  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "129373",
      "product_res_number": "Z-2644-2014",
      "event_date_initiated": "2014-08-13",
      "event_date_posted": "2014-09-15",
      "recall_status": "Terminated",
      "event_date_terminated": "2016-10-24",
      "res_event_number": "69032",
      "product_code": "NHT",
      "k_numbers": [
        "K131930"
      ],
      "product_description": "JBAIDS Instrument System Software version 3.5.0.72.\r\n\r\nThe JBAIDS instrument is a portable thermocycler and real-time fluorimeter capable of simultaneous and rapid identification of multiple biological warfare and clinical pathogens from a variety of biological specimens. The software and instrument provides an integrated environment for performing the required functions to control JBAIDS instruments and to perform PCR tests and analysis.",
      "code_info": "Model Number(s): JRPD-SFW-0001",
      "firm_fei_number": "3010770794",
      "recalling_firm": "Biofire Defense",
      "address_1": "79 W 4500 S",
      "city": "Salt Lake City",
      "state": "UT",
      "postal_code": "84107-2649",
      "additional_info_contact": "Amber Bawden\n801-262-3592 Ext. 1458",
      "reason_for_recall": "BioFire Defense, LLC is initiating a field action for JBAIDS Software version 3.5.0.72 because the wrong version software may have been installed.",
      "root_cause_description": "Software Manufacturing/Software Deployment",
      "action": "Consignees were notified on 8/13/14 via letter and telephone and were provider with instructions on how to install the correct software.",
      "product_quantity": "43 Units w/ laptops",
      "distribution_pattern": "Distributed USA (nationwide) and Guam.",
      "openfda": {
        "k_number": [
          "K192871",
          "K071188",
          "K262368",
          "K051713",
          "K140426",
          "K131930",
          "K072631"
        ],
        "device_name": "Assay, Nucleic Acid Amplification, Bacillus Anthracis",
        "medical_specialty_description": "Microbiology",
        "regulation_number": "866.3045",
        "device_class": "2",
        "registration_number": [
          "1050190"
        ],
        "fei_number": [
          "1050190"
        ]
      }
    }
  ]
}