  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "187938",
      "product_res_number": "Z-2081-2021",
      "event_date_initiated": "2021-06-02",
      "event_date_created": "2021-07-16",
      "recall_status": "Terminated",
      "event_date_terminated": "2022-04-28",
      "res_event_number": "88122",
      "product_code": "QMT",
      "product_description": "ANCILLARY ADULT 140-DOSE CONVENIENCE KIT, ITEM #1189482, for the administration of the Moderna COVID-19 vaccine\n\n convenience kit for the administration of the COVID-19 vaccine",
      "code_info": "All convenience kit lot numbers distributed 5/4/2021 - 5/24/2021, containing syringes as follows: a) SYRINGE/NDL, 3CC 25GX1 1/2\" BD, Lot numbers:  9252960, 0296288, 9252961 b) SYRINGE/NDL, 3CC 23GX1 1/2\" BD, Lot Numbers:  0324763, 0324761, 0324766, 0324770, 1015383, 1020431",
      "firm_fei_number": "3004386693",
      "recalling_firm": "Mckesson Medical-Surgical Inc. Corporate Office",
      "address_1": "9954 Maryland Drive Deep Run Iii",
      "address_2": "Ste. 4000",
      "city": "Richmond",
      "state": "VA",
      "postal_code": "23233",
      "additional_info_contact": "Mr. Frank Maritato\n804-553-2281",
      "reason_for_recall": "Two unsuitable 3mL syringes with conventional needles for vaccine administration included in some Moderna 140 Ancillary Convenience Kits.",
      "root_cause_description": "Process change control",
      "action": "McKesson issued an \"URGENT MEDICAL DEVICE SAFETY ALERT\" on 06/02/2021 by email to its consignees. The notice requested that the consignees do the following:\n-examine their inventory immediately to determine if you have any quantities of the product, \n-Immediately stop using these syringes with 3mI syringes with conventional needles for the administration of the Moderna COVID-19 vaccine.\n-As these 3mI syringes with conventional needles can be utilized for a variety of other uses, such as medication mixing in healthcare settings, you may continue to use these 3mI syringes with conventional needles in your healthcare facility for other appropriate uses.\n-Please fill out and return the Medical Device Alert Response Form attached.\n\nThe firm would be shipping replacement safety syringes.\n\nFor questions about this notification, please call 877-546-5661.",
      "product_quantity": "53,388 kits",
      "distribution_pattern": "US Nationwide Distribution",
      "openfda": {
        "registration_number": [
          "3003911670",
          "3005034064"
        ],
        "fei_number": [
          "3003911670",
          "3005034064"
        ],
        "device_name": "Prep Kit  Vaccine Injection",
        "medical_specialty_description": "General Hospital",
        "regulation_number": "880.5570",
        "device_class": "2"
      }
    }
  ]
}