1671522 1675068 1675256 01010209461677581 01010398071677582 089815786011677580 11012317441677001 12910147411677583 1299988631677002 1666101 1667344 1671521 1672581 1673932 1673934 1675484 0197579511677554 14329501677559 14581781677558 14726581677561 14764931677562 14852761677563 14868751677564 14924011677555 14979631677556 15089381677557 1672057 1673200 1673942 1673943 1675364 1675844 1676090 1668081 1671344 1671523 1676089 1676091 14534101677560 1675037 1675039 1676054 1668078 1672580 1673198 1673944 1675038 1675483 1675633 1676055 1676256 1676874 1677119 1672737 1675636 0498281091680030 1664428 1674844 1675635 1676877 1677120 1677590 1677876 1679283 16724701680464 1675366 1675522 1676052 1676056 1676876 1677589 1678141 1679281 1086576441674801 01010263791677584 1099641231680046 1672238 1673592 1673933 1675521 1675634 1676255 1677118 1678038 0285403231681908 0385638231681966 0586085001681914 0686314701681912 0785808691681911 0885808031681910 0885808131681913 14730291684180 14767151684181 14887821684182 1672684 1676053 1677591 1678037 1678759 03010828411684169 14401841684171 14581641684179 16452221684174 02113041321684168 12012592161684176 14395821684178 15033091684172 15093731684177 15147341684167 15147361684173 U121437911684170 X054183521684175 01010348161682762 0197693781682761 0274102001682763 0398146141682765 0398146141682766 0485942381682760 068634645101681909 07113786591682698 08011905061682759 08011925211682757 15375121682754 15375131682755 15375131682797 16585391682802 1671129 1672736 1673210 1677877 1678041 1678042 0385716691665636 0885808001683553 0885808141683410 0886707151683554 0886707321683413 0985808381683409 15176371682764 15176371682770 15375121682768 15375121682772 1673591 1677121 1678040 1678758 1678760 1678965 1681295 0285612261682767 15375111682769 15375121682771 1679001 1679282 1679725 1680163 0585992551683407 1680159 1680161 1681296 1675255 1679284 1679726 1681299 1681936 1682087\n1682919 1682920 1680463 1682921 078580819101683408 08011925151684145 089815786011684146 16711291683801 1676057 1676254 1678039 16780411683800 16780421683799 1678964 1680005 1680167 1681297 1681298 03111887311682699 03113192221682700 1678963 1678999 1679002 1680172 1681578 1681937 1682922 1682924 1683507 1683561 1684103 1684110 1684112 1684114 1684737 1674088 1680169 1681300 1681854 1681938 1682438 1682923 1683508 1684106 1684108 1684739 1685287 14210801688182 14240731688184 13972701688449 14118301688451 13986021688450 14122321688448 1681852 1681853 1683006 1684116 1684738 1684740 1684766 1685288 1685289 1685456 1685457 1685459 1686839 1675631 1677875 1680321 1682434 1682436 1683562 1684765 1684979 1685317 1686838 07113821111688932 079893176101689060 15454011689073 15526121689074 1685460 14047991689748 14940701689749 15335191689747 15335221689752 15335241689750 15356661689751 1684977 1685458 1685461 1685462 1685513 1686185 1686187 1686840 1686944 1674089 1683770 1684978 1685838 1685839 1686544 1687002 1687206 1688096 1688288 14349641689804 15382781688931 1683883 1688385 1686855 1687449 1687532 1687533 1687780 1688286 1688287 1688289 1688314 1688386 1687094 1689020 1689026 0886766761691782 1683882 1687536 1689025 1689347 1689850 02010198511691774 039829635101691775 0586085001689213 065209594101691776 0652095941691777 0686252491689128 0686314701689214 0785808191691778 0786619881691779 0786737671689179 0798931761691780 0885808031689215 0885808141691781 0886707151689129 11012317441691783 11012341471691784 1199795771691785 1275045211691786 1685318 1686854 1687779 1687781 1688012 1688315 1689027 1689349 1689849 1690446 0274106021691769 0298054651691770 0385530171691771 079889898101691772 15002251691767 15033071691763 15141551691764 15284491691765 1688011 R11976081691773 15107931692508 15471441691766 1688313 1689028 8200441691768 01112894491691844 1689029 1690832 0197711531689130 0398245821689132 04113450461691845 0498296371689133 05011317991691846 0798898941689134 0999202941689135 0999202951689136 1099329701689137 1099424871689181 15427091692524 15486691692512 1687535 1688405 1688406 1688408 1689827 1690447 1690829 1690830 1690831 11012247341691847 13939211691848 15263801692547 15319061691843 15405441692520 15405451692523 15427881692510 15521731692516 15660391692682 16849781691681 1689350 0197579511689149 0197615991689162 02010356061692278 03010712881692245 04010969641692272 04113306371692290 04113306371692298 05113498711692291 05113646921689112 0574251481689166 0686085051689152 0686252491689142 07113764191692279 0874662221689209 0886707151689153 0886707301689144 0886707321689143 0985808381689147 0985808391689150 0985808471689141 1074808051689170 1163865541689208 1174836031689210 1174908681689151 1174960541689140 12012592161689161 14003301689168 14025031689102 14076281692280 14147461692281 14212741692282 14370031689109 14730291689156 14848021689108 14887821689157 14924011689148 14924021692247 14924101689110 14979631689145 14979811692256 15040571692248 15053981689165 15053981689167 15089381689146 15370141692292 15419791689104 15438551689103 15480821689158 15487751692274 15507921689186 15528151689105 15552961692250 15562351689159 15609021689107 15612021689106 15794171692293 15810181692259 15850711689207 15860241689206 15964931689187 15993671689163 16094741689111 16094741689113 16103981689200 16113421689169 16154591689114 16231321689172 16282771689115 16308251689118 16313371689171 16320861689117 16325241689174 16346041689116 16380381689119 16383941692238 16403791689177 16408831689176 16420391689175 16443121689188 16449231689192 16449241689191 16451571689193 16470641689189 16482371689194 16487091689138 16526381689212 16533351689139 16535201689124 16540781689195 16551361689196 16564981689199 16567051689202 16567861689198 16569041689197 16570471689126 16570491692235 16572551689101 16574251692234 16576411689097 16588531689201 16604611689093 16612061692289 16612371689099 16612401689088 16617811689100 16623121689086 16623131689089 16627481689098 16627511689087 16633771689091 16829221689154 1688316 1688321 1689033 1689351 1690224 1690232 1690449 1691230 1692095 1692273 1692376 U121437911689183 X054183521689184 1689031 1689032 1689348 1689352 1690448 1691239 1692093 1692096 1692117 1692884 03010969651692246 0586085001692297 07113887941692299 12012613911689127 14224281689164 14395821689155 14600861692241 14689411692275 14745461689125 14761821692295 14979611692252 14979611692255 14979641692253 15266581692276 15312581692254 15322541692257 15322541692258 15510481692249 15552961689160 15796951692277 15835691692251 15856391692296 15989801692242 16340621689173 16400781692239 16410331689120 16439711689122 16475531689123 16483861689211 16516531689178 16555351692240 