  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 4
    }
  },
  "results": [
    {
      "cfres_id": "46607",
      "product_res_number": "Z-1219-06",
      "event_date_initiated": "2006-05-24",
      "event_date_posted": "2006-07-12",
      "recall_status": "Terminated",
      "event_date_terminated": "2006-07-11",
      "res_event_number": "35680",
      "product_code": "NFA",
      "k_numbers": [
        "K042040"
      ],
      "product_description": "TriActiv System Procedure Kit. Balloon Protected Flush Extraction System 190 cm (for embolism protection). The kit includes AutoStream Flow Control, Flush Kit, ShieldWire (temporary occlusion balloon guidewire), and Balloon Inflation Syringe). The system is used with a standard 7F guide catheter using the standard femoral approach.",
      "code_info": "Lot number 47303 exp 9/30/06",
      "firm_fei_number": "1000307073",
      "recalling_firm": "Kensey Nash Corp",
      "address_1": "735 Pennsylvania Dr",
      "city": "Exton",
      "state": "PA",
      "postal_code": "19341-1130",
      "additional_info_contact": "Robin Fatzinger, RAC\n484-713-2100",
      "reason_for_recall": "Volume control knob not able to go up to larger diameters",
      "root_cause_description": "Other",
      "action": "A letter was issued to the sales representative that have customers which received affected product, 5/24/2006. Sales representatives visited their sites within 3 days to determine if the affected product was still in inventory. If the product was still in inventory the sales representative provided the trained TriActiv System user(s) a Technical Bulletin that instructed the users how to correct the knob position.",
      "product_quantity": "10 kits",
      "distribution_pattern": "Nationwide.  The products are shipped to hospitals in AL, DE, Fl, IL, IN, MI, PA, SC, TN and TX.",
      "openfda": {
        "k_number": [
          "K013913",
          "K052280",
          "K061772",
          "K061332",
          "K023691",
          "K073563",
          "K073523",
          "K042040",
          "K053195",
          "K063785",
          "K051179",
          "K023303",
          "K060651",
          "K030201",
          "K003992",
          "K062870",
          "K023878",
          "K062201",
          "K052523",
          "K051984",
          "K014223",
          "K032884"
        ],
        "registration_number": [
          "3007695959",
          "3007738736",
          "2124215",
          "2183870",
          "3004859241",
          "3002807314",
          "2183744"
        ],
        "fei_number": [
          "3007695959",
          "3007738736",
          "2124215",
          "2183870",
          "3004859241",
          "3002807314",
          "2183744"
        ],
        "device_name": "Temporary Coronary Saphenous Vein Bypass Graft For Embolic Protection",
        "medical_specialty_description": "Cardiovascular",
        "regulation_number": "870.1250",
        "device_class": "2"
      }
    }
  ]
}