  "results": [
    {
      "cfres_id": "151938",
      "product_res_number": "Z-0940-2017",
      "event_date_initiated": "2016-12-15",
      "event_date_created": "2017-01-03",
      "recall_status": "Terminated",
      "event_date_terminated": "2018-10-09",
      "res_event_number": "76014",
      "product_code": "OGZ",
      "k_numbers": [
        "K141682"
      ],
      "product_description": "TVC Imaging System model TVC-MC9 , for the near-infrared examination of coronary arteries in patients undergoing invasive coronary angiography\nThe TVC Imaging System is an intravascular imaging device that simultaneously assesses vessel composition and structure using near-infrared spectroscopy (NIRS) and intravascular ultrasound (IVUS). This dual-modality instrument performs near-infrared spectroscopic analysis of the vessel to detect lipid core-containing plaques of interest (LCP) displayed in a map called a Chemogram, and simultaneously generates high resolution IVUS images that display structural details of the vessel and plaque in transverse and longitudinal views",
      "code_info": "Serial Numbers: MC9-15013, MC9-15006, MC9-15005, MC9-15008, MC9-15012, MC9-15007 and  12718M00009 (spare laser assembly)",
      "firm_fei_number": "3015551113",
      "recalling_firm": "Infraredx, Inc.",
      "address_1": "34 3rd Ave",
      "city": "Burlington",
      "state": "MA",
      "postal_code": "01803-4414",
      "additional_info_contact": "SAME\n781-221-0053",
      "reason_for_recall": "Near-infrared spectroscopic image (Chemogram) generated by the TVC Imaging System TVC-MC9 in some cases demonstrated an unusual appearance in that the yellow in the image (indicating the presence of lipid core plaques) was horizontally-oriented",
      "root_cause_description": "Device Design",
      "action": "Infraredx, Inc. sent an Urgent Product Recall Medical Device Field Correction letter dated December 15, 2016, describing the problem and replacing the defective lasers. A Field Service Representative will contact customers to schedule a replacement of the system's laser rack module. If customers receive this letter and are not contacted within 48 hours, please call Infraredx Technical Support@ 1-800-596-3104.\nFor questions regarding this recall call 781-221-0053.",
      "product_quantity": "7 units",
      "distribution_pattern": "Worldwide Distribution - US including MI and Internationally to South Korea",
      "openfda": {
        "k_number": [
          "K093993",
          "K123108",
          "K241576",
          "K163345",
          "K072932",
          "K133897",
          "K130719",
          "K183599",
          "K141682"
        ],
        "registration_number": [
          "8043817",
          "3017654247",
          "1220477",
          "1061124",
          "3015551113",
          "2245304"
        ],
        "fei_number": [
          "3017654247",
          "1220477",
          "1000512168",
          "3003094575",
          "3015551113",
          "2245304"
        ],
        "device_name": "Catheter, Intravascular, Plaque Morphology Evaluation",
        "medical_specialty_description": "Cardiovascular",
        "regulation_number": "870.1200",
        "device_class": "2"
      }
    }
  ]
}