16557431692237 16570481692236 16573091692243 16588521692244 01010231941692287 0197864801692270 0297983591692265 0498347181692260 05011284531692288 0698799341692263 0899202971692261 1198703541692269 12012689141692262 16646761694625 1688407 1689030 1690113 1690450 1690451 1691237 1691240 1691716 1692097 1692377 1692845 1692846 16630971694619 16631351694623 16633751694622 0398164801692264 0398230031692266 06011504761692285 0698898931692284 07011782091692286 0785904891693467 0799022011692283 1099641231692268 15306391694642 15718271694643 15761291694638 15963181694639 15984791694644 16042841694640 16408551694635 16430921694633 16430951694634 16470661694631 16473181694630 16537111694637 16568451694636 16585441693464 16623101694632 16681711693466 1691714 1691956 1692094 1692595 1692885 1693122 15516051694626 16057871694627 16456321694628 16604641694620 16629371694624 16633741694621 1689090 1691712 1691960 1692847 1693503 1693504 1694240 03010828411689182 16444781693465 1692098 1692385 1692848 1693123 1693428 1693502 1694068 1694553 16569031694629 0498336271692267 1087061741689180 16595041693462 16640781693463 1693505 U06716531682758 1694241 1694554 1692384 1693429 1693431 16877801691680 1693933 1693934 1694067 1694069 1694321 1694563 1694745 1694904 1694564 1694609 1694744 1694906 1695814 1691238 1694746 1694967 1695644 1695645 1695980 1695984 1696240 1694905 1694907 1694968 1692378 1693120 1694319 1694748 1696241 1696436 1696635 1694747 1694753 1694908 1694909 1695815 1695979 1696857 08011976211697410 14282961697408 16694711697366 1692379 16931201697841 1694318 1694749 1694751 1695986 02111887301697414 09012077821697864 15718341697677 16185421697678 16672151697363 1692883 1695157 1695246 1696867 02111887381697425 08012033661697432 10838881697415 11822771697421 14354491697429 15191221697423 15489741697430 16677641697365 16698101697364 1695968 1696858 1696876 02113022721697412 03113171971697420 13975431697418 16048201697427 16078611697679 1694320 1694754 1696868 1697117 1697844 1694242 1695245 1695982 1695983 1697116 1697845 1698366 16412481698372 1694066 1698367 0285361111698874 0385716691698862 09012062261699388 10686301697416 1199813321700185 14312431698900 15010871698839 15341711698865 15535861698881 15562351699389 15585631698898 15755731698882 15860241698864 16037701698857 16141361698858 16483861698866 16552921698834 16600491698838 16600501698836 16600551698884 16630581698897 16630601698887 1678041698872 16792841698837 1693121 1694755 1694756 1696438 1697118 15996601699390 15344051701045 1697842 1698365 1698688 0785904891698896 16414531698371 1695074 1695244 1695969 1695981 1696878 1692886 1697097 1698368 1698413 1699029 1699193 1699194 14936851698850 15029131698869 15249511698873 1692887 1695075 1697119 1699522 1699991 1699992 1693432 1693935 1694071 1695985 1697099 1698369 1698370 1698411 1698414 1699990 1699993 1700593 1700601 1698138 1699994 1699995 1700436 0786619891698868 1697322 1698575 1698689 1699534 1701134 1696437 1697323 1699521 1700602 1701555 1694750 1698576 1698578 1698581 1701556 1701558 1701787 1701790 1701136 1697098 1698577 1700762 1701820 1701823 1702504 1698579 1700597 1700763 1701344 1701788 1701789 1702503 1698580 1700766 1701822 1702657 1698412 1700594 1701821 1700434 1700764 1702867 1702868 1703291 1700595 1698140 1701792 1703816 1700435 1700437 1700598 1701345 1702885 1702886 1703678 1702656 1703290 1703293 1703304 1703677 1703864 1703954 1704421 1702866 1703289 1704475 1704500 1702505 1694072 1703292 1703294 1703813 1703814 1703815 1704090 1704091 1704092 0586023471705756 1695813 1699536 1700765 1702662 1703303 1703863 1704422 1704477 1706102 1697324 1704423 1704976 1705083 1705262 15344051707881 15375111707880 1699031 1700760 1702663 0297983581698870 0398164801706071 0498336261706072 14003361700149 16801691700131 1696866 1697843 1700596 1700761 1704672 1704675 1705134 1705497 U04705651706074 14210801708487 15392101707586 16959811703871 16959861703872 16973221704184 16983681706090 1699538 1705498 1705618 1706346 16940711708891 17004341708893 1702869 1704476 1704673 1704674 1704978 1706099 1706306 1706344 1706345 1707030 0297864871709804 066353694101709806 07113821111709802 075210935101709807 15344031709805 1704170 1704804 1704979 1705499 1706010 1707006 1707027 1087120521709988 1287297951709956 14802471707592 15024921707595 14082921710251 14141411710252 16195631707096 16627501707097 16754831707098 1701791 1702342 1705135 1705260 1705261 1707368 1707395 1707830 14940711707594 1695076 1695247 1698139 1701557 1702341 1705989 1707033 1708112 1708113 15282861703540 1694752 1705136 1705700 1706952 1707466 1707508 1707839 1708004 1708619 1705711 1705712 1706951 1707506 1708270 1708426 1708617 1702343 1705990 1705991 1707367 1707371 1710401 15205201711065 15582231711066 1708003 1707465 1707841 1709564 1706950 1707369 1710613 1699535 1707507 1707838 1708294 1708296 1708920 1709141 1709382 1709383 1709563 1709816 1710852 1711222 0485977971707519 1711037 0398164801710530 04011098971710369 0598611121710368 0698799341710367 0698843351710366 1086862891710362 15341691710294 16968571706542 16971181706091 1703302 1705698 1708618 1708710 1709380 0297991661710370 0398245821710363 0398262141710283 0498336261710361 08011873351710372 10828391698871 16451001700150 16552941703505 16640781703506 16771191703508 16991931706092 1702344 17063441710343 17075081711562 1709026 14850601713037 1709142 1710807 1710850 1708708 1709140 1709381 1709815 1704977 1708110 1708295 1709138 1709139 1709931 1709932 1709998 1710135 1710612 1705708 1707740 1710400 1710614 1710616 1711763 1711767 15375111715400 1705707 1705988 1707365 1709814 1710615 1710851 1712265 03113282321703144 0398192681703130 0398262141703134 0786606611703139 088672419101703138 1087298031703129 1187466931703142 16412481703137 16750371703133 16829241703128 U05705691703136 1711766 1712635 1710133 1706103 1706104 1711457 1711458 1712940 17057071716989 17073651716990 17077401716986 1708709 1709813 17106151716992 16861871716987 17057081716991 17099321716988 17099321716993 1712939 1707372 1707840 15341691703126 15341691703131 15341691703132 1710137 1712095 1713164 0197576031710287 0297967971710311 15344051710295 1705710 1708425 1709935 1710139 1710668 1711039 1711095 1711224 1711253 1711292 1711456 1711582 1711733 1712096 1712264 1712268 1712437 1712625 1712630 1712631 1712938 1713158 1713159 1713661 1713876 1714332 1714934 1715281 1715290 1715645 15554461718114 14205651718473 14226171718472 1714531 1705709 1709934 1711764 1713284 1713553 1715708 1713282 1713878 1711223 1712097 1712936 1713554 1713877 1714065 1714274 1714783 1715273 1716166 1710849 1713662 1714427 1714442 1714787 