  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 7
    }
  },
  "results": [
    {
      "cfres_id": "196168",
      "product_res_number": "Z-0104-2023",
      "event_date_initiated": "2022-09-19",
      "event_date_posted": "2022-10-20",
      "recall_status": "Open, Classified",
      "res_event_number": "90900",
      "product_code": "NDW",
      "k_numbers": [
        "K050374"
      ],
      "product_description": "Atellica CH Lithium_2 (LITH_2), Siemens Material Number 11532401. in vitro diagnostic assay",
      "code_info": "UDI-DI: 00630414287935\n\nAll lots impacted:\n\nLots \n110238 exp 10/18/22\n110302 exp 12/13/22\n120017 exp 4/18/23",
      "firm_fei_number": "2432235",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "511 Benedict Ave",
      "city": "Tarrytown",
      "state": "NY",
      "postal_code": "10591-5005",
      "additional_info_contact": "Jennifer Beecher\n914-631-8000",
      "reason_for_recall": "Potential for falsely elevated LITH_2 and Li results on the Atellica CH 930 Analyzer. In the worst case, there is a potential for erroneously elevated lithium patient results, which may lead to inappropriate dosage adjustment, potentially affecting treatment efficacy.",
      "root_cause_description": "Under Investigation by firm",
      "action": "On September 19, 2022, the firm issued an Urgent Medical Device Correction letter to affected customers. Customers should take the following actions:\n\n-Please review this letter with your Medical Director.\n-Discontinue use of and discard the affected product.\n\nSiemens is temporarily discontinuing production and distribution of the Atellica CH lithium reagents\nuntil the issue is resolved. In the interim, Siemens is recommending that customers transition to an\nalternate lithium assay.",
      "product_quantity": "1342 US; 2266 OUS",
      "distribution_pattern": "Worldwide distribution - US Nationwide.",
      "openfda": {
        "k_number": [
          "K050374",
          "K063684",
          "K243462",
          "K063705",
          "K070987",
          "K003583",
          "K112142"
        ],
        "registration_number": [
          "3006198300",
          "3003795116",
          "9681753",
          "2517506",
          "9610126",
          "1181121",
          "3005333358",
          "9610529",
          "3002809144",
          "2432235"
        ],
        "fei_number": [
          "3006198300",
          "3003795116",
          "2517506",
          "3002806559",
          "1181121",
          "3002806560",
          "3005333358",
          "3002809144",
          "3000303338",
          "2432235"
        ],
        "device_name": "Assay, Porphyrin, Spectrophotometry, Lithium",
        "medical_specialty_description": "Clinical Toxicology",
        "regulation_number": "862.3560",
        "device_class": "2"
      }
    }
  ]
}