  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 2
    }
  },
  "results": [
    {
      "cfres_id": "110377",
      "product_res_number": "Z-2001-2012",
      "event_date_initiated": "2012-06-18",
      "event_date_posted": "2012-07-12",
      "recall_status": "Terminated",
      "event_date_terminated": "2013-02-28",
      "res_event_number": "62381",
      "product_code": "OGK",
      "k_numbers": [
        "K052260"
      ],
      "product_description": "superDimension Systems AAS00016-xx with Software Version 4.0-4.9.\r\n\r\nIndicated for use for displaying images of the tracheobronchial tree to aid the physician in guiding endoscopic tools or catheters in the pulmonary tract and to enable marker placement within soft lung tissue. It does not make a diagnosis and is not an endoscopic tool. Not for pediatric use.",
      "code_info": "S030097\tS0030044\tS030065\tS030092\tS030579\tS420011\tS420102 I4202008\tS0030045\tS030066\tS030092\tS030586\tS420012\tS420105 I4202012\tS00312\tS030067\tS030094\tS030587\tS420013\tS420107 S0030011\tS030043\tS030069\tS030096\tS030588\tS420015\tS420108 S0030013\tS030046\tS030070\tS030098\tS030604\tS420019\tS420109 S0030014\tS030049\tS030072\tS030099\tS030605\tS420020\tS420110 S0030015\tS030052\tS030075\tS030102\tS030608\tS420021\tS420112 S0030017\tS030054\tS030076\tS030105\tS030609\tS420026\tS420114 S0030021\tS030056\tS030077\tS030449\tS300049\tS420036\tS420125 S0030022\tS030057\tS030080\tS030477\tS300082\tS420037\tS830001 S0030023\tS030058\tS030081\tS030492\tS310078\tS420077\t S0030024\tS030059\tS030083\tS030541\tS310100\tS420079\t S0030025\tS030060\tS030085\tS030546\tS420006\tS420080\t S0030026\tS030061\tS030087\tS030547\tS420008\tS420092\t S0030042\tS030064\tS030089\tS030573\tS420009\tS420100",
      "firm_fei_number": "3004962788",
      "recalling_firm": "superDimension, Inc",
      "address_1": "161 Cheshire Ln N Ste 100 161 Cheshire Ln N",
      "address_2": "Ste 100",
      "city": "Minneapolis",
      "state": "MN",
      "postal_code": "55441-5433",
      "additional_info_contact": "763-210-4000",
      "reason_for_recall": "superDimension, Inc., is conducting a recall of certain superDimension Systems operating Software Version  4.0 - 4.9. The affected systems are sold under the product code AAS00016-xx.\r\n\r\nThe products being recalled have been found to malfunction when connected to the superDimension Locatable Guide (LG). Specifically, the Systems using Software Version 4.0-4.9 will not recognize the Locatable Guide",
      "root_cause_description": "Device Design",
      "action": "superDimension sent an \"URGENT MEDICAL DEVICE CORRECTION\" letter dated June 15, 2012 to all affected customers.  The letter identifies the product, problem, and actions to be taken by the customers.    The letter advised customers to discontinue use of the system and requested consignees to complete and return the acknowledgement receipt form via fax to 763-210-4098.  Contact the firm at 763-210-4057 for questions regarding this recall.",
      "product_quantity": "99",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) including Puerto Rico and the states of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, Florida, Georgia, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, Ohio, Oklahoma, Oregon, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Vermont, Virginia, Washington, West Virginia, and Wisconsin and the countries of  Canada, Australia, Austria, Belgium, Czech Republic, Denmark, France, Germany, Hungary, Italy, Netherlands, South Africa, Spain, Switzerland, Turkey, and United Kingdom.",
      "openfda": {
        "registration_number": [
          "3003911670",
          "1420054",
          "3004111573",
          "3009888344",
          "3016548968",
          "1643817",
          "1054241",
          "1061124",
          "1423537",
          "1043214",
          "3005803389",
          "9616088",
          "3007329020",
          "3020460367"
        ],
        "fei_number": [
          "3002902735",
          "3003911670",
          "1420054",
          "3004111573",
          "3009888344",
          "3016548968",
          "1643817",
          "3001236905",
          "1000112877",
          "3005803389",
          "1000512168",
          "1043214",
          "3007329020",
          "3020460367"
        ],
        "device_name": "Orotracheal Intubation Guide Kit",
        "medical_specialty_description": "Anesthesiology",
        "regulation_number": "868.1400",
        "device_class": "2"
      }
    }
  ]
}