1715401 1716107 1716224 1716393 1711731 1714441 1714530 1716332 0197814481710307 0197864841710288 03113193751710301 03113253231719128 0498347171707091 0686252451707009 0886766371707011 0886766761707088 0985808381710374 10012252651710299 1086576401710371 1086607521707012 11012494201710365 128744180101710306 14740111707099 15071711710364 15346301710297 15346311710296 15358651710292 15544891719126 16402891710308 1712628 1713663 1714785 1714786 1715640 1715885 1716330 01010355971715154 03010896191715155 0498336281707008 0686375171715150 088676678201715156 1299961801715149 16626961715152 1705699 17057071716015 1714331 1715403 1716333 1716334 1716631 1716910 1717428 1714201 1716632 1716634 1716914 1717429 1718335 1718337 1718538 1711919 0197619521715151 0197864831710529 0574343451713959 058605175101718417 078672443101710289 0798931771711918 14011151710298 15340291718414 1709239 1712403 1716331 1718242 T101347631711915 15582231719241 14186751721272 088676678101718416 1712937 1714935 1714936 1715646 1715755 1716913 1717216 1717217 1718038 1718243 1718244 1718338 1719010 0297967511711914 0297991661711913 078661988101710305 0874675461713987 0886766761713879 16019681714300 1709565 1711920 1714937 1716559 1716633 1717218 1718093 1718094 1718811 1719452 1710141 1711225 1712633 1717430 1718533 1719011 1719133 1719290 1719548 089815787011719767 16630971715157 1711765 1712093 1712094 1712278 1712627 1713163 1713165 1714302 1715404 1715754 1715887 1716074 1716982 1718334 1718810 1719371 1719372 1719450 1718039 1718532 1720322 1708922 1718336 1719858 1719869 1720352 1715886 1718531 1719374 1720001 1720180 1720707 1720857 038553017101722034 16931201722031 16970981722044 17057081716285 17101371722032 1711921 1714064 1715402 1718037 1718539 1719291 1720296 1720461 1720987 1720988 1721005 1721230 1721249 1721397 1721398 1721677 1721678 1717961 1706953 1716520 1719980 1720443 1720460 1721248 1721679 1721790 1719373 1721549 14100061722055 1712083 1713162 1718809 1719130 1719979 1720407 1721794 02111887381722048 03113254191722037 0498296371722029 05113454841722054 088676636101722040 089772181011723131 089815780011722047 089815782011722041 14009451722049 14261191722052 14349691722046 15341681722026 15853991722043 16545431722033 16655141722030 1712077 1722005 1722244 1720157 1714784 1718540 1718722 1720986 1722006 1722007 1722241 1722754 0799050291723140 089771930011723137 089772180011723127 089772181011723134 089815782011723125 10783871724142 11460621723122 14011011724143 15792331723124 16588501723128 1719131 1721359 1722490 1723285 1723286 1722585 1720985 1721228 1711036 1711424 1722492 1723031 1723287 1724123 1724386 1724388 05113460951722027 089772182011722042 089815787111723123 1719132 1722690 1722961 1723033 1723289 1711922 1717215 1718100 1721539 1722248 1723280 1723602 1724045 1724390 1724614 17203521726885 17212301726884 17215491726887 1721791 17217941726882 1723029 17230331726883 1723414 1724044 1724382 1724612 1724197 1199885971725266 14107131722050 16094781725267 17136621727461 1722691 1723599 1725697 1722692 1723411 1723601 1724051 1724800 1725342 1722243 1724049 1721226 1722246 1722247 1722491 1722693 1723032 1720686 1721793 1723249 1723288 1723378 1724198 1724542 1725277 1725341 1725696 1725801 1725807 1725809 089771930011726583 089815782011726609 089815787011726590 089815787011726596 1299952531726577 1725528 1725952 089815786011726599 089815786111726601 1199783371726578 1723030 1725076 1725954 089772182011726593 089772182111726585 089815780011726606 089815782011726611 089815787111726587 12910114151726603 15583401728298 T041249611726580 1721792 1723600 1724613 15568731728132 15589531728131 1723247 1725276 1725953 1726156 1726664 1726794 1724798 1725527 1726075 1726306 1725075 1725802 1724794 0785904891725382 14868701725265 1725077 1725955 1726307 1726309 1726665 1726666 1726668 1727202 1727713 U10802261725065 0498437401728198 0798808501728200 089772179011728210 089815779011728202 0898703051728196 17131641728209 1725808 1723250 1724609 1725073 1727018 1727540 1727814 1728067 1728451 1728851 1720462 1723248 1726076 1727811 1727812 1728624 1728852 1728854 1729278 1730026 1730028 0598625401728197 0798908781728195 0798909941728199 0863683411729258 089771930011728201 089815780011728203 10209451728204 1099451591728211 15806611729248 1725526 1727839 1727840 1728675 1725803 1726903 1729375 1729203 1729989 1730178 1724792 1726667 1727350 1727548 1729492 1730027 1730177 1730361 1730564 1724050 1726157 1727336 1728406 1730609 1731040 1731056 1730345 1730482 01010411221732031 17284511731983 1729895 1730145 1731041 1731104 1731294 1732322 1731295 1731572 1731996 1732434 01010209461731937 089771930011732130 089815780011731935 089815782011732033 089815782011732034 089815787111732385 119994270101731936 119995012101731938 12910163101732032 12910186001732384 17046751732399 1728853 1729892 1731496 1731621 089815786011732383 16110531732382 1729816 1730346 1731135 1732063 089815779011732129 1726308 1727281 1727337 1730179 1730679 1731292 1731407 1731774 1731838 1731841 1731887 1733427 0109952641733384 12910087781733394 1730344 1732522 1732435 0109952641733378 0698876431733371 089815782011733395 10209591734810 1099329711734818 14908731722053 1732217 1732349 1733436 1733856 01010433461733360 0598611141734817 0698611191733387 10147431734814 10209451734803 1730668 1731111 1731799 1731839 1732436 1732514 1733186 1733476 1734259 1734337 1734490 1734942 1735035 01010185531733381 01010220801733396 01010262141733383 0285403251733369 0285604581733379 03010783901733361 04010931061733386 089815779011734819 089815786011733397 089815786011733398 089815786011734813 1199741491733388 1199741491733389 12910087781734811 17220071735666 17317741735669 1733271 1734349 1735036 1735303 01010355971733359 089815780011734805 10147421734807 10185991734812 14655101733367 1730343 1735304 1735672 1735875 0285403251733377 03113249431734816 08011913901734815 14707701733374 15344031738001 17080031735667 17263071735665 1730337 1733187 1733420 1733433 1733434 1733435 1733971 1735294 1735461 1733088 1734489 1734864 1734876 1734943 1735733 1736041 1736042 1731840 1731842 1732433 1732520 1733091 1733475 1734686 1735656 1735874 1735877 1735878 1736040 01010209461736269 089815780011736271 089815780111736272 089815787111736267 12910114151736270 1735034 1735125 1735734 1736043 1730362 1730746 1731527 1734257 1735033 1735037 1735911 1736044 1736347 1736484 1737037 01010271781736266 1199813321737075 119989240101737079 119995009101736273 12910147411737088 16949061737519 1731528 1732866 1733679 1734258 1735650 1735879 1736039 1736349 1736483 1737033 1737034 1737036 1737044 1737046 1737757 1737759 10095511737080 