{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-10-24",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "68298",
      "product_res_number": "Z-1315-2008",
      "event_date_initiated": "2008-02-05",
      "event_date_posted": "2008-07-16",
      "recall_status": "Terminated",
      "event_date_terminated": "2008-09-09",
      "res_event_number": "46806",
      "product_code": "BSZ",
      "k_numbers": [
        "K042419"
      ],
      "product_description": "Fabius GS Anesthesia Machine; Gas machine for anesthesia or analgesia; Catalog Number: 8604700; Drager Medical Inc, Telford, PA 18969.",
      "code_info": "Serial Numbers: ARWM0008, ARWM0009, ARWM0010, ARWM0011, ARWM0012, ARWM0013, ARWM0014, ARWM0015, ARWM0029, ARWM0030, ARWM0031, ARWM0032, ARWM0033, ARWM0034, ARWM0035, ARWM0036, ARWM0037, ARWM0050, ARWM0051, ARWM0052, ARWM0053, ARWM0057, ARWM0058, ARWM0059, ARWM0061, ARWM0062, ARWM0063, ARWM0064, ARWM0065, ARWM0066, ARWM0067, ARWM0068, ARWM0069, ARWM0070, ARWM0071, ARWM0072, ARWM0073, ARWM0074, ARWM0075, ARWM0076, ARWM0077, ARWM0078, ARWM0079, ARWM0080, ARWM0081, ARWM0082, ARWM0083, ARWM0084, ARWM0085, ARWM0092, ARWM0093, ARWM0097, ARWM0098, ARWM0099, ARWM0100, ARWM0101, ARWM0102, ARWM0103, ARWM0104, ARWM0105, ARWM0108, ARWM0112, ARWM0113, ARWM0114, ARWM0115, ARWM0116, ARWM0117, ARWM0118, ARWM0119, ARWM0120, ARWM0121, ARWM0122, ARWM0123, ARWM0124, ARWM0125, ARWM0126, ARWM0127, ARWM0128, ARWM0129, ARWM0130, ARWM0131, ARWM0132, ARWM0133, ARWM0134, ARWM0135, ARWM0136, ARWM0137, ARWM0138, ARWM0139, ARWM0140, ARWM0157, ARWM0159, ARWM0160, ARWM0161, ARWM0162, ARWM0163, ARWM0164, ARWM0165, ARWM0167, ARWM0168, ARWM0169, ARWM0170, ARWM0171, ARWM0172, ARWM0176, ARWM0177, ARWM0178, ARWM0179, ARWM0180, ARWM0181, ARWM0184, ARWM0215, ARWM0216, ARWM0217, ARWM0219, ARWM0224, ARWM0232, ARWM0284, ARWM0285, ARWM0286, ARWM0287, ARWM0288, ARWM0289, ARWM0308, ARWM0309, ARWM0310, ARWM0311, ARWM0313, ARWM0314, ARWM0315, ARWM0323, ARWM0326, ARWM0327, ARWM0328, ARWM0329, ARWM0330, ARWM0331, ARWM0333, ARWM0335, ARWM0337, ARWN0001, ARWN0002, ARWN0003,  ARWN0005, ARWN0014, ARWN0015, ARWN0016, ARWN0017, ARWN0018, ARWN0019, ARWN0022, ARWN0023, ARWN0024, ARWN0025, ARWN0026, ARWN0027, ARWN0028, ARWN0029, ARWN0030, ARWN0031, ARWN0033, ARWN0034, ARWN0035, ARWN0036, ARWN0037, ARWN0038, ARWN0039, ARWN0040, ARWN0041, ARWN0042, ARWN0043, ARWN0044, ARWN0045, ARWN0046, ARWN0058, ARWN0060, ARWN0061, ARWN0062, ARWN0063, ARWN0064, ARWN0065, ARWN0066, ARWN0067, ARWN0076, ARWN0078, ARWN0080, ARWN0081, ARWN0082, ARWN0083, ARWN0084, ARWN0085, ARWN0087, ARWN0088, ARWN0176, ARWN0184, and ARWN0214.",
      "firm_fei_number": "2517967",
      "recalling_firm": "Draeger Medical, Inc.",
      "address_1": "3135 Quarry Rd",
      "city": "Telford",
      "state": "PA",
      "postal_code": "18969-1042",
      "additional_info_contact": "Technical Support\n800-543-5047",
      "reason_for_recall": "Detachment of Component: The chassis on some machines were manufactured out of specifications.  As a result, the caster may break loose from the chassis.",
      "root_cause_description": "Nonconforming Material/Component",
      "action": "Consignees were notified by an Urgent Medical Device Recall letter sent on 2/7/08.  The letter informed users that the recalling firm will contact them to schedule appointments to inspect any potentially affected machines.  The letter instructed users to continue using the machines, before the inspection, making sure to carefully follow the warnings in the instruction manual.  If a caster on the machine appears unstable before the inspection is conducted, users were advised to contact the recalling firm.  For additional information, contact 1-800-543-5047.",
      "product_quantity": "196 UNITS",
      "distribution_pattern": "Nationwide Distribution.",
      "openfda": {
        "k_number": [
          "K112722",
          "K981845",
          "K923172",
          "K983635",
          "K171292",
          "K791530",
          "K092864",
          "K864806",
          "K000859",
          "K930492",
          "K973971",
          "K923519",
          "K061609",
          "K884879",
          "K864590",
          "K133886",
          "K022031",
          "K102182",
          "K080175",
          "K062155",
          "K771989",
          "K041622",
          "K101439",
          "K081941",
          "K071097",
          "K102471",
          "K864589",
          "K120075",
          "K253330",
          "K844849",
          "K071142",
          "K122063",
          "K901713",
          "K894309",
          "K800801",
          "K830161",
          "K850648",
          "K844563",
          "K862831",
          "K945463",
          "K251352",
          "K841274",
          "K000706",
          "K862085",
          "K151954",
          "K853139",
