{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-05",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "180473",
      "product_res_number": "Z-2135-2020",
      "event_date_initiated": "2019-07-29",
      "event_date_created": "2020-05-27",
      "recall_status": "Open, Classified",
      "res_event_number": "85301",
      "product_code": "BXB",
      "product_description": "Body Weight Support (BWS) Light; Model No. RB011; a component of C-Mill, Grail, M-Gait - Product Usage: is designed to facilitate functional gait training of subjects with gait and/or balance impairment by providing dynamic partial vertical unloading during walking on a treadmill. The device will be used as an assistive device for gait deficiency rehabilitation and research purposes.",
      "code_info": "AS9924-00-0002, RB011-00-0005,RB011-00-0006, RB011-00-0010, RB011-00-0012, RB011-00-0015, RB011-00-0017, RB011-00-0018, RB011-00-0019, RB011-00-0020, RB011-00-0021, RB011-0022, RB011-0023, RB011-0024, RB011-0025, RB011-0026, RB011-0027, RB011-0028, RB011-0029, RB011-0030, RB011-0031,  RB011-0032, RB011-0033, RB011-0034, RB011-0035, RB011-0036, RB011-0037, RB011-0038, RB011-0039, RB011-0040, RB011-0041, RB011-0042, RB011-0044, RB011-0045, RB011-0046, RB011-0047, RB011-0048, RB011-0049, RB011-0050, RB011-0051, RB011-0052, RB011-0053, RB011-0054, RB011-0055, RB011-0056, RB011-0057, RB011-0058, RB011-0059, RB011-0060, RB011-0061, RB011-0062, RB011-0063, RB011-0064, RB011-0065, RB011-0066, RB011-0067, RB011-0068, RB011-0069, RB011-0070, RB011-0071, RB011-0072,  RB011-0073, RB011-0074, RB011-0075, RB011-0076, RB011-0077, RB011-0078, RB011-0079, RB011-0080, RB011-0081, RB011-0082, RB011-0083, RB011-0084, RB011-0085, RB011-0086, RB011-0087, RB011-0088, RB011-0089, RB011-0090, RB011-0091, and RB011-0092.",
      "firm_fei_number": "3006750942",
      "recalling_firm": "MOTEK MEDICAL B.V.",
      "address_1": "Hogehilweg 18 unit c 18 hogehilweg",
      "city": "Amsterdam",
      "country": "Netherlands",
      "additional_info_contact": "Kevin Walls\n720-9625412",
      "reason_for_recall": "The firm discovered a potential manufacturing deviation of the yoke of the Body Weight Support (BWS) Light that in rare occasions may result in the yoke detaching from the cable guide.",
      "root_cause_description": "Process control",
      "action": "On 07/29/2019 the firm notified their customers of the Field Safety Notice: BWS-L Yoke 2019  letter sent via email.  The firm informed their customers to share the notice with all those within their organizations who need to be aware.  Customers were asked to confirm receipt of the Field Safety Notice.  Additionally, customers were asked to reply by email to support@motekforcelink.com and attach a completed and signed copy of the Advisory Notice Confirmation (last page of the notice) within 10 business days of receipt of the notice.  \nThe firm provided the following information corrective and/or prevention action:\n\"In all cases, if you are treating patients with cranial deficiencies, stop using the system for these patients immediately\".\nThree possible options were provided. Customers were asked to select one that  is most appropriate for their institute and send them this information via the Advisory Notice Confirmation:\nA. You request a Yoke Replacement Kit to be sent to you directly, such that a skilled person from your institute can do the replacement under the firm's remote supervision. The replacement consists of removing two easily accessible bolts, replacing the affected components, and reinserting two new bolts. Verification will require  to take two pictures of the reinserted bolts, and confirmation by the firm is needed before you can use the system again for patients with cranial deficiencies. Detailed instructions will be provided. \nB. Or, you do not treat patients with cranial deficiencies, and  do not want to order a Yoke Replacement Kit. If so, you can continue using the system as normal and can consider this Field Safety Notice to be not applicable for your usage (even though we still need you to send in the Advisory Notice Confirmation). In the next service visit, we will replace the yoke to further minimize any risk.\nC. Or, you do treat patients with cranial deficiencies, but you do not want to order a Yoke Replacement Kit. In this case, you",
      "product_quantity": "80 units",
      "distribution_pattern": "Worldwide distribution - US Nationwide distributions including in the states of AZ, CA, FL, KY, MA, MI, NJ, NY, NC, and OH. The countries of Argentina, Austria, Belgium, China, Curacao, Czech Republic,  Denmark, Ecuador, Egypt,  Finland, France, Germany, Japan, Mexico, Netherlands, Russia, South Korea, Spain,  Switzerland, and United Arab Emirates.",
