  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "169028",
      "product_res_number": "Z-2295-2019",
      "event_date_initiated": "2015-03-16",
      "event_date_created": "2019-08-21",
      "recall_status": "Terminated",
      "event_date_terminated": "2022-02-09",
      "res_event_number": "81594",
      "product_code": "PDT",
      "k_numbers": [
        "K121659"
      ],
      "product_description": "Brand Name Burn Navigator\nModel Numbers:\nCatalog 1120, Burn Navigator H2\nCatalog 1125, Burn Navigator H2 Transport Package\n\nBurn Navigator User Manual\nManual-1120 Burn Navigator User's Manual, Rev. F\n\nBattery:\nPanasonic Model CF-H2 Part Number CMP-1523",
      "code_info": "The Battery Packs are referred to as component number CMP-1523 in the Burn Navigator User's Manual).  The Model/Catalog Numbers 1120 and 1125 are associated with the Burn Navigator H2.   Model 1120 (Qty 20)                   Model 1125 (Qty 8) 2EKSA25422                              3HKSB54129 2GKSA30512                              3HKSB54132 2GKSA30513                              3HKSB54136 2GKSA31162                              3HKSB54143 2JKSA37073                               4BKSA60753 2JKSA37077                               4BKSA60758 2JKSA37465                               4BKSA60776 2JKSA37887                               4BKSA60786 2LKSA43100 2LKSA43540 2LKSA43541 3FKSB51215 3FKSB51217 3HKSB54006 3HKSB54008 3HKSB54115 3HKSB54140 3HKSB54142 4CKSA61744 4CKSA61752  Battery Models and Lot Numbers: Model: CF-VZSU53AW CF-VZSU53AR CF-VZSU53AQ  4 digit alphanumeric Panasonic Lot Numbers beginning with:  B**** C1** C2** C3** C4** C5** C6** C7** C8** C93* C94* C96* C9C* C9J* C9K*",
      "firm_fei_number": "3010283477",
      "recalling_firm": "Arcos, Inc",
      "address_1": "5819 Highway 6 Ste 150 5819 Highway 6",
      "address_2": "Ste 150",
      "city": "Missouri City",
      "state": "TX",
      "postal_code": "77459-4061",
      "additional_info_contact": "Christopher Meador\n713-397-3030",
      "reason_for_recall": "Battery packs  may heat up abnormally and ignite in rare instance.",
      "root_cause_description": "Component design/selection",
      "action": "On March 20, 2015, the firm sent an email with a Battery Replacement Program Attachment.  The Attachment informed customers that it main hardware supplier had notified them that some batteries used with tablet computers have in rare instances, overheated and caught fire. \n\nThe Attachment instructed customers to check the batteries in the computer tablets to determine if they matched the serial numbers for the batteries being replaced.  Customer were instructed to: 1) Remove Battery Pack - shut down their PC and remove the battery pack (noting that there are two battery slots, one on each the left and right side of the Tablet Computer. 2) Review Battery Pack Model Number and Manufacturing Lot Number - by reviewing battery pack nameplate. The following Battery Pack Model Numbers and Manufacturing Lot Numbers must be replaced:\n\nBattery Pack Model Numbers:                 Manufacturing Lot Number\nCF-VZSU53AW                                           Beginning with B, C1, C2, C3, C4, C5, CF-VZSU53AR                                            C6, C7, C8, C93, C94, C96, C9C, C9J,\nCF-VZSU53AQ                                            C9K\n\nIf the customer has the affected batteries, they are to contact the firm whether or not  the batteries match the criteria so that the firm will know the status of the customer's batteries.  If batteries match the criteria, the firm will replace them for free.\n\nContact Information:\nToll Free Phone Number:  877-542-8025\n\nIn addition Tablet Computers can continue to be operated with wall power or with batteries not affected by the recall.",
      "product_quantity": "28",
      "distribution_pattern": "US: AZ, LA, OR, MD, TX, and WA.\n\nOUS: No products were distributed outside the US.",
      "openfda": {
        "k_number": [
          "K121659",
          "K140387"
        ],
        "device_name": "Burn Resuscitation Decision Support Software",
        "medical_specialty_description": "Anesthesiology",
        "regulation_number": "868.1890",
        "device_class": "2"
      }
    }
  ]
}