  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 2
    }
  },
  "results": [
    {
      "cfres_id": "213090",
      "product_res_number": "Z-1571-2025",
      "event_date_initiated": "2025-03-04",
      "event_date_posted": "2025-04-15",
      "recall_status": "Open, Classified",
      "res_event_number": "96519",
      "product_code": "QWR",
      "k_numbers": [
        "K221925"
      ],
      "product_description": "Brand Name: ID NOW COVID-19 2.0 24T\nProduct Name: ID NOW COVID-19 2.0 24T\nModel/Catalog Number: 192-000\nSoftware Version: N/A\nProduct Description: Simple Point-Of-Care Device To Detect Sar-Cov-2 Nucleic Acid Targets\nComponent: No",
      "code_info": "Model No 192-000; GTIN/DI 10811877011351; Lot 000M884910 \nModel No 192-000; GTIN/DI 10811877011351; Lot 000M885121 \nModel No 192-000; GTIN/DI 10811877011351; Lot 000M911157 \nModel No 192-000; GTIN/DI 10811877011351; Lot 000M913227 \nModel No 192-000; GTIN/DI 10811877011351; Lot 000M913286 \nModel No 192-000; GTIN/DI 10811877011351; Lot 000M913303 \nModel No 192-000; GTIN/DI 10811877011351; Lot 000M913641 \nModel No 192-000; GTIN/DI 10811877011351; Lot 000M914188 \nModel No 192-000; GTIN/DI 10811877011351; Lot 000M914221 \nModel No 192-000; GTIN/DI 10811877011351; Lot 000M914852 \nModel No 192-000; GTIN/DI 10811877011351; Lot 000M915373 \nModel No 192-000; GTIN/DI 10811877011351; Lot 000M916267 \nModel No 192-000; GTIN/DI 10811877011351; Lot 000M921404 \nModel No 192-000; GTIN/DI 10811877011351; Lot 000M922204 \nModel No 192-000; GTIN/DI 10811877011351; Lot 000M922578 \nModel No 192-000; GTIN/DI 10811877011351; Lot 000M922601 \nModel No 192-000; GTIN/DI 10811877011351; Lot 000M922634 \nModel No 192-000; GTIN/DI 10811877011351; Lot 000M924048 \nModel No 192-000; GTIN/DI 10811877011351; Lot 000M925615 \nModel No 192-000; GTIN/DI 10811877011351; Lot 000M938692 \nModel No 192-000; GTIN/DI 10811877011351; Lot 00M922634A \nModel No 192-000; GTIN/DI 10811877011351; Lot 00M924048A\n\n****Updated 5/8/2025****\n04/29/2025 (Adding Additional Lots):\nModel No 192-000;  GTIN/DI 10811877011351;  Lot 000M913147\nModel No 192-000;  GTIN/DI 10811877011351;  Lot 000M913159\nModel No 192-000;  GTIN/DI 10811877011351;  Lot 000M922950\nModel No 192-000;  GTIN/DI 10811877011351;  Lot 000M922988\nModel No 192-000;  GTIN/DI 10811877011351;  Lot 000M926188",
      "firm_fei_number": "1221359",
      "recalling_firm": "Abbott Diagnostics Scarborough, Inc.",
      "address_1": "10 Southgate Rd",
      "city": "Scarborough",
      "state": "ME",
      "postal_code": "04074-8303",
      "additional_info_contact": "Vicki Assardo\n224-6681505",
      "reason_for_recall": "The impacted lots have a higher occurrence of invalid rates when compared to the product Instructions for Use.",
      "root_cause_description": "Process change control",
      "action": "On March 4, 2025 FIELD CORRECTION NOTICE letters were sent to end user customers. \nPlease complete the following actions, as applicable.\n1. If you have impacted inventory in stock, then:\n\" Discontinue use of and destroy any remaining inventory of the impacted lots according to your procedures.\n\" Complete and return the Customer Reply Form (Form must be completed, signed and returned to receive a replacement lot).\n\" Please retain this letter for your records.\n2. If you have forwarded the product listed above to others in your network, then:\n- Inform them of this Field Correction Notice, provide to them a copy of this notice and request they take the necessary action.\n3. If you do not have impacted inventory in stock, then:\n- All product lots not identified in Appendix I can continue to be used.\nIf you have questions regarding this information, please contact Sedgwick by phone at 888 671 8858 (available 8am   5pm EST, Monday - Friday).\nAdverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online (http://www.fda.gov/MedWatch/report.htm), by mail (http://www.fda.gov/MedWatch/getforms.htm), by phone (1-800-332-1088), or by fax (1-800-FDA-0178).\nIf you have experienced any patient or user injury associated with this Field Correction Notice, please immediately report the event to your local area Customer Service.\n\n****Update added 5/8/2025****\nOn April 29, 2025 FIELD CORRECTION NOTICE letters were sent to customers informing them of additional lots included in this recall.",
      "product_quantity": "339456 eaches (337560 US, 1896 OUS)",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Australia; Austria; Belgium; Bulgaria; Canada; Costa Rica; France; French Polynesia; Germany; Italy; Luxembourg; Malaysia; Netherlands; Paraguay; Poland; Qatar; Saudi Arabia; Slovenia; South Korea; Spain; Switzerland; Taiwan; United Kingdom; Uruguay.",
      "openfda": {
        "k_number": [
          "K223783",
          "K243346",
          "K221925"
        ],
        "registration_number": [
          "3023372670",
          "1221359",
          "3004141078",
          "2243471"
        ],
        "fei_number": [
          "3023372670",
          "1221359",
          "3004141078",
          "2243471"
        ],
        "device_name": "Simple Point-Of-Care Device To Detect Sar-Cov-2 Nucleic Acid Targets From Clinical Specimens In Near-Patient Settings",
        "medical_specialty_description": "Microbiology",
        "regulation_number": "866.3982",
        "device_class": "2"
      }
    }
  ]
}