{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 2
    }
  },
  "results": [
    {
      "cfres_id": "220343",
      "product_res_number": "Z-2390-2026",
      "event_date_initiated": "2026-04-21",
      "event_date_posted": "2026-06-13",
      "recall_status": "Open, Classified",
      "res_event_number": "98793",
      "product_code": "OFK",
      "product_description": "Description/REF:\nSAC KIT: 18 GA X 12CM W BIO/SAC-01218-B;\nSAC KIT: 18 GA X 12CM W CHG/SAC-01218-C;\nSAC KIT: 18 GA X 16CM W BIO/SAC-01618-B",
      "code_info": "REF(Material)/UDI-DI/Lot(Batch):\nSAC-01218-B/10801902225303/33F26C0515;\nSAC-01218-C/10801902225273/33F26C0517;\nSAC-01618-B/10801902225310/33F26C0518",
      "firm_fei_number": "3013162291",
      "recalling_firm": "ARROW INTERNATIONAL, LLC",
      "address_1": "3015 Carrington Mill Blvd",
      "city": "Morrisville",
      "state": "NC",
      "postal_code": "27560-5437",
      "additional_info_contact": "Customer Service\n866-396-2111",
      "reason_for_recall": "Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.",
      "root_cause_description": "Nonconforming Material/Component",
      "action": "On 4/21/2026, correction notices (Firm Event Number: EIF-000608, lidocaine and bupivacaine) were mailed to customers who were asked to do the following:\n\nArrow International LLC is voluntarily recalling product due to the Huons Co. Ltd. recall of Lidocaine and Bupivacaine, which is included in impacted Arrow Kits & Sets.\n\nMedical facilities\n1. Provide this correction notice to all persons who need to be aware within your organization or to any organization where the potentially affected product has been transferred.\n2. Check your inventory for product within the scope of this Medical Device Notification and place a copy of this notice with all affected product.\n3. Complete and return the response form virtually via:\nhttps://fca.myteleflex.com/en/recall/000608\n\nDistributors:\n1. Provide Medical Device Notification to all customers who have received impacted product. Each of your customers is then required to complete the Acknowledgement Form and return it to you as their distributor.\n2. Check your inventory for affected product. Place a copy of this notification with all affected product and ensure that the notification is included with future distributions of affected product.\n3. Complete and return the response form via link: https://fca.myteleflex.com/en/recall/000608\n4. If you have further distributed product outside of your country, please notify Teleflex Customer Service by return email.\n\nIf you have questions contact firm's Customer Service: \nLink: https://fca.myteleflex.com/en/recall/000608,\nCustomer Service: Telephone: 1-866-396-2111,\nFAX: 1-855-419-8507 Email: Recalls@teleflex.com\n\nOn 5/29/2026, additional amended notices (Firm Event Number: EIF-000608-01) were mailed to include additional affected devices.\n\nOn 5/1/2026, additional notices (Firm Event Number: EIF-000609, sodium chloride) were mailed and emailed to customers who have 0.9% sodium chloride (10mL saline solution ampule) in kits and sets.",
      "product_quantity": "0",
      "distribution_pattern": "US Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR.",
      "openfda": {
        "registration_number": [
          "3010041511",
          "9616088",
          "1220477",
          "3004111573",
          "3007695959",
          "1000121056",
          "3007329020",
          "3009888344",
          "3011137372",
          "3015859709",
          "1061124",
          "1223004",
          "3025341484",
          "2245270",
          "3011836011",
          "9680794",
          "1423537",
          "3019751610",
          "2124215",
          "1643817"
        ],
        "fei_number": [
          "3010041511",
          "3003737899",
          "1220477",
          "3004111573",
          "3007695959",
          "1000121056",
          "3007329020",
          "3002902735",
          "3001236905",
          "3009888344",
          "1000512168",
          "3015859709",
          "1000120555",
          "3025341484",
          "2245270",
          "3011836011",
          "3005747797",
          "3019751610",
          "2124215",
          "1643817"
        ],
        "device_name": "Percutaneous Atrial Catheter Kit",
        "medical_specialty_description": "Cardiovascular",
        "regulation_number": "870.1250",
        "device_class": "2"
      }
    }
  ]
}