          "K971425",
          "K982137",
          "K833335",
          "K131945",
          "K934669",
          "K760228",
          "K133958",
          "K803052",
          "K973949",
          "K855022",
          "K920944",
          "K930351",
          "K946127",
          "K884991",
          "K112114",
          "K801100",
          "K081447",
          "K050626",
          "K050055",
          "K852175",
          "K905635",
          "K854213",
          "K061102",
          "K110213",
          "K900937",
          "K973985",
          "K771479",
          "K042419",
          "K853435",
          "K980553",
          "K142552",
          "K864447",
          "K171311",
          "K033026",
          "K073707",
          "K945234",
          "K882489",
          "K965160",
          "K830160",
          "K901117",
          "K123125",
          "K861170",
          "K050676",
          "K771057",
          "K833147",
          "K050883",
          "K963481",
          "K890410",
          "K030624",
          "K851621",
          "K090892",
          "K000995",
          "K872178",
          "K131790",
          "K040743",
          "K935546",
          "K042086",
          "K032803",
          "K884802",
          "K132903",
          "K030939",
          "K172045",
          "K972848",
          "K123211",
          "K901119",
          "K143530",
          "K122445",
          "K042607",
          "K981530",
          "K003663",
          "K151570",
          "K863207",
          "K191027",
          "K001727",
          "K230931",
          "K864644",
          "K893780",
          "K001315",
          "K900329",
          "K921742",
          "K081844",
          "K970163",
          "K801103",
          "K003251",
          "K974193",
          "K213867",
          "K844008",
          "K090233",
          "K992421",
          "K011404",
          "K083098",
          "K170872",
          "K132530",
          "K963994",
          "K933285",
          "K830159",
          "K993410",
          "K001488",
          "K160665",
          "K913791",
          "K060193",
          "K802591",
          "K965041",
          "K062388",
          "K031400",
          "K072884",
          "K841370",
          "K844421",
          "K790456",
          "K042154",
          "K875350",
          "K103035",
          "K945722",
          "K003579",
          "K033498",
          "K113633",
          "K252746",
          "K051629",
          "K903991",
          "K083413",
          "K902962",
          "K840356",
          "K102389",
          "K844012",
          "K925580",
          "K001988",
          "K201957",
          "K052335",
          "K010613",
          "K041430",
          "K822218",
          "K072420",
          "K830680",
          "K903135",
          "K012446",
          "K971030",
          "K770554",
          "K833151",
          "K801102",
          "K900328",
          "K973896",
          "K781492",
          "K140264"
        ],
        "registration_number": [
          "1054713",
          "3014579161",
          "3007603826",
          "3005783425",
          "9610531",
          "3007238566",
          "1066427",
          "3013188547",
          "9617604",
          "3021559257",
          "3049808762",
          "9710602",
          "2112667",
          "2221819",
          "3010703925",
          "3022543477",
          "2518157",
          "3003589956",
          "3035111606",
          "8043316",
          "1000122786",
          "1066270",
          "9611500",
          "2020813",
          "8010042",
          "3009156722",
          "9611295",
          "3000126629",
          "3009129579",
          "3016618143",
          "3005803793",
          "3010157426",
          "1530958"
        ],
        "fei_number": [
          "1000639866",
          "3014579161",
          "3007603826",
          "3005783425",
          "3002558692",
          "3007238566",
          "3002808128",
          "3013188547",
          "3002807907",
          "3007513035",
          "3021559257",
          "3049808762",
          "2112667",
          "2221819",
          "3010703925",
          "3022543477",
          "2518157",
          "3003513129",
          "3003589956",
          "3035111606",
          "1000122786",
          "3002807992",
          "2020813",
          "3009156722",
          "3000126629",
          "3009129579",
          "3001104093",
          "3016618143",
          "3005803793",
          "3010157426",
          "1000114111",
          "3000957797",
          "3003319165"
        ],
        "device_name": "Gas-Machine, Anesthesia",
        "medical_specialty_description": "Anesthesiology",
        "regulation_number": "868.5160",
        "device_class": "2"
      }
    }
  ]
}