  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 2
    }
  },
  "results": [
    {
      "cfres_id": "86866",
      "product_res_number": "Z-1196-2010",
      "event_date_initiated": "2009-11-02",
      "event_date_posted": "2010-03-23",
      "recall_status": "Terminated",
      "event_date_terminated": "2013-05-23",
      "res_event_number": "53816",
      "product_code": "NSA",
      "k_numbers": [
        "K040904"
      ],
      "product_description": "Heartstart HS1 Defibrillator Model M5068A",
      "code_info": "DOMESTIC UNIT SERIAL NUMBERS - A09J-02644, A09J-02180, A09J-02643, A09J-02499, A09J-02169, A09J-02175, A09J-02554, A09J-02645, A09J-01891, A09J-02382, A09J-01903, A09J-02361, A09J-02380, A09J-02033, A09J-02409, A09J-01904, A09J-02351, A09J-02384, A09J-02675, A09J-02625, A09J-02628, A09J-01901, A09J-02360, A09J-02485, A09J-02631, A09J-02642, A09J-02641, A09J-02383, A09J-02647, A09J-02693, A09J-02269, and A09J-02552., A09J-02493, and A09J-01974.",
      "firm_fei_number": "1000524572",
      "recalling_firm": "Philips Medical Systems",
      "address_1": "2301 5th Ave Ste 200 2301 5th Ave",
      "address_2": "Ste 200",
      "city": "Seattle",
      "state": "WA",
      "postal_code": "98121",
      "additional_info_contact": "206-664-5290",
      "reason_for_recall": "Potential for failure of a capacitor during use which would prevent the AED from delivering effective defibrillation therapy when indicated.",
      "root_cause_description": "Nonconforming Material/Component",
      "action": "The firm notified their US distributors of the recall and instructed them to contact customers and provide information how to return the product to Philips for a free replacement.  Firm contacted distributors via phone and email on 11/2/09 and by letter on 11/9/09.  A web site providing information related to the recall was activated on 11/3/09.",
      "product_quantity": "34 units distributed domestically",
      "distribution_pattern": "Worldwide Distribution.",
      "openfda": {
        "pma_number": [
          "P160029",
          "P160008"
        ],
        "registration_number": [
          "3012607778",
          "3009077524",
          "3006108746",
          "3030677",
          "3000126629",
          "1930027",
          "3011823740",
          "3004123209"
        ],
        "fei_number": [
          "3012607778",
          "1000524572",
          "3009077524",
          "3040313479",
          "3000126629",
          "1930027",
          "3011823740",
          "3004123209"
        ],
        "device_name": "Over-The-Counter Automated External Defibrillator",
        "medical_specialty_description": "Cardiovascular",
        "regulation_number": "870.5310",
        "device_class": "3",
        "k_number": [
          "K040904"
        ]
      }
    }
  ]
}