      "cfres_id": "208441",
      "product_res_number": "Z-2339-2024",
      "event_date_initiated": "2024-06-13",
      "event_date_posted": "2024-07-11",
      "recall_status": "Open, Classified",
      "res_event_number": "94825",
      "product_code": "FLL",
      "k_numbers": [
        "K152748"
      ],
      "product_description": "Braun Thermoscan PRO 6000 Ear Thermometer",
      "code_info": "UDI: 00732094309003 (product code 06000-200) and 00732094309027 (product code 06000-300); All serial numbers.",
      "firm_fei_number": "1417572",
      "recalling_firm": "Baxter Healthcare Corporation",
      "address_1": "1 Baxter Pkwy",
      "city": "Deerfield",
      "state": "IL",
      "postal_code": "60015-4625",
      "additional_info_contact": "Center for One Baxter\n800-422-9837",
      "reason_for_recall": "The Braun Thermoscan PRO 6000 ear thermometers may have been shipped with a compact disc (CD) containing an outdated version of the Instructions for Use (IFU).",
      "root_cause_description": "Process control",
      "action": "A communication was sent to affected customers via U.S.P.S., first class\nmail on 06/13/2024. Baxter is asking customers to: 1. Clinicians may continue to use the PRO 6000 ear thermometers according to the current IFU (CD material Number 421032). Please discard all versions of the outdated IFU (material number 419450). The material number is printed on the CD. 2. Acknowledge the receipt of this notification by completing a reply form on the customer portal.\n\nUPDATE PER EMAIL DATED 7/24/2024:\n\nThe firm issued a follow-up letter dated 7/24/2024 via USPS first class mail.  The follow-up letter flagged Urgent Medical Device Correction provides the following additional actions to be taken by the customer as a result of the firm expanding the scope of the recall strategy and actions to be taken by the customer: (The content of the other numbered actions remains the same as the initial letter dated 6/13/2024.)\n\n2. Please share this communication with all potential users in your organization\nand instruct them to follow cleaning instructions in the IFU Maintenance and\nService section for proper cleaning. For convenience, the cleaning\ninstructions are summarized in the enclosed Cleaning Guide.\n3. Do not use the device if the ring around the measurement button shows a\ngreen blinking or flashing light instead of a ready state (solid green light).\nContact Baxter Technical Support to report this issue.\n4. Do not use the device if the device requires multiple power-ups prior to going\nto ready state (solid green light). Contact Baxter Technical Support to report\nthis issue.\n5. If you experience an overheating probe tip, do not use the device. Contact\nBaxter Technical Support to report the issue.\n\nA Cleaning Guide was enclosed with the letter as well as another response form.",
      "product_quantity": "265,647 units",
      "distribution_pattern": "Distributed to United States, Canada, Australia, Malaysia, New Zealand, Philippines,\nAustria, Belgium, France, Germany, Italy, Netherlands, Norway, Spain, Sweden, and United Kingdom.\n\nUPDATED PER EMAIL DATED 7/24/2024:\n\nUnited States (including Puerto Rico and Guam), Canada, Australia, Brunei, Hong Kong, India, Korea, Malaysia, Maldives, Myanmar, New Zealand, Philippines, Singapore, Taiwan, Thailand, Vietnam, Afghanistan, Austria, Bahrain, Belgium, Croatia, Cyprus, Czech Republic,\nDenmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Iraq, Ireland, Israel, Italy, Jordan, Kuwait, Luxemburg, Namibia, Netherlands, Norway, Oman, Poland, Portugal, Qatar, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom, Yemen, Bahamas, Bolivia, Chile, Colombia, Ecuador, Guatemala, Mexico, and Panama.",
      "openfda": {
        "k_number": [
          "K121428",
          "K873010",
          "K961879",
          "K050137",
          "K222248",
          "K912742",
          "K820409",
          "K102934",
          "K010866",
          "K780861",
          "K023538",
          "K971866",
          "K801366",
          "K203332",
          "K760118",
          "K002712",
          "K180207",
          "K063542",
          "K890823",
          "K201161",
          "K883793",
          "K112189",
          "K211352",
          "K925791",
          "K792257",
          "K101043",
          "K003326",
          "K010383",
          "K041516",
          "K200531",
          "K171455",
          "K160544",
          "K212945",
          "K821436",
          "K051873",
          "K203215",
          "K984506",
          "K842199",
          "K864089",
          "K152739",
          "K190984",
          "K203656",
          "K201980",
          "K833711",
          "K821886",
          "K102508",
          "K851075",
          "K811672",
          "K971994",
          "K161735",
          "K162509",
          "K110956",
          "K103097",
          "K102718",
          "K240220",
          "K922344",
          "K131155",
          "K983406",
          "K201977",
          "K791558",
          "K943362",
          "K012136",
          "K021851",
          "K203170",
          "K852454",
          "K033820",
          "K810173",
          "K203454",
          "K020433",
          "K832090",
          "K030658",
          "K062784",
          "K952765",
          "K844273",
          "K172870",
          "K211931",
          "K232010",
          "K180355",
          "K791209",
