  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "208756",
      "product_res_number": "Z-2559-2024",
      "event_date_initiated": "2024-06-11",
      "event_date_posted": "2024-08-12",
      "recall_status": "Open, Classified",
      "res_event_number": "94797",
      "product_code": "ODY",
      "k_numbers": [
        "K063048"
      ],
      "product_description": "Various Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd. syringes:\n\nOutake, REF CET200 and CET300, \n\nThe Catheter Extractor is a catheter accessory device intended for minimally invasive removal of tunneled long term catheters. The Catheter Extractor is indicated for use with tunneled, double lumen hemodialysis catheters with cuff sizes ranging from 10.8 French to 14.5 French.",
      "code_info": "Catalog Number: CET200\nUDI-DI code: 00884450052587\nLot Numbers:\nT2651502\nT2700122\nT2712304\nT2748973\nT2758209\nT2779867\nT2833064\nT2850758\nT2903817\n\nCatalog Number: CET300\nUDI-DI code: 00884450045497\nT2626011\nT2662043\nT2701699\nT2726893\nT2750399\nT2758210\nT2805965\nT2832137\nT2849258\nT2886083\nT2921078",
      "firm_fei_number": "1721504",
      "recalling_firm": "Merit Medical Systems, Inc.",
      "address_1": "1600 W Merit Pkwy",
      "city": "South Jordan",
      "state": "UT",
      "postal_code": "84095-2416",
      "additional_info_contact": "801-253-1600",
      "reason_for_recall": "Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.",
      "root_cause_description": "No Marketing Application",
      "action": "On 06/11/2024, the firm sent an \"URGENT MEDICAL DEVICE RECALL NOTICE\" Letter to customers informing them that specific Merit Medical convenience kits contain plastic syringes manufactured by Jiangsu Shenli Medical Production Co. Ltd. These syringes lack the appropriate/sufficient FDA market authorization.  In addition, Merit Medical posted the Urgent Medical Device Recall letter and complete list of affected products/kits on their website at https://www.merit.com/products/documents/product-notices/.\n\nCustomer are instructed to:\n1. Immediately determine if any of the products identified in the attached Customer Response Form (CRF) are within their facility, quarantine them, and discontinue use and distribution.\n2. Ensure that appropriate personnel within their organization are made aware of this field action.\n3. If the product has been further distributed to other facilities, institutions, or manufacturers, ensure this notice is immediately shared with them. Additional distribution details may be required by health authorities.\n4. Follow the attached instructions for placing the recall labels on the products containing\nJiangsu Shenli syringes, which inform the users to not use the affected syringes. If additional recall labels are needed, contact Customer Service at (801) 208-4381. Product may be returned to inventory once properly over-labeled.\n5.At the point of use, the Jiangsu Shenli syringes are to be discarded. Photographs of Jiangsu Shenli syringes are enclosed to assist in identification.\n6.Post this  URGENT MEDICAL DEVICE RECALL NOTICE  and Jiangsu Shenli Syringe photographs on or near the affected products.\n7.If replacement syringes are needed, contact Customer Service at (801) 208-4381 for recommended syringe suppliers.\n\nFor questions contact Merit Sales Representative or Merit Customer Service at 801-208-4381, Hours: 6 am to 6 pm MST Mon.-Fri.",
      "product_quantity": "3631 kits",
      "distribution_pattern": "U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.",
      "openfda": {
        "device_name": "Tunneled Catheter Remover",
        "medical_specialty_description": "General Hospital",
        "regulation_number": "880.5970",
        "device_class": "2",
        "k_number": [
          "K063048",
          "K122263"
        ],
        "registration_number": [
          "3015309643",
          "1721504",
          "3006950086",
          "1125782",
          "1061124",
          "3012536737",
          "1123137",
          "3015225571",
          "3033589330"
        ],
        "fei_number": [
          "3015309643",
          "1721504",
          "3006950086",
          "1000123180",
          "1000512168",
          "3012536737",
          "3015225571",
          "3033589330",
          "3002975287"
        ]
      }
    }
  ]
}