      "cfres_id": "178655",
      "product_res_number": "Z-1359-2020",
      "event_date_initiated": "2020-02-04",
      "event_date_posted": "2020-03-05",
      "recall_status": "Open, Classified",
      "res_event_number": "84579",
      "product_code": "FRN",
      "k_numbers": [
        "K133532"
      ],
      "product_description": "CareFusion Alaris System, Alaris PC unit models 8000 and 8015 (UDI: 10885403801518)",
      "code_info": "Model 8000, software versions 9.5 and prior; Model 8015, software versions 12.1.0, and 9.33 and prior",
      "firm_fei_number": "2016493",
      "recalling_firm": "CareFusion 303, Inc.",
      "address_1": "10020 Pacific Mesa Blvd",
      "city": "San Diego",
      "state": "CA",
      "postal_code": "92121-4386",
      "reason_for_recall": "1)Error 255-XX-XXX results in inability to edit settings 2)Delay options programming may result in no KVO rate/therapy interruption 3)Software errors results in no low battery alarm/infusion stopping 4)Medium priority KVO/End of Infusion alarms may result in unrecognized infusion completion 5)Custom concentration data entry errors results in concentrations being lower/higher than medication orders",
      "root_cause_description": "Device Design",
      "action": "On 02/04/20, Urgent Medical Device Recall Notifications were mailed and posted on the firm's website: www.bd.com/alaris-system-software-recall  Two notifications were posted: 1) Software versions 9.33 and prior, and 2) software version 12.1.0. On 02/05/20. Recall letters can be found at https://www.bd.com/en-us/recall-notifications/detail-mms-20-1953\nFirm will contact all customers to initiate the scheduling process for the software update.  Before the update customers were asked to so the following: \n\n1) Software Errors related to System Error 255-XX-XXX: If the error occurs while administering a critical medication(s), continue the infusion while you expedite a replacement pump if one is readily available, or restart and reprogram the PC unit.\n\n2) KVO Rate Not Available When Using Delay Options Programming: For software versions 9.19 and prior: Set a Callback alert of \"After\" or \"Before and After.\" For all software versions: Do not use Delay Options when a KVO rate is required.\n\n3) Low Battery Alarm Failure: Whenever possible, keep the PC unit plugged into AC power.\n\n4) Keep Vein Open (KVO) / End of Infusion Alarms Priority: Check and increase the audio volume.\n\n5) Use Errors related to Custom Concentration Programming: Only use custom concentration when the medication label does not match any of the drug concentration selections on the programming screen.\n\nCustomers can call the following phone numbers for assistance: Customer Advocacy at 888-812-3266, Clinical and Pharmacy Support Center at 858-617-1316, Recall Support Center 888-562-6018, and Technical Support 888-812-3229.\n\nOn 07/29/21, Follow up Medical Device Recall Notifications were mailed to customers informing them that a new software version, v12.1.2., is available to remediate this recalls issues. Complete the acknowledgement form and contact the Recall Support Center to schedule remediation.",
      "product_quantity": "870070",
      "distribution_pattern": "Worldwide Distribution: (U.S. Nationwide) to states of: CO, SC, NY, CA, LA, TX, AK, ID, VA, DE, FL, PA, KY, KS, TN, WA, MS, IL, GA, AR, NC, NV, NE, OR, RI, WI, HI, GU, PR, MA, IA, CT, MI, AZ, ME, MT, IN, OH, WV, WY, MN, AL, NH, OK, DC, NJ, MO, SD, UT, NM, MD, VT, MP, ND, AS; and countries of: Canada, Australia, New Zealand, the Middle East, South Africa, Taiwan, Southeast Asia and the European Union.",
      "openfda": {
        "k_number": [
          "K912589",
          "K900467",
          "K123822",
          "K830177",
          "K781171",
          "K883653",
          "K813562",
          "K915407",
          "K832167",
          "K930459",
          "K083689",
          "K141102",
          "K896075",
          "K900888",
          "K072105",
          "K905610",
          "K141389",
          "K790814",
          "K875347",
          "K864138",
          "K990259",
          "K890457",
          "K921050",
          "K844792",
          "K041714",
          "K902835",
          "K950419",
          "K050433",
          "K242660",
          "K880857",
          "K953896",
          "K111275",
          "K832410",
          "K921072",
          "K980378",
          "K023491",
          "K931549",
          "K952799",
          "K820219",
          "K782020",
          "K811774",
          "K905713",
          "K830605",
          "K910758",
          "K214045",
          "K860086",
          "K925449",
          "K842479",
          "K910934",
          "K925091",
          "K914589",
          "K981776",