          "K243000",
          "K202111",
          "K160091",
          "K121999",
          "K070339",
          "K872584",
          "K820948",
          "K131210",
          "K923189",
          "K003313",
          "K874521",
          "K180714",
          "K101747",
          "K072635",
          "K193338",
          "K071019",
          "K051699",
          "K813468",
          "K181015",
          "K823493",
          "K032612",
          "K030673",
          "K091589",
          "K000969",
          "K051566",
          "K883005",
          "K210118",
          "K954434",
          "K011254",
          "K163603",
          "K172861",
          "K251160",
          "K033534",
          "K182597",
          "K964286",
          "K210564",
          "K890484",
          "K935267",
          "K830819",
          "K023069",
          "K902624",
          "K771999",
          "K051422",
          "K181215",
          "K183393",
          "K951285",
          "K100953",
          "K182629",
          "K002562",
          "K962369",
          "K840696",
          "K063418",
          "K103407",
          "K163256",
          "K871465",
          "K202420",
          "K974195",
          "K920713",
          "K003308",
          "K881909",
          "K201536",
          "K152905",
          "K061800",
          "K210937",
          "K833158",
          "K861100",
          "K955142",
          "K820670",
          "K203731",
          "K871529",
          "K050303",
          "K771997",
          "K951279",
          "K874520",
          "K842132",
          "K810474",
          "K832042",
          "K120412",
          "K873205",
          "K962520",
          "K832635",
          "K201600",
          "K152640",
          "K992468",
          "K121427",
          "K090031",
          "K011585",
          "K191668",
          "K153146",
          "K832135",
          "K033082",
          "K971045",
          "K791371",
          "K243082",
          "K180387",
          "K081445",
          "K102554",
          "K041694",
          "K233192",
          "K832392",
          "K873833",
          "K191004",
          "K821556",
          "K030219",
          "K032658",
          "K882304",
          "K771996",
          "K061007",
          "K013779",
          "K031503",
          "K812037",
          "K083419",
          "K893391",
          "K160999",
          "K132761",
          "K223044",
          "K092958",
          "K981337",
          "K110559",
          "K050837",
          "K172733",
          "K820153",
          "K203707",
          "K023510",
          "K850395",
          "K121696",
          "K913083",
          "K760720",
          "K200781",
          "K791323",
          "K844201",
          "K070878",
          "K161211",
          "K830552",
          "K910718",
          "K172508",
          "K040377",
          "K072108",
          "K162137",
          "K210924",
          "K800512",
          "K843609",
          "K833568",
          "K831089",
          "K032714",
          "K213079",
          "K041692",
          "K896952",
          "K101387",
          "K170262",
          "K231625",
          "K041056",
          "K952541",
          "K854706",
          "K981456",
          "K190242",
          "K852954",
          "K951284",
          "K181081",
          "K221211",
          "K951402",
          "K170723",
          "K013876",
          "K935603",
          "K203583",
          "K240648",
          "K951283",
          "K002727",
          "K201582",
          "K113761",
          "K232473",
          "K143267",
          "K134043",
          "K130231",
          "K200943",
          "K823066",
          "K202591",
          "K051835",
          "K090294",
          "K850244",
          "K062445",
          "K081788",
          "K853551",
          "K010502",
          "K782043",
          "K191829",
          "K100843",
          "K103680",
          "K851873",
          "K780800",
          "K822939",
          "K962386",
          "K131243",
          "K954792",
          "K160308",
          "K202687",
          "K163516",
          "K984497",
          "K222057",
          "K813333",
          "K862333",
          "K951282",
          "K172889",
          "K081162",
          "K802147",
          "K951404",
          "K031556",
          "K091040",
          "K051100",
          "K951277",
          "K190873",
          "K851379",
          "K800849",
          "K951274",
          "K933642",
          "K223807",
          "K030361",
          "K781312",
          "K802607",
          "K093795",
          "K033817",
          "K874514",
          "K100495",
          "K881415",
          "K896779",
          "K801930",
          "K864397",
          "K831718",
          "K854598",
          "K120712",
          "K813306",
          "K951939",
          "K070976",
          "K895703",
          "K180298",
          "K202481",
          "K863646",
          "K122412",
          "K841105",
          "K962991",
          "K062016",
          "K123361",
          "K961365",
          "K050230",
          "K172874",
          "K203636",
          "K082266",
          "K935280",
          "K935466",
          "K121118",
          "K101322",
          "K132140",
          "K813402",
          "K113159",
          "K103617",
          "K152975",
          "K853392",
          "K210171",
          "K132457",
