      "cfres_id": "151287",
      "product_res_number": "Z-0880-2017",
      "event_date_initiated": "2016-11-09",
      "event_date_created": "2016-12-22",
      "recall_status": "Terminated",
      "event_date_terminated": "2019-03-27",
      "res_event_number": "75792",
      "product_code": "NDE",
      "product_description": "GAM Kirschner Wire, Sterile R, T2 K-Wire, Sterile R and T2 K-Wire Recon, Stryker Trauma GmbH, Prof.-Kuntscher-Strasse 1-5, Schonkirchen, Germany 24232; T2 K-Wire, Sterile R  \nIntended for the temporary stabilization of bone segments or fragments",
      "code_info": "GAM Kirschner Wire - Manufacturing Part Number 12106450S, Lot #'s K0800BB, K0800BC, K0800BD, K0800BE, K0800BF, K0800C0, K081720, K081721, K0810872, K081723, K081727, K082C8B, K084F89, K084F98, K08683D, K08683E, K086841, K08820B, K089AF5, K08E1E1, K0911F4, K09379B, K096A26, K096A2A, K09A2C, K098213, K098215, K099AA0, K09AD33, K09BA4F, K09D564, K09F026, K0A1EF8, K0A1EFA, KoA1EFB, K0A38FB, K0A63AB, K0A63AC and K0A7BC1  T2 K-Wire - Manufacturing Part Number 18060050S, Lot #'s K084FBF, K0920A4, K0A1EFD, K093747, K0937C4, K09BA53, K09BA54, K09BA55 and K09BA56  T2 K-Wire Recon - Manufacturing Part Number 18063030S, Lot #'s K086847, K08E1E9 and K099AA8",
      "firm_fei_number": "3003070421",
      "recalling_firm": "Stryker Howmedica Osteonics Corp.",
      "address_1": "325 Corporate Dr",
      "city": "Mahwah",
      "state": "NJ",
      "postal_code": "07430-2006",
      "additional_info_contact": "Mr. William Van Ryn\n201-831-5000",
      "reason_for_recall": "A review of packaging revealed the seal integrity of the pouch may be compromised. More specifically, there is a potential that the sterile pouch is not sealed at one end due to a manufacturing error.",
      "root_cause_description": "Packaging process control",
      "action": "Stryker sent an Urgent Medical Device Removal letter dated November 9, 2016, to all affected customers via UPS (with return receipt). Stericycle will be handling the returns.  The letter identified the product, the problem, and the action to be taken by the customer.  Customers were instructed to inform users of the Medical Device Recall and forward the notice to all those individuals who need to be aware within their organization.  Return all affected products available at their location to:\nStryker C/O Stericycle\n2670 Executive Dr., Suite A\nIndianapolis, IN  46241\nAttn:  RA 2016-169 - Event XXX\nCustomers were also instructed to complete and sign the enclosed Business Reply Form and return to Stericycle.\nCustomers with questions should call +1-201-851-6683.\nFor questions regarding this recall call 201-831-5000.",
      "product_quantity": "2,084 units",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) Internationally to AU, CA, CL, CH, ES, FR, GB, SE, JP, FR, ZA, and NL.",
      "openfda": {
        "device_name": "Rod, Fixation, Intramedullary And Accessories, Metallic And Non-Collapsible",
        "medical_specialty_description": "Orthopedic",
        "regulation_number": "888.3020",
        "device_class": "2"
      }
    }
  ]
}