          "K760370",
          "K943952",
          "K821661",
          "K944351",
          "K875348",
          "K802820",
          "K800629",
          "K910844",
          "K921073",
          "K813473",
          "K831258",
          "K950766",
          "K002126",
          "K843615",
          "K023189",
          "K882047",
          "K895644",
          "K855215",
          "K883609",
          "K850669",
          "K230041",
          "K854157",
          "K024190",
          "K992672",
          "K994374",
          "K063288",
          "K052594",
          "K831897",
          "K031249",
          "K050107",
          "K013855",
          "K830389",
          "K000328",
          "K860162",
          "K831935",
          "K915646",
          "K242752",
          "K860229",
          "K130690",
          "K883993",
          "K902891",
          "K924865",
          "K830017",
          "K843311",
          "K001829",
          "K841018",
          "K851564",
          "K032642",
          "K830809",
          "K864809",
          "K051252",
          "K030739",
          "K963574",
          "K897096",
          "K991363",
          "K790315",
          "K830544",
          "K772094",
          "K863532",
          "K933506",
          "K002679",
          "K843773",
          "K140783",
          "K910919",
          "K812339",
          "K896593",
          "K990518",
          "K820073",
          "K830102",
          "K960957",
          "K894028",
          "K870524",
          "K930494",
          "K863864",
          "K864997",
          "K030576",
          "K911769",
          "K032005",
          "K861208",
          "K872774",
          "K133801",
          "K023083",
          "K052219",
          "K211218",
          "K873161",
          "K070529",
          "K842289",
          "K940147",
          "K980092",
          "K082182",
          "K973593",
          "K890786",
          "K831444",
          "K891583",
          "K060119",
          "K871728",
          "K900375",
          "K880625",
          "K210074",
          "K982020",
          "K841630",
          "K875332",
          "K930495",
          "K874649",
          "K934671",
          "K911289",
          "K870279",
          "K862295",
          "K905696",
          "K920681",
          "K924527",
          "K790332",
          "K962709",
          "K911964",
          "K875316",
          "K961263",
          "K811272",
          "K830486",
          "K003616",
          "K013776",
          "K043590",
          "K881759",
          "K983005",
          "K223607",
          "K933652",
          "K864233",
          "K852645",
          "K900217",
          "K873522",
          "K062592",
          "K844627",
          "K250357",
          "K870991",
          "K920680",
          "K190182",
          "K865058",
          "K003543",
          "K032147",
          "K802301",
          "K770730",
          "K860353",
          "K072603",
          "K042432",
          "K905783",
          "K925196",
          "K982170",
          "K932895",
          "K841827",
          "K900742",
          "K882106",
          "K926385",
          "K043600",
          "K841729",
          "K832289",
          "K820063",
          "K832008",
          "K011096",
          "K972284",
          "K801725",
          "K822761",
          "K982159",
          "K945942",
          "K090736",
          "K012383",
          "K242114",
          "K830038",
          "K912031",
          "K003966",
          "K933144",
          "K851360",
          "K173084",
          "K080954",
          "K040061",
          "K830454",
          "K112067",
          "K812035",
          "K100704",
          "K803078",
          "K896422",
          "K871806",
          "K081412",
          "K052817",
          "K020881",
          "K820860",
          "K811917",
          "K790373",
          "K895710",
          "K833983",
          "K875364",
          "K861539",
          "K843949",
          "K944750",
          "K890736",
          "K944733",
          "K803183",
          "K812898",
          "K895709",
          "K910428",
          "K061128",
          "K003029",
          "K983321",
          "K900699",
          "K823406",
          "K823204",
          "K813455",
          "K812442",
          "K042121",
          "K121613",
          "K100705",
          "K801441",
          "K062700",
          "K791528",
          "K843548",
          "K833723",
          "K850488",
          "K854073",
          "K993387",
          "K892210",
          "K923068",
          "K872850",
          "K874181",
          "K902432",
          "K792142",
          "K830656",
          "K812824",
          "K063696",
          "K861324",