16787591737532 17318871735668 1733855 1735735 1735876 1737520 1738056 1730378 1735873 1737048 1737758 1738069 1738072 1738701 10147421739786 1737045 1738068 1738934 1738936 01010209461740249 01010306751740284 01010398071739797 01010411211739787 02010726561739796 08011897251739795 10123891740246 10147421740254 10209451739790 10398071739789 10398081739791 10598201740245 10726571739792 11460611740291 1199813321739788 1199885971739793 1299961801740292 14103891739794 15239061740582 15360491740273 16016541739799 16078601739800 16973241739798 17311111735670 1738057 1738058 1738059 1738070 1738073 1738450 1738480 1738703 1738935 01010209461737083 01010284621740265 01010284621740266 01010398071740250 02010526301737084 089815787011740282 10163101740255 10185991740294 10209451740248 10209611740259 10271781740280 10306761740261 10471261740289 10471281740285 10726571740260 10828391740298 11012272721740296 1199950091737087 12910114151737086 12910114151740252 12910198351740257 12910209591740258 16032991740287 1737035 1738071 1739202 1739450 1736485 1738449 1720162 1740224 15344031741948 1740417 1738448 1738937 1739454 1740049 1740463 1740465 1720159 1720160 1731055 1732867 1738702 1739452 1740050 1740455 1740709 1740710 1740961 01010209611737078 01010411221737076 1199950121737090 12910163101737081 16006791742489 1741230 1741930 01010411221742468 02010598211742520 10262871741791 10262871742518 12910163101742496 15887041742501 16854571741792 1737047 1740580 1740580 1741056 1741231 1741597 02010608581741788 1739203 1738060 1739451 1740962 1741378 1741598 1742050 1742101 15470161740616 15650491740615 1739453 1741857 1742255 01010209461742491 01010306761742502 01010398051742486 01010411191742493 01010433461742523 02010518961742508 02010518961742513 02010583981742519 08011976221742510 08011976221742511 0999220741742907 10012317681742507 10496941741793 11233561742504 1199783371742903 1299867991742910 15166741742911 15299521742514 1738704 1738933 1739654 1742100 1742400 1742885 14240741745471 15341691745663 15344051745671 15344051745798 1739456 1742102 1742602 1742604 1742651 1742869 1742870 1743030 1743360 1743809 15344031746279 15346301746278 14783001740614 01010356111742509 03010829301744747 03113263451742481 05113539141743308 10566171744752 14574851743346 14689411743292 15158681743328 15982881743368 17093811744775 17246121744779 1739455 1741800 1743812 1744028 01010355971744745 01010411221744759 03010979501744736 03010991661743353 0598610241743298 059861113101743309 079861886101744767 0798931761744765 0863722581743310 0999220691744770 10012327691743289 10017461744741 10398081744748 10538581744737 10686301744749 10712481744757 11005511744754 11012546351743291 11198371744761 11232791744753 1199795791744772 12012606851744738 1286531651743313 1287431821744768 14295021743342 14315481743316 14512161743345 14561231743320 14561241743321 14638491743348 14855621743273 14925391743355 14979871743349 14980111743363 15051911743350 15322541743364 15335011743294 15374731743295 15374741743283 15410011743351 15419751743331 15541411743332 15573631743280 15573641743284 15573651743333 15655451743285 16150351743330 16485331743336 16549481743337 16557421743338 16780421743276 16985771744763 17046721744781 17161071744780 1720158 1741623 1742238 1742603 1742884 1743031 1743604 1744937 15344051747105 1733232 1742145 1742883 1744924 1742919 1743023 1744222 1744828 0297983581743302 11012393371743277 14925381743354 14979611743359 15511301743279 15799151743324 16039941743297 16731991743339 17192901744778 1740416 01010411191745927 01010433461745925 01112787091743301 02010518961745923 03010783871745924 14311831745926 14730301743278 15322551743366 15362721743322 15487751743296 15521711743275 15595471743335 15638111743281 15853881743325 16037701743329 17131631744776 1743032 1744498 1744499 1745123 15166711746200 15230041746188 15324321746199 15554461746190 1742871 1743811 1744029 1744827 1745445 1746035 17421001746891 1745240 1745849 1746181 02010598211744740 10263791744746 1740460 1744224 1744825 1744951 1745260 1746186 1746322 1746714 01010411221744755 10496941744760 17172161744773 17424001746889 1746033 1746506 17389331750108 0999202951743341 10598211744739 10655801744758 10838891744756 11179131744744 11704961747951 14357011743318 14969471743274 17377571745098 1744496 1744936 1744956 1745109 1745669 1745846 1745957 1746029 U02667071743326 15546441743270 01010185531744771 01010398081745928 10583981747953 10896211747955 10950141747957 11211071747336 15583401746198 1744824 1747106 01010263791747948 02010458361749536 02010538661749272 02010538671749273 02010598201749302 04011005511749245 04011198421749539 10198341749307 10598201749299 10896211747959 11005511749246 11505571749264 16325191749260 1740581 1744225 1745239 1745672 1745674 1746323 1746326 1746715 1747379 1747539 1740462 15583401750694 1744955 1746754 1747380 1745444 1748829 1087061741743290 17430311749129 1748663 1749288 14393241751499 1749881 1748475 02010598201751166 03010768381751165 10838881751164 1748208 1748664 03111887331752894 14247171752880 15344031752893 17394531751921 17404161751922 17404171751923 1744030 17456741751920 17460351751924 17465061751925 1746717 1748655 1748665 1749053 1749384 1749579 1750643 1749568 1745848 1747570 1748009 1749370 1749882 1749914 1750681 1750739 1748504 1748830 14217991753488 1749569 1750180 1742886 1745673 1750110 1750344 1750738 1751626 15308161753779 15533651753780 16325251753781 0298005771752561 04011129681752568 04113398251753076 10950141751171 16673411753062 16695971753063 17226901752686 1746417 1746504 1747107 1747513 1747619 1748006 1748010 1749383 1750181 1750343 1750380 1750455 1750500 1750502 1750961 1752160 1752439 15548431753409 02010566171749304 05113433571753539 06011515671753529 11471291753528 16551391753075 1744826 1749055 1751346 1752018 1752523 1748011 1752438 1753112 10950141753744 17082941753702 17314071753723 17323221753725 1746718 1746740 1747381 1748224 1749051 1749117 1749382 1749580 1750355 1750456 1750499 1750736 1750960 1751628 1751859 1752169 1752715 1752716 1752717 1752941 10819021751174 15317841746919 15344031755239 15366291746918 15405451746917 1746716 1741055 1747644 1749052 1749371 1749915 1749916 1750737 1752019 1752942 1753039 1753291 1753366 1753452 1753684 1751858 1751861 1752440 1753931 1747775 1750959 1753472 1753884 1743529 1745847 17499141755207 1750076 1750734 1752020 1752463 1752948 1754059 1754430 1754437 1754560 1754797 02010518961749250 03010783901751170 03010796731751169 10953111751168 11249251749247 14658321754667 14702991754659 14715181754673 15119311754675 15308171754651 15321141754653 15321141754677 15321161754649 15410031754650 