  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "199353",
      "product_res_number": "Z-1600-2023",
      "event_date_initiated": "2023-04-06",
      "event_date_posted": "2023-05-17",
      "recall_status": "Open, Classified",
      "res_event_number": "91982",
      "product_code": "QJQ",
      "k_numbers": [
        "K903654"
      ],
      "product_description": "Fisher Wallace Stimulator Models FW-100 and FW-200",
      "code_info": "FW-100: UDI-DI 00860000919005, Serial Numbers 2020200000-2022278000;\n\nFW-200: UDI-DI 00860000919012, Serial Numbers 2021140129T-2022030528T",
      "firm_fei_number": "3006258094",
      "recalling_firm": "Fisher Wallace Laboratories Inc.",
      "address_1": "630 Flushing Ave Ste 104 630 Flushing Ave",
      "address_2": "Ste 104",
      "state": "NY",
      "postal_code": "11206-5775",
      "additional_info_contact": "Diana Aghabekian\n800-692-4380",
      "reason_for_recall": "If the batteries are installed incorrectly, the product may overheat, possibly leading to pain, discomfort, irritation, or even burns during use of the device.",
      "root_cause_description": "Nonconforming Material/Component",
      "action": "An URGENT MEDICAL DEVICE CORRECTION dated 4/6/23 was sent to customers.\n\nShort term actions to be taken by the User:\n1. Always ensure the batteries are inserted correctly into the device (see above diagram).\n2. Do not leave batteries in the device when the device is not in use (e.g. in between a morning and evening treatment).\n3. Do not fall asleep while using the device because corrective actions cannot be taken when the user is asleep.\n4. If the device exhibits excess heat, burning smell, or odor, turn off the device, and discontinue use. Contact Fisher Wallace Laboratories immediately.\n\nThe long term corrective action by the company is still being determined.\n\nWe request your action by replying to this email to confirm that you understand the information in this medical device correction and the subsequent actions needed to be taken.\n\nFisher Wallace apologizes for any inconvenience this may cause. If you have any questions about what to do with your product, please contact Fisher Wallace by telephone Monday - Friday, 10:00 AM to 6:00 PM, Eastern Time at (800) 692-4380 or by email at postmarketsurveillance@fisherwallace.com.",
      "product_quantity": "43,489 total devices",
      "distribution_pattern": "US Nationwide. Global Distribution.",
      "openfda": {
        "k_number": [
          "K024377",
          "K903014",
          "K090052",
          "K873920",
          "K883812",
          "K182311",
          "K060158",
          "K894515",
          "K062284",
          "K895175",
          "K903654",
          "K070412",
          "K771534",
          "K910616",
          "K932050",
          "K894097",
          "K232253",
          "K252951",
          "K230826"
        ],
        "registration_number": [
          "3014498660",
          "3005026995",
          "3027997270",
          "3006258094",
          "2954783",
          "2020648",
          "8040537",
          "3007738866",
          "3008312728",
          "3002809311",
          "3014773746",
          "3010900508",
          "3013526742"
        ],
        "fei_number": [
          "3003186357",
          "3014498660",
          "3005026995",
          "3027997270",
          "3006258094",
          "2020648",
          "3007738866",
          "3008312728",
          "3002809311",
          "3005154439",
          "3014773746",
          "3010900508",
          "3013526742"
        ],
        "device_name": "Cranial Electrotherapy Stimulator To Treat Insomnia And/Or Anxiety",
        "medical_specialty_description": "Neurology",
        "regulation_number": "882.5800",
        "device_class": "2"
      }
    }
  ]
}