          "K811933",
          "K882102",
          "K894292",
          "K933825",
          "K844718",
          "K003582",
          "K101194",
          "K071545",
          "K883796",
          "K912193",
          "K831761",
          "K012044",
          "K050474",
          "K070718",
          "K083019",
          "K010937",
          "K141789",
          "K172653",
          "K211125",
          "K051641",
          "K791293",
          "K915522",
          "K922909",
          "K915523",
          "K040899",
          "K813644",
          "K955231",
          "K873884",
          "K894037",
          "K933545",
          "K922074",
          "K050800",
          "K932847",
          "K022869",
          "K912914",
          "K990101",
          "K813403",
          "K070398",
          "K973133",
          "K030885",
          "K843616",
          "K903782",
          "K151255",
          "K832285",
          "K941595",
          "K040149",
          "K860694",
          "K905778",
          "K892906",
          "K770087",
          "K243855",
          "K991203",
          "K052052",
          "K994275",
          "K860969",
          "K980089",
          "K904925",
          "K915649",
          "K953270",
          "K844401",
          "K974778",
          "K830490",
          "K810252",
          "K942161",
          "K980738",
          "K885083",
          "K931318",
          "K830570",
          "K913895",
          "K873938",
          "K940446",
          "K872514",
          "K925112",
          "K042696",
          "K020947",
          "K810850",
          "K841481",
          "K021988",
          "K860857",
          "K915566",
          "K052973",
          "K041241",
          "K850098",
          "K853664",
          "K930781",
          "K141654",
          "K852965",
          "K842027",
          "K060479",
          "K771664",
          "K761151",
          "K013840",
          "K801704",
          "K952967",
          "K010924",
          "K890862",
          "K944669",
          "K850183",
          "K874564",
          "K994039",
          "K822571",
          "K990425",
          "K844768",
          "K920588",
          "K811238",
          "K852073",
          "K880569",
          "K923981",
          "K897013",
          "K854074",
          "K924866",
          "K033425",
          "K873776",
          "K910637",
          "K051568",
          "K031600",
          "K865051",
          "K830452",
          "K052218",
          "K992044",
          "K022575",
          "K823193",
          "K811783",
          "K030459",
          "K913720",
          "K902309",
          "K904619",
          "K843772",
          "K842421",
          "K822842",
          "K914988",
          "K991275",
          "K895426",
          "K960512",
          "K242115",
          "K042228",
          "K923601",
          "K120830",
          "K924870",
          "K903783",
          "K913719",
          "K903789",
          "K231918",
          "K905029",
          "K024056",
          "K870992",
          "K041191",
          "K872741",
          "K861207",
          "K871084",
          "K111386",
          "K823063",
          "K850729",
          "K893488",
          "K904403",
          "K842905",
          "K183311",
          "K900204",
          "K871881",
          "K923224",
          "K924868",
          "K882107",
          "K860222",
          "K873444",
          "K832470",
          "K895797",
          "K791720",
          "K830906",
          "K001828",
          "K910007",
          "K032771",
          "K844730",
          "K770849",
          "K883003",
          "K091441",
          "K895436",
          "K926191",
          "K870330",
          "K912591",
          "K210073",
          "K922670",
          "K852168",
          "K914854",
          "K994274",
          "K042405",
          "K790945",
          "K910523",
          "K153193",
          "K890282",
          "K832166",
          "K042085",
          "K073051",
          "K980751",
          "K983975",
          "K953382",
          "K870242",
          "K010566",
          "K965230",
          "K162613",
          "K941151",
          "K881079",
          "K933005",
          "K023062",
          "K972678",
          "K800570",
          "K874546",
          "K830936",
          "K161469",
          "K844915",
          "K943692",
          "K873875",
          "K841512",
          "K850997",
          "K812042",
          "K880733",
          "K791729",
          "K011975",
          "K873847",
          "K020251",
          "K981816",
          "K864808",
          "K903787",
          "K844091",
          "K903343",
          "K991437",
          "K994375",
          "K033325",
