      "cfres_id": "97616",
      "product_res_number": "Z-1925-2011",
      "event_date_initiated": "2010-12-21",
      "event_date_posted": "2011-04-08",
      "recall_status": "Terminated",
      "event_date_terminated": "2012-08-31",
      "res_event_number": "57866",
      "product_code": "NDR",
      "product_description": "Inhibin A ELISA kit, Part Number: DSL-10-28100-4\nThe Inhibin A ELISA kit is intended for the quantitative measurement of Dimeric Inhibin A in human serum or plasma. It is intended strictly for in vitro use as an aid in the diagnosis and monitoring of various hormonal reproductive disorders.",
      "code_info": "Lot #: 090779 (EXP: 02/04/2011), 091046 (EXP: 02/16/2011), 091053(EXP: 11/09/2011)",
      "firm_fei_number": "2050012",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "200 S Kraemer Blvd",
      "city": "Brea",
      "state": "CA",
      "postal_code": "92821-6208",
      "additional_info_contact": "Clair K. O'Donovan, Ph.D.\n714-961-4483",
      "reason_for_recall": "The recall was initiated because the lots of Inhibin A identified above may contain microbial contamination in the conjugate diluent bottles\nincluded in the kit.",
      "root_cause_description": "Pending",
      "action": "Beckman Coulter sent an Urgent Product Corrective Action letter dated December 21, 2010, to all affected customers.  The letter identified the product, the problem, and the action to be taken by the customer.  \nCustomers were instructed to discontinue use of the affected product and discard all remaining inventory.  \n  \nIn addition customers were requested to share the information with their laboratory staff and retain the notification as part of their laboratory\nQuality System documentation. Customers were instructed to complete and return the enclosed response form within 10 days so\nBeckman could be assured that customers received this important notification.\nFor product replacement:\n1.  In the United States, please contact Customer Service at 1-800-526-3821, option 1.\n2.  In Canada, please contact Customer Service at 1-800-463-7828.\n3.  Outside of the United States and Canada, contact your local Beckman Coulter representative.\nCustomers who needed assistance or had any questions regarding this notification, were instructed to contact Technical Support \nat 1-800-854-3633 in the United States and Canada. For those customers outside the United States and Canada, instructions were given to contact their local Beckman Coulter representative.",
      "product_quantity": "740 total units for both products",
      "distribution_pattern": "Worldwide Distribution - USA including NY, IN, PA, NM, OH, TX, AZ, NC, TN, MI and CT and the countries Denmark, Germany, Spain, Turkey, United Arab Emirates, United Kingdom.",
      "openfda": {
        "registration_number": [
          "2517506",
          "3014325803",
          "3008198807",
          "2122870",
          "3022178699",
          "8043909",
          "3003741796"
        ],
        "fei_number": [
          "2517506",
          "3014325803",
          "3008198807",
          "2122870",
          "3002808498",
          "3022178699",
          "3003741796"
        ],
        "device_name": "Enzyme Immunoassay, Inhibin-A",
        "medical_specialty_description": "Clinical Chemistry",
        "regulation_number": "862.1260",
        "device_class": "1",
        "k_number": [
          "K002128"
        ]
      }
    }
  ]
}