15419781754672 15439711754670 15552971754656 15592481754666 15619441754664 15619441754678 15683121754654 15984791754662 16002601754661 16019681754657 16103971754680 16470671754669 17521601755208 1754139 1754799 1754949 1754950 1749713 1752718 1753040 1754833 1746961 1749712 1750187 03010783871754648 03010829301754643 05113646941754638 0598526011754640 11145131754173 16018101756231 1744497 17467161755941 17480091755205 1748229 1748231 17484751755344 1748505 1750735 17507391755206 1750832 1752021 17524381755346 1752529 17527161755345 1753041 1753368 1753736 1753742 1754138 1755240 1755241 1755607 1755738 1755800 1756173 1756288 17410551757975 17410561757976 17421021757980 17430321757977 17480111757978 17544371757982 01010355971757973 01010433461757972 02010598211757971 10829311756680 10896211757974 10950141757970 17409621757979 1751627 1755945 1756097 1756174 1756249 1756348 1756349 1756628 1756649 1756834 1750353 1753116 1753290 1753719 1756695 1756931 02010695601757463 07011651411758383 07011803961758384 07011819651758388 08011913551758236 08012000831758239 10209611758243 11033121753065 11466421758244 11547841758385 15029141758387 15582231758382 15642471758386 16439691757462 1750501 1755464 1756251 1756627 15346301759843 1754043 1749054 1755622 1755480 1757011 17428841760588 17490541761035 17537421760587 1099451591759253 1754946 1757224 1758042 02010695601760996 16865441761001 17334271760988 17463261761000 17504991760992 17560971760990 1757824 1758041 1757155 1757563 1757684 1757686 1758947 1752531 1752943 1753685 1754800 1754802 1755463 1755946 1756253 1757225 1757434 1757435 1757564 1757565 1757566 1757823 1758044 1758267 1758481 1758652 1758656 1758949 1758975 1758976 1760036 1760057 17490541761589 17490541761593 17490541761600 14868811761549 15409991761577 16731991761541 14924031761563 14979521761564 14979641761574 14979921761565 15002271761559 15060311761556 15289391761540 15334961761547 15335021761550 15335031761548 15341711761560 15346321761557 15370131761561 15853881761555 15979471761575 16155931761566 16403751761558 16414831761576 16731991761542 1754441 1759282 1759620 1760037 1760056 1760711 1752161 1759279 1759280 1759760 1759761 1760311 1761389 01010355941761583 03010796731761585 0799050301761545 0999220741761572 1099428191761553 1199885971761581 12910087781761580 12910095511761582 1299867981761573 14141441761570 1753948 1756817 1758040 1758043 1758046 1758047 1758649 1758974 1759445 1759621 1759969 1759970 1760706 1760707 1761205 1761208 15341691764754 15495431764568 1753740 1759610 1760708 1761207 1761755 1758655 1760307 1762236 1760709 1762838 1760312 1761388 1761960 1763116 1761754 1762360 15346301754548 15322621766986 1759762 1760308 1761917 1763245 1757226 1758946 1758948 1759281 1760710 1762059 1762598 1763112 1763114 1763465 1763466 1763605 1763606 1763607 1763909 1763996 1763998 1764386 1764387 1764388 1764438 1764761 15596061766991 08012006841766119 11608881766139 11704961766138 16159331766137 1756256 1761758 1761916 1762087 1763111 1763244 1763458 1763459 1764439 1764767 1764776 1764854 1764966 1765193 01010433451766118 06011537741766121 07011756031766143 09012035171766116 09012042821766141 10012252301766129 10012317681766132 11528341766124 11555561766142 11818791766149 16402371766135 17544371768392 1755794 17559461768389 17562531768391 1757227 17580431768393 17582671768390 1758993 1761387 1762361 1762597 1763168 1763608 1763908 1764853 1764972 1765186 1765252 1765366 1765367 1765371 1765605 1765729 1765764 1766234 1766235 15346301769478 1765604 05011169641767343 06011505631767346 08011913531767361 08012006771767363 09012049591767362 11382031767360 1759369 1762061 1762534 1764396 1764965 1765762 1766519 1766520 1767392 01010185521767348 06011504761767350 07011642401767349 07011718111768532 08012006771768534 11012376931767347 1299952531767358 15361871767526 1761915 1762793 1765185 1765253 1765731 1766757 1766787 1767176 15335211767543 15360401767544 15364311767527 15367331767528 15370121767529 1758045 1758972 1759766 1762060 1763113 1763464 1764394 1766236 1766237 1766758 1766761 1766762 1757000 1761204 1763609 1765370 1765990 1767163 1767187 1767484 1767486 1767726 1768084 1768094 1768378 1755623 1763115 1764763 1765991 1767162 1767581 14370501755761 15072551755758 15131701755759 1762836 1764964 1766514 1768783 1769437 1765730 1766756 1768379 15126801770714 15395831770740 17612041773572 17639081773571 1767177 1767483 1767729 1768090 1769090 1769661 1765985 1767728 1769089 1770132 15191221767058 15284541771067 15317841769908 15344051774490 15554461756055 15570961770713 15577661769896 1747773 1765732 1766765 1768784 1769091 1769261 1769660 1769664 1769683 1769865 1770130 1770131 1770397 1770400 1770614 1770849 1771032 1771245 1765984 1766921 1767168 1768367 1768368 1768672 1768787 1769435 1769436 1770398 1770004 1770396 1770619 1771544 1771309 1762794 1768788 1769439 1769659 1769866 1769867 1770399 1770850 1770914 1771547 1772410 1772415 1772610 1772855 1773362 1771243 1771308 1772073 1772414 1772576 1772798 1772801 1772852 1773722 0586051751776406 1287431851776407 1287441801776405 15344031776404 15344051776510 15346301776509 15375111776403 15375111776409 09012104841775788 1771246 1771545 1772857 1772858 1773072 1773074 1773361 1773364 1773892 1774197 1774477 07011756031775790 08011925141775796 09011713651775792 09012049611775798 09012077811775797 09012078351775789 09012135521775793 11589121775791 11606171775795 1759272 1772856 1773593 1773724 1773738 1773743 1774064 1774657 17597661777537 17620871777591 1765388 1771241 1771242 1772660 1773363 1773742 1774066 1774273 1774856 1774857 1775427 17556231777592 17680901777597 1769868 1772113 1773073 1773405 1774063 1774469 1774658 1767393 1769864 1770395 1771030 1771118 1772151 1774198 1774271 1774371 1774467 1774659 1775280 1775282 1772577 11460621775794 1769440 1769663 1771035 1771248 1771642 1774065 1774276 1774767 1774847 1775631 1775674 1775826 1775827 1769262 1769438 1775373 03111887421778758 04113321771778757 07113942331778761 13991831778763 14002981778755 14016251778752 1768145 1768673 1771033 1771546 1772564 1774468 1775630 1775825 1775829 1776493 1776496 1776667 1776829 1771967 1774449 1776402 1776882 1776917 1777061 1777300 1777390 15362661779457 1761525 1770580 1771040 1772799 1775094 1775097 1776185 1776271 1776276 1776492 1776658 1776720 1776949 1777058 1777059 1777855 1774272 1771593 1775824 1776401 1776828 1777068 1777301 1777500 1777502 1778096 1778345 15002161775307 1778097 1773723 1776272 1777072 1777302 1777303 1777904 1777932 1778095 1778195 1778252 12011846281779285 1774190 1774660 1775828 1776013 1776831 1776881 1776883 1777306 1777501 1777656 1777657 