  "results": [
    {
      "cfres_id": "87457",
      "product_res_number": "Z-1170-2010",
      "event_date_initiated": "2009-05-01",
      "event_date_posted": "2010-04-12",
      "recall_status": "Terminated",
      "event_date_terminated": "2012-04-30",
      "res_event_number": "54087",
      "product_code": "RDW",
      "product_description": "LM Aculight Corporation Perseus Pulsed Fiber Laser, Model number: ML36L-1541.  Additional Model numbers with non-defective type\"A\" fiber: M22L-1540, M27L-1541, M281-1540, M30PL-1541, & M33L-1541.  Lockheed Martin Aculight, Bothell, WA, 98021.\n\nPulsed laser energy generator intended for use as an OEM component for the LiDAR/Rangefinding, Surveillance, Mapping, and Sensing.",
      "code_info": "Part Number: 35-001466-032, Corresponding Unit Serial Numbers: 00001,00002, and 00003.",
      "firm_fei_number": "3003346474",
      "recalling_firm": "Lockheed-Martin Aculight",
      "address_1": "22121-20th Ave SE Ave SE",
      "city": "Bothell",
      "state": "WA",
      "postal_code": "98021",
      "reason_for_recall": "Defect of optical fiber - The Perseus device with type \"B\" fiber has been found to emit potentially hazardous laser radiation.",
      "root_cause_description": "Radiation Control for Health and Safety Act",
      "action": "A Radiation Warning Notification was sent to all customers.  The notification stated the purpose and problem.  In addition, a Notification of Potential Perseus Laser Safety Issue letter was sent to the only customer who received the defective type \"B\" fiber product.  The letter identified the affected product and stated that the product was not considered eye-safe.  The firm recommended that the customer cease using the laser with the Part Number: 35-001466-032 and Corresponding Unit Serial Numbers: 00001,00002, and 00003.  The customer is to return the affected product for additional testing and warranty repair.  Questions should be directed to Rick Whitty on 425-482-1100 or at Richard.j.whitty@lmco.com.",
      "product_quantity": "3",
      "distribution_pattern": "Worldwide Distribution -- USA, including states of PA, FL, MD, & CO and the countries of Austria and Canada.",
      "openfda": {
        "device_name": "General Purpose Laser Products",
        "medical_specialty_description": "Unknown",
        "regulation_number": "",
        "device_class": "N"
      }
    }
  ]
}