  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-28",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "213895",
      "product_res_number": "Z-2046-2025",
      "event_date_initiated": "2025-05-07",
      "event_date_posted": "2025-06-30",
      "recall_status": "Open, Classified",
      "res_event_number": "96892",
      "product_code": "SDE",
      "product_description": "Brand Name: Valeda Light Delivery System\nProduct Name: Light Based Device for Dry Age Related Macular Degeneration\nProduct Description: Light Based Device for Dry Age Related Macular Degeneration\nComponent: No",
      "code_info": "Lot Code: UDI-DI +B749200001030/$$+720172/16D20250311T; Serial Number 20172 \nUDI-DI +B749200001030/$$+720166/16D20250311W; Serial Number 20166 \nUDI-DI +B749200001030/$$+720167/16D20250311X; Serial Number 20167 \nUDI-DI +B749200001030/$$+720168/16D20250311Y; Serial Number 20168 \nUDI-DI +B749200001030/$$+720169/16D20250311Z; Serial Number 20169 \nUDI-DI +B749200001030/$$+720175/16D20250311W; Serial Number 20175 \nUDI-DI +B749200001030/$$+720177/16D20250311Y; Serial Number 20177 \nUDI-DI +B749200001030/$$+720178/16D20250311Z; Serial Number 20178 \n\nUpdate 14 May 2025:\nUDI-DI +B749200001030/$$+720170/16D20250311R; Serial Number 20170",
      "firm_fei_number": "3016232028",
      "recalling_firm": "LUMITHERA INC",
      "address_1": "19578 10th Ave Ne Ste 200 19578 10th Ave Ne",
      "address_2": "Ste 200",
      "city": "Poulsbo",
      "state": "WA",
      "postal_code": "98370-7332",
      "additional_info_contact": "Tracy Puckett\n844-3423333",
      "reason_for_recall": "U.S. customers were shipped devices that were configured for the European Union and were unable to plug the device in the electrical outlet.",
      "root_cause_description": "Under Investigation by firm",
      "action": "On 05/08/2025, the firm initiated sending customers an \"Valeda - Return\" email to inform them that the Valeda system that was distributed to them was sent with the wrong configuration.\n\nCustomer were instructed to use the attached return label and return the systems back to LumiThera.",
      "product_quantity": "8",
      "distribution_pattern": "US Nationwide distribution in the states of AZ, CA, FL, IL, ND, TN, TX.",
      "openfda": {
        "k_number": [
          "DEN230083"
        ],
        "registration_number": [
          "",
          "3016232028",
          "2133810",
          "3004186703"
        ],
        "fei_number": [
          "",
          "3016232028",
          "3000203066",
          "3004186703"
        ],
        "device_name": "Light Based Device For Dry Age-Related Macular Degeneration",
        "medical_specialty_description": "Ophthalmic",
        "regulation_number": "886.5520",
        "device_class": "2"
      }
    }
  ]
}