  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "84481",
      "product_res_number": "Z-1404-2010",
      "event_date_initiated": "2009-07-27",
      "event_date_posted": "2010-04-16",
      "recall_status": "Terminated",
      "event_date_terminated": "2011-01-08",
      "res_event_number": "53031",
      "product_code": "NVJ",
      "k_numbers": [
        "K042132"
      ],
      "product_description": "Renessa RF System, model PR0918, manufactured by Novasys Medical, Newark, CA",
      "code_info": "Issue affects all lots in distribution.",
      "firm_fei_number": "3003647794",
      "recalling_firm": "Novasys Medical Inc",
      "address_1": "39684 Eureka Dr",
      "city": "Newark",
      "state": "CA",
      "postal_code": "94560-4805",
      "additional_info_contact": "Charles S. Carignan, MD\n510-226-4060",
      "reason_for_recall": "Instructions for Use (IFU) revision-- Because of complaints received by the firm of worsened stress incontinence symptoms 3-4 weeks after treatment, the instructions for use were revised to emphasize potential side effects.  Although Novasys has not, to date, identified a device malfunction associated with the symptoms discussed above, the investigation over time shows that there is a correlation",
      "root_cause_description": "Labeling design",
      "action": "A Field Action notification letter was sent to all current customers (physicians) along with the revised instructions for use, and a small supply of the revised patient information brochures.  The company also updated physician information brochures and patient/physician DVDs to include the information.  This was shipped to physicians in August/September of 2009.  Additional copies of the Patient Brochures can be ordered by contacting Novasys Customer Service at 866-784-4777.  Questions should be addressed to the firm at the same telephone number.",
      "product_quantity": "As of 6/12/09 there were 436 Probes at customer accounts.",
      "distribution_pattern": "US nationwide only. (TX, FL, CA, LA, KY, MO, GA,OR WA, HI, AZ, OK AL, TN) .",
      "openfda": {
        "k_number": [
          "K042132"
        ],
        "device_name": "Applicator, Transurethral, Radio Frequency, For Stress Urinary Incontinence In Women",
        "medical_specialty_description": "General, Plastic Surgery",
        "regulation_number": "878.4400",
        "device_class": "2"
      }
    }
  ]
}