          "K192860",
          "K974332",
          "K011442",
          "K040114",
          "K853584",
          "K232793",
          "K911401",
          "K041414",
          "K811918",
          "K230083",
          "K803190",
          "K122215",
          "K862927",
          "K853134",
          "K960826",
          "K841922",
          "K913907",
          "K032789",
          "K961975",
          "K821148",
          "K897086",
          "K070125",
          "K820726",
          "K875365",
          "K944125",
          "K792050",
          "K760718",
          "K053226",
          "K851447",
          "K903784",
          "K941984",
          "K023863",
          "K912092",
          "K902092",
          "K873495",
          "K953098",
          "K133532",
          "K944190",
          "K904922",
          "K071588",
          "K920013",
          "K932544",
          "K923875",
          "K880479",
          "K870656",
          "K895867",
          "K852060",
          "K080743",
          "K882594",
          "K110535",
          "K991013",
          "K953660",
          "K910335",
          "K812499",
          "K831259",
          "K875174",
          "K861001",
          "K973917",
          "K771478",
          "K991501",
          "K915231",
          "K042980",
          "K830151",
          "K863997",
          "K911782",
          "K851351",
          "K830657",
          "K901065",
          "K832120",
          "K925311",
          "K251640",
          "K872515",
          "K913891",
          "K842568",
          "K903786",
          "K923829",
          "K853393",
          "K812021",
          "K910350",
          "K894320",
          "K830423",
          "K811739",
          "K252015",
          "K780086",
          "K912738",
          "K802901",
          "K904924",
          "K051045",
          "K021350",
          "K841017",
          "K931375",
          "K953731",
          "K222048",
          "K873746",
          "K822989",
          "K123049",
          "K896544",
          "K910024",
          "K801546",
          "K903571",
          "K890783",
          "K041738",
          "K913661",
          "K853881",
          "K812823",
          "K870444",
          "K914855",
          "K913718",
          "K000842",
          "K912875",
          "K904923",
          "K902613",
          "K905611",
          "K902374",
          "K870801",
          "K820764",
          "K880204",
          "K110901",
          "K002211",
          "K832641",
          "K860218",
          "K873961",
          "K050324",
          "K901550",
          "K023264",
          "K100798",
          "K161667",
          "K211206",
          "K911103",
          "K921378",
          "K841373",
          "K832350",
          "K955180",
          "K881590",
          "K853629",
          "K844851",
          "K993524",
          "K043204",
          "K090087",
          "K896023",
          "K833306",
          "K982640",
          "K991936",
          "K864106",
          "K861598",
          "K834461",
          "K894360",
          "K833655",
          "K990447",
          "K900057",
          "K771300",
          "K810295",
          "K901978",
          "K883645",
          "K790559",
          "K872952",
          "K973609",
          "K865060",
          "K864517",
          "K850792",
          "K052171",
          "K872821",
          "K230022",
          "K900468",
          "K041152",
          "K802400",
          "K973851",
          "K913056",
          "K820038",
          "K854318",
          "K952794",
          "K060806",
          "K913931",
          "K870768",
          "K944352",
          "K091855",
          "K000821",
          "K053554",
          "K850674",
          "K031432",
          "K014091",
          "K982836",
          "K851017",
          "K143369",
          "K891237",
          "K821069",
          "K931458",
          "K922702",
          "K853814",
          "K832923",
          "K863470",
          "K832755",
          "K052217",
          "K861209",
          "K933006",
          "K914122",
          "K924872",
          "K914034",
          "K912192",
          "K023973",
          "K884505",
          "K041550",
          "K903879",
          "K931132",
          "K844719",
          "K861436",
          "K091308",
          "K863683",
          "K031185",
          "K904979",
          "K242390",
          "K812917",
          "K843878",
          "K013234",
          "K882598",
          "K200176",
          "K050301",
          "K904518",
          "K880623",
          "K061325",
          "K153599",
          "K842263",
          "K080297",
          "K920773",
          "K093332",
          "K251636",
          "K800569",
