  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-09-03",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "138532",
      "product_res_number": "Z-1991-2015",
      "event_date_initiated": "2015-06-09",
      "event_date_posted": "2015-07-31",
      "recall_status": "Open, Classified",
      "res_event_number": "71649",
      "product_code": "RBS",
      "product_description": "Olympus Scientific Solutions Americas  Analytical X-ray System.  Delta, Delta Element, Delta Plus, DPO-4000-XX-U-EN-EN-PM-SP12, DPO-6000-XX-U-EN-EN-AP-SP10.\n\nMobile X-ray System.",
      "code_info": "Delta, Delta Element, Delta Plus, DPO-4000-XX-U-EN-EN-PM-SP12, DPO-6000-XX-U-EN-EN-AP-SP10.",
      "firm_fei_number": "3006241029",
      "recalling_firm": "Olympus Scientific Solutions Americas",
      "address_1": "48 Woerd Ave",
      "city": "Waltham",
      "state": "MA",
      "postal_code": "02453-3824",
      "reason_for_recall": "Higher than published scatter around the probe head when testing some samples. At the point the customer would interact directly with the instrument (Trigger position), most readings were lower.  However, measurements around the probe head were higher particularly on samples know to be highly scattering (plastics, soils) with minimal increases on high density samples (steel, aluminum).",
      "root_cause_description": "Component change control",
      "action": "Olympus Scientific Solutions planned action:\n\n1. Notify product managers as to potential issues \n2. Distribute new table and notice to instrument owners 30 JUL 15\n3. Post new table on website for access 15 JUN 15",
      "product_quantity": "1047",
      "distribution_pattern": "Nationwide Distribution.",
      "openfda": {
        "device_name": "Industrial X-Ray",
        "medical_specialty_description": "Unknown",
        "regulation_number": "",
        "device_class": "N"
      }
    }
  ]
}