  "results": [
    {
      "cfres_id": "220376",
      "product_res_number": "Z-2416-2026",
      "event_date_initiated": "2026-04-27",
      "event_date_posted": "2026-06-12",
      "recall_status": "Open, Classified",
      "res_event_number": "98951",
      "product_code": "OGI",
      "product_description": "BLOCK TRAY\nDYNJRA2119A\n\nINDWELLING NERVE BLOCK\nDYNJRA1011A",
      "code_info": "DYNJRA2119A\nUDI-DI 10198459194702\nlot\n25LBB738\n\nDYNJRA1011A\nUDI-DI 10888277783614\nlots\n25KBB849\n25LBB734",
      "firm_fei_number": "1417592",
      "recalling_firm": "Medline Industries, LP",
      "address_1": "3 Lakes Dr",
      "city": "Northfield",
      "state": "IL",
      "postal_code": "60093-2753",
      "additional_info_contact": "Haley Barclay\n800-633-5463",
      "reason_for_recall": "Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.",
      "root_cause_description": "Process control",
      "action": "Firm began notifying consignees on 4/27/2026. Customers were instructed to request stickers to over-label affected convenience kits with a warning label. Warning label indicates for user to remove the affected component and replace it with product from supply. No kits are to be returned.",
      "product_quantity": "49,654 kits total",
      "distribution_pattern": "US Nationwide distribution. OUS distribution pending.",
      "openfda": {
        "device_name": "Nasal Endotracheal Tube Holder Kit",
        "medical_specialty_description": "Anesthesiology",
        "regulation_number": "868.5770",
        "device_class": "1",
        "registration_number": [
          "3009888344",
          "9616088",
          "3004111573",
          "1061124",
          "1423537",
          "3007329020",
          "1643817"
        ],
        "fei_number": [
          "3009888344",
          "3004111573",
          "1000512168",
          "3007329020",
          "1643817",
          "3002902735",
          "3001236905"
        ]
      }
    }
  ]
}