  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-28",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "223338",
      "product_res_number": "Z-0465-2027",
      "event_date_initiated": "2026-10-01",
      "event_date_posted": "2026-11-13",
      "recall_status": "Open, Classified",
      "res_event_number": "100062",
      "product_code": "OIO",
      "product_description": "Presource Convenience kits containing recalled WelMed Extension Tubing Sets\n\nIR SPECIAL PACK               \t\n\tSAN73IRJC1\n\tSAN73IRJC2\n\tSAN73IRJC3\n\tSAN73IRJC5\n\tSAN73IRJC6\n\tSAN73IRJC7\n\tSAN73IRJC8\n\tSAN73IRJC9\n\tSAN73IRJCA\n\tSAN73IRJCB\n\t\nMYELOGRAM ARTHROGRAM TRAY     \t\n\t27-7531E\n\nMYELOGRAM II PACK             \t\n\tSAN41MTHMA\n\tSAN41MTHMB\n\t\nMYELOGRAM TRAY                \t\n\t27-0722G\n\t27-0722H\n\t\nRADIOLOGY PROCEDURE PACK      \t\n\t41-0532E\n\t41-0532F",
      "code_info": "SAN73IRJC1\t\t\n\tUDI-DI 10197106940396\t\n\t\tKit lot 451417\nSAN73IRJC2\t\t\n\tUDI-DI 10198956116245\t\n\t\tKit lot 548858\nSAN73IRJC3\t\t\n\tUDI-DI 10198956131668\t\n\t\tKit lot 560690\nSAN73IRJC5\t\t\n\tUDI-DI 10198956252851\t\n\t\tKit lot 651915\nSAN73IRJC6\t\t\n\tUDI-DI 10198956267190\t\n\t\tKit lot 662732\nSAN73IRJC7\t\t\n\tUDI-DI 10198956282452\t\n\t\tKit lot 675388\nSAN73IRJC8\t\t\n\tUDI-DI 10198956315761\t\n\t\tKit lot 702876\nSAN73IRJC9\t\t\n\tUDI-DI 10198956373471\t\n\t\tKit lot 723744\nSAN73IRJCA\t\t\n\tUDI-DI 10197106799253\t\n\t\tKit lots 395797\n\t\t425148\n\t\t529837\n\t\t606168\n\t\t615098\n\t\t620447\n\t\t688228\nSAN73IRJCB\t\t\n\tUDI-DI 10198956327498\t\n\t\tKit lots 743507\n\t\t757083\n\t\t757832\n\t\t\n27-7531E\t\t\n\tUDI-DI 10195594712495\t\n\t\tKit lots 286748\n\t\t368357\n\t\t396790\n\t\t399893\n\t\t412923\n\t\t419276\n\t\t432662\n\t\t458844\n\t\t543120\n\t\t597861\n\t\t629382\n\t\t645409\n\t\t746339\n\t\t762283\n\t\t\nSAN41MTHMA\t\t\n\tUDI-DI 10888439635645\t\n\t\tKit lots 359818\n\t\t442460\n\t\t463684\n\t\t465389\n\t\t487989\n\t\t596461\n\t\t628974\n\t\t672810\n\t\t702964\nSAN41MTHMB\t\t\n\tUDI-DI 10198956409613\t\n\t\tKit lot 744495\n\t\t\n27-0722G\t\t\n\tUDI-DI 10197106493151\t\n\t\tKit lots 230488\n\t\t248267\n\t\t310130\n\t\t374292\n\t\t404288\n\t\t443048\n\t\t483867\n\t\t517251\n\t\t605745\n\t\t631531\n\t\t665068\n27-0722H\t\t\n\tUDI-DI 10198956326170\t\n\t\tKit lots 713307\n\t\t752697\n\t\t\n41-0532E\t\t\n\tUDI-DI 10195594724368\t\n\t\tKit lots 269719\n\t\t294015\n\t\t346479\n\t\t382489\n\t\t384109\n41-0532F\t\t\n\tUDI-DI 10197106793022\t\n\t\tKit lots 408335\n\t\t417340\n\t\t441388\n\t\t569445\n\t\t599562\n\t\t628149\n\t\t641723\n\t\t690286\n\t\t706157\n\t\t713238\n\t\t732036\n\t\t745511",
      "firm_fei_number": "3001236905",
      "recalling_firm": "Cardinal Health 200, LLC",
      "address_1": "3651 Birchwood Dr",
      "city": "Waukegan",
      "state": "IL",
      "postal_code": "60085-8337",
      "additional_info_contact": "Janna Parks\n847-887-1114",
      "reason_for_recall": "Kits include Extension Tubing Sets recalled by supplier WelMed. Affected tubing sets may leak.",
      "root_cause_description": "Component design/selection",
      "action": "Firm began notifying customers on October 1, 2026.\n\nPrior to use, users should visually inspect the tubing connection area, including bonded junctions and connection points, for any visible irregularities, gaps, deformation, incomplete bonding, leakage, or other abnormalities consistent with the examples shown above.\n\nIf a defect is observed, or if leakage or connection difficulties are encountered during setup or priming, the product should not be used for patient care and discarded. Customers should report affected product through their normal complaint reporting process and retain the product for evaluation if requested.\n\nDirect consignees were instructed to notify retail customers of the market action.",
      "product_quantity": "1,155,430",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the country of Canada.",
      "openfda": {
        "registration_number": [
          "1061124",
          "3009888344",
          "1643817",
          "3004111573",
          "1423537",
          "3007329020",
          "9616088"
        ],
        "fei_number": [
          "3002902735",
          "3009888344",
          "1643817",
          "3004111573",
          "1000512168",
          "3007329020",
          "3001236905"
        ],
        "device_name": "Radiographic Contrast Tray",
        "medical_specialty_description": "Radiology",
        "regulation_number": "892.1680",
        "device_class": "2"
      }
    }
  ]
}