          "K062699",
          "K830355",
          "K841374",
          "K833958",
          "K040985",
          "K903788",
          "K903781",
          "K030323",
          "K220417",
          "K142596",
          "K861040",
          "K894041",
          "K861007",
          "K902409",
          "K921901",
          "K832349",
          "K864276",
          "K872516",
          "K042081",
          "K761307",
          "K052221",
          "K883587",
          "K851821",
          "K961862",
          "K924867",
          "K955258",
          "K843331",
          "K912590",
          "K831534",
          "K023827",
          "K893769",
          "K840263",
          "K831203",
          "K024084",
          "K843168",
          "K823035",
          "K821047",
          "K820946",
          "K882159",
          "K213744",
          "K241736",
          "K891289",
          "K942162",
          "K865059",
          "K001629",
          "K042609",
          "K813590",
          "K850653",
          "K854964",
          "K840616",
          "K912458",
          "K924871",
          "K253703",
          "K982838",
          "K894506",
          "K882556",
          "K924603",
          "K841784",
          "K780503",
          "K896546",
          "K782034",
          "K782019",
          "K923290",
          "K961703",
          "K124053",
          "K820482",
          "K211122",
          "K890062",
          "K860502",
          "K896717",
          "K800627",
          "K924869",
          "K914349",
          "K832606",
          "K903785",
          "K944721",
          "K864227",
          "K905251",
          "K780705",
          "K983838",
          "K910425",
          "K901755",
          "K915499",
          "K061108",
          "K904048",
          "K002380",
          "K883577",
          "K091386",
          "K954826",
          "K070268",
          "K031749",
          "K822454",
          "K942378",
          "K894321",
          "K904890",
          "K873494",
          "K872764",
          "K043466",
          "K883902",
          "K843274",
          "K890120",
          "K852358",
          "K884014",
          "K242735",
          "K172831",
          "K013767",
          "K844078",
          "K070235",
          "K801904",
          "K871807",
          "K882249",
          "K912928",
          "K091042",
          "K811884",
          "K003611",
          "K860492"
        ],
        "registration_number": [
          "2016493",
          "3002808022",
          "3009754410",
          "2032862",
          "9612030",
          "3005627859",
          "3013557562",
          "2520313",
          "3029973819",
          "1061124",
          "3003144120",
          "3004082685",
          "3000240707",
          "3008717264",
          "3012494290",
          "9710593",
          "3016798778",
          "2027804",
          "1220973",
          "3013652261",
          "3015158050",
          "1319639",
          "3008058400",
          "9680741",
          "3006575795",
          "3008810357",
          "3006158088",
          "3011883032",
          "3010564249",
          "3047799708",
          "1721686",
          "9615908",
          "1220477",
          "1055927",
          "3042027705",
          "2528981",
          "1526611",
          "9614279",
          "1416980",
          "1058584",
          "3020777588",
          "9617594",
          "2411512",
          "1319660",
          "3014732157",
          "3016509758",
          "9710049",
          "3006534321",
          "3009746061",
          "2024064",
          "1423507",
          "1219343",
          "2032098",
          "3013298431",
          "1423537",
          "3004056159",
          "3012302888",
          "9680787",
          "2244478",
          "3010273872",
          "3008513006",
          "3009211636",
          "3000204189",
          "9617787",
          "3004548776",
          "2523676",
          "2523167",
          "1018470",
          "3008692839",
          "1220908",
          "9680425",
          "3004112621",
          "2032112",
          "9617604",
          "3005778453",
          "2024022",
          "3012337868",
          "3015276780",
          "2521402",
          "9611004",
          "3007939342",
          "3010293992",
          "3024122953",
          "3005700264",
          "1314492",
          "1047843",
          "3004906872",
          "1643817",
          "1320894",
          "3013319212",
          "3010082909",
          "3009456923",
          "3015972753",
          "9681819",
          "2022994",
          "1713468",
          "3003098013",
          "3024706751",
          "3004530184",
          "3005768810",
          "9610847",
          "1314417",
