      "cfres_id": "186432",
      "product_res_number": "Z-1503-2021",
      "event_date_initiated": "2021-03-17",
      "event_date_created": "2021-04-27",
      "recall_status": "Terminated",
      "event_date_terminated": "2022-06-29",
      "res_event_number": "87637",
      "product_code": "PCL",
      "k_numbers": [
        "K140455"
      ],
      "product_description": "Trinity Biotech Captia Measles IgM Kit, Catalog No. 2326060 - Product Usage: intended for the qualitative detection of Measles igM antibodies in human serum of patients suspected of measles (rubeola) infection.",
      "code_info": "Kit Lot:  2326060-060 Expiration Date: 2021-06-28 (UDI): 05391516743655",
      "firm_fei_number": "1318354",
      "recalling_firm": "Clark Laboratories, Inc.",
      "address_1": "2823 Girts Rd",
      "city": "Jamestown",
      "state": "NY",
      "postal_code": "14701-9666",
      "additional_info_contact": "Rich Eckloff\n716-485-4900 Ext. 2416",
      "reason_for_recall": "Positive Control may run high out of its assigned ISR range and result in an invalid run and necessitate retesting",
      "root_cause_description": "Under Investigation by firm",
      "action": "Trinity Biotech issued Product Notification Letter on March 17, 2021 stating reason for recall, health risk and action to take:\nPlease note, if a laboratorys test resulted in a valid run, such as all kit Controls and Calibrator met the Quality Control Criteria as stated in the package insert, the patient results can be reported and a re-evaluation of patient results is not necessary.\nCustomers are instructed to discontinue the use of any Measles IgM Lot #2326060-060 kits that they may have in their inventory. Please complete the attached fax back form informing us of how many of these kits you and your customers have in your inventory that will require replacement.\nA thorough investigation into the root cause of this issue has been initiated and any corrective/preventive actions will be implemented as appropriate. If you have any questions or comments arising from this customer communication, the Trinity Biotech Help Desk is ready to answer your questions (800-325-3424, option 2).",
      "product_quantity": "80 kits",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution and the country of Ireland.",
      "openfda": {
        "k_number": [
          "K140455"
        ],
        "registration_number": [
          "2029372",
          "1318354"
        ],
        "fei_number": [
          "1318354",
          "1000306617"
        ],
        "device_name": "Enzyme Linked Immunoabsorbent Assay, Rubeola Igm",
        "medical_specialty_description": "Microbiology",
        "regulation_number": "866.3520",
        "device_class": "1"
      }
    }
  ]
}