      "openfda": {
        "k_number": [
          "K852065",
          "K903851",
          "K924513",
          "K830793",
          "K810238",
          "K950682",
          "K850030",
          "K921323",
          "K944819",
          "K920860",
          "K852977",
          "K831910",
          "K900466",
          "K923857",
          "K871012",
          "K882916",
          "K945690",
          "K834203",
          "K880865",
          "K885182",
          "K842431",
          "K890607",
          "K830316",
          "K924484",
          "K820120",
          "K935587",
          "K935188",
          "K914394",
          "K881314",
          "K851331",
          "K933059",
          "K850332",
          "K950755",
          "K842426",
          "K905718",
          "K926313",
          "K872815",
          "K902452",
          "K931723",
          "K913227",
          "K920380",
          "K924654",
          "K862093",
          "K923217",
          "K921348",
          "K930780",
          "K920270",
          "K931736",
          "K851177",
          "K905299",
          "K935520",
          "K854354",
          "K933311",
          "K850274",
          "K910150",
          "K874884",
          "K912086",
          "K935509",
          "K925977",
          "K830649",
          "K893924",
          "K945121",
          "K900795",
          "K925270",
          "K890200",
          "K874196",
          "K935467",
          "K910149",
          "K953456",
          "K844779",
          "K841413",
          "K842432",
          "K842882",
          "K935011",
          "K910026",
          "K842433",
          "K820083",
          "K072887",
          "K945221",
          "K822031",
          "K945733",
          "K800007",
          "K830122",
          "K903467",
          "K842430",
          "K872586",
          "K003272",
          "K852399",
          "K844995",
          "K842428",
          "K850871",
          "K923494",
          "K925666",
          "K934175",
          "K821016",
          "K862314",
          "K891345",
          "K823495",
          "K895720",
          "K894907",
          "K873658",
          "K864427",
          "K922493",
          "K926136",
          "K870402",
          "K925807",
          "K844118",
          "K925043",
          "K926262",
          "K952975",
          "K920686",
          "K951029",
          "K931555",
          "K933394",
          "K925271",
          "K912537",
          "K890838",
          "K871981",
          "K951028",
          "K880092",
          "K882406",
          "K945824",
          "K872639",
          "K823539",
          "K931128",
          "K940455",
          "K820476",
          "K842425",
          "K842881",
          "K935216",
          "K872638",
          "K822212",
          "K904963",
          "K923805",
          "K830369",
          "K862730",
          "K923216",
          "K904735",
          "K895015",
          "K902707",
          "K935808",
          "K870510",
          "K842429",
          "K955106",
          "K841674",
          "K843493",
          "K925926",
          "K910151",
          "K952369",
          "K842427",
          "K955105",
          "K946169",
          "K942618",
          "K920519",
          "K843309",
          "K885095",
          "K863639",
          "K935061",
          "K914860",
          "K922185"
        ],
        "registration_number": [
          "9680413",
          "3011171243",
          "3011539161",
          "3005404127",
          "3008960494",
          "3044777120",
          "3015131233",
          "3006116732",
          "3006085220",
          "9616789",
          "3030379943",
          "2000019555",
          "3034728798",
          "9612147",
          "3006400320",
          "3010452421",
          "3007797030",
          "3008887767",
          "2311923",
          "3010675856",
          "8020836",
          "9615023",
          "1124832",
          "3016678950",
          "3007615665",
          "1043214",
          "3012841710",
          "3014660626",
          "3003454229",
          "3010050274",
          "3013846071",
          "3013730933",
          "3032587540",
          "2029188",
          "3010885565",
          "3014527921",
          "8021957",
          "3030643364",
          "3017899143",
          "1530917",
          "2510425",
          "3000286776",
          "3011806255",
          "3013697488",
          "3009684923",
          "3034993050",
