  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 2
    }
  },
  "results": [
    {
      "cfres_id": "180413",
      "product_res_number": "Z-1790-2020",
      "event_date_initiated": "2020-03-02",
      "event_date_created": "2020-04-23",
      "recall_status": "Terminated",
      "event_date_terminated": "2020-10-05",
      "res_event_number": "85282",
      "product_code": "NAO",
      "k_numbers": [
        "K060677"
      ],
      "product_description": "Siemens N Latex CDT Kit- IVD quantitative determination of carbohydrate-deficient transferrin (CDT) in human serum.\nSiemens Material Number (SMN):  10445997",
      "code_info": "US Lot number: 49822  UDI:  842768018534 OUS: Lot numbers: 49729, 49849, 49985, 50084  Exp. Date: 2020-04-11",
      "firm_fei_number": "3003098680",
      "recalling_firm": "CSL Behring GmbH",
      "address_1": "Emil-Von-Behring-Str. 76 76 emil-von-behring-str.",
      "city": "Marburg",
      "country": "Germany",
      "reason_for_recall": "Negative bias for the carbohydrate-deficient transferrin (CDT) measurement  in comparison to the HPLC method, observed to occur after 15 months based on shelf life monitoring",
      "root_cause_description": "Under Investigation by firm",
      "action": "Siemens Healthcare Diagnostics issued Urgent Medical Device Correction (UMDC) letter PP-20-002.A.US to the 5 affected customers in the U.S. on March 5, 2020 via FedEx overnight delivery. ( Urgent Field Corrective Action (UFCA) PP20-002.A.OUS was issued to all OUS Siemens sites with affected customers electronically for distribution and regional reporting to their local competent authorities). \nLetter states reason for recall, health risk, and action to take by the Customer Please review this letter with your Medical Director.\nDiscontinue use of and discard the kit lots listed in Table 1.\nReview your inventory of these products to determine your laboratorys replacement needs and to provide information to Siemens Healthineers for reporting to the authorities. Please retain this letter with your laboratory records and forward this letter to those who may have received this product.\nContact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.",
      "product_quantity": "297 PCK US; 1247 PCK OUS",
      "distribution_pattern": "IN, KS, MO, OH, VA\nForeign: Australia, Austria, Bangladesh, Belgium, Canada, China\nCzech Republic, Egypt, Finland, France, Germany, Great Britain, Hungary\nIndia, Italy, Japan, Mexico, Netherlands, Norway, Poland, Portugal, Republic Korea, Russia Federation, Slovakia, Slovenia, South Africa, Spain\nSweden, Switzerland, Taiwan",
      "openfda": {
        "k_number": [
          "K992502",
          "K060677",
          "K001943"
        ],
        "registration_number": [
          "9610806",
          "8023024",
          "3014325803"
        ],
        "fei_number": [
          "6387",
          "3001644026",
          "3014325803"
        ],
        "device_name": "Test, Carbohydrate Deficient Transferrin",
        "medical_specialty_description": "Clinical Chemistry",
        "regulation_number": "862.1360",
        "device_class": "1"
      }
    }
  ]
}