          "K910118",
          "K920469",
          "K220276",
          "K092447",
          "K890202",
          "K010235",
          "K875234",
          "K771994",
          "K182437",
          "K840648",
          "K232128",
          "K821504",
          "K102187",
          "K201032",
          "K850363",
          "K180148",
          "K841322",
          "K830039",
          "K823333",
          "K221039",
          "K984551",
          "K183663",
          "K002182",
          "K854035",
          "K034023",
          "K212598",
          "K160680",
          "K970271",
          "K160802",
          "K182751",
          "K073366",
          "K180314",
          "K982140",
          "K223522",
          "K830429",
          "K880639",
          "K951281",
          "K020725",
          "K250401",
          "K951278",
          "K111463",
          "K162083",
          "K970011",
          "K912280",
          "K822435",
          "K200578",
          "K841780",
          "K160816",
          "K211744",
          "K850362",
          "K874524",
          "K180131",
          "K201879",
          "K874159",
          "K090386",
          "K952001",
          "K202368",
          "K871888",
          "K781661",
          "K051233",
          "K150160",
          "K853713",
          "K844056",
          "K884765",
          "K131771",
          "K792116",
          "K180767",
          "K031907",
          "K921190",
          "K033790",
          "K951400",
          "K111703",
          "K801423",
          "K200631",
          "K915633",
          "K180385",
          "K021601",
          "K952073",
          "K971831",
          "K033077",
          "K020504",
          "K172795",
          "K253162",
          "K874404",
          "K033069",
          "K943695",
          "K163601",
          "K032364",
          "K820024",
          "K112289",
          "K885308",
          "K882127",
          "K874523",
          "K861771",
          "K030299",
          "K951280",
          "K201843",
          "K043110",
          "K080593",
          "K182667",
          "K112116",
          "K182652",
          "K240333",
          "K990168",
          "K902386",
          "K903501",
          "K903590",
          "K181013",
          "K172894",
          "K020471",
          "K133113",
          "K772217",
          "K110776",
          "K113095",
          "K001008",
          "K201579",
          "K080759",
          "K874694",
          "K853317",
          "K860436",
          "K122520",
          "K881813",
          "K831733",
          "K042013",
          "K934482",
          "K222311",
          "K060006",
          "K203431",
          "K830353",
          "K832391",
          "K141505",
          "K230889",
          "K023500",
          "K011059",
          "K010238",
          "K211583",
          "K830561",
          "K885187",
          "K202009",
          "K801924",
          "K133111",
          "K063185",
          "K813639",
          "K854349",
          "K771998",
          "K062010",
          "K193253",
          "K003223",
          "K161933",
          "K232678",
          "K870956",
          "K122912",
          "K220084",
          "K233280",
          "K854007",
          "K900048",
          "K092959",
          "K181967",
          "K120711",
          "K822831",
          "K202741",
          "K964851",
          "K200599",
          "K964605",
          "K102947",
          "K813459",
          "K011704",
          "K023711",
          "K821555",
          "K913082",
          "K190990",
          "K160306",
          "K042653",
          "K810814",
          "K122221",
          "K021614",
          "K171214",
          "K954523",
          "K931747",
          "K021753",
          "K220657",
          "K063291",
          "K896914",
          "K850564",
          "K202296",
          "K821244",
          "K880759",
          "K111282",
          "K905025",
          "K905652",
          "K021028",
          "K952159",
          "K193213",
          "K100488",
          "K851010",
          "K983887",
          "K964643",
          "K111637",
          "K090564",
          "K761244",
          "K850196",
          "K770777",
          "K091676",
          "K240109",
          "K023412",
          "K914005",
          "K771995",
          "K930930",
          "K203643",
          "K170662",
          "K830530",
          "K983619",
          "K852087",
          "K781136",
          "K831150",
          "K854800",
          "K881368",
          "K202603",
          "K850424",
          "K101244",
          "K760421",
          "K051251",
          "K021052",
          "K791208",
          "K081061",
          "K925674",
          "K902912",
          "K210014",
          "K955846",
          "K201864",
          "K051269",
          "K031928",
          "K851146",
          "K981708",
          "K171578",
          "K812387",
          "K852903",
          "K111974",
          "K171182",
          "K222588",
          "K213100",
          "K170219",
          "K130361",
          "K212257",
          "K243136",
          "K203662",
          "K823077",
          "K103800",
          "K240223",
          "K182755",
          "K850195",
          "K021048",
          "K841260",
          "K161728",
          "K800396",
          "K120004",
          "K840675",
          "K160358",
          "K780131",