          "3008264111",
          "3006750942",
          "1723702",
          "3005180774",
          "3010898730",
          "3024538051",
          "3013201082",
          "2431314",
          "3012651177",
          "3017264244",
          "3027669735",
          "3017488401",
          "1223754",
          "3009767053",
          "3004751498",
          "9681904",
          "3007740591",
          "3004905643",
          "3042285056",
          "3030220267",
          "1119903",
          "3038285234",
          "3033511104",
          "3010595009",
          "3008309560",
          "3009411363",
          "3015142701",
          "3008381107",
          "3002761514",
          "3014209670",
          "3015537297",
          "3008351005",
          "3013517454",
          "3012187972",
          "3008812560",
          "3021233816",
          "3015280549",
          "3013551163",
          "3004733458",
          "3018993530",
          "3013515165",
          "3017555752",
          "1911273",
          "3014739961",
          "3013383022",
          "3010483249",
          "3016099060",
          "3009649442",
          "3007084636",
          "3012316249",
          "3045857302",
          "3012251300",
          "3036471706",
          "3036496458",
          "3015005470",
          "3010128811",
          "3003238222",
          "3010370470",
          "2132111",
          "3014246565",
          "3015495690",
          "3007080244",
          "3004096515",
          "3013497538",
          "3004116008",
          "3006153348",
          "3017609569",
          "3002104323",
          "3021373887",
          "3014830562"
        ],
        "fei_number": [
          "3005676952",
          "3011171243",
          "1000220171",
          "3011539161",
          "3005404127",
          "3008960494",
          "3044777120",
          "3015131233",
          "3006116732",
          "3006085220",
          "1000140026",
          "3030379943",
          "2000019555",
          "3034728798",
          "3006400320",
          "3010452421",
          "3007797030",
          "3008887767",
          "2311923",
          "3010675856",
          "3006734242",
          "3016678950",
          "3007615665",
          "1043214",
          "3012841710",
          "3014660626",
          "3003454229",
          "3010050274",
          "3013846071",
          "3013730933",
          "3032587540",
          "3010885565",
          "3014527921",
          "3000215572",
          "3030643364",
          "3017899143",
          "3000204844",
          "2510425",
          "3000286776",
          "3768",
          "3011806255",
          "3002806913",
          "3013697488",
          "3004704224",
          "3009684923",
          "3034993050",
          "3008264111",
          "3006750942",
          "3005180774",
          "3010898730",
          "3024538051",
          "3013201082",
          "2431314",
          "3012651177",
          "3002520594",
          "3017264244",
          "3027669735",
          "3017488401",
          "3009767053",
          "3004751498",
          "3007740591",
          "3004905643",
          "3042285056",
          "3004531930",
          "3030220267",
          "1119903",
          "3038285234",
          "3033511104",
          "3010595009",
          "3008309560",
          "3009411363",
          "3004373978",
          "3015142701",
          "3008381107",
          "3002761514",
          "3014209670",
          "3015537297",
          "1000119345",
          "3008351005",
          "3013517454",
          "3012187972",
          "3008812560",
          "1000220847",
          "3021233816",
          "3015280549",
          "3013551163",
          "3004733458",
          "3018993530",
          "3013515165",
          "3017555752",
          "3014739961",
          "3013383022",
          "3010483249",
          "3016099060",
          "3009649442",
          "3007084636",
          "3012316249",
          "3045857302",
          "3012251300",
          "3036471706",
          "3036496458",
          "3015005470",
          "3010128811",
          "3003238222",
          "3010370470",
          "3014246565",
          "3015495690",
          "3007080244",
          "3004096515",
          "3013497538",
          "3004116008",
          "3006153348",
          "3017609569",
          "3002104323",
          "3021373887",
          "3014830562"
        ],
        "device_name": "Exerciser, Powered",
        "medical_specialty_description": "Physical Medicine",
        "regulation_number": "890.5380",
        "device_class": "1"
      }
    }
  ]
}