1777714 1777715 1777902 1777931 1778344 1778535 1778544 1778722 1779074 1771038 1772968 1775281 1775301 1776666 1776719 1777520 1777936 1778112 1778333 1778536 1779026 04113343381780927 11012559041779431 12011846241779272 1774593 1778725 1778814 1779726 1779800 1779843 1778175 1778537 1778724 1779189 1779222 1779575 1777713 1778889 1779001 1779367 1779369 1779574 1779975 1778726 1779197 1779375 1780152 1780508 1780607 17721131783164 17724141783176 17725641783165 17728521783166 1763610 1776014 1777504 1779477 1779503 1779806 1779945 1779958 1780605 1780632 1781096 12012602181779429 1771244 1771247 1779025 1779368 1779891 1779980 1780660 1781718 1781766 1777903 1778891 1780258 1781362 1779931 1780606 03113192221784094 12011846011784101 1771039 1775098 1775285 1778890 1779974 1780509 1780648 1781371 1782030 1782032 1782678 1782687 1783105 16183251784425 16952471784423 0498305411784104 08011913521784105 08012020671784093 09012077821784102 10012211581784106 10012371731784109 11012559021784103 11608881784107 11725511784110 1776497 1777307 1778723 1781214 1781370 1781437 1782035 1782683 1782688 1776230 1778542 1779679 1780507 1780650 15346301786718 1777958 1779799 1781097 1781215 1781372 1782682 1782876 1783299 1783475 1783476 1783962 1784126 1784306 1784307 1784308 1784536 1781797 1778063 1782679 1782680 1782681 1782874 1783152 1783534 1783581 1783663 1783748 1783972 1784827 1784843 1785299 1783302 1783749 1783963 1784174 1780649 1782034 1782872 1784160 1784537 1785232 1779878 1783533 1783753 1785159 R11976081788316 U08814851788319 U11852411788305 U121355681788320 V092442911788315 V092442911788318 W09896821788303 W09919681788304 W123831761788302 17673931787461 17703971787462 1762532 1767725 1779002 1779979 1782033 1782036 1782686 1783518 1784175 1784746 1786157 1780142 1780661 1782031 1782685 1782873 1783151 1783300 1784128 1785978 1786030 1786159 1786970 1787089 15340291789141 1784309 1784314 1784539 1785009 1785533 1786097 1786394 1786519 17674841787941 17677291787940 17720731787939 17744681787945 1785536 1785678 1785108 1786939 1787088 1787690 W023143311787969 W093590691787968 1785614 1787139 1787198 1787230 1787285 1787286 1787602 1784127 1785233 1787812 1787284 1787298 1787603 1787928 1787929 1787971 1785980 1788517 15158491777150 15282921783900 15563231775313 1784312 1785612 1785722 1785846 1785979 1786520 1788140 1788287 1788745 1788746 1770848 1779302 1781361 1784918 1786158 1786965 1787528 1787604 1787838 1787927 1788020 1788192 1788288 1788634 1788635 1788636 1788930 1789250 1789251 1789518 1789951 1784745 1785107 1785535 1785737 1786382 1786940 1787814 1787930 1788141 1788396 1788931 1789165 1789252 1789475 1789646 1789647 1789648 15232581773603 15382751773599 15466731773605 15466731773608 1759368 1772072 1780257 1781765 1784037 1784535 1787811 1787926 1788019 1788139 1788749 1789136 1789644 1789754 1790188 1786383 1790265 1774697 1776525 1779119 1789769 1789952 1790441 1790520 1790556 1790650 1790981 1791142 1775095 1778347 1778543 1779196 1779584 1779585 1781349 1789649 1790311 1790799 1790805 15532481773610 1789645 1774846 1776830 1779807 1779947 1782315 1782316 1782684 1782875 1785534 1789476 1790238 1763463 1783890 1784313 1785300 1788940 1788941 1790829 1791187 1783301 1790187 01111846131791653 01111846191791669 01111846251793255 01112739671791658 01112778821791668 01112778841793254 01112819161793251 01112831231793243 02111887321793248 02112994421791663 02113041301791670 03111887391793257 03113084321793239 03113109101793241 03113132541793256 03113159791793246 03113159801793240 03113171971793247 03113193761793236 03113264581793258 04113382631793245 04113398241791672 08011913531793253 08011952601793238 09012049591793235 10012317681793249 11012484491791662 11477631793250 12012689131793252 16593181793234 17599701793242 1784747 1787634 1788286 1789213 1789650 1790980 1791065 1791502 1791584 1783754 1789466 1790558 1790808 1790809 1791056 1791343 1791583 1791585 1791742 1791895 1791899 1792121 1792869 1792870 1790640 01112792371793244 1780141 1786938 1789257 04113450471791671 1781381 1783535 1785231 1787137 1790312 1790442 1791162 1791345 1791741 1791894 1792428 1792429 1792564 1792999 1793345 1793800 1793801 1793839 1794034 1790967 1791145 1791900 1793836 1793838 1793840 1794264 1794265 1794266 1794268 1794935 1791587 1792123 1793207 1793908 1794039 1794467 1794716 1794720 1795155 1795406 17634631797745 1792192 1793550 1793837 1794717 1778346 1786702 1791896 1791999 1792295 1794466 1794468 1794471 1794926 1794927 1795249 1795879 1788844 1789953 1792296 1792868 1794267 1794465 1795467 1795644 1796073 1796074 1796312 1791161 1793206 1793551 1795578 1795825 1795975 17834751798890 1784538 1786518 1786717 1791588 1792330 1794472 03113159801798904 1783755 1785613 1786611 1788289 1789961 1792041 1792432 1794033 1795576 1793346 1791740 1790498 1790639 1790802 1791739 1792122 1793262 1793344 1797076 1791346 1791898 01112787571798918 02111887321799453 02111887381799485 02112999051799575 02113022021798924 02113033801798913 02113070181798916 02113091331798914 02113091341799495 02113100641798915 03113142371798903 03113193571799304 03113210171798921 03113275531799346 04113247871798923 04113325541798919 04113366901799448 04113390301799273 04113394621798920 04113398241799272 05113404711798912 05113433571799470 05113454841799458 06113645521798911 12012748281799576 14271861798917 17728581800250 17783441800252 1783973 17848431800251 1787287 17876041799221 1792430 1795405 1796314 1787635 1787813 1788142 1788285 1788748 1789214 1790649 1791586 1797476 03113164361798900 04113390301799265 04113450471799046 1787138 1789519 1790807 1796076 03113084321798910 03113136191798905 03113234661798902 04113306371800583 1790519 1790806 1792298 1799645 1795577 01111846211800587 03111887421800579 03113172021800580 04113375681800585 05113531111800588 05113533891800582 05113571011800581 06113604041800586 07113838881800578 1790189 1790518 1791163 1792878 1792997 1799646 17752851802697 1788932 1790618 1789258 17899511802701 1790724 1792431 1792871 1795880 03113210191800589 04113375651798922 07011803961800584 1793263 1793956 1794936 1795877 1795925 1796320 1796855 15985761796348 16928861796334 17194521796363 17317991796382 17356501796333 17552411796335 11012341471784095 1762237 1796850 08011818551784108 01112819111784097 04113253301784098 17770611806245 17865181806243 1789520 1792297 1792998 1794269 1797072 1802766 1803350 1803545 1803848 1804528 17793691806244 17833021806605 17887451806242 1788843 1790443 17915831806608 1792867 17928701806606 1796075 1796851 1802765 1802767 1803317 1803348 1803349 1803691 1803851 1804015 1804016 17865191806603 13971331805658 