  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "148049",
      "product_res_number": "Z-2743-2016",
      "event_date_initiated": "2016-06-17",
      "event_date_created": "2016-09-07",
      "recall_status": "Terminated",
      "event_date_terminated": "2017-11-07",
      "res_event_number": "74613",
      "product_code": "NOV",
      "product_description": "Achillon Minimally Invasive Device for Achilles Tendon Rupture Treatment\nThe Achillon Suture System is a single use, sterile device used to treat acute Achilles tendon ruptures. The Achillon Suture System is a mini-invasive procedure indicated to teat acute Achilles tendon ruptures: Acute rupture of the Achill tendon(<10 days); Rupture located between 2 and 8 cm above calcaneum; Open or closed ruptures",
      "code_info": "Model #'s 19700 and 19700ND",
      "firm_fei_number": "3003418325",
      "recalling_firm": "Integra LifeSciences Corp.",
      "address_1": "311 Enterprise Dr",
      "city": "Plainsboro",
      "state": "NJ",
      "postal_code": "08536-3344",
      "additional_info_contact": "Dr. Patricia Kihn\n717-840-3431",
      "reason_for_recall": "Data from internal studies suggests there may be a potential that the secondary package seal integrity does not provide the level of sterility intended for Minimally Invasive Achilles Tendon Suture System Achillon.",
      "root_cause_description": "Packaging",
      "action": "Integra sent an Urgent Voluntary Medical Device Recall letter and recall acknowledgment and return form dated June 17, 2016 to their customers via traceable courier.  Customers were asked to review their inventory and complete the attached form.  Customers with questions should contact Customer Service at 1-800-654-2873.\nFor questions regarding this recall call 717-840-3431.",
      "product_quantity": "6,729 units",
      "distribution_pattern": "Nationwide distribution",
      "openfda": {
        "k_number": [
          "K032078"
        ],
        "registration_number": [
          "3034676720",
          "3008868758",
          "3008114965",
          "1221934"
        ],
        "fei_number": [
          "3034676720",
          "3008868758",
          "3008114965",
          "1221934"
        ],
        "device_name": "Anchor, Suture, Bone Fixation, Metallic",
        "medical_specialty_description": "Orthopedic",
        "regulation_number": "888.3030",
        "device_class": "2"
      }
    }
  ]
}