          "K853824",
          "K090395",
          "K210565",
          "K132514",
          "K042575",
          "K925792",
          "K872527",
          "K213746",
          "K983576",
          "K232338",
          "K833981",
          "K202440",
          "K933832",
          "K250878",
          "K152748",
          "K200946",
          "K912650",
          "K961357",
          "K042202",
          "K853130",
          "K060240",
          "K102935",
          "K190717",
          "K951399",
          "K080981",
          "K951276",
          "K893418",
          "K934964",
          "K112376",
          "K032434",
          "K090361",
          "K983295",
          "K101736",
          "K070342",
          "K112703",
          "K202753",
          "K172427",
          "K842317",
          "K192210",
          "K213485",
          "K200710",
          "K031905",
          "K170769",
          "K211050",
          "K033818",
          "K011291",
          "K882949",
          "K861623",
          "K193625",
          "K851704",
          "K101232",
          "K821833",
          "K881446",
          "K081256",
          "K760948",
          "K100942",
          "K171888",
          "K042906",
          "K083185",
          "K012217",
          "K140681",
          "K070028",
          "K100563",
          "K103839",
          "K060649",
          "K152508",
          "K771935",
          "K972259",
          "K041646",
          "K844124",
          "K201508",
          "K191251",
          "K200471",
          "K221800",
          "K001281",
          "K212346",
          "K851224",
          "K043331",
          "K060173",
          "K844624",
          "K031958",
          "K062867",
          "K833755",
          "K113141",
          "K250470",
          "K203497",
          "K052849",
          "K904955",
          "K852825",
          "K830592",
          "K193621",
          "K843381",
          "K061942",
          "K181239",
          "K813409",
          "K153149",
          "K993116",
          "K821070",
          "K022542",
          "K951401",
          "K070590",
          "K010224",
          "K830839",
          "K780593",
          "K200802",
          "K211629",
          "K202791",
          "K191978",
          "K230298",
          "K030126",
          "K101253",
          "K012974",
          "K152690",
          "K092557",
          "K070491",
          "K100016",
          "K101912",
          "K001279",
          "K963276",
          "K173113",
          "K992601",
          "K790416",
          "K843040",
          "K173048",
          "K191283",
          "K081160",
          "K984553",
          "K240023",
          "K895875",
          "K023956",
          "K850083",
          "K213084",
          "K112929",
          "K161325",
          "K852246",
          "K100226",
          "K021612",
          "K851147",
          "K832404",
          "K163518",
          "K925675",
          "K992327",
          "K211776",
          "K010462",
          "K213038",
          "K191570",
          "K884119",
          "K951403",
          "K042627",
          "K152680",
          "K200159",
          "K211752",
          "K042631",
          "K943270",
          "K082192",
          "K923289",
          "K173730",
          "K050463",
          "K213082",
          "K761255",
          "K172121",
          "K213011",
          "K214077",
          "K152402",
          "K100660",
          "K180047",
          "K843157",
          "K861473",
          "K993950",
          "K991914",
          "K240077",
          "K120671",
          "K183431",
          "K020520",
          "K864195",
          "K221170",
          "K962070",
          "K896959",
          "K203747",
          "K032362",
          "K874568",
          "K932204",
          "K895137",
          "K182133",
          "K181242",
          "K930680",
          "K983355",
          "K880578",
          "K030580",
          "K070578",
          "K030324",
          "K864044",
          "K173194",
          "K940686",
          "K841740",
          "K951275",
          "K910868",
          "K223037",
          "K083299",
          "K962497",
          "K072641"
        ],
        "registration_number": [
          "2936999",
          "3008246752",
          "1420054",
          "3016866092",
          "3008808082",
          "3016719957",
          "3007007790",
          "3003622639",
          "3003887292",
          "3010521050",
          "1450758",
          "3027344406",
          "3043226252",
          "1036863",
          "3009831876",
          "3038610665",
          "1218950",
          "3011353843",
          "3008361498",
          "9611500",
          "9613304",
          "3015142815",
          "3007048291",
          "3008559238",
          "3005877899",
          "3038259592",
          "3027847716",
          "3030447506",
          "3011191255",
          "3013298431",
          "3003072803",
          "3003174284",
          "2132111",
          "3009973336",
          "3007770159",
          "3005905524",
          "3021422553",
          "3012302888",
          "8022077",
          "9610105",
          "2086043",
          "3009124884",
          "3008507279",
          "3037724692",
          "3011898560",
          "3009191551",
          "3004415095",
          "1417592",
          "3011651103",
          "3008058135",
