{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-28",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "189439",
      "product_res_number": "Z-0176-2022",
      "event_date_initiated": "2021-09-21",
      "event_date_created": "2021-10-26",
      "recall_status": "Terminated",
      "event_date_terminated": "2023-09-08",
      "res_event_number": "88750",
      "product_code": "PMU",
      "k_numbers": [
        "K192898"
      ],
      "product_description": "ASC PneumoLiner device, Part No. WA90500US (US market only)\n\nThe Pneumoliner device is intended for use as a multiple instrument port and tissue containment system during minimally invasive gynecologic laparoscopic surgery to enable the isolation and containment of tissue, considered benign, resected during single-port or multi- site laparoscopic surgery during power morcellation and removal.",
      "code_info": "Lots 647572 and 667060",
      "firm_fei_number": "2429304",
      "recalling_firm": "Olympus Corporation of the Americas",
      "address_1": "3500 Corporate Pkwy",
      "address_2": "PO Box 610",
      "city": "Center Valley",
      "state": "PA",
      "postal_code": "18034-0610",
      "additional_info_contact": "Laura Storms\n484-896-5688",
      "reason_for_recall": "Due to manufacturing error, the Pneumoliner Bag Distal Tab that exits the Introducer shaft is in the wrong orientation. This results in the user deploying the bag upside down, which will make tissue encapsulation and bag closure more difficult, introducing the risk of trapping the small bowel/viscera in the bag at closure resulting in patient injury.",
      "root_cause_description": "Nonconforming Material/Component",
      "action": "On September 21, 2021, the firm, Olympus, sent an \"URGENT: MEDICAL DEVICE FIELD REMOVAL ACTION\" letters to affected customers. Customers were instructed to do the following:\n\n1. Immediately assess any affected product you have in stock and quarantine any affected product.\n2. Call your Olympus customer service representative at 1-888-524-7266 option 1. Olympus will issue a Return Material Authorization in order to return any affected product at no charge to you. Olympus will issue a credit or\nreplacement to your facility for your affected product.\n3. Access the Olympus recall portal to indicate that you have received this notification. Go to https://olympusamerica.com/recall. Enter the recall number Recall-0401 and provide your contact information as indicated in the portal.\n\nIn addition, if you may have further distributed this product, please identify your customers, notify them at once of this field removal action and appropriately document your notification process.\n\nIf you require additional information, please contact Global V.P.at (484)896-5688 from Monday till Friday 8AM ET to 5PM ET or by e-mail at laura.storms@olympus.com.",
      "product_quantity": "350 pieces/70 boxes",
      "distribution_pattern": "US Nationwide distribution to states of:  AL, CA, CT, FL, IL, MD, MI, MN, MO, NE, NY, OH, OK, OR, PA, SC, UT, VA, WA, WI,  and WV ; and International (foreign) distribution to: Europe.",
      "openfda": {
        "k_number": [
          "K192898",
          "K212659",
          "K250212",
          "DEN150028"
        ],
        "registration_number": [
          "9610773",
          "3009039068",
          "3005049530",
          "2219920",
          "3023620966",
          "3025598220"
        ],
        "fei_number": [
          "3009039068",
          "3005049530",
          "2219920",
          "3023620966",
          "3025598220",
          "3002807798"
        ],
        "device_name": "Containment System, Laparoscopic Power Morcellation, With Instrument Port",
        "medical_specialty_description": "Obstetrics/Gynecology",
        "regulation_number": "884.4050",
        "device_class": "2"
      }
    }
  ]
}