  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-11-26",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "cfres_id": "110870",
      "product_res_number": "Z-2171-2012",
      "event_date_initiated": "2012-02-01",
      "event_date_posted": "2012-08-09",
      "recall_status": "Terminated",
      "event_date_terminated": "2013-02-08",
      "res_event_number": "62533",
      "product_code": "MYT",
      "k_numbers": [
        "K051433"
      ],
      "product_description": "i-Stat CK-MP cartridges\r\nAbbott Point of Care Inc.\r\n\r\nThe i-Stat CKMB test is an in vitro diagnostic test for the quantitative measurement of creatinine kinase MB mass in whole blood or plasma samples. CK-MB measurements can be used as an aid in the diagnosis and treatment of myocardial infarction (MI).",
      "code_info": "List number 06F25-01, 06F25-02, 03P92-25,   All lots",
      "firm_fei_number": "2245578",
      "recalling_firm": "Abbott Point Of Care Inc.",
      "address_1": "400 College Rd E",
      "city": "Princeton",
      "state": "NJ",
      "postal_code": "08540-6607",
      "additional_info_contact": "Mr. Larry Krasley\n609-454-9272",
      "reason_for_recall": "The frequency of suppressed results for i-Stat cTnl, BNP and CK-MP cartridges is affected by atmospheric pressure.",
      "root_cause_description": "Device Design",
      "action": "Abbott Point of Care (APOC) sent an URGENT RECALL NOTICE dated February 2012 to all affected customers via e-mail and also contacted by phone.  The letter identified the product, the problem, and the action to be taken by the customer.  Customers were advised that if their facility was performing testing at more than approximately 7500 feet above sea level, they should have an alternate test method available should they experience an issue.  If they have forwarded any cartridges to another facility, they should provide a copy of the Urgent Recall Notice to them.  For questions customers were instructed to call 800-366-8020, Option 1 or contact their Abbott Point of Care representative.\r\nFor questions regarding this recall call 609-454-9000.",
      "product_quantity": "368,149 cartridges US; 32,725 OUS",
      "distribution_pattern": "Worldwide Distribution - USA including AZ and CO and internationally to  Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Chile, China, Colombia, Costa Rica, Dominican Republic, El Salvador, Finland, France, Georgia,  Germany, Greece, Honduras, Hong Kong, India, Ireland, Italy, Korea, Kuwait, Mauritius, Mexico, Netherland (Antilles), Nigeria, Norway, Oman, Paraguay, Peru, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, R.O.C., Thailand, UAE, United Kingdom, and Yemen.",
      "openfda": {
        "k_number": [
          "K984433",
          "K051433",
          "K033155"
        ],
        "registration_number": [
          "3005643513",
          "2245578",
          "3006198300"
        ],
        "fei_number": [
          "3005643513",
          "2245578",
          "3006198300"
        ],
        "device_name": "Biosensor, Immunoassay, Cpk Or Isoenzymes",
        "medical_specialty_description": "Clinical Chemistry",
        "regulation_number": "862.1215",
        "device_class": "2"
      }
    }
  ]
}