          "3010262192",
          "3014579161",
          "2320762",
          "3007154442",
          "3008841035",
          "3004871905",
          "3001570235",
          "9617543",
          "3010405299",
          "3010370454",
          "9681837",
          "3008879776",
          "3021343210",
          "8032009",
          "1220908",
          "3008756507",
          "3011654863",
          "3017422354",
          "2133641",
          "1060680",
          "3024512705",
          "3027285900",
          "2410808",
          "3011200334",
          "3016670149",
          "3035111606",
          "9681384",
          "3008770708",
          "3008717264",
          "9680961",
          "9615140",
          "3019737388",
          "3045891669",
          "3017446092",
          "1250045",
          "3012311599",
          "3029871637",
          "3003630387",
          "3009437315",
          "3005785090",
          "3015768317",
          "3012104670",
          "3006733253",
          "3007836437",
          "3043479346",
          "3009518322",
          "3000130829",
          "3012725363",
          "3004145393",
          "3012494290",
          "9680225",
          "3014167550",
          "3011182134",
          "3010552191",
          "1000137876",
          "3003687189",
          "1066427",
          "3004727197",
          "3033536319",
          "3006176286",
          "3004977335",
          "3010097171",
          "3003965134",
          "3005583357",
          "3005951243",
          "3015142935",
          "3018094310",
          "1721676",
          "3016867013",
          "2184009",
          "3010536822",
          "3017059349",
          "3007616088",
          "3003898360",
          "3006852457",
          "3007407626",
          "9617601",
          "3007703436",
          "1217805",
          "3006636961",
          "3016864576",
          "3006845683",
          "3049808762",
          "3038295706",
          "3026032283",
          "1419898",
          "3042272797",
          "2221819",
          "3010390468",
          "3032948391",
          "3008307705",
          "3008716327",
          "3011862780",
          "3006792041",
          "3010419931",
          "2032273",
          "3015529350",
          "3009499478",
          "3009156722",
          "3008820594",
          "3013496867",
          "3010047454",
          "3010476117",
          "3016075711",
          "3027510187",
          "3011359729",
          "3006082230",
          "2082313",
          "3002736372",
          "3005031124",
          "3017894657",
          "3013557562",
          "3028019184",
          "3008956776",
          "3017581269",
          "3022028624",
          "2011171",
          "3027605322",
          "8020605",
          "3006140792",
          "3011137372",
          "9681834",
          "3017643152",
          "3011167030",
          "3006537699",
          "3005699784",
          "3011919098",
          "1221195",
          "8043323",
          "3023031046",
          "3001084743",
          "3005841768",
          "3012933969",
          "3011243696",
          "3012416077",
          "3005569927",
          "3009756153",
          "3020848330",
          "2133786",
          "1928237",
          "3003993690",
          "3031841433",
          "3012834717",
          "3013689228",
          "3030671608",
          "1720839",
          "3018514222",
          "3015134004",
          "3004112621",
          "3034405508",
          "3008729899",
          "8030673",
          "3015876",
          "3016452327",
          "3008830133",
          "3011077581",
          "2029275",
          "3003903031",
          "3015276780",
          "3009328501",
          "3014886121",
          "3016850322",
          "3034993050",
          "3011835632",
          "3003971136",
          "3008384335",
          "3017858797",
          "2182318",
          "3012528160",
          "3005798905",
          "3013053058",
          "3010041511",
          "3030626857",
          "1718850",
          "3044174020",
          "9613558",
          "1423537",
          "3013826848",
          "3000206696",
          "3009327812",
          "3008167616",
          "1000122786",
          "3007502227",
          "3026753375",
          "3026264",
          "9617604",
          "9610816",
          "3033530347",
          "3014541700",
          "2434115",
          "3004035727",
          "3009540516",
          "3016088707",
          "3006648320",
          "2183924",
          "2031508",
          "3010383847",
          "3016904853",
          "3017896172",
          "3010822231",
          "3003591740",
          "3010288146",
          "3010317211",
          "3009308092",
          "9710602",
          "3001124136",
          "3017406171",
          "3017825461",
          "3034603279",
          "3006169981",
          "3004026659",
          "2111760",
          "3003589956",
          "1836518",
          "1043214",
          "1066444",
          "3030089838",
          "3003399181",
          "3016618143",
          "3012678246",
          "3000163458",
          "3007307487",
          "3003981912",
          "3013481815",
          "8043685",
          "3016509757",
          "3018783553",
          "9613793",
          "3018008670",
          "3009487317",
          "3012421607",
          "3029824200",
          "3014015501",
