  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 13
    }
  },
  "results": [
    {
      "cfres_id": "101555",
      "product_res_number": "Z-0835-2012",
      "event_date_initiated": "2010-11-08",
      "event_date_posted": "2012-01-23",
      "recall_status": "Terminated",
      "event_date_terminated": "2012-01-31",
      "res_event_number": "59181",
      "product_code": "NDJ",
      "k_numbers": [
        "K953385"
      ],
      "product_description": "Package of cross-drive bone screws labeled in part: \"***REF #95-1504***\"1.5MM\" SYSTEM HIGH TORQUE (HT) CROSS-DRIVE SCREW, 5/ PK***1.5 X4MM TITANIUM***CAUTION: FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN***BIOMET MICROFIXATION***1520 TRADEPORT DRIVE***JACKSONVILLE, FL 32218 USA***www.biometmicrofixation.com***\"\n\n\"***REF: 91-1504***\"1.5MM\" SYSTEM HIGH TORQUE (HT) CROSS-DRIVE SCREW***1.5 X4MM TITANIUM***CAUTION: FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN.***BIOMET MICROFIXATION***1520 TRADEPORT DRIVE***JACKSONVILLE, FL 32218 USA***www.biometmicrofixation.com***\"\n\nProducts subject to recall:\nREF #95-1504***\"1.5MM\" SYSTEM HIGH TORQUE (HT) CROSS-DRIVE SCREW, 5/ PK, Lot 622120.          \n\nREF: 91-1504***\"1.5MM\" SYSTEM HIGH TORQUE (HT) CROSS-DRIVE SCREW, 1/PK, Lot 622120.    \n\nInternal fixation screw intended to aid in the alignment and stabilization of bone in the oral cranio-maxillofacial skeletal system.",
      "code_info": "Lot # 622120",
      "firm_fei_number": "1032347",
      "recalling_firm": "Biomet Microfixation, Inc.",
      "address_1": "1520 Tradeport Dr",
      "city": "Jacksonville",
      "state": "FL",
      "postal_code": "32218-2480",
      "additional_info_contact": "Christy Cain\n904-741-9498",
      "reason_for_recall": "On 11/08/2010 Biomet Microfixation, Jacksonville, FL initiated a recall of the Biomet 1 count (REF 91-1504, Lot 622120) and 5 count (REF 95-1504, Lot 622120)   packages of 1.5MM System High Torque (HT) Cross-Drive Screw 5/Pk and the Biomet 1 count (REF 99-7204, Lot 202450) and 5 count packages (REF 01/7204, Lot 202450) of the 1.5MM System Center-Drive Screw, 5/Pk because of a possibility that thes",
      "root_cause_description": "Mixed-up of materials/components",
      "action": "Biomet Microfixation, Jacksonville, FL, sent an Urgent Medical Device Safety Alert letter dated November 8, 2010, to all affected consignees.  The letter identified the product, the problem, and the action to be taken by the consignee.  Consignees were instructed to fill out the attached Inventory Reconciliation sheet and FAX to 904-741-9425.  If consignees further distributed this product, they must notify the recipient of this action by providing the recipient a copy of the notice.    \nFor questions call 800-874-7711 or 904-741-4400, ext 468.",
      "product_quantity": "295",
      "distribution_pattern": "Worldwide Distribution - USA including CA, GA, OH, TX, and VA and the countries of S Africa and Japan",
      "openfda": {
        "registration_number": [
          "3004369035"
        ],
        "fei_number": [
          "3004369035"
        ],
        "device_name": "Screw, Fixation, Bone, Non-Spinal, Metallic",
        "medical_specialty_description": "Orthopedic",
        "regulation_number": "888.3040",
        "device_class": "2"
      }
    }
  ]
}