  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 2
    }
  },
  "results": [
    {
      "cfres_id": "213143",
      "product_res_number": "Z-1744-2025",
      "event_date_initiated": "2024-07-12",
      "event_date_posted": "2025-05-09",
      "recall_status": "Open, Classified",
      "res_event_number": "96551",
      "product_code": "NCB",
      "k_numbers": [
        "K150986"
      ],
      "product_description": "Brand Name: Contour\nProduct Name: Contour Gold Eyelid Weight\nModel/Catalog Number: LL3010\nProduct Description: Contour Gold Eyelid Weight, 1.0 gm\n\n\nMedDev Contour and ThinProfile Eyelid Weight Implants are designed for the gravity-assisted treatment of protracted permanent lagophthalmos, usually resulting from facial paralysis.",
      "code_info": "Model No LL3010;\nUDI code: 010081314202038217260727107227\nLot Number: 7227",
      "firm_fei_number": "2921577",
      "recalling_firm": "Meddev Corp",
      "address_1": "730 N Pastoria Ave",
      "city": "Sunnyvale",
      "state": "CA",
      "postal_code": "94085-3522",
      "additional_info_contact": "Catherine Leong\n+1-408-7309702",
      "reason_for_recall": "Due to mislabeling of products",
      "root_cause_description": "Process control",
      "action": "On 07/12/2024, the firm sent a customer notification letter via first class mail to inform customers that the labels on the Contour Gold Eyelid Implants LL3010/Lot # 7227 potentially do not match the weight; the implants labeled LL3010 (1.0 gm) may contain LL3012 (1.2 gm).\n\nCustomers are instructed to check their inventory for any/all of the affected products and to contact MedDev Corporation to request for an RGA number in or return affected products and receive replacement products.\n\nFor questions/concerns contact Quality Compliance Specialist at 1-800-543-2789 or email catherine@meddev-corp.com\n\nAn updated \"URGENT: MEDICAL DEVICE RECALL\" Letter was distributed to customers on 04/17/2025.  Updated Customer Letter is informing customer that the potential product mix-up may include an additional product and lot number (Model Number LL3012 and Lot Number 7228)",
      "product_quantity": "34 weights",
      "distribution_pattern": "U.S. Nationwide distribution in the states of CA, CO, CT, FL, IA, IL, MA, ME, MI, NE, NY, OH, OK, OR, SC, TX, and WV.",
      "openfda": {
        "k_number": [
          "K203569"
        ],
        "device_name": "Weights, Eyelid, Implantable",
        "medical_specialty_description": "Ophthalmic",
        "regulation_number": "886.5700",
        "device_class": "2",
        "registration_number": [
          "3002991496",
          "9615745",
          "3004467263"
        ],
        "fei_number": [
          "3002991496",
          "3004467263",
          "3000314405"
        ]
      }
    }
  ]
}