          "1835242",
          "3014774380",
          "9615817",
          "3013823570",
          "3008011951",
          "3016701404",
          "3003832357",
          "2030624",
          "3013596024",
          "1720734",
          "3033715634",
          "3012295567",
          "3015967131",
          "3007305624",
          "1721686",
          "2032112",
          "3014449259",
          "3004215929",
          "3015058854",
          "3023412068",
          "1055808",
          "1316463",
          "3036108847",
          "2381321",
          "3008975047",
          "3015188281",
          "3015697152",
          "3042579404",
          "3024999438",
          "3017619536",
          "2025816",
          "3015136472",
          "3008483389",
          "3006845687",
          "3014848734",
          "3013500228",
          "3004009431",
          "3014856521",
          "3006157842",
          "3037341611",
          "3014271659",
          "3016640651",
          "9616787",
          "3032109181",
          "3009888765",
          "3014263632",
          "1054421",
          "3009001657",
          "3042156936",
          "3003657019",
          "3013679558",
          "3001650535",
          "3010157426",
          "3014235520",
          "3009551875",
          "3007731000",
          "3003896222",
          "1018233",
          "3008973759",
          "3044854644",
          "3026585241",
          "3003697998",
          "3017973191",
          "3013896610",
          "1651104",
          "1832894",
          "3006610790",
          "3007499349",
          "3000126629",
          "1034116",
          "3023272766",
          "3008729547",
          "3013426279",
          "3005613467",
          "3012359877",
          "3012804266",
          "2437780",
          "3009673645",
          "8031509",
          "3008164770",
          "3003660459",
          "3016693518",
          "3010106598",
          "3021635975",
          "3008064428",
          "3015214304",
          "3032109",
          "3012316249",
          "3003263092",
          "3004832480",
          "3010713600",
          "3004111573",
          "3012312061",
          "1650347",
          "3008318433",
          "3027256461",
          "3016901080",
          "2434008",
          "3007603826",
          "3003674698",
          "3000268902",
          "1054241",
          "3009685433",
          "3015140207",
          "3012251275",
          "3012134907",
          "3013874291",
          "3010838917",
          "3005621440",
          "3024491912",
          "1120986",
          "3016595211",
          "3012840029",
          "3010852191",
          "3044754336",
          "3004835896",
          "3008048783",
          "3005783425",
          "3024790242",
          "3017520431",
          "3013517171",
          "3017493034",
          "3028172072",
          "3011065399",
          "3009617411",
          "3019470217",
          "9616843",
          "1123010",
          "3038761440",
          "1066270",
          "3007134734",
          "9616085",
          "3007748817",
          "3021559257",
          "1055236",
          "1450057",
          "3037007069",
          "1062836",
          "3010968402",
          "3000215556",
          "3008754120",
          "3007137643",
          "9681377",
          "3017026146",
          "3018188848",
          "1723241",
          "9710548",
          "3015109923",
          "3018783526",
          "3013783488",
          "3008771621",
          "2518422",
          "1915484",
          "1216677",
          "8022257",
          "3015422337",
          "1422443",
          "3015542362",
          "3021186226",
          "3005536872",
          "3004911384",
          "3030597679",
          "3015616170",
          "3013019728",
          "3016705853",
          "3011792045",
          "3043513123",
          "3003587684",
          "3011713222",
          "3014468707",
          "3008733163",
          "1921846",
          "3007838093",
          "3005951403",
          "1314417",
          "3005083505",
          "3027135885",
          "3006142663",
          "3013545943",
          "3015997711",
          "8010026",
          "2081040",
          "3015455735",
          "3016674095",
          "3012309743",
          "3009077524",
          "3010984687",
          "3011875311",
          "3008687634",
          "3017887928",
          "3004101980",
          "3009889471",
          "1222304",
          "2648045",
          "3041276591",
          "3007413079",
          "9710641",
          "3006979678",
          "3032916632"
        ],
        "fei_number": [
          "2936999",
          "3008246752",
          "1420054",
          "3016866092",
          "3008808082",
          "3016719957",
          "3007007790",
          "3003622639",
          "3003887292",
          "3010521050",
          "1450758",
          "3027344406",
          "3043226252",
          "1036863",
          "3009831876",
          "3038610665",
          "1218950",
          "3011353843",
          "3008361498",
          "3015142815",
          "3007048291",
          "3008559238",
          "3005877899",
          "3038259592",
          "3003081151",
          "3027847716",
          "3003215851",
          "3001236905",
          "3030447506",
          "1000518606",
          "3002253377",
          "3011191255",
          "3005961978",
