  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "209671",
      "product_res_number": "Z-3262-2024",
      "event_date_initiated": "2024-08-07",
      "event_date_posted": "2024-09-25",
      "recall_status": "Open, Classified",
      "res_event_number": "95173",
      "product_code": "QAV",
      "k_numbers": [
        "K221338"
      ],
      "product_description": "PT301US Airvo 3 Respiratory Support Device\nwith software version 1.2.0 and/or 1.5.1\n\nThe Airvo 3 is intended to provide high flow warmed and humidified respiratory gases for administration to spontaneously breathing infant, child, adolescent and adult patients in hospitals and sub-acute facilities.",
      "code_info": "Model Number: PT301US\nUDI-DI code: 09420012466662\nLot/Serial Numbers:",
      "firm_fei_number": "3002806988",
      "recalling_firm": "Fisher & Paykel Healthcare, Ltd.",
      "address_1": "15 Maurice Paykel Place",
      "city": "Auckland",
      "state": "East Tamaki",
      "country": "New Zealand",
      "additional_info_contact": "649-574-0100",
      "reason_for_recall": "Due to a software issue, affected devices that are set up with High Pressure Oxygen (HPO), if the flow alignment alarm occurs, the device will deliver room air only.  If this happens, a patient my experience oxygen desaturation that could lead to hypoxia.",
      "root_cause_description": "Software change control",
      "action": "On 08/13/2024, the firm sent an \"Urgent: Medical Device Correction\" Letter via email, phone, letter or face to face communication to inform customers that due to a software issue with the Airvo 3 (software version 1.5.1 or earlier) when the flow alignment alarm 3.2.2 occurs, Airvo 3 devices set up with high pressure oxygen the device will deliver room air only.  Patients may experience oxygen desaturation that could lead to hypoxia.\n\nFor Distributors - Actions required for affected products are:\nStep1-Identify affected products using Model Number reference (REF), Serial Number (SN) and Manufacturing Date (YYYY-MM-DD).  If the device is still within your inventory, put the product in quarantine to prevent ongoing distribution until Fisher&Paykel  (F&P) Healthcare has contacted you with next steps.\nStep2-Complete and return the attached response form\nStep3-Await instructions from Fisher&Paykel Healthcare\n\nFor healthcare facilities:\nUntil the software is updated, the Airvo 3 device can continue to be used by your customers. When using the device, all instructions, including warnings and cautions in the Airvo 3 User Manual must be followed, particularly those in Sections 1 and 2.\nIf alarm 3.2.2 occurs, follow the onscreen instructions.\n\nFor questions/assistance - contact F&P Field Action Team via email at FieldAction.USA@fphcare.com or directly at +1 (800) 446 3908 ext 5003 or +1 (949) 453 4000 ext 5003.",
      "product_quantity": "944 units",
      "distribution_pattern": "U.S.: AZ, CA, CO, CT, DC, FL, GA, IA, IL, KY, MA, MD, MI, MN, MO, MT, ND, NH, NJ, NV, NY, OH,  OR, PA, SD, TX, UT, VA and WA.\n\nO.U.S.: Not provided",
      "openfda": {
        "k_number": [
          "K221338",
          "DEN170001",
          "K221318"
        ],
        "device_name": "High Flow/High Velocity Humidified Oxygen Delivery Device",
        "medical_specialty_description": "Anesthesiology",
        "regulation_number": "868.5454",
        "device_class": "2",
        "registration_number": [
          "3008264065",
          "9611451",
          "3023209222",
          "3012971441",
          "2085602",
          "3015928685"
        ],
        "fei_number": [
          "3008264065",
          "3023209222",
          "3012971441",
          "3002806988",
          "3015928685",
          "3000210616"
        ]
      }
    }
  ]
}