01112792371791667 03111887331805665 03113142381805666 04113290741805656 05113450481805667 08011913541805657 13986211805662 13999951805664 14003351805663 02113022021806691 03113190061806692 04113253301806875 07113826921806685 07113931961806686 11462001788524 1805597 1805596 0485942391804717 15415721804719 15575421804715 02113033811805694 03113171961805689 04113325541805691 1805574 05113433561807945 05113590161807942 06113625221807930 06113742691807943 07113789371807928 07113862301806869 07113893171806867 07113911091807933 13937301807944 13986011807932 1787090 1788747 1789129 1791897 1793549 1796507 1796853 1805134 01010245661805679 03111887311807937 03113109111807947 03113159801807938 04113380731807939 07113815431806874 1808351 01112914351805687 02111887301807940 02113022741807946 02113111321807948 03113109121806877 03113164361805692 03113234591806879 04113190051807941 04113360391806880 04113366881807935 04113375641806878 05113489251805684 05113498711805682 06113602461807929 07113789351807931 07113795141806690 07113878311806870 07113899051807934 12012658521791660 14007141805677 14009451806883 14147461806689 1796852 1807949 01112838651806881 03113225321807936 04113331661806866 05113521421805681 05113531111805686 1804180 1796392 1797074 1807950 1807996 1808352 1808674 1809813 1810038 17813721811214 17903121811215 17920411811213 1805091 1805092 1807995 1810998 1810456 1810839 1810037 1811180 1811181 1811486 1811842 1812257",
      "firm_fei_number": "3010536692",
      "recalling_firm": "MicroPort Orthopedics Inc.",
      "address_1": "5677 Airline Rd",
      "city": "Arlington",
      "state": "TN",
      "postal_code": "38002-9501",
      "additional_info_contact": "Rachael Wise\n901-867-4771",
      "reason_for_recall": "MicroPort Orthopedics Inc. is voluntarily recalling any existing inventory of PROFEMUR Titanium and Cobalt Chrome modular necks with previous package insert versions to replace the PROFEMUR Hip System Package Insert with the most recent revision (150803-8).",
      "root_cause_description": "Labeling Change Control",
      "action": "The firm, MicroPort Orthopedics Inc., disseminated a \"URGENT ACTION REQUIRED\" VOLUNTARY DEVICE PRODUCT RECALL  notice dated 7/31/2020 by mail to its consignees. The letter  directed the consignee to doe the following: \n\nPlease check your stock, including all consignment stock to immediately determine if you have any affected product.\n\nPLEASE NOTE: No other MicroPort Orthopedics Inc. products are involved in this recall.\n\nIf product(s) is(are) located, please return immediately to the MicroPort Orthopedics Inc. Distribution Center at: 11481 Gulf Stream, Arlington, TN 38002. Mark all return shipping boxes with RECALL on several sides for better identification and processing.\n\nShould you have a question regarding returns, please contact MPOs Customer Experience Department available Monday through Friday 7:30 am to 7:30 pm CST, at 1-866-872-0211.\n\nPlease complete and return the enclosed form immediately even if you do not have any affected product to return. This information is essential in order to carry out effectiveness checks required per FDA guidelines. Return the completed form by fax to: 901-867-4771, or by e-mail to: PostMarket@ortho.microport.com. \n\nPlease feel free to 901-867-4771 should you have any questions.",
      "product_quantity": "12,348 units",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) to states of: AL, AZ, CA, CO, FL, GA,  IA, ID, IL, IN, KS, MA, MD, MI, MO, MT, NC, NH, NJ, NV, OH, PA, TN, TX, UT, VA,  WA, WI including PR; and countries of: Brazil, Canada, China, Costa Rica, France, Germany, Japan, The Netherlands and United Kingdom.",
      "openfda": {
        "k_number": [
          "K090764",
          "K002941",
          "K881983",
          "K961378",
          "K062383",
          "K900262",
          "K883921",
          "K954854",
          "K962034",
          "K960908",
          "K960633",
          "K905186",
          "K024134",
          "K953268",
          "K952297",
          "K963155",
          "K923911",
          "K030122",
          "K994146",
          "K895702",
          "K070061",
          "K032709",
          "K894051",
          "K901594",
          "K883794",
          "K062978",
          "K952776",
          "K833970",
          "K914343",
          "K994126",
          "K933359",
          "K030971",
          "K955553",
          "K001745",
          "K002575",
          "K010243",
          "K033313",
          "K040685",
          "K921382",
          "K973077",
          "K003016",
          "K022966",
          "K003649",
          "K001942",
          "K032110",
          "K002796",
          "K893584",
          "K982936",
          "K883661",
          "K910010",
          "K993601",
          "K891262",
          "K061699",
          "K972931",
          "K890331",
          "K934909",
          "K950541",
          "K042344",
          "K962268",
          "K002141",
          "K971900",
          "K891539",
          "K082146",
          "K873797",
          "K051826",
          "K931423",
          "K881288",
          "K900435",
          "K964178",
          "K001471",
          "K905254",
          "K913708",
          "K103518",
          "K023692",
          "K893785",
          "K903563",
          "K120595",
          "K050262",
          "K870088",
          "K910664",
          "K901687"
        ],
        "registration_number": [
          "3010065126",
          "3007120646",
          "3009241418",
          "3009546990",
          "3010206549",
          "1057425",
          "3015542154",
          "3013176080",
          "9681642",
          "1000577893",
          "3036756245",
          "3004142400",
          "2134470",
          "9616680",
          "1827096",
          "1226544",
          "9613369",
          "2245304",
          "3010363503",
          "1834331",
          "3007353762",
          "3043543260",
          "1220477",
          "3008544874",
          "3012275107",
          "3011461101",
          "1221053",
          "3006395932",
          "3026771806",
          "2528981",
          "2648920",
          "9616671",
          "3012392319",
          "3010536692",
          "3004513872",
          "3001278222",
          "1020279",
          "3010039400",
          "3006809628",
          "3007913880",
          "3006017180",
          "3005718816",
          "3010057495",
          "1000547811",
          "3010288357",
          "2134285",
          "3009888740",
          "3004748528",
          "3011683674",
          "3010287737",
          "3009882462",
          "3009959868",