          "3013298431",
          "3003072803",
          "3003174284",
          "3009973336",
          "3007770159",
          "3005905524",
          "3021422553",
          "3012302888",
          "1000208624",
          "3009124884",
          "3008507279",
          "3037724692",
          "3011898560",
          "3009191551",
          "3004415095",
          "1417592",
          "3011651103",
          "3008058135",
          "3010262192",
          "3014579161",
          "3007154442",
          "3008841035",
          "3004871905",
          "2828",
          "3001570235",
          "3001708019",
          "3010405299",
          "3010370454",
          "3008879776",
          "3021343210",
          "1220908",
          "3008756507",
          "3011654863",
          "3017422354",
          "3024512705",
          "3002737156",
          "3027285900",
          "2410808",
          "3011200334",
          "3016670149",
          "3035111606",
          "3008770708",
          "3008717264",
          "3019737388",
          "3045891669",
          "3017446092",
          "1250045",
          "3012311599",
          "3029871637",
          "3003630387",
          "3009437315",
          "3005785090",
          "3015768317",
          "3002808327",
          "3005176386",
          "3012104670",
          "3006733253",
          "3007836437",
          "3043479346",
          "3009518322",
          "3000130829",
          "3012725363",
          "3004145393",
          "3012494290",
          "3014167550",
          "3011182134",
          "3003513129",
          "3003405786",
          "1000137876",
          "1000112968",
          "3003687189",
          "3004727197",
          "3033536319",
          "3006176286",
          "3001697252",
          "3004977335",
          "3010097171",
          "3003965134",
          "1000517719",
          "3005583357",
          "3005951243",
          "3015142935",
          "3018094310",
          "1721676",
          "3004631140",
          "3016867013",
          "3003515005",
          "3010536822",
          "3017059349",
          "3007616088",
          "3003898360",
          "3002558692",
          "3006852457",
          "1000121041",
          "3007407626",
          "3007703436",
          "1217805",
          "3006636961",
          "3016864576",
          "3003334361",
          "3006845683",
          "3049808762",
          "3038295706",
          "3026032283",
          "1419898",
          "3042272797",
          "2221819",
          "3010390468",
          "3032948391",
          "3008307705",
          "3008716327",
          "3011862780",
          "3006792041",
          "3010419931",
          "1000306431",
          "3015529350",
          "3009499478",
          "3009156722",
          "3001746544",
          "3008820594",
          "3013496867",
          "3010047454",
          "3010476117",
          "3016075711",
          "3027510187",
          "3006469671",
          "3001104093",
          "3011359729",
          "3006082230",
          "2082313",
          "3005031124",
          "3003807346",
          "3017894657",
          "3013557562",
          "3028019184",
          "3008956776",
          "3017581269",
          "3022028624",
          "2011171",
          "1000123297",
          "3027605322",
          "3006140792",
          "1000116132",
          "1000140026",
          "3017643152",
          "3011167030",
          "3006537699",
          "3005699784",
          "3011919098",
          "1221195",
          "3023031046",
          "3001084743",
          "3005841768",
          "3001238683",
          "3012933969",
          "3012416077",
          "3005569927",
          "3009756153",
          "3020848330",
          "1928237",
          "3003993690",
          "3031841433",
          "3000203466",
          "3012834717",
          "3013689228",
          "3030671608",
          "3018514222",
          "3015134004",
          "3004112621",
          "3034405508",
          "3008729899",
          "3015876",
          "3016452327",
          "3008830133",
          "3011077581",
          "3003903031",
          "3000206435",
          "3015276780",
          "3009328501",
          "3014886121",
          "3016850322",
          "3034993050",
          "3003565456",
          "3011835632",
          "3003971136",
          "3008384335",
          "3017858797",
          "2182318",
          "3012528160",
          "3005798905",
          "3013053058",
          "3010041511",
          "3030626857",
          "1718850",
          "3044174020",
          "3006078362",
          "3013826848",
          "3000206696",
          "3009327812",
          "3008167616",
          "1000122786",
          "3007502227",
          "3026753375",
          "3033530347",
          "3014541700",
          "2434115",
          "3003079707",
          "3004035727",
          "3009540516",
          "3016088707",
          "3003040625",
          "3006648320",
          "3003188183",
          "3003261020",
          "3010383847",
          "3016904853",
          "3017896172",
          "3017147080",
          "3010822231",
          "3003591740",
          "3010288146",
          "3010317211",
          "1000110974",
          "3009308092",
          "3001124136",
          "3017406171",
          "3017825461",
          "3034603279",
          "3006169981",