          "2029015",
          "3045809869",
          "3003927385",
          "3017636737",
          "3017907746",
          "2032227",
          "3007738736",
          "3010432890",
          "1928237",
          "2650143",
          "3005053560",
          "3008570882",
          "9610825",
          "1921846",
          "1123137",
          "3005382983",
          "3010291427",
          "2030624",
          "3011270181",
          "3004491689",
          "2320968",
          "1723170",
          "1121996",
          "1722139",
          "1219702",
          "1066270",
          "9615050",
          "3010381606",
          "3004167615",
          "3018229072",
          "9612051",
          "3011688490",
          "1222313",
          "3018094310",
          "3013722099",
          "1125615",
          "3003879246",
          "3036773433",
          "3015192982",
          "3002879653",
          "1219602",
          "2029275",
          "1054713",
          "1045254",
          "3013764800",
          "2011171",
          "3008132411"
        ],
        "fei_number": [
          "2016493",
          "3003282181",
          "3002808022",
          "3009754410",
          "3005627859",
          "3013557562",
          "2520313",
          "2648209",
          "1000136478",
          "3029973819",
          "3004082685",
          "3000240707",
          "3008717264",
          "3012494290",
          "3016798778",
          "3003807466",
          "3003513491",
          "1220973",
          "3013652261",
          "3015158050",
          "3008058400",
          "3006575795",
          "1000123180",
          "3008810357",
          "1000121050",
          "3006158088",
          "3011883032",
          "3003767881",
          "3010564249",
          "3047799708",
          "1721686",
          "3003346138",
          "3002558692",
          "1220477",
          "3042027705",
          "3001236905",
          "2528981",
          "1526611",
          "3003166194",
          "1416980",
          "3002826093",
          "3020777588",
          "2411512",
          "1000512168",
          "3014732157",
          "3016509758",
          "3006534321",
          "1000125732",
          "3003318268",
          "1000306431",
          "3009746061",
          "2024064",
          "1219343",
          "3013298431",
          "3004056159",
          "3012302888",
          "3010273872",
          "3008513006",
          "3009211636",
          "3000204189",
          "3002807099",
          "1000113359",
          "3004099094",
          "3004548776",
          "2523676",
          "2523167",
          "1018470",
          "3008692839",
          "1000138054",
          "1220908",
          "3004112621",
          "3005778453",
          "2024022",
          "3012337868",
          "3015276780",
          "2521402",
          "3000957797",
          "3007939342",
          "3010293992",
          "3024122953",
          "3005700264",
          "3002930098",
          "3002806934",
          "1314492",
          "3000206435",
          "1047843",
          "3004906872",
          "1643817",
          "3008002452",
          "3002879021",
          "1000114111",
          "1000219637",
          "3013319212",
          "3010082909",
          "3009456923",
          "3015972753",
          "1713468",
          "2022994",
          "3003098013",
          "3024706751",
          "3004530184",
          "3002806469",
          "1000136269",
          "3005768810",
          "1000160458",
          "1314417",
          "3001451571",
          "3045809869",
          "3003927385",
          "3017636737",
          "3017907746",
          "3001155023",
          "3007738736",
          "3010432890",
          "1928237",
          "3005053560",
          "1921846",
          "3005382983",
          "3010291427",
          "1000517638",
          "3011270181",
          "3005601686",
          "3004491689",
          "1000117172",
          "1121996",
          "1219702",
          "3010381606",
          "3004167615",
          "3018229072",
          "3018094310",
          "3011688490",
          "3013722099",
          "3003879246",
          "3036773433",
          "3003058448",
          "3000147578",
          "3015192982",
          "3002879653",
          "1219602",
          "3002999813",
          "3002806457",
          "1045254",
          "3007596547",
          "3003327833",
          "3003076831",
          "2937446",
          "3013764800",
          "2011171",
          "3008132411"
        ],
        "device_name": "Pump, Infusion",
        "medical_specialty_description": "General Hospital",
        "regulation_number": "880.5725",
        "device_class": "2"
      }
    }
  ]
}