          "3008102042",
          "3009973336",
          "1645151",
          "3004371426",
          "3002907620",
          "3008556682",
          "2183744",
          "3009732568",
          "3010468876",
          "3002807314",
          "1822565",
          "3003387384",
          "1720929",
          "3005551626",
          "3003477135",
          "1219655",
          "3010178296",
          "3014302784",
          "3006801265",
          "3013302242",
          "3035366890",
          "3005562917",
          "3003761012",
          "1424263",
          "1450662",
          "1218882",
          "3033509898",
          "3010123256",
          "3038503932",
          "1064858",
          "3005061536",
          "3013014058",
          "3010173425",
          "3015207155",
          "9613350",
          "3002806470",
          "1722511",
          "3002806535",
          "1221763",
          "1644408",
          "3025603301",
          "3008021110",
          "1526534",
          "3003541440",
          "1531174",
          "1643264",
          "1818910",
          "8044172",
          "3010120135",
          "3020155054",
          "1529009",
          "1833986",
          "3009760038",
          "1835831",
          "3012916784",
          "3004153896",
          "3013055499",
          "9610612",
          "1038671",
          "3008868758",
          "3002806603",
          "1836357",
          "3012329926",
          "3002807090",
          "3007923096",
          "3005323511",
          "1043653",
          "2249615",
          "3010097171",
          "3004635447",
          "3002807315",
          "3008812251",
          "2246552",
          "9615674",
          "3003595343",
          "3002807310",
          "3021008900",
          "3015516266",
          "9616944",
          "3010314800",
          "3009973505",
          "3011295718",
          "3006001176",
          "1226188",
          "2249697",
          "3009871135",
          "8043792",
          "9614209",
          "1649518",
          "2133928",
          "9610726",
          "3006128100",
          "2242737",
          "3000236920",
          "1833506",
          "3007740680",
          "1721676",
          "1225838",
          "3023852420",
          "3012725451",
          "1828464",
          "1825034",
          "3030412764",
          "1828288",
          "8010379",
          "3006356043",
          "3010009693",
          "3002807295",
          "3000196378",
          "3007366790",
          "3016438694"
        ],
        "fei_number": [
          "3010065126",
          "1000395711",
          "3007120646",
          "3009241418",
          "3002806653",
          "3002808293",
          "3009546990",
          "3002806725",
          "3010206549",
          "3015542154",
          "3013176080",
          "1000577893",
          "3036756245",
          "1000221916",
          "3004142400",
          "1827096",
          "3003414678",
          "3003064967",
          "2245304",
          "3010363503",
          "1000306340",
          "3001814066",
          "3007353762",
          "3043543260",
          "3002459190",
          "1220477",
          "3008544874",
          "3012275107",
          "3011461101",
          "1221053",
          "3006395932",
          "3026771806",
          "2528981",
          "1000139987",
          "3002807479",
          "3002807172",
          "3012392319",
          "3010536692",
          "3005233524",
          "3004513872",
          "3001278222",
          "1020279",
          "3010039400",
          "3006809628",
          "3002807036",
          "1000116912",
          "3007913880",
          "3006017180",
          "3002993297",
          "3005718816",
          "1000118610",
          "3010057495",
          "1000547811",
          "3010288357",
          "3009888740",
          "3004748528",
          "3011683674",
          "3010287737",
          "3009882462",
          "3009959868",
          "3009973336",
          "3003015552",
          "3002907620",
          "3008556682",
          "2183744",
          "3009732568",
          "3002807314",
          "3003387384",
          "1720929",
          "3005551626",
          "3003477135",
          "1219655",
          "1000220733",
          "3014302784",
          "3006801265",
          "3013302242",
          "3001644153",
          "3035366890",
          "3003386935",
          "3005562917",
          "3003761012",
          "1450662",
          "1218882",
          "3033509898",
          "3010123256",
          "3038503932",
          "3006003617",
          "3005061536",
          "3001049282",
          "3013014058",
          "3015207155",
          "3002806470",
          "3008342029",
          "3002806535",
          "1221763",
          "3025603301",
          "3008021110",
          "1526534",
          "3003541440",
          "1643264",
          "1818910",
          "3010120135",
          "3002462840",
          "3020155054",
          "3002854251",
          "3002924544",
          "1000086912",
          "3001548076",
          "3009760038",
          "3012916784",
          "3004153896",
          "3003128516",
          "3013055499",
          "3001236277",
          "1038671",
          "3008868758",
          "3002806603",
          "3012329926",
          "3000717902",
          "3007923096",
          "3005323511",
          "1043653",
          "1000199976",
          "3002807893",
          "1000143502",
          "3010097171",
          "3004635447",
          "3002807315",
          "3008397795",
          "2246552",
          "3003070421",
          "3003595343",
          "3002807310",
          "3021008900",
          "3015516266",
          "3007282753",
          "3010314800",
          "3009973505",
          "3011295718",
          "1832656",
          "3006001176",
          "3009871135",
          "1000132037",
          "3006128100",
          "3002808269",
          "3002806315",
          "3007740680",
          "1721676",
          "1828464",
          "3023852420",
          "3012725451",
          "1825034",
          "3030412764",
          "3002287318",
          "1828288",
          "3003110073",
          "3006356043",
          "3010009693",
          "3002807295",
          "3005601732",
          "1000149660",
          "3000196378",
          "3016438694"
        ],
        "device_name": "Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented",
        "medical_specialty_description": "Orthopedic",
        "regulation_number": "888.3360",
        "device_class": "2"
      }
    }
  ]
}