          "2111760",
          "3005479955",
          "3002807626",
          "3003589956",
          "1043214",
          "3030089838",
          "3004213967",
          "3003399181",
          "3016618143",
          "1000126561",
          "3012678246",
          "3011343761",
          "3000163458",
          "3007307487",
          "3003981912",
          "3013481815",
          "3000957797",
          "3016509757",
          "3003076831",
          "3018783553",
          "1000112877",
          "3018008670",
          "3009487317",
          "3012421607",
          "3029824200",
          "3014015501",
          "1000143485",
          "3014774380",
          "3013823570",
          "3008011951",
          "3016701404",
          "3003832357",
          "3013596024",
          "1720734",
          "3033715634",
          "3012295567",
          "3015967131",
          "3010687142",
          "3015191855",
          "3007305624",
          "1721686",
          "1000065475",
          "3014449259",
          "3004215929",
          "3015058854",
          "3023412068",
          "1316463",
          "3036108847",
          "2381321",
          "3008975047",
          "3015188281",
          "3015697152",
          "3042579404",
          "3000718682",
          "3024999438",
          "3017619536",
          "2025816",
          "1000110823",
          "3015136472",
          "3008483389",
          "3006845687",
          "3014848734",
          "3004034544",
          "3013500228",
          "3003240904",
          "3004009431",
          "3014856521",
          "3006157842",
          "3000171642",
          "3037341611",
          "3014271659",
          "3003010579",
          "3016640651",
          "3032109181",
          "3009888765",
          "3014263632",
          "3009001657",
          "3042156936",
          "3003657019",
          "3013679558",
          "3001650535",
          "3010157426",
          "3014235520",
          "3009551875",
          "3003757840",
          "3007731000",
          "3003272793",
          "1018233",
          "3008973759",
          "3044854644",
          "3026585241",
          "3003697998",
          "3017973191",
          "3013896610",
          "3000204692",
          "1832894",
          "3006610790",
          "3000126629",
          "1034116",
          "3023272766",
          "3008729547",
          "3013426279",
          "3005613467",
          "3012359877",
          "3012804266",
          "3009673645",
          "3008164770",
          "3003660459",
          "3016693518",
          "3010106598",
          "3011774813",
          "3021635975",
          "3008064428",
          "3015214304",
          "3012316249",
          "3003263092",
          "3004832480",
          "3010713600",
          "3004111573",
          "3012312061",
          "3008318433",
          "1000300598",
          "3027256461",
          "3016901080",
          "2434008",
          "3007603826",
          "3002807167",
          "3003674698",
          "3000268902",
          "3009685433",
          "3015140207",
          "3012251275",
          "3012134907",
          "3013874291",
          "3010838917",
          "3005621440",
          "3024491912",
          "1120986",
          "3016595211",
          "3012840029",
          "3010852191",
          "3044754336",
          "3004835896",
          "3008048783",
          "3005783425",
          "3024790242",
          "3000719964",
          "3017520431",
          "3013517171",
          "3017493034",
          "3028172072",
          "3011065399",
          "3009617411",
          "3019470217",
          "3038761440",
          "3007134734",
          "3007748817",
          "3021559257",
          "1450057",
          "3037007069",
          "3010968402",
          "3000215556",
          "3008754120",
          "3003753439",
          "3007137643",
          "3017026146",
          "3018188848",
          "3006654321",
          "1000116158",
          "3015109923",
          "3018783526",
          "3013783488",
          "3008771621",
          "2518422",
          "1915484",
          "1216677",
          "3015422337",
          "3015542362",
          "3021186226",
          "3005536872",
          "3004911384",
          "3030597679",
          "3015616170",
          "3013019728",
          "3016705853",
          "3011792045",
          "3043513123",
          "3003587684",
          "3011713222",
          "3014468707",
          "3008733163",
          "1921846",
          "3007838093",
          "3005951403",
          "1314417",
          "3002534070",
          "3005083505",
          "3027135885",
          "3006142663",
          "3013545943",
          "3015997711",
          "2081040",
          "3015455735",
          "3016674095",
          "3012309743",
          "3009077524",
          "3010984687",
          "3011875311",
          "3008687634",
          "3017887928",
          "3004101980",
          "3005747797",
          "3002806783",
          "3009889471",
          "2648045",
          "3041276591",
          "3007413079",
          "3006979678",
          "3032916632",
          "3007513035"
        ],
        "device_name": "Continuous Measurement Thermometer",
        "medical_specialty_description": "General Hospital",
        "regulation_number": "880.2910",
        "device_class": "2"
      }
    }
  ]
}